IChoice OX200 User manual

Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO2)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10.The sensor will not alert you if your readings are out of normal range.
There is no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
11.Explosion hazard: Do not use the sensor in an explosive atmosphere.
1. Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2. The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3. The sensor requires no routine calibration or maintenance.
4. Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note: Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings: 1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the ichoiceRelax app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH – Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX – Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP - Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR - Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter - Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check - Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO2and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO2: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO2: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by FACTORS INCLUDING BUT
NOT LIMITED TO:
12.Ethylene oxide sterilizing or immersing the sensors in liquid.
13.Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14.Intravascular dyes such as indocyanine green or methylene blue.
15.SpO2measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16.Excessive user movement.
17.High-frequency electrosurgical interference and defibrillators.
18.Venous pulsations.
19.Hypotension, severe vasoconstriction, severe anemia or hypothermia. My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11 Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse oximeter +
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO2) and your pulse rate.
The SMART pulse oximeter + relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse oximeter + relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol Symbol
Definition Definition
SpO2%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTINGCONNECTING TO APP (optional)PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO2.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.

Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO2)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10.The sensor will not alert you if your readings are out of normal range.
There is no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
11.Explosion hazard: Do not use the sensor in an explosive atmosphere.
1. Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2. The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3. The sensor requires no routine calibration or maintenance.
4. Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note: Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings: 1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the ichoiceRelax app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH – Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX – Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP - Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR - Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter - Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check - Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO2and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO2: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO2: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by FACTORS INCLUDING BUT
NOT LIMITED TO:
12.Ethylene oxide sterilizing or immersing the sensors in liquid.
13.Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14.Intravascular dyes such as indocyanine green or methylene blue.
15.SpO2measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16.Excessive user movement.
17.High-frequency electrosurgical interference and defibrillators.
18.Venous pulsations.
19.Hypotension, severe vasoconstriction, severe anemia or hypothermia. My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11 Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse oximeter +
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO2) and your pulse rate.
The SMART pulse oximeter + relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse oximeter + relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol Symbol
Definition Definition
SpO2%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTINGCONNECTING TO APP (optional)PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO2.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.

Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO2)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10.The sensor will not alert you if your readings are out of normal range.
There is no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
11.Explosion hazard: Do not use the sensor in an explosive atmosphere.
1. Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2. The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3. The sensor requires no routine calibration or maintenance.
4. Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note: Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings: 1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the ichoiceRelax app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH – Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX – Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP - Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR - Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter - Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check - Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO2and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO2: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO2: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by FACTORS INCLUDING BUT
NOT LIMITED TO:
12.Ethylene oxide sterilizing or immersing the sensors in liquid.
13.Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14.Intravascular dyes such as indocyanine green or methylene blue.
15.SpO2measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16.Excessive user movement.
17.High-frequency electrosurgical interference and defibrillators.
18.Venous pulsations.
19.Hypotension, severe vasoconstriction, severe anemia or hypothermia. My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11 Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse oximeter +
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO2) and your pulse rate.
The SMART pulse oximeter + relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse oximeter + relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol Symbol
Definition Definition
SpO2%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTINGCONNECTING TO APP (optional)PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO2.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.

Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO2)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10.The sensor will not alert you if your readings are out of normal range.
There is no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
11.Explosion hazard: Do not use the sensor in an explosive atmosphere.
1. Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2. The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3. The sensor requires no routine calibration or maintenance.
4. Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note: Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings: 1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the ichoiceRelax app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH – Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX – Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP - Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR - Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter - Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check - Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO2and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO2: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO2: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by FACTORS INCLUDING BUT
NOT LIMITED TO:
12.Ethylene oxide sterilizing or immersing the sensors in liquid.
13.Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14.Intravascular dyes such as indocyanine green or methylene blue.
15.SpO2measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16.Excessive user movement.
17.High-frequency electrosurgical interference and defibrillators.
18.Venous pulsations.
19.Hypotension, severe vasoconstriction, severe anemia or hypothermia. My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11 Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse oximeter +
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO2) and your pulse rate.
The SMART pulse oximeter + relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse oximeter + relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol Symbol
Definition Definition
SpO2%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTINGCONNECTING TO APP (optional)PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO2.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.

Bluetooth Indicator
Heart Rate (PR)
Signal Strength
Pulse Bar
Oxygen Level (SpO2)
Battery Indicator
OX200 User Manual
1. Read the manual carefully before use.
2. The fingertip sensor is sensitive to motion. Keep your hand still while
in use.
3. The fingertip sensor requires a sufficient blood flow to obtain proper
readings. Poor blood circulation can result in inaccurate readings.
If your hands are cold or you have poor circulation, warm up your
hands before attempting to obtain a reading.
4. A tourniquet, blood pressure cuff or other blood flow hindrances may
also result in inaccurate readings.
5. Nail polish and/or acrylic nails may also affect the readings. We
reccommend turning your finger on it's side when in use.
6. Your finger must be clean for a proper reading.
7. The fingertip sensor must be clean for a proper reading.
8. If a reading is difficult to obtain, switch to another finger or to the
other hand.
9. There are a number of other conditions which may lead to an
inaccurate reading including but not limited to recent medical tests
that included an injection of dyes, use of arterial catheters, a weak
pulse, low levels of hemoglobin in the blood, low perfusion (the
quality of your pulse), elevated levels of dysfunctional hemoglobin,
the strength and type of light that you are in while using the pulse
oximeter and the existence of transmitters within certain ranges of
the sensor during use.
10.The sensor will not alert you if your readings are out of normal range.
There is no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
11.Explosion hazard: Do not use the sensor in an explosive atmosphere.
1. Clean the device with a soft cloth dampened with isopropyl
alcohol. Do not pour or spray any liquids onto the sensor. Do
not allow any liquid to enter any openings in the device. Allow
the sensor to dry thoroughly before use.
2. The rubber inside of the sensor is composed of medical grade
rubber. It is Latex Free, non-toxic, and not harmful to the skin.
3. The sensor requires no routine calibration or maintenance.
4. Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
1. Using your nail, easily open the small rubber cover to expose
the micro USB port. Insert the micro USB end (See A) into the
oximeter's charging port
2. Place the USB end (See B) into any
USB power adapter (not included)
or a USB port on your laptop or computer.
3. When charging, the battery icon
on the display will blink. It is fully charged
when the icon stops blinking.
4. The battery can support 2 hours of continuous
use or approximately 400 measurements.
Note: Make sure the power adapter’s output
voltage is DC5V and output current is ≤1000mAh
Warnings: 1. Keep the oximeter away from young children - small items
such as the USB cable are a choking hazard. 2. Please do not take
a measurement when the device is charging.
1. Open the clamp as shown in the picture above. Using your
thumb and pointer, gently squeeze the back end of the
unit. Place one of your fingers into the opening. Your finger
should be inserted fully.
2. The unit will automatically turn on when the unit is opened
and a finger is inserted. It will automatically begin taking a
reading.
Remember to keep your hands still for the reading. Do not
shake your finger during the test. It is recommended that
you do not move your body while taking a reading.
3. The display will automatically show your pulse rate and
oxygen level readings after a few seconds. Note: Your
pulse rate will not stabilize. It will constantly change as
your pulse rate is based on your heart beats.
4. When there is no signal, a low signal is detected, or a
finger is removed, the unit will power off automatically.
1. Search for the ichoiceRelax app in the App store or Google Play
Store. Download the App.
2. Launch the app and fill out the proper fields (You will need a
valid email address).
3. The app will automatically look for a compatible heart rate
sensor within range. You must place the heart rate sensor on
your finger to turn it on. Once connected, the app will load.*
*You cannot access the app without a heart rate sensor. If there
is no signal from the sensor and/or you remove the sensor from
your finger, the app will lock and begin seaching again for a unit
within range.
ichoiceRelax App Features
COACH – Exercises to help guide your breathing with on-screen
pacers driven by your live heart rate readings.
RELAX – Various calming videos with relaxing sounds. Follow the
on-screen pacer driven by your live heart rate readings or close
your eyes and allow the sounds to guide you.
SLEEP - Choose between calming videos with relaxing sounds to
help you fall asleep. Follow the on-screen pacers driven by your
live heart rate readings or close your eyes and allow the sounds to
guide you.
MONITOR - Optional meters help you take readings via the
fingertip sensor and compare the results in charts and graphs.
Filter results by last 7 days, week, month, and year.
3 Minute Baseline Meter - Shows your current relaxation state
and generates an inner balance trend based on your calm vs.
stress readings.
30 Second Spot Check - Easily obtain your heart rate and
oxygen level at any given time.
The ichoiceRelax App uses continuous and accurate heart rate
readings to guide you into a state of relaxation to help reduce your
stress, calm you down and improve health and focus.
The fingertip sensor sends continuous heart rate readings to the
app to help coach and guide you to breathe at your ideal relaxation
breathing rate.
Everyone’s physiology is different and each person will have their
own ideal relaxation breathing rate based on their current mindset.
Your live heart rate readings (obtained by the heart rate sensor) will
help personalize exercises and guide you to breathe at the rate that
is ideal for you to achieve maximum calmness. Our app synchronizes
with your physiology giving you a personalized result.
1. Display: LED display
Low power indication: Battery icon will blink
2. Power Requirement: Lithium battery (250mAh) for power supply
Battery Life: Up to 2 hours (250mAh, 3.7V lithium battery)
Input Voltage: AC100V-240V, Input frequency: 50Hz-60Hz
Output Voltage: DC5V, Output current: 1000mA (Max)
3. Resolution: ±1% for SpO2and ±1 BPM for Pulse Rate
4. Measurement Range: Measuring range for SpO2: 70-100%
Measuring range for PR: 30-250 BPM
5. Measurement Accuracy:
SpO2: 70-100%, ±2 digits; <70%, no definition
PR: 30~99 BPM, ±2 BPM; 100~250 BPM, ±2%
6. Environment Requirement:
Operation Temperature: 41°F~104°F (5°C~40°C)
Storage Temperature: -13°F~158°F (-25°C~70°C )
Ambient Humidity: 15%~93% no condensation in operation;
≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Automatic Power Off: The fingertip sensor will power off automatically
after 8 seconds when the finger is removed.
8. It is recommended that the product be kept in a dry place. A damp
ambient might affect its lifetime and even might damage the product.
9. Declaration: EMC of this product complies with IEC60601-1-2 standard.
The materials which users can come into contact with are non-toxic
and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Inaccurate measurements may be caused by FACTORS INCLUDING BUT
NOT LIMITED TO:
12.Ethylene oxide sterilizing or immersing the sensors in liquid.
13.Significant levels of dysfunctional hemoglobin (such ascarbonxy-
hemoglobin or methemoglobin).
14.Intravascular dyes such as indocyanine green or methylene blue.
15.SpO2measurements may be adversely affected in the presence of
high ambient light such as direct sunlight. In bright light conditions,
cover the sensor area if necessary.
16.Excessive user movement.
17.High-frequency electrosurgical interference and defibrillators.
18.Venous pulsations.
19.Hypotension, severe vasoconstriction, severe anemia or hypothermia. My display shows
an error message.
My sensor will not
connect to the app
The app reads
"heart rate error"
I cannot navigate
the app
The sensor will not
give a reading.
The readings are
not accurate.
The sensor will not
power on.
Issue Possible Reason
Compliance
Test
Solution
Electromagnetic Environment - Guidance
1. You're wearing fingernail polish and/or
arcylic nails
2. Your hands are too cold / there is not
enough blood flowing to the fingertip
3. The battery on your sensor is low
4. Too much movement
1. The battery are low
2. The pulse oximeter is damaged
1. There is an internal error with your unit
1. You are not in the correct app
2. Your fingertip sensor is off
3. Your finger is not inside the sensor
4. The sensor is having trouble getting a
reading
5. Your sensor is not within Bluetooth range
6. Your phone's bluetooth is off
1. Contact customer service
1. Make sure you downloaded
ichoiceRelax app.
2. Make sure your finger is inside the
sensor and the unit is powered on.
3. Rub your hands together to generate
heat and get blood flowing to your
fingertips.
4. Make sure the unit and phone are
close proximity
5. Turn your phone's bluetooth on
1. Charge the unit
2. Contact customer service
04222019_v5
Guidance and manufacturer’s declaration – electromagnetic emission
RF emissions
CISPR 11 Group 1
Class B
RF emission
CISPR 11
The pulse oximeter uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The pulse oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the pulse oximeter should assure that it is used in such an
environment.
Distributed by:
ChoiceMMed America Corp.
2558 Pearl Buck Rd: Suite 8A
Bristol, PA 19007
Phone: (215) 874-0458
www.ichoicerelax.com
ChoiceMMed America Corp.
© 2019 ALL RIGHTS RESERVED
1. Keep the fingertip sensor away from young children. Small parts
may be hazardous if swallowed.
Thank you for purchasing the iChoice SMART pulse oximeter +
relaxation coach. With proper use, this system will teach you how to
breathe at your Personal Best Relaxation Breathing Rate to help reduce
stress, calm you down, and improve your health and focus. The fingertip
sensor transmits vital readings in real-time to the free ichoiceRelax app.
Using these live readings, the app will personalize breathing exercises to
help you achieve optimum relaxation.
The smart fingertip sensor can also be used as a pulse oximeter to spot
check your blood oxygen saturation (in % SpO2) and your pulse rate.
The SMART pulse oximeter + relaxation coach is for general wellness
and health applications including sports, fitness, and relaxation
management. It is not intended to be used in prevention, diagnosis, or
treatment of any disease.
Box Contents
1. SMART pulse oximeter + relaxation coach
2. User Manual
3. Carrying Pouch
4. Micro USB Charger
A
B
ichoiceRelax App
Phone Compatibility
iOS 12.0 + and Android 5.0+
1. Turn your finger on it's side
2. Rub your hands together to generate
heat and get blood flowing to your
fingertips
3. Charge your unit
4. Make sure your hand is resting and
there is little to no movement
Symbol Symbol
Definition Definition
SpO2%
PR bpm
SN
IP22
Oxygen Level
Pulse Rate (beats per minute)
Serial Number
Protected against dripping
water Follow instructions
for use
Conformity to WEEE
Directive
Type BF applied part
Storage temperature and
relative humidity
Battery Indicator
GENERAL DESCRIPTION
PRECAUTIONS FOR USE
MAINTENANCE
TROUBLE SHOOTINGCONNECTING TO APP (optional)PULSE OXIMETER DISPLAY
GENERAL USE INSTRUCTIONS
CHARGING THE UNIT
TECHNICAL SPECIFICATIONS
WARNINGS
SYMBOL DEFINITIONS
A normal SpO2reading typically ranges from 95%-99%.
A normal at rest Pulse Rate reading ranges between 60-100
beats per minute.
If you have questions regarding your SpO2or Pulse Rate readings
we recommend speaking with your healthcare professional.
Note:
1. The pulse bar indicates the adequacy of Pulse Rate and SpO2.
If the bar shows less than 30%, readings may be inaccurate.
2. If the screen displays the signal strength symbol (?), the unit
is having trouble getting a reading. Keep your hands still and
retry.

OX200 QUICK START GUIDE
1. Place your finger inside the unit
2. Readings appear automatically
after a few seconds
3. Connect to App (Optional)
To obtain full warranty coverage, please be sure to complete your
product registration within 30 days from date of purchase.
If you choose not to register your product, this Limited Warranty
will be void.
Register via Website
You can register your device online at www.ichoicerelax.com/register
Register via App
When you download the ichoiceRelax app for the first time, you will
be prompted to enter in your personal information. The information
entered here will automatically save to our database and constitute
as your product registry.
Call Customer Service
Monday - Friday
10am - 4:00pm EST
Phone: (215) 874-0458
STATEMENT OF CHOICEMMED AMERICA CORP. / LIMITED WARRANTY
ChoiceMMed America Corp. warrants to the original purchaser
that this equipment will be free from defects in materials and
workmanship for a period of two years from the date of purchase.
This warranty shall be limited to the repair adjustment and/or
replacement at the manufacturer's option of defective parts free of
charge except the cost of transportation to ChoiceMMed America
Corp. Manufacturer's responsibility is limited to actual cost of
item. When returning equipment for warranty service, all shipping
and insurance charges must be prepaid and the proof of purchase
must be enclosed. Returns will only be shipped to locations within
the USA. ChoiceMMed America Corp. will not be responsible for
expenses or inconveniences, or consequential damages occasioned
by equipment or by breach of any expressed or implied warranty
with respect thereto.
Limitations
This warranty covers all defects encountered in normal use of the
equipment and does not apply in the following cases.
1. If the equipment has been serviced by other than a
certified center.
2. Damage to the equipment due to mishandling, abuse, accident
or not following operating instructions.
3. INSTRUMENTS: Warranty does not extend to display face,
batteries, carrying case or charging cable.
For troubleshooting assistance please contact Consumer Support.
Our representatives are trained to provide you assistance over the
phone and may be able to resolve your problem without returning.
For the best possible assistance please have your unit available
when calling us.
www.ichoicerelax.com/help
ChoiceMMed America Corp.
2558 Pearl Buck Rd, Suite 8A, Bristol
PA 19007
1. Search for the ichoiceRelax app
in the App store or Google Play
Store. Download the App to your phone.
ichoiceRelax is compatible with iOs 12.0 and
Android 5.0+
2. Launch the app and fill out the proper fields
(You will need a valid email address).
3. The app will automatically look for a
compatible heart rate sensor within range.
You must place the heart rate sensor on your
finger to turn it on. Once connected, the app
will load.*
*You cannot access the app without a heart rate
sensor. If there is no signal from the sensor
and/or you remove the sensor from your
finger, the app will lock and seach for a unit
within range again.
What is a Pulse Oximeter?
A pulse oximeter is a non-invasive device that
monitors blood oxygen saturation (SpO2) and pulse
rate (heart rate). Pulse oximeters provide an easy
way of assessing your blood oxygen level and
pulse rate.
What is SpO2?
SpO2is also known as oxygen saturation. Oxygen
saturation is a measure of how much oxygen the
blood is carrying as a percentage of the maximum
it could carry.
What is a normal reading?
Pulse Rate - The normal resting range for pulse
rate is typically considered from 60-100 beats per
minute. Ask your health professional this question
as it pertains to you.
Oxygen Saturation - The normal range for SpO2
is typically considered from 95%-99%. The SpO2
measurement may be lower for people who live
at high altitudes. Ask your health professional this
question as it pertains to you.
What kind of conditions may cause an
inaccurate reading?
Cold hands, poor circulation, weak pulse,
movement, fingernail polish and acrylic nails may
cause inaccurate results.
The SpO2is not changing – is it stuck?
Your SpO2reading will not change rapidly like the
pulse rate reading. It may vary by 1%-2%.
The pulse rate is changing rapidly – is it broken?
Your heart rate changes with emotions, excitement
and exercise. It is normal for the readings to
fluctuate.
When do I charge the battery?
The battery indicator on the display will blink when
charging is needed. A stable battery icon indicates
the unit is fully charged.
Warranty Information
Heart Rate
(PR)
Bluetooth
Indicator
Signal
Strength
Battery
Indicator
Pulse Bar
Oxygen Level
(SpO2)
This device omplies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including
interference that may cause undesired operation.
Attention that changes or modification not expressly approved
by the party responsible for compliance could void the user’s
authority to operate the equipment.
Note: This product has been tested and found to comply
with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to prodvide
reasonal protection against harmful 41 interference in a
residential installation. This product generates, uses, and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is
no guarantee that interference will not occur in a particular
installation. If this product does cause harmful interference to
radio communications or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of
the following measures:
a) Reorient or relocate the receiving antenna.
b) Increase the separation between the equipment and receiver.
c) Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
d) Consult the dealer or an experienced radio/TV technician for
help.
FCC DECLARATION

OX200 QUICK START GUIDE
1. Place your finger inside the unit
2. Readings appear automatically
after a few seconds
3. Connect to App (Optional)
To obtain full warranty coverage, please be sure to complete your
product registration within 30 days from date of purchase.
If you choose not to register your product, this Limited Warranty
will be void.
Register via Website
You can register your device online at www.ichoicerelax.com/register
Register via App
When you download the ichoiceRelax app for the first time, you will
be prompted to enter in your personal information. The information
entered here will automatically save to our database and constitute
as your product registry.
Call Customer Service
Monday - Friday
10am - 4:00pm EST
Phone: (215) 874-0458
STATEMENT OF CHOICEMMED AMERICA CORP. / LIMITED WARRANTY
ChoiceMMed America Corp. warrants to the original purchaser
that this equipment will be free from defects in materials and
workmanship for a period of two years from the date of purchase.
This warranty shall be limited to the repair adjustment and/or
replacement at the manufacturer's option of defective parts free of
charge except the cost of transportation to ChoiceMMed America
Corp. Manufacturer's responsibility is limited to actual cost of
item. When returning equipment for warranty service, all shipping
and insurance charges must be prepaid and the proof of purchase
must be enclosed. Returns will only be shipped to locations within
the USA. ChoiceMMed America Corp. will not be responsible for
expenses or inconveniences, or consequential damages occasioned
by equipment or by breach of any expressed or implied warranty
with respect thereto.
Limitations
This warranty covers all defects encountered in normal use of the
equipment and does not apply in the following cases.
1. If the equipment has been serviced by other than a
certified center.
2. Damage to the equipment due to mishandling, abuse, accident
or not following operating instructions.
3. INSTRUMENTS: Warranty does not extend to display face,
batteries, carrying case or charging cable.
For troubleshooting assistance please contact Consumer Support.
Our representatives are trained to provide you assistance over the
phone and may be able to resolve your problem without returning.
For the best possible assistance please have your unit available
when calling us.
www.ichoicerelax.com/help
ChoiceMMed America Corp.
2558 Pearl Buck Rd, Suite 8A, Bristol
PA 19007
1. Search for the ichoiceRelax app
in the App store or Google Play
Store. Download the App to your phone.
ichoiceRelax is compatible with iOs 12.0 and
Android 5.0+
2. Launch the app and fill out the proper fields
(You will need a valid email address).
3. The app will automatically look for a
compatible heart rate sensor within range.
You must place the heart rate sensor on your
finger to turn it on. Once connected, the app
will load.*
*You cannot access the app without a heart rate
sensor. If there is no signal from the sensor
and/or you remove the sensor from your
finger, the app will lock and seach for a unit
within range again.
What is a Pulse Oximeter?
A pulse oximeter is a non-invasive device that
monitors blood oxygen saturation (SpO2) and pulse
rate (heart rate). Pulse oximeters provide an easy
way of assessing your blood oxygen level and
pulse rate.
What is SpO2?
SpO2is also known as oxygen saturation. Oxygen
saturation is a measure of how much oxygen the
blood is carrying as a percentage of the maximum
it could carry.
What is a normal reading?
Pulse Rate - The normal resting range for pulse
rate is typically considered from 60-100 beats per
minute. Ask your health professional this question
as it pertains to you.
Oxygen Saturation - The normal range for SpO2
is typically considered from 95%-99%. The SpO2
measurement may be lower for people who live
at high altitudes. Ask your health professional this
question as it pertains to you.
What kind of conditions may cause an
inaccurate reading?
Cold hands, poor circulation, weak pulse,
movement, fingernail polish and acrylic nails may
cause inaccurate results.
The SpO2is not changing – is it stuck?
Your SpO2reading will not change rapidly like the
pulse rate reading. It may vary by 1%-2%.
The pulse rate is changing rapidly – is it broken?
Your heart rate changes with emotions, excitement
and exercise. It is normal for the readings to
fluctuate.
When do I charge the battery?
The battery indicator on the display will blink when
charging is needed. A stable battery icon indicates
the unit is fully charged.
Warranty Information
Heart Rate
(PR)
Bluetooth
Indicator
Signal
Strength
Battery
Indicator
Pulse Bar
Oxygen Level
(SpO2)
This device omplies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including
interference that may cause undesired operation.
Attention that changes or modification not expressly approved
by the party responsible for compliance could void the user’s
authority to operate the equipment.
Note: This product has been tested and found to comply
with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to prodvide
reasonal protection against harmful 41 interference in a
residential installation. This product generates, uses, and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is
no guarantee that interference will not occur in a particular
installation. If this product does cause harmful interference to
radio communications or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of
the following measures:
a) Reorient or relocate the receiving antenna.
b) Increase the separation between the equipment and receiver.
c) Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
d) Consult the dealer or an experienced radio/TV technician for
help.
FCC DECLARATION

OX200 QUICK START GUIDE
1. Place your finger inside the unit
2. Readings appear automatically
after a few seconds
3. Connect to App (Optional)
To obtain full warranty coverage, please be sure to complete your
product registration within 30 days from date of purchase.
If you choose not to register your product, this Limited Warranty
will be void.
Register via Website
You can register your device online at www.ichoicerelax.com/register
Register via App
When you download the ichoiceRelax app for the first time, you will
be prompted to enter in your personal information. The information
entered here will automatically save to our database and constitute
as your product registry.
Call Customer Service
Monday - Friday
10am - 4:00pm EST
Phone: (215) 874-0458
STATEMENT OF CHOICEMMED AMERICA CORP. / LIMITED WARRANTY
ChoiceMMed America Corp. warrants to the original purchaser
that this equipment will be free from defects in materials and
workmanship for a period of two years from the date of purchase.
This warranty shall be limited to the repair adjustment and/or
replacement at the manufacturer's option of defective parts free of
charge except the cost of transportation to ChoiceMMed America
Corp. Manufacturer's responsibility is limited to actual cost of
item. When returning equipment for warranty service, all shipping
and insurance charges must be prepaid and the proof of purchase
must be enclosed. Returns will only be shipped to locations within
the USA. ChoiceMMed America Corp. will not be responsible for
expenses or inconveniences, or consequential damages occasioned
by equipment or by breach of any expressed or implied warranty
with respect thereto.
Limitations
This warranty covers all defects encountered in normal use of the
equipment and does not apply in the following cases.
1. If the equipment has been serviced by other than a
certified center.
2. Damage to the equipment due to mishandling, abuse, accident
or not following operating instructions.
3. INSTRUMENTS: Warranty does not extend to display face,
batteries, carrying case or charging cable.
For troubleshooting assistance please contact Consumer Support.
Our representatives are trained to provide you assistance over the
phone and may be able to resolve your problem without returning.
For the best possible assistance please have your unit available
when calling us.
www.ichoicerelax.com/help
ChoiceMMed America Corp.
2558 Pearl Buck Rd, Suite 8A, Bristol
PA 19007
1. Search for the ichoiceRelax app
in the App store or Google Play
Store. Download the App to your phone.
ichoiceRelax is compatible with iOs 12.0 and
Android 5.0+
2. Launch the app and fill out the proper fields
(You will need a valid email address).
3. The app will automatically look for a
compatible heart rate sensor within range.
You must place the heart rate sensor on your
finger to turn it on. Once connected, the app
will load.*
*You cannot access the app without a heart rate
sensor. If there is no signal from the sensor
and/or you remove the sensor from your
finger, the app will lock and seach for a unit
within range again.
What is a Pulse Oximeter?
A pulse oximeter is a non-invasive device that
monitors blood oxygen saturation (SpO2) and pulse
rate (heart rate). Pulse oximeters provide an easy
way of assessing your blood oxygen level and
pulse rate.
What is SpO2?
SpO2is also known as oxygen saturation. Oxygen
saturation is a measure of how much oxygen the
blood is carrying as a percentage of the maximum
it could carry.
What is a normal reading?
Pulse Rate - The normal resting range for pulse
rate is typically considered from 60-100 beats per
minute. Ask your health professional this question
as it pertains to you.
Oxygen Saturation - The normal range for SpO2
is typically considered from 95%-99%. The SpO2
measurement may be lower for people who live
at high altitudes. Ask your health professional this
question as it pertains to you.
What kind of conditions may cause an
inaccurate reading?
Cold hands, poor circulation, weak pulse,
movement, fingernail polish and acrylic nails may
cause inaccurate results.
The SpO2is not changing – is it stuck?
Your SpO2reading will not change rapidly like the
pulse rate reading. It may vary by 1%-2%.
The pulse rate is changing rapidly – is it broken?
Your heart rate changes with emotions, excitement
and exercise. It is normal for the readings to
fluctuate.
When do I charge the battery?
The battery indicator on the display will blink when
charging is needed. A stable battery icon indicates
the unit is fully charged.
Warranty Information
Heart Rate
(PR)
Bluetooth
Indicator
Signal
Strength
Battery
Indicator
Pulse Bar
Oxygen Level
(SpO2)
This device omplies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including
interference that may cause undesired operation.
Attention that changes or modification not expressly approved
by the party responsible for compliance could void the user’s
authority to operate the equipment.
Note: This product has been tested and found to comply
with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to prodvide
reasonal protection against harmful 41 interference in a
residential installation. This product generates, uses, and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is
no guarantee that interference will not occur in a particular
installation. If this product does cause harmful interference to
radio communications or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of
the following measures:
a) Reorient or relocate the receiving antenna.
b) Increase the separation between the equipment and receiver.
c) Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
d) Consult the dealer or an experienced radio/TV technician for
help.
FCC DECLARATION

OX200 QUICK START GUIDE
1. Place your finger inside the unit
2. Readings appear automatically
after a few seconds
3. Connect to App (Optional)
To obtain full warranty coverage, please be sure to complete your
product registration within 30 days from date of purchase.
If you choose not to register your product, this Limited Warranty
will be void.
Register via Website
You can register your device online at www.ichoicerelax.com/register
Register via App
When you download the ichoiceRelax app for the first time, you will
be prompted to enter in your personal information. The information
entered here will automatically save to our database and constitute
as your product registry.
Call Customer Service
Monday - Friday
10am - 4:00pm EST
Phone: (215) 874-0458
STATEMENT OF CHOICEMMED AMERICA CORP. / LIMITED WARRANTY
ChoiceMMed America Corp. warrants to the original purchaser
that this equipment will be free from defects in materials and
workmanship for a period of two years from the date of purchase.
This warranty shall be limited to the repair adjustment and/or
replacement at the manufacturer's option of defective parts free of
charge except the cost of transportation to ChoiceMMed America
Corp. Manufacturer's responsibility is limited to actual cost of
item. When returning equipment for warranty service, all shipping
and insurance charges must be prepaid and the proof of purchase
must be enclosed. Returns will only be shipped to locations within
the USA. ChoiceMMed America Corp. will not be responsible for
expenses or inconveniences, or consequential damages occasioned
by equipment or by breach of any expressed or implied warranty
with respect thereto.
Limitations
This warranty covers all defects encountered in normal use of the
equipment and does not apply in the following cases.
1. If the equipment has been serviced by other than a
certified center.
2. Damage to the equipment due to mishandling, abuse, accident
or not following operating instructions.
3. INSTRUMENTS: Warranty does not extend to display face,
batteries, carrying case or charging cable.
For troubleshooting assistance please contact Consumer Support.
Our representatives are trained to provide you assistance over the
phone and may be able to resolve your problem without returning.
For the best possible assistance please have your unit available
when calling us.
www.ichoicerelax.com/help
ChoiceMMed America Corp.
2558 Pearl Buck Rd, Suite 8A, Bristol
PA 19007
1. Search for the ichoiceRelax app
in the App store or Google Play
Store. Download the App to your phone.
ichoiceRelax is compatible with iOs 12.0 and
Android 5.0+
2. Launch the app and fill out the proper fields
(You will need a valid email address).
3. The app will automatically look for a
compatible heart rate sensor within range.
You must place the heart rate sensor on your
finger to turn it on. Once connected, the app
will load.*
*You cannot access the app without a heart rate
sensor. If there is no signal from the sensor
and/or you remove the sensor from your
finger, the app will lock and seach for a unit
within range again.
What is a Pulse Oximeter?
A pulse oximeter is a non-invasive device that
monitors blood oxygen saturation (SpO2) and pulse
rate (heart rate). Pulse oximeters provide an easy
way of assessing your blood oxygen level and
pulse rate.
What is SpO2?
SpO2is also known as oxygen saturation. Oxygen
saturation is a measure of how much oxygen the
blood is carrying as a percentage of the maximum
it could carry.
What is a normal reading?
Pulse Rate - The normal resting range for pulse
rate is typically considered from 60-100 beats per
minute. Ask your health professional this question
as it pertains to you.
Oxygen Saturation - The normal range for SpO2
is typically considered from 95%-99%. The SpO2
measurement may be lower for people who live
at high altitudes. Ask your health professional this
question as it pertains to you.
What kind of conditions may cause an
inaccurate reading?
Cold hands, poor circulation, weak pulse,
movement, fingernail polish and acrylic nails may
cause inaccurate results.
The SpO2is not changing – is it stuck?
Your SpO2reading will not change rapidly like the
pulse rate reading. It may vary by 1%-2%.
The pulse rate is changing rapidly – is it broken?
Your heart rate changes with emotions, excitement
and exercise. It is normal for the readings to
fluctuate.
When do I charge the battery?
The battery indicator on the display will blink when
charging is needed. A stable battery icon indicates
the unit is fully charged.
Warranty Information
Heart Rate
(PR)
Bluetooth
Indicator
Signal
Strength
Battery
Indicator
Pulse Bar
Oxygen Level
(SpO2)
This device omplies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including
interference that may cause undesired operation.
Attention that changes or modification not expressly approved
by the party responsible for compliance could void the user’s
authority to operate the equipment.
Note: This product has been tested and found to comply
with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to prodvide
reasonal protection against harmful 41 interference in a
residential installation. This product generates, uses, and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is
no guarantee that interference will not occur in a particular
installation. If this product does cause harmful interference to
radio communications or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of
the following measures:
a) Reorient or relocate the receiving antenna.
b) Increase the separation between the equipment and receiver.
c) Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
d) Consult the dealer or an experienced radio/TV technician for
help.
FCC DECLARATION
Other manuals for OX200
1
Table of contents
Other IChoice Medical Equipment manuals