manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Idaho
  6. •
  7. Laboratory Equipment
  8. •
  9. Idaho LightScanner Instruction Manual

Idaho LightScanner Instruction Manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Connecting the LightScanner
1. Make sure the power is OFF on the LightScanner.
2. Connect the cables to the instrument (all on the back panel) in the following order:
a. Power cord
b. Camera cable (snap on, make sure the connection is secure)
c. Serial cable
3. Next plug in the power cord to the wall power source.
4. Connect the camera cable to the computer tower (connection on the
bottom of the tower).
5. Connect the serial cable to the computer.
6. Connect the remainder of the computer cables (monitor, mouse,
power etc).
7. Turn on the computer.
8. Turn on the power switch on the back of the LightScanner instrument. You will hear two beeps, the sound of
a fan and the RED power light on the front of the instrument will turn on.
9. You are now ready to run the LightScanner.
Camera Check
A quick check to confirm that the camera is communicating with the computer should be performed.
Open the camera software found under:
Programs >> Roper Scientific >> RS config
1. A table will open displaying a NAME in the Camera 1 box, usually “Camera 1” confirming that the camera is
communicating with the computer.
2. If there is not a name displayed and the boxes are gray, it indicates that the computer is not communicating
with the camera. If this is the case, please contact Idaho Technology technical support.
3. Open the following program:
Programs >> Roper Scientific >> Pvcam test
4. Select Collect Image. This will generate an image of the heat block confirming that the camera is working
correctly.
5. Once you see the image click Done.
LightScanner® Start-Up and
Installation Instructions for Users
Idaho Technology Inc. | www.idahotech.com
LightScanner® Start-Up and Installation Instructions for Users
Power
Switch
Serial
Port
Camera
Port
LSCN-PRT-0021 Rev 01
Calibration
After an instrument is installed, a plate position calibration will need to be performed. This calibration is performed
in the Run window and requires a full plate of samples. For 96-well instruments, this can be done with the plate in
the Start-up Kit.
1. Double-click on the LightScanner icon on the desktop.
2. Choose the Run tab.
3. Insert the prepared plate into the instrument, notched edge first. The instrument should beep when the plate
is loaded and the door closed.
4. Press the Calibrate Map button.
5. An image of a plate with
cross-hairs will appear. The
cross-hairs can be moved
using the mouse.
Note: Instructions for calibra-
tion are also at the top of the
screen.
a. First move the cross hairs
over the center of the well
at the top left corner (A1)
making sure the cross is
in the middle of the well
(sometimes this is not the
brightest part). Right-click
and click Enter.
b. Next, repeat with the top
right well (A12)
c. Lastly, repeat for the
bottom left well (H1)
d. The software should
produce an image of the
plate where every well is
ringed by a pink line.
e. Confirm that the positions
of the wells are not
skewed.
6. If you are not happy with
the results of the calibration,
repeat with a different plate.
Sometimes adhesive material
from the film may produce
pink streaks on the lip of a well, if this happens repeat the calibration using a different plate.
7. Calibration should be repeated if the instrument is moved to a new location.
8. Weekly or monthly re-calibrations are not necessary. It is only to be done the first time the instrument is set-
up or after it has been moved.
Performing a Run
To perform the first run follow the instructions in the LightScanner Operator’s Manual.
Remember, if using bar-coded plates the bar code must be read prior to starting the run.
Customer and Technical Support
Reach Us On the Web
Idaho Technology’s Web site is:
http://www.idahotech.com
We strongly encourage users to visit our Web site
for answers to frequently asked questions, updated
information and additional insights into operating
the LightScanner Instrument.
Reach Us By E-mail
[email protected] - Customer Support
[email protected] - General Support
Reach Us By Phone
Technical support is available during:
8 a.m. to 5 p.m. - Mountain Standard Time
For technical support call:
1-800-735-6544 - United States and Canada
(801) 736-6354 - Utah
+1-801-736-6354 - International*
*Note: The IDD prefix (represented by a “+”)
for the United States is the international prefix
number needed to place a call from the coun-
try listed to another country. This is followed
by the country code for the country you are
calling; The country code is “1”.
Idaho Technology Inc. | www.idahotech.com
LightScanner® Start-Up and Installation Instructions for Users
1 2
3
Cross hatch is
over the center of
the circle, and not
on the darkest part
of the well
Calibrate in the
following order
When finished,
the pink circles
need to be
centered around
each well. This
image shows a
good example of
a calibrated plate