Idex HALE SPV Manual

09/01/95
Model SPV
Semi-AutomaticPrimingValve
INSTALLATIONAND
MAINTENANCE MANUAL
MANUALP/N029-0020-33-0,REVA,©1995HALEPRODUCTS,INC.,PRINTEDINU.S.A.
All Hale products are quality components: ruggedly designed, accurately machined, precision inspected, carefully assembled
and thoroughly tested. In order to maintain the high quality of your unit, and to keep it in a ready condition, it is important to
follow the instructions on care and operation. Proper use and good preventive maintenance will lengthen the life of your unit.
ALWAYS INCLUDE THE UNIT SERIAL NUMBER IN CORRESPONDENCE.
Hale Products cannot assume responsibility for product failure resulting from improper maintenance
or operation. Hale Products is responsible only to the limits stated in the product warranty. Product
specifications contained in this material are subject to change without notice.
HALE PRODUCTS, INC. A Unit of IDEX Corporation 700 Spring Mill Avenue Conshohocken, PA 19428 TEL: 610-825-6300 FAX: 610-825-6440


Semi-Automatic Priming Valve
i
NOTE TO SYSTEM INSTALLER
IMPORTANT: Please provide two copies of the Hale SPV manual to the end user of
the equipment. For additional manuals, contact Hale Products, Inc at (610) 825-
6300. Ask for Manual P/N 029-0020-33-0.
TABLE OF CONTENTS
SECTION PAGE
1 SAFETY ............................................................................................................1
2 DESCRIPTION..................................................................................................2
3 INSTALLATION .................................................................................................4
4 MAINTENANCE ............................................................................................ 10
5 PARTSLIST .....................................................................................................12
6 WARRANTY ................................................................................................... 14


Semi-Automatic Priming Valve
1
WARNING: Parts under spring tension can
become projectiles and cause serious injury.
When removing or installing these parts make
sure they are restrained and spring tension is
released slowly.
WARNING: Disconnect or turn-off master
battery switch prior to service of the Hale SPV
electrical components. Make sure there is no
power at the primer solenoid before starting
service procedures.
WARNING: Any electrical system has the
potential to cause sparks during installation,
service or repair. Take care to eliminate
explosive or hazardous environments during
installation, service or repair.
WARNING: Check all hoses for weak or worn
conditions after each use. Ensure that all
connections and fittings are tight and secure.
WARNING: Use only pipe, hose, and fittings
from the priming pump vacuum connection to
the Hale SPV vacuum connection rated for 29 in
(737 mm) Hg vacuum.
CAUTION: Make sure the Hale SPV is located
higher than the pump body allowing water to
drain from the vacuum hose.
SAFETY
IMPORTANT NOTICE: Before attempting installation and maintenance of the Hale SPV read and
follow all safety precautions listed below. The warnings and cautions listed are necessary for the
safe installation or maintenance of the Hale SPV. When developing departmental apparatus
maintenance procedures make sure the warnings and cautions are incorporated as written.

Semi-Automatic Priming Valve
2
Figure 1. Hale SPV Valve and Push-Button Switch Assembly
DESCRIPTION
(19 mm) ID vacuum hose connects the Hale
SPV to the priming pump, eliminating potential
leak points in the priming hose. The short
length of hose also reduces the volume that the
priming pump must evacuate allowing faster
priming time and longer primer life. The priming
pump motor operates by a sealed momentary
contact push-button located on the pump
operator panel of the apparatus. There are NO
hose connections at the pump operator panel
thus saving valuable space.
Depressing and holding the PRIME push-button
will energize the priming pump motor creating a
vacuum in the 3/4inch (19 mm) ID hose. As
sufficient vacuum is created the diaphragm will
depress the spring allowing the priming valve to
open. Once the Hale SPV opens, the priming
pump creates a vacuum in the main pump body,
priming the main pump. Once primed, as
indicated by water discharging to the ground
from the priming pump outlet, the push-button is
released. Once push-button is released, the
priming pump stops running allowing the Hale
SPV to close for normal pump operation.
The Hale Semi-Automatic Priming Valve (Hale
SPV) (shown in figure 1) is a diaphragm
operated type valve that opens using the
vacuum generated by the priming pump. The
Hale SPV mounts directly to the priming
connection of a Hale midship pump.
Additionally, a universal adapter that bolts to an
installer supplied bracket is available to permit
use of the Hale SPV on booster and volute type
pumps. The installer must also supply hose
connections from the adapter to the pump
priming connection. The Hale SPV operates
from the vacuum generated by the priming
pump when the momentary contact push-button
located on the pump operator panel is
depressed. The Hale SPV conforms to NFPA
requirements for priming time and hydrostatic
test when bolted to Hale midship type pumps.
Each valve is factory tested to 24 in (610 mm)
hg vacuum and 600 PSIG (41 BAR).
When used on a midship fire pump, the valve
mounts to the pump body with two studs and
nuts, standard strainer and seal ring. Mounted
on the midship fire pump, in close proximity to
the priming pump, only a short length of 3/4inch

Semi-Automatic Priming Valve
3
The Hale SPV is available from the factory in four different configurations. A description of the
different configurations follows.
Hale SPV Retrofit kit: Kit contains the components necessary for installing the Hale SPV on a
midship fire pump with Hale PVG installed. The kit contains a Hale SPV Valve assembly, Panel
Placard, Push-button Switch and Installation Manual.
Hale SPV Universal Mount Installation Kit: Kit contains the components necessary for installing
the Hale SPV on a volute type or booster pump. The kit contains a Hale SPV Valve assembly,
Universal Mounting Adapter, Panel Placard, Push-button Switch and Installation Manual.
Hale SPV Mounted on Midship Pump: This available option on new midship fire pumps has the
Hale SPV mounted to the priming flange connection on the midship pump body. Components that
are included with this option and shipped with the midship pump are the Push-button Switch, Panel
Placard and Installation Manual. When completing the apparatus installation of the Hale SPV with
this option the priming pump must be mounted on the apparatus and vacuum hose connected in
addition to the push-button switch and panel placard being installed on the operator panel.
Hale SPV and Primer Mounted on Midship Pump: This available option on new midship fire
pumps has the Hale SPV mounted to the priming flange connection on the midship pump body and
the priming pump mounted to the pump gearbox. When this option is ordered the vacuum hose is
connected from the Hale SPV to the priming pump. Components that are included with this option
and shipped with the midship pump are the Push-button Switch, Panel Placard and Installation
Manual. When completing the apparatus installation of the Hale SPV with this option the push-
button switch and panel placard must be installed on the operator panel.

Semi-Automatic Priming Valve
4
INSTALLATION
Figure 2. Hale SPV General System Layout Diagram
When ordered as an option on a new midship fire pump the Hale SPV valve assembly is already
attached to the pump body with the vacuum hose connected to the priming pump when the primer
is ordered mounted to the gearbox. All that remains after the midship pump is mounted on the
apparatus frame is to install the panel placard and push-button on the pump operator panel and
make the electrical connections from the switch to the battery and priming pump solenoid. If the
midship pump was ordered with the primer shipped loose then the primer must be attached to the
apparatus and priming hose must be attached to the Hale SPV hose connection. Figure 2 shows
the relative location of components of the Hale SPV.

Semi-Automatic Priming Valve
5
Figure 3. Push-Button Switch Installation
FACE
NUT
PANEL
PLACARD LOCK
NUT
PUSH-BUTTON
SWITCH TO PRIMING
PUMP SOLENOID
CONNECTION
(USE MINIMUM 14
AWG WIRE)
TO + BATTERY
POWER
(USE MINIMUM 14
AWG WIRE)
Panel Placard and Push-Button Switch
Installation:
The following procedures shall be used to install
the panel placard and push-button switch.
WARNING: Disconnect or turn-off master
battery switch prior to service of the Hale
SPV electrical components. Make sure
there is no power at the primer solenoid
before starting service procedures.
WARNING: Any electrical system has the
potential to cause sparks during installation,
service or repair. Take care to eliminate
explosive or hazardous environments during
installation, service or repair.
1. Carefully locate the position of the push-
button switch on the operator panel.
2. Drill or punch a 0.64 inch (16 mm) diameter
hole in the panel.
3. Remove the push-button switch from the
packaging and install the locknut on the
mounting neck.
4. Refering to figure 3, Insert the push-button
switch through the 0.64 inch (16 mm)
diameter hole from the back of the operator
panel.
5. Place PRIME placard over exposed end of
push-button switch on outside of operator
panel.
6. Install face nut on push-button switch to hold
switch and placard in place. Tighten locknut
and face nut.
7. Using #10 ring terminals on the wire ends,
install minimum 14 AWG type SXL or GXL
wire (SAE J1128) from one switch terminal
to the priming pump solenoid connection.
Secure the ring terminal to the push-button
switch using the screws provided.
8. Using #10 ring terminals on the wire ends,
install minimum 14 AWG type SXL or GXL
wire (SAE J1128) from the other switch
terminal to the positive battery power.
Secure the ring terminal to the push-button
switch using the screws provided with the
switch.
9. Test operation of the Hale SPV and priming
pump. Conduct vacuum and hydrostatic
tests in accordance with department
procedures, NFPA1901 or NFPA 1911.

Semi-Automatic Priming Valve
6
Retrofit Installation:
When installing the Hale SPV as a retrofit on a
midship fire pump the following procedures shall
be followed. (Refer to figure 5)
1. Place the apparatus out of service in
accordance with departmental procedures.
Drain all water from pump body.
WARNING: Disconnect or turn-off master
battery switch prior to service of the Hale
SPV electrical components. Make sure
there is no power at the primer solenoid
before starting service procedures.
2. Open the pump operator panel and gain
access to the existing Hale PVG.
3. Disconnect the electrical leads and vacuum
hoses to the Hale PVG.
4. If the Hale PVG is mounted at the operator
panel, remove the Hale PVG and placard.
5. Install the panel placard and momentary
contact push-button provided with the Hale
SPV on the operator panel.
NOTE: A spacer is required where the Hale
PVG was removed to allow mounting of the
push-button switch. Manufacture a spacer
using the template in figure 4 as a guide.
The outer surface of the spacer must be flat
to allow proper installation of the panel
placard push-button switch. Use material
thickness and fastening devices that leave a
flat surface.
6. Connect electrical leads to the momentary
contact push-button.
7. Locate the priming valve flange on the
midship pump. (Refer to figure 5) The
priming valve flange is located next to the
gearbox on the midship pump.
8. Remove the 7/16-14 x 1-1/4inch long cap
screws that hold the priming flange, hose
and fittings on the midship pump body.
Remove the flange assembly from the pump
body.
9. Remove the strainer and seal ring from the
pump body. Clean the strainer of all debris.
DO NOT discard the strainer as it must be
used with the Hale SPV.
10. Apply a light coat of thread locking
compound (Loctite #290 or equal) to one
end of the 7/16-14 x 1-3/4inch long studs.
Insert studs into the pump body where the
cap screws were removed then tighten
studs.
11. Insert strainer and seal ring in pump body
then install Hale SPV on the pump body.
Figure 4. Panel Placard Spacer
0.64 IN. (16 mm)
THRU HOLE
2.32 IN.
(59 mm)
1.16 IN.
(29 mm)
0.25 IN. (6 mm)
RAD TYP
(2) THRU HOLES FOR
SPACER FASTENERS
2.44 IN. (62 mm)
1.22 IN. (31 mm)
1 IN. (25 mm)
NOTE: MATERIAL THICKNESS AND SPACER
FASTENERS MUST ALLOW FOR A FLAT SURFACE
WHEN THIS SPACER IS MOUNTED ON THE
OPERATOR PANEL. USE COUNTERSUNK
FLATHEAD SCREWS OR PEM TYPE FASTENERS

Semi-Automatic Priming Valve
7
Make sure the vacuum hose connection is
facing in the down direction when the Hale
SPV is installed.
12. Apply a light coating of thread locking
compound (Loctite #290 or equal) to the
exposed stud threads. Secure Hale SPV in
place using 7/16-14 nuts.
WARNING: Use only pipe, hose, and
fittings from the priming pump vacuum
connection to the Hale SPV vacuum
connection rated for 29 in (737 mm) Hg
vacuum.
13. Attach 3/4inch (19 mm) ID vacuum hose
from the hose connection on the Hale SPV
to the priming pump connection. Use hose
with 3/4inch (19 mm) inside diameter that is
rated for 29 in (737 mm) Hg vacuum
(Aeroquip 2556-12 or equal). Make sure the
hose is routed and properly secured in place
to prevent chaffing and abraision.
14. Test operation of the Hale SPV and priming
pump. Conduct vacuum and hydrostatic
tests in accordance with department
procedures, NFPA1901 or NFPA 1911.
Figure 5. Hale SPV Valve Installation on Midship Fire Pump
MIDSHIP PUMP
BODY (REF.)
HALE SPV
ASSEMBLY
7/16-14 NUT (2)
7/16-14 x 1-3/4INCH
LONG STUD (2)
PRIMING
STRAINER
SEAL
RING
DIAPHRAGM
COVER
DRAIN PRIMING HOSE CONNECTION
USE ¾ INCH (19 mm) INSIDE DIAMETER
HOSE

Semi-Automatic Priming Valve
8
Figure 6. Universal Mounting Adapter
Dimensions
3-7/8IN.
(98 mm)
2-1/4IN.
(57 mm)
3-3/4IN.
(95 mm)
2-1/2IN.
(64 mm)
Figure 7. Universal Mounting Adapter Mounting Positions
NOTE: STUDS MUST BE INSTALLED IN HOLES THAT LIE ON THE
HORIZONTAL CENTERLINE OF THE UNIVERSAL ADAPTER.
The Universal Mounting Adapter can be mounted to a secure point on the
apparatus with the attaching holes (flat portion) located in the down position or
as shown below.
end of the 7/16-14 x 1-3/4inch long studs.
Install the studs in horizontal holes on the
universal adapter. Refer to figure 8.
6. Insert strainer and seal ring in universal
adapter then install Hale SPV on the
universal adapter over the studs. Make sure
the vacuum hose connection is facing in the
down direction when the Hale SPV is
installed. Additionally, the small drain hole
on the diaphragam cover will be in the down
Hale SPV Installation with Universal
Mounting Adapter:
When installing the Hale SPV assembly with a
universal mounting adapter use the following
procedures.
1. Place the apparatus out of service in
accordance with departmental procedures.
CAUTION: Make sure the Hale SPV is
located higher than the pump body allowing
water to drain from the vacuum hose.
2. Determine location of Hale SPV Universal
Mounting adapter on the apparatus. When
choosing location of universal mounting
adapter make sure there is minimum of 5
inches (127 mm) clearance for hose
connections.
3. Refering to the dimensions in figure 6, drill
two 29/64 inch (12 mm) holes 2-1/2inches (64
mm) apart to attach the universal mounting
adapter to a secure point on the apparatus.
4. Select the proper mounting adapter
orientation (refer to figure 6 and 7) and
using 7/16-14 x 1 inch long cap screws attach
the universal mounting adapter to the
apparatus.
5. Apply a light coat of thread locking
compound (Loctite #290 or equal) to one

Semi-Automatic Priming Valve
9
Figure 8. Hale SPV Valve Mounting Using Universal Adapter
HALE SPV
ASSEMBLY
UNIVERSAL
MOUNTING
ADAPTER
PRIMING
STRAINER
SEAL
RING
DIAPHRAGM
COVER
DRAIN
7/16-14 NUT (2)
7/16-14 x 1-3/4INCH
LONG STUD (2)
NOTE: STUDS MUST BE INSTALLED IN
HOLES THAT LIE ON THE
HORIZONTAL CENTERLINE OF
THE UNIVERSALADAPTER.
position.
7. Apply a light coating of thread locking
compound (Loctite #290 or equal) to the
exposed stud threads. Secure the Hale
SPV valve in place on the universal
mounting adapter using 7/16-14 nuts.
WARNING: Use only pipe, hose, and
fittings from the priming pump vacuum
connection to the Hale SPV vacuum
connection rated for 29 in (737 mm) Hg
vacuum.
8. Attach 3/4inch (19 mm) ID vacuum hose
from the hose connection on the Hale SPV
to the priming pump connection. Use hose
with 3/4inch (19 mm) inside diameter that is
rated for 29 in (737 mm) Hg vacuum
(Aeroquip 2556-12 or equal). Make sure the
hose is routed and properly secured in place
to prevent chaffing and abraision.
9. Install a fitting on the side opposite the Hale
SPV on the universal adapter. Connect a 3/4
inch (19 mm) inside diameter hose rated for
29 in (737 mm) Hg vacuum (Aeroquip 2556-
12 or equal) from the priming tap on the
pump body to the hose fitting on the
universal mounting adapter.
10. Test operation of the Hale SPV and priming
pump. Conduct vacuum and hydrostatic
tests in accordance with department
procedures, NFPA1901 or NFPA 1911.
PRIMING HOSE CONNECTION
USE ¾ INCH (19 mm) INSIDE
DIAMETER HOSE

Semi-Automatic Priming Valve
10
Hale SPV Test Procedures:
Performance of the Hale SPV shall be tested
monthly. The following procedures shall be
used to test the Hale SPV. If proper
performance as indicated below cannot be
achieved then the valve must be repaired.
1. Close all valves and drains on the pump.
Cap all suction openings and outlet of
suction side relief valve (if so equipped).
2. Connect a test vacuum gauge to the intake
test gauge connection on the pump panel.
3. Depress and hold the prime push-button to
energize the priming pump motor. Observe
the diaphragm on the Hale SPV while the
priming pump is running. The diaphragm
should draw in when sufficient vacuum is
generated in the vacuum hose.
4. Continue operation of the priming pump
while observing the vacuum test gauge.
When the vacuum gauge stops dropping at
approximately 22 in (559 mm) Hg. release
the push-button. Observe the vacuum
gauge. If the vacuum falls more than 10
inches (254 mm) in 5 minutes it is certain
there is a leak in the system. Check for
leakage of the system.
5. Hydrostatic test the Hale SPV to 250 PSIG
(17 BAR) in accordance with departmental
procedures.
Repair Procedures:
The following procedures are provided for repair
of the Hale SPV. Refer to the exploded view in
section 5 of this manual to determine the
relative location of parts as the Hale SPV is
being disassembled.
1. Disconnect the vacuum hose from the hose
The Hale SPV is designed to provide trouble free service with a minimum of preventive
maintenance. Performance degradation can result from a ruptured diaphragm or worn seat on the
valve. A Hale SPV repair kit that contains a manual, seal ring, valve and diaphragm is available
from Hale Products, Inc. The repair kit part number is 546-1680-00-0
MAINTENANCE
connection on the valve body.
2. Remove the 7/16-14 nuts that hold the valve
assembly to the pump body or universal
adapter.
3. Remove the valve assembly, strainer and
seal ring from the pump.
4. Matchmark the diaphragm cover and valve
body to permit proper alignment of the drain
hole during re-assembly of the valve.
5. Remove the 5/16-18 x 3/4inch long cap
screws that hold the diaphragm cover and
diaphragm to the valve body.
6. Carefully pry the cover and diaphragm from
the valve body being careful to not damage
the diaphragm. Separate the cover and
diaphragm.
WARNING: Parts under spring tension can
become projectiles and cause serious injury.
When removing or installing these parts
make sure they are restrained and spring
tension is released slowly.
7. Support the valve on a stable work surface
to prevent accidental release of spring
tension. While pressing in on the
diaphragm plate unscrew the #10-24 x 1/2
inch long screw. Once the screw is
removed slowly release the spring tension
and remove the diaphragm plate.
8. Remove valve spring from valve body.
9. Remove the valve from the valve body.
10. Remove vacuum hose connection fitting and
service elbow from the valve body.
11. Carefully inspect all components of the valve
for damage and wear. The Hale SPV repair
kit (Hale P/N 546-1680-00-0) contains a new
diaphragm, valve and seal ring. Always

Semi-Automatic Priming Valve
11
replace these components when the valve is
disassembled. If the brass seat on the valve
body is damaged the Hale SPV must be
replaced.
12. During re-assembly of the valve use thread
locking compound (Loctite #290 or equal) to
lock threads in place.
13. Apply a light coating of thread sealing
compound and install the service elbow and
vacuum hose connection fitting into the
valve body.
14. Insert the new valve into the valve body from
the side that attaches to the pump or
universal adapter.
15. Place the small end of the compression
spring over the boss in the valve body
around the valve.
WARNING: Parts under spring tension can
become projectiles and cause serious injury.
When removing or installing these parts
make sure they are restrained and spring
tension is released slowly.
16. With the Hale SPV on a stable surface,
place the diaphragm plate over the spring
and press down to compress spring. While
holding diaphragm plate down with spring
compressed, install #10-24 x 1/2inch long
screw into the valve.
17. Place a new diaphragm on the valve body
and align screw holes.
18. Place the diaphragm cover over diaphragm
aligning matchmarks ensuring the cover
drain hole will be in the correct position
when the valve is installed on the pump.
19. Install the 5/16-18 x 3/4inch long cap screws
to hold the cover in place.
20. Apply a light coating of grease to the seal
ring then place strainer and new seal ring
into valve body. Install seal ring, strainer
and valve assembly on pump (or universal
adapter) over studs. Secure in place using
7/16-14 nuts. Tighten nuts.
21. Connect the 3/4inch (19 mm) ID vacuum
hose from the priming pump to the hose
connection fitting on the Hale SPV
assembly.
22. Test operation of the Hale SPV. Conduct
vacuum and hydrostatic tests in accordance
with department procedures.

Semi-Automatic Priming Valve
12
PARTS LIST
ITEM PART NUMBER DESCRIPTION QTY UM
1 340-0640-00-1 3/4 ID HOSE (AEROQUIP 2556-12 or EQUAL) 14 IN
2 110-1800-02-0 NUT, 7/16-14 ZINC PLATED 2 EA
3 018-8040-00-0 STUD, 7/16-14 X 1-3/4 IN. LG. ZINC PLATED 2 EA
4 010-0040-00-0 ZMQ-329H PRM VALVE STRAINER 1 EA
5 040-2260-00-0 40-4N200 RING-SEAL 1 EA
6 018-1406-07-0 SCREW, 5/16-18 X 3/4 IN. LG. 8 EA
7 044-0231-00-0 DIAPHRAGM COVER (PAINTED) 1 EA
8 046-0121-00-0 DIAPHRAGM, 5-3/8 IN. DIA. 1 EA
9 018-1004-32-0 SCREW, #10-24 X 1/2 IN. LG. 1 EA
10 005-0021-00-0 SPV DIAPHRAGM PLATE 1 EA
11 042-0081-00-0 VALVE RETURN SPRING, STAINLESS 1 EA
12 038-0151-00-0 SPV PRIMING VALVE VALVE 1 EA
13 082-4027-00-0 STREET ELBOW, 3/4 MALE NPT
X 3/4 FEMALE NPT 1 EA
14 082-4027-01-0 CONNECTION, 3/4 MALE NPT
X 3/4 HOSE BARB 1 EA
15 038-0141-00-0 SPV PRIMING VALVE BODY 1 EA
16 101-0050-01-0 SPV PRIME PLACARD 1 EA
17 200-0120-04-0 STARTER SWITCH, MOMENTARY PUSH 1 EA
18 007-3370-00-0 SPV UNIVERSAL MOUNTING ADAPTER 1 EA

Semi-Automatic Priming Valve
13
Hale SPV Repair Kit
A repair kit is available from Hale Products, Inc. The repair kit contains components that require
replacement when the Hale SPV is removed from the pump and disassembled.
REPAIR KIT PART NUMBER: 546-1680-00-0
The repair kit contains the following components. Refer to the exploded view on the previous page
for parts identification.
ITEM PART NUMBER DESCRIPTION QTY UM
5 040-2260-00-0 40-4N200 RING-SEAL 1 EA
8 046-0121-00-0 DIAPHRAGM, 5-3/8 IN. DIA. 1 EA
12 038-0151-00-0 SPV PRIMING VALVE VALVE 1 EA
029-0020-33-0 INSTALLATION MANUAL (NOT SHOWN) 1 EA

Semi-Automatic Priming Valve
14
WARRANTY
LIMITED WARRANTY
EXPRESSWARRANTY. HaleProductsInc.(“Hale”) herebywarrantsto theoriginalbuyer thatproducts
manufactured by it are free of defects in material and workmanship for two (2) years or 2000 hours
usage, whichever shall first occur. The “Warranty Period” commences on the date the original buyer
takes delivery of the product from the manufacturer.
LIMITATIONS. HALE’S obligation is expressly conditioned on the Product being.
— Subjected to nominal use and service;
— Properly maintained in accordance with HALE’S Instruction Manual as to
recommended services and procedures;
— Not damaged due to abuse, misuse, negligence or accidental causes;
— Notaltered, modified, serviced (non-routine) or repaired other than by anAuthorized
Service Facility;
— Manufactured per design and specifications submitted by the original Buyer.
THE ABOVE EXPRESS LIMITED WARRANTY IS EXCLUSIVE. NO OTHER EXPRESS
WARRANTIES ARE MADE. SPECIFICALLY EXCLUDED ARE ANY IMPLIED WARRANTIES,
INCLUDING WITHOUT LIMITATIONS, THE IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE OR USE; QUALITY; COURSE OF DEALING; USAGE
OF TRADE; OR PATENT INFRINGEMENT FOR A PRODUCT MANUFACTURED TO ORIGINAL
BUYER’S DESIGN AND SPECIFICATIONS.
EXCLUSIVE REMEDIES. If Buyer promptly notifies HALE upon discovery of any such
defect (within the Warranty Period), the following terms shall apply:
— Any notice to HALE must be in writing, identifying the Product (or component)
claimed defective and circumstances surrounding its failure;
— HALE reserves the right to physically inspect the Product and require Buyer to
return same to HALE’S plant or other Authorized Service Facility;
— Insuch event, Buyer must notify HALE for a Returned GoodsAuthorizationnumber
and Buyer must return the Product F.O.B. within (30) days thereof;
— If determined defective, HALE shall, at its option, repair or replace the Product, or
refund the purchase price (less allowance for depreciation),
— Absentproper notice withinthe WarrantyPeriod, HALE shall haveno furtherliability
or obligation to Buyer therefore.
THE REMEDIES PROVIDED ARE THE SOLE AND EXCLUSIVE REMEDIES AVAILABLE. IN NO
EVENTSHALLHALE BE LIABLEFOR INCIDENTALORCONSEQUENTIALDAMAGES INCLUDING,
WITHOUT LIMITATION, LOSS OF LIFE; PERSONAL INJURY; DAMAGE TO REALOR PERSONAL
PROPERTY DUE TO WATER OR FIRE; TRADE OR OTHER COMMERCIAL LOSSES ARISING,
DIRECTLY OR INDIRECTLY, OUT OF PRODUCT FAILURE.
Table of contents
Other Idex Control Unit manuals
Popular Control Unit manuals by other brands

Kohler
Kohler K-304 Installation and care guide

Siemens
Siemens 3RK2 Series Original operating instructions

Optex
Optex VXI-CMOD Installation and operation manual

Phytec
Phytec phyCORE-TC1775 Start-up instructions

Mitsubishi Electric
Mitsubishi Electric AJ65VBTCU-68ADVN user manual

Western
Western 56462 operating instructions

TJERNLUND
TJERNLUND UC1 UNIVERSAL CONTROL (VERSION X.06) ADDENDUM 8505017 REV B... manual

Asco
Asco 290 Series Maintenance manual

Lamtec
Lamtec ETAMATIC quick reference

Bticino
Bticino KW8013 User and installation manual

Torex
Torex VAD 050 manual

Midcontinent
Midcontinent MD302-3 Installation manual and operating instructions