ImageWorksVeterinary ImageVet DC User manual

ImageVet DC
Operation, Assembly & Installation Manual
Rev. 4.1 June 2015

2
OPERATION INSTRUCTIONS
1. General precautions 4
1.1. Description of the symbols used 4
1.2. Purpose and use of the equipment 5
1.2.1. Classication 5
1.2.2. Environmental conditions 5
1.2.3. Guarantee 5
2. Description of the system 6
2.1. Description of the X-ray unit 6
2.2. Operating mode 6
3. Operation 7
3.1. Turning the X-ray unit on and o 7
3.1.1. Turning on the basic X-ray unit 7
3.1.2. Turning on the remote 7
3.1.3. Automatic remote shut o 8
3.1.4. Remote stand-by according to time 8
3.2. Remote display functions 8
3.3. Control panel 8
3.4. Checking the parameters 9
3.5. Factory settings 10
3.6. Batteries and charge level indication 10
4. Using the X-ray unit 11
4.1 Position of the patient 11
4.2. Positioning the X-ray head 11
4.3. Position of the X-ray plate or sensor 11
4.4. Setting the exposure mode and time 12
4.5. Setting the mode and exposure time in USER mode 12
4.6. Procedure to be followed when taking the X-ray 13
5. Advanced options 14
5.1. Setting the operating mode 14
5.2. Setting the type of movable collimator 14
5.3. Restoring factory settings 15
6. Error messages 15
7. Routine maintenance 16
8. Cleaning and disinfection 16
9. Dismantling the equipment when no longer used 17
10. Technical specications 17
11. Dimensions 19
12. Identication nameplates 21
INSTALLATION & ASSEMBLY INSTRUCTIONS (WALL MOUNT)
Drawings 23-24
1. General safety precautions 25
2. Packaging 25
2.1. Dimensions and contents 25
2.2. Handling and storage 25
3. Pre-Installation 26
3.1. Mechanical specications required 26
3.2. Central control unit power supply 26
3.3. Wiring connection between central control unit and “In-Use” light 26

3
INSTALLATION & ASSEMBLY INSTRUCTIONS (WALL MOUNT - CONT’D)
4. Installation 27
4.1. Positioning the X-ray unit’s yoke 27
4.2. Wall-mounted plate for supporting the X-ray unit 27
4.2.1. Vertical single stud installation with wooden post 28
4.2.2. Horizontal double stud installation with wooden post 28
4.2.3. Vertical single/double stud installation with iron post(s) 28
4.3. Extension arm 29
4.3.1. Bottom mount installation extension arm (vertical) 30
4.4. Installing the folding arm 31
4.5. Installing the tubehead 33
4.6. Installing the collimator 36
4.7. Balancing the folding arm 37
4.8. Adjusting the folding arm end-stops 37
4.9. Wall-mounted plate wiring connections 38
4.10. Completion of wall-mounting plate and holder for remote 39
5. Setup & Technician Setup Menus 40
5.1. Restoring factory settings 41
5.2. Calibrating the x-ray head 41
6. Central Control Unit Card P/N 97660515 42
7. Basic Head Control Card P/N 97660514 43
8. X-ray Tubehead 45
8.1. Pairing the Remote with the X-ray Tubehead 46
8.2. Check occupation of radio frequencies for remote 47
8.3. Error codes given on the remote 49
Block Diagrams 51
MOBILE SYSTEM ASSEMBLY INSTRUCTIONS
1. Legs, Base & Column 54
2. Installing Folding Arm & Adaptor Plate 55
3. Hanging the Timer Plate Assy 56
4. Electrical Connections 57
5. Installing Covers 58
6. Installing the Tubehead 59
7. Installing the Collimator 62
8. Balancing the Folding Arm 63
9. Installing the Remote Holster 65
10. Installing the Handles 66

4
1. General precautions
These instructions explain how to properly use the X-ray unit. Please carefully read and become familiar with the content of this manual before attempting
to use the apparatus.
NOTE: This manual does not specify the rules and regulations for acquiring, possessing or using a source of ionizing radiation as
each country has its own laws. Only the most common ones shall be mentioned and this means that it is the user’s responsibility
to check local standards and observe the relevant laws.
No part of this manual is to be reproduced, stored in a retrieval system or transmitted in any form or by any means, i.e. electronic, mechanical, photocopying,
translation or otherwise, without the prior written permission of the manufacturer.
The manufacturer has a policy of ongoing development and is committed to improving its products. Therefore, some instructions, specications and
illustrations given in this manual may dier slightly from the product actually purchased. The manufacturer reserves the right to make changes to this
manual without giving prior notice.
1.1. Description of the symbols used
The following symbols are used:
WARNING!
Indicates a situation where failure to observe the instructions could cause damage to the
equipment or injury to the user and/or patient.
NOTE:
Indicates important information for the user and/or technical support personnel.
Earth connection.
Alternating current.
On.
O.
Ionizing radiation.
F.C.C. mark (Federal Communication Commission)
Equipment conforms to the essential requirements established by USA and Canada.
FCC ID

5
1.2. Purpose And Use of the Equipment
This X-ray unit is designed for use in veterinary dentistry to provide intra-oral X-rays for diagnostic purposes.
This equipment can be used to produce lm X-rays developed with chemicals, or digital images from photo-phospher plates, or digital X-ray sensors.
1.2.1. Classication
RADIO EQUIPMENT AND TELECOMMUNICATIONS TERMINAL EQUIPMENT Classication
Equipment classication in accordance with directive 99/05/EEC art.12: Class I.
EMC classication
Classication of the equipment according to standard CEI EN 55011: Group 1 - Type B.
1.2.2. Environmental Conditions
•The equipment should be installed in an area where the following conditions are present:
• Temperature from +50° to +104° F
• Relative humidity from 25 to 75% without condensate
• Atmospheric pressure from 700 to 1060 hPa
• The electrical wiring in the room where the equipment is installed must conform to the I.E.C. 60364-7-710;V2 specication (i.e. the regulations
concerning the electrical wiring to be used in surgeries) or equivalent standards in force in the country where the equipment is installed.
• ELECTRICAL CONNECTIONS : The electrical wiring must have an eective ground conductor as set forth by I.E.C. – US National Electrical
Codes and C.E.I. standards. These rules require that a ground fault circuit interrupter is installed upstream. The ground fault circuit interrupter
must be as specied below:
- Contact capacity: 120V 15A in compliance with standards IEC 60898-1 and IEC 60947-2.
- Dierential sensitivity: 0.03A.
- Power supply: 3 x 14AWG (single conductor) OR 3 x 16AWG (stranded).
The color of the 3 wires should be as specied in the standards (BROWN/BLACK power, BLUE/WHITE neutral, YELLOW/GREEN ground).
1.2.3. Guarantee
•The manufacturer guarantees the safety of the operator as well as the reliability and high performance of the equipment. The guarantee shall be
voided if the following precautions are not taken:
• Observe the conditions specied in the guarantee certicate itself.
• The equipment must only be used by following the instructions contained herein.
• Equipment installation, expansion and technical support must be performed exclusively by personnel authorized by the manufacturer to carry out
these operations.
• Never open the equipment casing. Installation, repairs and, in general, any other operations requiring the casing to be opened are to be performed
exclusively by personnel authorized by the manufacturer to carry out these operations.
• The equipment must be installed in areas which comply with that stated above at point 1.2.2 “Environmental Conditions”.
• The area where the X-ray unit is installed must comply with ocial regulations regarding protection against radiation in the country where the
equipment is used.
SAFETY PRECAUTIONS
• If any person who is not an authorized technician changes the product in any way by replacing parts or components with others not used by
the manufacturer they shall assume responsibility for the product.
• Do not forget to turn o the main switch on the equipment before leaving the surgery.
• This equipment is not water-proof (risk of electrocution).
• This equipment is not suitable for use in the presence of a mix of inammable anaesthetic gas with oxygen or nitrous oxide.
• This equipment must be stored properly so that it is kept in top working order at all times.
• Use of electric apparatus that do not comply to standard I.E.C. 60601-1-2, in the doctor’s oce or in the immediate vicinity
of it may cause electromagnetic interferences or interferences of other types, resulting in equipment malfunctions. In these cases it’s strictly
recommended to shut o the power supply to the equipment before hand.
• The manufacturer shall not be held responsible for misuse, carelessness or improper use of the equipment.
• This equipment is to be used exclusively by qualied personnel with the proper veterinary training.
• The user must be present at all times when the equipment is turned on or ready for start-up. In particular, never leave the equipment unattended
in the presence of children, mentally disabled or other unauthorized personnel in general.
• If the X-ray equipment is damaged or oil leaks, do not use the equipment and contact customer service immediately.
PROTECTION AGAINST RADIATION
•X-rays are hazardous and adequate precautions must be taken when using them. Areas where it is possible to be exposed to the X-rays
shall be clearly indicated by using this symbol which should remind personnel to observe the safety rules laid down by the laws in force
in the country where the equipment is used.
• Control the emission of X-rays from the greatest distance possible (6 feet) from the focal spot and the X-ray irradiation beam in the
opposite direction from where the rays are emitted.
• Only authorized personnel and the patient should remain in the area when X-rays are being emitted.

a
b
c
e
d
h
i
f
g
6
2. Description of the system
2.1. Description of the X-ray unit
Description of the parts:
a X-ray tubehead.
Depending on the operating mode, the high frequency constant
potential X-ray tubehead works at 60KV 7mA (En60 mode), 63KV 6mA
(En63 mode) or 65KV 6mA (En65 mode).
Rotation is limited by mechanical end-stops.
b Removable collimator/cone.
The tubehead can work with dierent types of collimators that are
automatically recognized:
• 8” round collimator (incorporated in the tube head): minimum
skin/focal distance 20 cm and 60 mm output beam
• Removable 12” round collimator (standard in RX DC): minimum
source/skin distance 30 cm and diameter of X-ray beam exiting cone
55 mm (with cone attached).
c Focal spot.
d Folding arm.
e Extension arm (available in three lengths with the wall-mounted
version): - 40 cm, 60 cm and 90 cm.
f Remote controller
The remote can be placed either near the central central control unit or in a
remote position. As a result, the doctor can move conveniently around the
room and move out of the area where X-rays are emitted.
g Hand held remote holder.
h Central central control unit.
i Master switch.
2.2. Operating mode
The X-ray unit can function in dierent modes (MULTI-MODE technology):
1) En60 operating mode
X-ray emission at 60KV and 7mA. The X-ray unit automatically calculates the
best exposure time (from 0.01s to 1.00s) based on the selected tooth and
patient size.
2) En63 operating mode
X-ray emission at 63KV and 6mA. The X-ray unit automatically calculates
the best exposure time (from 0.01s to 1.00s) based on the selected tooth and
patient size.
3) En65 operating mode
X-ray emission at 65KV and 6mA. The X-ray unit automatically calculates the
best exposure time (from 0.01s to 1.00s) based on the selected tooth and
patient size.
4) AUTO operating mode
The X-ray unit automatically suggests the best operating mode (En60, En63,
En65) and exposure time (from 0.01s to 1.00s) based on the selected tooth
and patient size.
The suggested exposure time can be corrected from the remote.
In addition, the X-ray unit has a special USER mode that can be set by the
user. The user can select the best combination of load factors (operating
mode and exposure time) for each tooth and patient size.

7
3. Operation
3.1.Turning the X-ray unit on and o
3.1.1.Turning on the X-ray unit
The central control unit is turned on and shut o with switch (A). The switch lights up to
signal the central control unit is energized.
3.1.2. Turning on the remote
The remote is turned on by pressing any key, except for the exposure key. The unit
will be in the standard conguration described in detail in paragraph 3.1.3 and then search
for the base it works with. If the base is o, the remote will not indicate the eld or status
“READY”. If the base is turned on, the remote will detect it within thirty seconds.
NOTE: The technical specications of the switch are given in paragraph 1.2.2.
The central control unit is turned on and o with the main switch (A), as illus-
trated in the gure below. The switch lights up when the central control unit is
energized. Whenever turned on, the equipment performs an operational test
that takes a few seconds. A beep is provided at the end of the test.
NOTE: The exposure time and the parameters displayed when the unit is
turned on are the last ones set before the central control unit was turned
o. If the remote controller is left untouched for a few minutes it will go into
standby mode. Simply press any key on the control panel to reactivate it.
NOTE: To optimize the range of the remote while it is being used, keep it away
from walls and metal instruments and above all, do not cover its antenna found
on top of the screen. In addition, performance may be reduced if the remote is
moved too quickly while X-rays are being taken. Error E 31 may be displayed if
out of range.

Cool Down Time
Field Strength
Tooth Selection
Exposure Time
and Dose
Battery Level
Patient Size
Removable Collimator
Not Present
Thermal Load Bar
8
3.1.3. Automatic remote shut o
Once the central control unit has been turned o, the remote automatically shuts o after approximately one minute. The remote also
automatically shuts o when it is further than the maximum allowable distance from the control until.
3.1.4. Remote stand-by according to time
The entire X-ray unit will switch over to stand-by (even if the base is on) and the remote will automatically shut o after approximately ve minutes
of non-use to save battery power. The remote turns back on displaying the last selection made by the user whenever any key, except for the X-ray
exposure button, is pressed. To edit the standby time, refer to chapter 4 that deals with the remote’s “Advanced options”.
3.3. Control panel
As illustrated in the gure below, the remote has four function keys and a single X-ray exposure button
Tooth Selection Size Selection
Increase Time
Decrease Time
X-ray Exposure
3.2. Remote display functions

9
BRIEFLY PRESSED (less than 3 sec.)
Changes over from LARGE to SMALL and vice
versa (takes place when key is released).
Changes amongst the various types of teeth
to select the area to be examined.
Increases the exposure times in steps, according
to the set scale.
Decreases the exposure times in steps, according
to the set scale.
NO EFFECTS ARE OBTAINED IF THE KEY
IS PRESSED LESS THAN A SECOND.
NOTE: “Dead man” function: When the exposure button is depressed on the wireless remote, X-rays will be emitted only when the user presses
and holds down the button. X-ray emission will terminate if the key is released too soon.
NOTE: Warm-up: When the machine has not been used for a prolonged period (more than 3 months) or when turned on for the rst time, a num-
ber of exposures with short times (0.01-0.02 seconds) is recommended to stabilize operation.
NOTE: After the removable collimator (12”) has been attached or detached, exposure times will automatically change. Allow a couple of sec-
onds before the display is refreshed, showing the new exposure time and the collimator icon.
3.4. Checking the parameters
Before actually taking an exposure, make sure the exposure parameters for the examination in progress are correctly set.
• Controlling the type of collimator selected.
The X-ray unit is normally used in conjunction with its 12” collimator.
When the collimator is removed, an icon appears on the display.
ICON ON: indicates the X-ray unit is using the built-in round cone (diameter 60mm, focal spot-skin distance 8”).
ICON OFF: indicates the removable round cone is attached (source-skin distance 12”).
Refer to paragraph 5.2 to set the type of movable collimator used.
PRESSED LONGER (more than 3 sec.)
Saves the selected setting (exposure time, sensitivity, etc…). The
memo icon lights up when the data item can be saved.
It displays the values corresponding to the tooth exposure times in
mGy and in mGy*cm2 if pressed again.
Increases the scroll speed of the values in increasing order.
Decreases the scroll speed of the values in decreasing order.
STARTS X-ray EXPOSURE (the button has to be held down while the
X-rays are being emitted, “dead man” function).
NOTE: The main functions of the keys on the remote vary according to how they are pressed:
• Checking the selected patient size.
KEY
NOTE: The function related to pressing the key briey is performed by pressing the key which will activate the function assigned to it. On the other
hand, to perform the function carried out when the key is held down longer, press the key until the relative function is started. A short beep will
signal the function has started.
NOTE: Whenever a change has been made, the preset exposure times will automatically be modied.
Small Patient
Selected
Large Patient
Selected

10
3.5. Factory settings
The X-ray unit is supplied with the following factory settings:
• Operating mode: AUTO
• Sensitivity: Level 17
• Remote stand by: 5 minutes
• Exposure times as per standard R20: 0,010 - 0,011 - 0,012 - 0,014 - 0,016 - 0,018 - 0,020 - 0,022 - 0,025 - 0,028 - 0,032 - 0,036 - 0,040 - 0,045 -
0,050 - 0,056 - 0,063 - 0,071 - 0,080 - 0,090 - 0,100 - 0,110 - 0,125 - 0,140 - 0,160 - 0,180 - 0,200 - 0,220 - 0,250 - 0,280 - 0,320 - 0,360 - 0,400
- 0,500 - 0,560 - 0,630 - 0,710 - 0,800 - 0,900 - 1,000
NOTE: These times comply with current standards I.E.C. 60601-2-7 (1999) and the ISO 497 series R20 recommendations and
CANNOT BE MODIFIED.
3.6. Batteries and charge level indication
The remote runs on two AA alkaline batteries to assure sufcient stand-alone operation. The charge level of the batteries is given on the screen as follows:
Battery fully charged (a symbol does not appear in the area that shows the battery charge level).
Battery half-charged.
Battery charge level low or almost dead (causing the remote to automatically shut o).
NOTE: The batteries should be removed from the remote if it is not going to be used for an extended period.
• Checking the selected type of intraoral exam.
Upper PreMolar/Molars Lower PreMolar/MolarsUpper Canines Lower CaninesIncisors

90°
NOTE: X-RAY BEAM
IS PERPENDICULAR
TO THE BISECTING
ANGLE
X-RAY CONE
CENTRAL X-RAY
BEAM
TOOTH ANGLE
BISECTING ANGLE
SENSOR PLANE
MAXILLARY
11
4. Using the X-ray unit
4.1 Position of the patient
4.2. Positioning the X-ray head
4.3. Position of the X-ray plate or sensor
Patients may be in sternal, dorsal or lateral recumbancy on a wet table.
The parallel technique, where applicable, provides more accurate images in terms of size compared to the bisecting technique. However, in most
cases this is only possible for the most caudal of mandibular molar/premolars. A 12” focal-skin distance (collimator installed) is preferable to obtain
better quality pictures. To avoid exposing the image receiver only partly (whether it is a sensor or photo-phosphor plate system) proper alignment
should be observed.
Position the X-ray head so that the cone is aligned at the desired angle. There is an angulation bezel provided for those who prefer to work with
specic angles (either sternal or dorsal recumbancy).
• The X-ray emission axis is perpendicular to bisecting angle between the image receiver (for example a sensor or photo-phosphor plate) and the
tooth’s long axis.
• As a result, the picture of the tooth will only be distorted by the divergence of the X-rays in relation to the focal spot.

12
4.4. Setting the exposure mode and time
The exposure parameters are set by following the directions given below:
1) select the tooth to be examined
2) select the patient size
The exposure time is automatically shown on the remote screen.
NOTE: Each operating mode (En60, En63 or En65) is displayed for approximately 1 second according to the tooth or patient size just selected.
Use keys and to change the value from 3 to 25. If the displayed value diers from the one previously saved, icon comes on. To quit this
mode, press key or . The change made to the sensitivity factor is applied to all the teeth and both patient sizes.
The selected operating mode is always used for each tooth and patient size combination in modes En60, En63 and En65.
In AUTO mode, each tooth and patient size combination is associated to the best mode from amongst the ones available. In this mode it is not possible to
assign a mode other than the default for each combination. To set the mode, refer to paragraph 4.5 “Setting the mode and exposure time in USER mode”.
To change the mode amongst En60, En63, En65 and AUTO, refer to paragraph 5.1 “Setting the operating mode”.
4.5. Setting the mode and exposure time in USER mode
In USER mode, it is possible to assign an exposure time and a mode from amongst En60, En63 and En65 to each tooth-patient size combination. The default
setting corresponds to the AUTO mode settings with sensitivity factor F=17.
To activate USER mode, regardless of the mode currently being used, press keys and at the same time. Icon will come on to signal USER
mode is active.
NOTE: If the exposure time is not saved, the change made will be lost after a new entry or as soon as the remote changes over to stand-by.
The suggested exposure time can be changed with keys and . Exposure times ranging from 0.01s and 1.00s belonging to the R20 scale can be
set. Random exposure times dierent from the ones provided in the R20 scale cannot be set.
When the exposure time displayed diers from the default setting, icon comes on.
To save the new setting, make sure icon is on and then press and hold key for approximately 2 seconds. The remote will beep to conrm the setting
has been saved. At this point, make sure icon is o.
Important: after customized settings have been made, the “Original exposure value charts” are no longer valid.
If icon is displayed while the exposure time is changed, it means the set time cannot be saved for the selected tooth-patient size combination.
Either way, the X-rays may be taken with the time displayed.
Important: when the suggested exposure time is changed, the sensitivity factor is also modied (by default set to F=17). Once this change has
been saved, it is applied to all the teeth and both patient sizes
The exposure time can also be modied by changing the sensitivity factor. Press keys and at the same time , the actual sensitivity factor will
be displayed.

13
4.6. Procedure to be followed when taking the X-ray
• Pick up the remote and move to a safe distance away (at least 6 feet) from the X-ray unit during the exposure. Make sure “READY” is indicated.
• Press and hold down the “Exposure” key on the remote until the audible warning sound (beep) stops and the yellow light goes o.
“X-ray Exposure” key
Light on control panel illuminated during X-ray emission
• Once exposure has been completed, it is possible to proceed with the next exposure unless the X-ray unit has reached the maximum allowable tempera-
ture. The percentage of the maximum allowable temperature is always shown on the screen (see icon below).
To deactivate USER mode, press keys and again (icon goes o).
The exposure parameters are set as directed below:
1) select the tooth under examination.
2) select the patient size.
The exposure time is automatically displayed on the remote.
NOTE: It is not possible to access the sensitivity factor menu in USER mode. In addition, keys and are
inoperative in USER mode.
The exposure times and mode assigned to the tooth – patient size combinations are custom set by following the directions given below:
1) press and hold down key about two seconds. Customized settings can be entered and icon comes on.
2) select the desired tooth-patient size combination.
3) change the exposure time with keys and .
NOTE: It is possible to set exposure times ranging from 0.01s and 1.00s that are part of the R’20 scale.
4) press keys and simultaneously to open the menu used to select the operating mode.
5) select the operating mode with keys and .
6) quit the menu and press key to make the entry operative (if key is pressed, the menu will quit without changing the previous setting).
7) press and hold down key for approximately two seconds to conrm the entry and disable customized settings (icon goes out).
NOTE: It is possible to set the exposure parameters for several combinations. To do this, repeat steps 2 through 6 before going on to step 7.
NOTE: If the Exposure key is released at any time before exposure has terminated, exposure will be interrupted and error code E01 will appear on
the display. If this occurs, press the “Patient Size” key to reset.

14
5.1.Setting the operating mode
The X-ray unit features the following operating modes:
• En60: all the X-rays are taken at 60KV and 7mA.
• En63: all the X-rays are taken at 63KV and 6mA.
• En65: all the X-rays are taken at 65KV and 6mA.
• AUTO: the system automatically selects the best setting from amongst En60, En63 and En65 for each tooth-patient size combination.
NOTE: the current setting is displayed on the remote for approximately 1 second for each tooth-patient size selected before the
relative exposure time is shown.
To set the operating mode, press the and keys to go to the set up menu. Scroll the parameters up to parameter P06 and then press key.
Scroll the options to nd the desired operating mode and then press key. Press key to quit the set up menu.
5.2. Setting the type of movable collimator
NOTE: In order to obtain top performance of the X-ray unit, set the cone according to the type used.
To set the type of cone, press the and keys to go to the set up menu. Scroll the parameters up to parameter P07 and then press the
key . Scroll the options to nd the type of cone used ( Round Ø55 mm ) and press the key . Press to quit the set up menu.
• Once the maximum temperature has been reached, the pause time for cooling will be signaled by symbol .
• At this point the exposure function will be disabled until the screen shows “READY” again.
• As soon as “READY” appears on the remote, another exposure may be taken.
Hold down these two keys to go to the set up menu (from P 01 to P 07).
Press key “Patient Size” to make the selection. Once within the individual congurations, they can be scrolled
with keys “+” and “-” and selected by pressing key “Patient Size” again. Key “tooth” quits set up without saving
the setting.
The congurations are given in detail below:
P 01: Sets the standby time (from a minimum of 5 to a maximum of 30 minutes).
P 02: Assigns an identication tag to the X-ray unit’s base (from 1 to 5 or none).
P 03: Shows the list of software versions.
P 04: Remote code display.
P 05: Reserved.
P 06: Selects the operating mode (En60, En63, En65 and AUTO).
P 07: Sets the type of removable cone used.
Press these two keys to adjust the sensitivity levels (determined based on the table given below and type of
sensor/receiver used), modifying the current value from the minimum to the maximum allowable one (on a scale
from 3 to 25), with keys “+” and “-”. Press key “size” to conrm the desired level and go back to the main screen.
This menu is not available in USER mode.
Activating/deactivating the USER mode. Icon comes on to signal USER mode is activated.
5. Advanced options
The remote allows the user to view, edit and set some operating parameters by simply combining the keys provided. Follow the steps given below to access:
+
+
+

15
5.3. Restoring factory settings
To restore the factory settings (see paragraph 3.5) press keys and to go to the set up menu. Press keys and simultaneously. "rESS" will
briey appear and the remote will be rebooted.
6. Error messages
With regards to other error codes, contact the technical support department.
FAULT CAUSE SOLUTION
E01 X-RAY KEY RELEASED TOO EARLY Hold down the key until the image has been captured.
E02 SHOOTING SEQUENCE NOT COMPLETED Remote most likely lost the signal. Try to repeat
exposure. If the problem persists, contact technical
service.
E03 REMOTE INTERNAL TEST ERROR Take out the batteries and then put them back in after
waiting a few seconds. If the problem persists, contact
technical service.
E04
E05
E08 REMOTE AUTO DIAGNOSIS TEST FAILED Contact technical service.
E06 GENERAL REMOTE ERROR Try to repeat exposure. If the problem persists, contact
technical service.
E07 RF SIGNAL TOO LOW Remote lost the signal. Try to repeat exposure. If the
problem persists, contact technical service.
E09 REMOTE SERIAL NUMBER INCORRECT OR NOT
INITIALIZED Contact technical service.
E10
E12
E13
E16
X-RAY UNIT INTERNAL ERROR Contact technical service.
E11 COLLIMATOR SELECTION NOT CONSISTENT After turning the rectangular collimator on or off, wait a few
seconds to allow the icon on the handheld to be updated.
E14
E15 GENERIC TUBEHEAD FAULT Contact technical service.
E17 DEVICE OVERHEATED
HEAD RELEASED Wait approximately 15 minutes forautomatic system reset.
E18
E19 SUPPLY VOLTAGE TOO HIGH/LOW Check the supplysystem.
If the problem persists, contact technical support.
E30 TRACKING ERROR (See sub codes on page 50) Refer to page 50.
E31
E32 REMOTE CONTROL ERROR Reduce the distance between the remote
controland X-ray head and then repeat the
X-ray. Follow the information given on how
to properly use the remote’s antenna.
E33 NO POWER SUPPLIED TO X-RAY TUBEHEAD X-ray Tubehead or arm cord may be faulty. Contact
technical support service.
NOTE: In presence of strong wireless communication traffic, the connection between the remote and the main control may
break down.To restore the connection, run the “Restoring factory settings” procedure.
DSP ERROR (See sub codes on page 50) Refer to page 50.
DSP ERROR (See sub codes on page 50) Refer to page 50.

16
The X-ray unit may be a source of infection passed from one patient to another if it is not cleaned properly. For this reason it should be disinfected on the
outside after each use.
If digital X-ray sensors are used, make sure they are always used with disposable hygienic sheaths. When disinfecting the X-ray unit use soft disposable paper
towels and avoid using corrosive substances or immersing the unit in liquids. Do not use sponges or any materials that can be reused.
Wipe down the unit with paper towels dampened with warm water. Alternatively, Lysol or Clorox wipes will work well.
Compatibility tests between plastics and the following products have been carried out with no negative consequences:
• Incidin Spezial (Henkel Ecolab).
• Omnizid (Omnident)
• Plastisept ( ALPRO ) (not tuberculocide as not an alcohol-based disinfectant)
• RelyOn Virkosept ( DuPont).
• Green & Clean SK ( Metasys ) (not tuberculocide since it is not an alcohol-based disinfectant)
7. Routine maintenance
ATTENTION!
Any technical maintenance work required must be carried out by qualied personnel or by a specialized technician authorized by the
manufacturer.
It is the user’s responsibility to check that routine maintenance is carried out by an authorized technician at least once every 2 years. The maintenance
methods are specied in the Technical Service Manual in possession of an authorized technician.
8. Cleaning and disinfection
ATTENTION!
Do not use products containing isopropyl alcohol (2-propanol, iso-propanol).
Do not use products that contain sodium hypochlorite (bleach).
Do not use cleaners that contain phenol.
Do not spray anything directly on the surfaces.
Never combine products with each other or with liquids other than the products listed above.
All products must be used as directed by the manufacturer.
ATTENTION!
Shut o the unit prior to cleaning and disinfecting the external parts.
All material used to clean and disinfect should be thrown away.

17
SPECIFICATIONS
• Rated voltage: 230Vac/115Vac (according to the model).
• Max. mains voltage uctuation: ±10%.
• Nominal current: 6A for the 230Vac version; 10A for the 115Vac version.
• Frequency: 50/60Hz.
• Maximum absorbed power: 1,4KVA.
• Apparent line resistance:: 0,5Ω(230Vac), 0,2Ω(115Vac).
• Fuses: 6.3A T for the 230Vac version; 10A T for the 115Vac version
• Generator: constant potential type.
• High nominal voltage: 60 / 63 / 65KV.
• Nominal current: 6 / 7mA.
• Power requirements at 0.1 sec: 420W (60KV 7mA), 378W (63KV 6mA), 390W (65KV 6mA).
• Current reference time: 0.7 mAs (7mA – 0.1s) / 0.6 mAs (6mA – 0.1s).
• Focal spot : 0.7mm
• Total ltration: 2.5mm Al @ 65KV.
• Half-value layer (HVL): >2mm Al @ 65KV.
• Leaked radiation: <0.25mGy / h at 3 feet from focal spot.
• Reproducability: 0.05.
• Electrical classication: Class I - Type B, intermittent service.
• Set exposure time: from 0.010 to 1.000 seconds.
• Accuracy of times indicated: ±10%.
• mGy display precision: ±30%.
WEIGHTS
• Weight of the unit with packaging: 77 lbs max.
• Weight of the hand-held remote: approximately 9 oz.
9. Dismantling the equipment when no longer used
10. Technical specications
As set out in Directives 2002/95/ EC, 2002/96/ EC and 2003/108/ EC, on the restrictions of the use of certain hazardous substances in electrical and electronic
equipment along with collection, treatment, recycling and disposal of waste, electrical and electronic equipment, the latter must be treated as municipal
waste, therefore sorted and collected separately. When new equipment of equivalent type is purchased, the waste equipment should be returned to the
distributor on a one-to-one basis for disposal.
As far as reuse, recycling and other forms of waste recovery mentioned above are concerned, the manufacturer is responsible for the actions specied by
individual local laws. Efcient collection of sorted waste separately to recycle and treat waste electrical and electronic equipment aids in preventing negative
environmental impacts while protecting human health. In addition, it facilitates recycling of the materials used to construct the equipment.
The crossed out garbage bin(s) placed on the identication nameplates of the equipment indicates that the waste equipment must be collected separately
from other waste.
ATTENTION!
Illegal waste disposal may involve nes dened by local laws.

18
X-RAY TUBE:
• X-ray tube: TOSHIBA D-0711.
• Focal spot: 0.7mm in accordance with IEC 336 / 1993.
• Tolerance of focal spot position on reference axis : ± 2%.
• Nominal high voltage and maximum current: (65KV, 7mA) ± 10%.
• Anode construction material: Tungsten (W).
• Anode inclination: 16.0°
• Thermal load on anode: 7.0 KJ (10 KHU).
• Maximum continuous heat dissipation: 210 W.
• Work cycle: 1:60 (1 second exposure - 60 second pause).
CONE TECHNICAL SPECIFICATIONS:
• With round cone: SSD=30cm (12”), X-ray beam less than or equal to Ø55mm,
• Without round cone: SSD=20cm (8”), X-ray beam less than or equal to Ø60mm.
A – REFERENCE AXIS
B – FOCAL SPOT IDENTIFICATION
REMOTE BATTERIES:
• Type 2 x AA Alkaline 1.5 V
TECHNICAL MEASUREMENT FACTORS:
The high voltage value is measured with a non-invasive instrument.
The anode current is controlled inside with measurement resistors and circuits to obtain very precise measurements. Operation of the circuits is checked at
the time of testing. Once assembled, the anode current can no longer be directly measured.
The exposure time should be evaluated by measuring the time that elapses from the moment in which high voltage exceeds 75% of the nominal value to the
moment in which it drops below this value. Considering the high gradient of the rising and trailing edges of the anode voltage and squaring due to inherent
ltration, use of a threshold ranging from 25% to 75% may be considered non-inuential.
A – REFERENCE AXIS
B – FOCAL SPOT IDENTIFICATION

19
11. Dimensions
All dimensions are expressed in millimeters (inches). SIDE VIEW
SIDE VIEW
400/600/900
880
1145200 95
60 1340
(34.6)
(45.1)(7.9) (3.7)
(15.7/23.6/35.4)(2.4)(52.8)
150°
150°
1125
1490 /1690 / 1990
890 / 1090 / 1390 400
525
(44.3)
(20.7)
(58.7 / 66.5 / 78.3)
(35.0 / 42.9 / 54.7)(15.7)
265
(10.4)
880
735
1145200 95
2185
115
(34.6)
(28.9)
(4.5)
890 / 1090 / 1390
615 / 815 / 1115
8601310
265
400
1720
)6.15()9.33(
(67.7)
(35.0 / 42.9 / 54.7)
(24.2 / 32.1 / 43.9)
(10.4)
(15.7)
(45.1) (7.9) (3.7)
(86.0)

20
TOP VIEW
400/600/900
(15.7/23.6/35.4)
165°
260°
2020 / 2220 / 2520
15°
450°
1490 / 1690 / 1990
(79.5 / 87.4 / 99.2)
(58.7 / 66.5 / 78.3)
Table of contents
Popular Medical Equipment manuals by other brands

Delft Imaging
Delft Imaging Delft Light user manual

Atmos
Atmos C 451 operating instructions

Ventec Life Systems
Ventec Life Systems V-Home Clinical and Technical Manual

Flo
Flo Vigaro instruction manual

Sirona
Sirona Intego Brief operating instructions

Chiroform
Chiroform StandGOSystem PARALLEL BARS PLUS 2M User and maintenance manual

Kinetec
Kinetec Spectra user manual

OPTIKON
OPTIKON ASSISTANT Service manual

Quantum
Quantum QT-740 Operator's manual

Intersurgical
Intersurgical i-view Instructions for use

Swiss Bionic Solutions
Swiss Bionic Solutions OMNIUM 1 Quick reference guide

St. Jude Medical
St. Jude Medical Optisure LDA220 user manual