IMR EM 201 User manual

Environmental Equipment, Inc.
3634 Central Ave. St. Petersburg Fl, 33711 * Phone: 727-328-2818 or 1 800 –RING- IMR Fax: 727-328-2826
E-mail: info@imrusa.com * Web: www.imrusa.com
INSTRUCTION MANUAL
ELECTRONIC MANOMETER
EM 151 / EM 201

TABLE OF CONTENTS
Safety Rules............................................................................... 2
Features...................................................................................... 3
Operating Instructions ............................................................. 4
EM151/EM201 Specifications ................................................ 6
Pressure Conversion Chart..................................................... 7
Warranty..................................................................................... 8
INTRODUCTION
The EM151 and EM201 provide a simple, time-
saving alternative to U-tube pressure measurements.
Use the UEi EM151 & EM201 to measure gas supply
and draft pressures in residential or light commercial
properties. The EM201 differential mode provides a
convenient method to measure pressure drops.

SAFETY RULES
WARNING!
Before every use, check to be sure the meter and its
tubing are free from breaks or blockage. While using
the meter, if you suspect gas is leaking, or if you can
smell gas, close off the gas supply and ventilate the
occupied space. Identify and repair the source of the
leak before continuing with the test.
In the interest of safety, this instrument should only
be used by trained, competent professionals who
know the hazards and consider the risks of working
on and with industrial equipment.
These meters do not measure water pressures;
under these conditions they will fail. Do not exceed
10 PSI input pressure
2

3
FEATURES
EM151
• Singleinputtomeasurepressureto±20”wg
• TripleScale:inWG,mBarandPSI
• Ruggedized,includingrubberbootwithintegralmagnet
• Simpletwo-buttonoperation
• Measuresinincheswatergauge
• Autopower-offafter20minutes
• 3yearlimitedwarranty
EM201
• Dualinputdifferentialmeasurementto±60”wg
• TripleScale:inWG,mBarandPSI
• Ruggedized,includingrubberbootwithintegralmagnet
• MeasuresinincheswatergaugeandmBar
• Autopower-offafter20minutes
• Min/Maxvaluecapture
• DataHold
• Back-litdisplay
• 3yearlimitedwarranty

OPERATING INSTRUCTIONS
1. Zero the meter
Priortouse,themetermustbezeroed.Todothis,ensurethe
meterisinanon-pressurizedarea,andpressandholdthe
‘ZERO’buttonforapproximatelyfourseconds.Thedisplaywill
show‘888’toindicatethatthezerofunctionhasoccurred.
2. Measure Pressure
Connecttheflexibletubebetweenthemeterandthe
pressuresource.
TheEM201’spositiveinputistotherightwhenlookingatthe
frontoftheinstrument.Applyingapositivepressuretothis
portwilldisplayapositivereading.Theinputtotheleftisthe
negativeport.(Applyingapositivepressureherewilldisplaya
positivereading).Useeitherinputtomeasureasinglesource
byventingtheunusedinputtoatmosphere.Theinstrumentwill
continuallyprovideadifferentialreadingofthetwoinputs.
4

5
3. SCALE
ToselectmBarpressthe“SCALE”buttonontheEM201.Press
“Scale”againtoreturntoinchesofwatergauge.Toselectscale
ontheEM151brieflypressthegreenbutton.
4. MIN/MAX (EM201 Only)
TheEM201willcapturevaluesforboththeminimumand
maximumvaluesread.Toresetzerotheinstrumentorsimply
powerthemeteroff,thenbackon.Capturedvaluescanbe
displayedineitherincheswatergaugeormBar.
5. HOLD (EM201 only)
TheEM201hasadisplayholdfeature.Pressthisduring
readingtofreezethedisplay.Pressagaintoreturnto
actualmeasurement.
6. BACKLIGHT (EM201 only)
TheEM201hasabacklitdisplay.Press toturnonoroffthe
backlightfeature.
7. Over-range
Iftheinputexceedsthemeterrange“OL”willbedisplayed.
Immediatelydisconnectfromthepressuresupplysafelyto
reducechancesofdamagetotheinstrument.

6
SPECIFICATIONS
EM151
Measurementrange: -20”to+20”WG
Minimumpressuredisplayed: 0.01”WG
Accuracy: ±0.02”<2”
±1%rdg>2”
Resolution: 0.01”-9.99”to+9.99”WG
0.1”<-10”,>+10”wg
Operatingrange: 0-40˚C,10-90%RH
noncondensing
Batterylife: 50hoursminwith
alkalinebattery
EM201
Measurementrange: -60to+60”WG
Minimumpressuredisplayed: 0.01”WG
Accuracy: ±0.02”<2”WG
±1%rdg>2”WG
Resolution: 0.001”-9.999”to+9.999”WG
0.01”<-10”,>+10”wg
Operatingrange: 0-40˚C,10-90%RH
noncondensing
Batterylife: 40hoursminwith
alkalinebattery

7
Multiply By To Get
In.ofH2O 1.868 mmHG
In.ofH2O 0.074 inHG
In.ofH2O 0.249 kPa
In.ofH2O 249.1 Pascals
In.ofH2O 2.491 mBar
In.ofH2O 0.036 PSI
In.ofH2O 25.40 mmH2O
PRESSURE CONVERSIONS

IMR Environmental Equipment, Inc.
Warranty
IMR Environmental Equipment ,Inc. states the following:
IMR hereby grants the following worldwide IMR warranty for an IMR analyzer purchased from
an authorized dealer.
1. The IMR warranty shall entitle every IMR customer to demand a free replacement or repair
of the defective parts from any IMR dealer authorized for the respective IMR unit.
2. The IMR warranty shall be granted on the factory new unit and shall commence on the date
of the delivery of the original IMR unit to the customer. It shall last for a period of twelve
months regardless of the type and the intensity of use and regardless of any change of
owner, which may occur during this warranty period.
3. The IMR warranty shall refer to absence of faults with respect to the state of the art nature
of the sold unit in terms of material and finish. The warranty for all parts fitted during the
twelve-month warranty period shall end with the unit warranty.
4. After the establishment of a material or production fault by IMR or the authorized IMR
dealer, the faults will be eliminated by means of free repair or replacement. Replaced parts
shall become the property of IMR.
5. No warranty claims may be made for maintenance and setting work, cleaning or other utility
materials required for the function of the unit and other wear parts unless they have a direct
bearing on work performed under the warranty.
6. The terms and conditions for the acknowledgement of this warranty shall be the
presentation of the fully completed warranty card, which must contain the confirmation from
the authorized IMR dealer on its delivery and, if applicable, the prescribed maintenance work.
7. The IMR warranty shall only be applicable if
a. The analyzer has been maintained in accordance with the instructions issued by the
manufacturers and the operating instructions by an authorized IMR dealer.
b. Only original IMR spare parts have been used for any repairs.
c. The unit has been used properly, the operating instructions observed and the unit has not
been used for a purpose other than the one for which it has been designed.
d. The IMR unit has been left in its original design and meets the original IMR specifications.
e. The fault is not due to external influences or use for a purpose other than the one for which
it has been designed.
f. Exclusively authorized IMR dealers have made repairs to the IMR unit.
g. The IMR unit has been sent to an authorized IMR dealer immediately after the fault was
discovered.
8. Warranty time for the analyzer, including electrochemical sensors is 12 months.
IMR Environmental Equipment, Inc.
3634 Central Ave.
USA – St. Petersburg, FL 33711
Ph: +1-727-328-2818
Fax: +1-727-328-2826
Web: www.imrusa.com