USING THE AC ADAPTER /
BATTERY CHARGER
UNIT POWERED OFF
Insert the supplied power supply plug by into the port at the bottom end of the unit
and connect the power supply adapter to the wall socket. The GREEN“Charging”
A full charge of a depleted battery should take about4hours. When the unit is
Once the battery reaches full charge the power button will be green and remain
solid.
UNIT POWERED ON
For extended operation periods, or to charge the battery during treatment sessions,
the AC adapter CAN be connected while the device is in use. The power button
charging. The power button WILL REMAIN SOLID GREEN once the battery is
fully charged.”
CHARGING THE BATTERY: USE ONLY THECHARGER PROVIDED BY
INNOVAMED HEALTH.
Use of the wrong charger can cause excessive heat,
damage to the charging circuit and shorten the life of the battery.
ALARMS:
“Battery Critical”
–
If battery voltage drops below critical level, cycling stops,
audible alarm will sound, and power button WILL FLASH RED. Alarm will continue
for 30 seconds (unless unit is powered OFF) and automatically turn the unit OFF.
Plug in power supply immediately.
“Low Pressure or Leak”
–
When activated, if pressure limit is NOT reached
within 20 seconds after pump is energized, cycling stops, ALARM sounds and BLUE
(unless unit is powered OFF) and automatically turn the unit power OFF. Make sure
wrap is attached snuggly to the leg. Turn the unit OFF, and then turn the unit back
customer service for a replacement unit at 877-475-7050.
ALARM RESET:
To reset analarm condition after the“operation inhibit”stage is reached, the unit
must be turned OFF. If notmanually turned OFF within 30 seconds of such an alarm
condition occurring, unit automatically turns itself OFF. Once the unit is turned back
on the alarm is RESET.
OPTIONS:
By Prescribing Physician Only:
At the top edge of the pump/controller unit is access to amini USB port. This is
NOT to be directly connected to a computer. In order to download use data, a data
expander interface must be purchased from InnovaMed Health.
An additional function available through the USB port is to add a remote control
module for ON/OFF operation from a hand-held transmitter. These modules aid in
controlling the system for patients unable to access the device operations.
CLEANING AND DISINFECTING
NOTE: Inspect the VenaPro unit and follow the cleaning and disinfecting procedures prior to eachuse.
WARNING:
cleaning or disinfecting.
DO NOT IMMERSE UNIT IN ANY LIQUID FOR ANY REASON
•
Clean the outer surface of the pump unit using a soft cloth, moistened withsoapy water or 70%
isopropyl alcohol.
•
Do not use abrasive or volatile cleaners.
WARNINGS AND
PRECAUTIONS
• T
use only.
• Medical Electrical Equipment needs special
precautions regarding EMC. Portable and mobile
other Medical Electrical Devices.
•
may cause skin irritation. Regulary check for patent
discomfort, compliance, and skin irritation.
• To prevent extremity compartment syndrome,
special attention should be given to patients who
are positioned in the supine lithotomy position for
extended lengths of time. This includes patients
• Do not open or remove covers. No user serviceable
parts inside. Direct all unit issues to InnovaMed
Health.
•
wrapped too tightly. Loosen Immediately.
• Stop using device if swelling occurs; consult a
Physician.
• Device is to be used only by the patient prescribed,
and only for its intended use.
•
unplugged from the wall outlet prior to and while
cleaning or disinfecting.
• Equipment should not be used in the presence of
or nitrous oxide.
• Do not immerse in any liquid for any reason.
• Do not operate device in a wet environment.
•
• Do not subject the unit to extreme shocks, such as
dropping the pump.
• Contains no user serviceable parts. Contact
InnovaMed health Customer Service.
• Do not place any items in an autoclave.
• Operation of this device can be done by the patient.
• No Service is to be attempted while the device is in
use.
•
•
•
•
water or 70% isopropyl alcoholand let air dry.
•
To ensure the unit IS completely dry prior to use, leave unit in the OFF
condition and disconnected from the wall outlet for 30 minutes after cleaning
or disinfecting.