Intex Oxicare User manual

The finger pulse oximeter produced by our company is a
kind of noninvasive and continuous medical instrument for
detecting the arterial oxygen saturation and pulse rate of
human body. It is portable and can accurately and rapidly
measure the blood oxygen, so that you know whether you
are lack of oxygen at present.
USER MANUAL

Blood oxygen saturation is the percentage of the capacity of
HBO2 in the blood which is combined with oxygen in the total
capacity of Hb. That is, the concentration of blood oxygen,
which isavery important physiological parameter of the
respiratory and circulatory system. Many respiratory
diseases can cause the decrease of oxygen saturation. In
addition, the dysfunction of the body's automatic regulation
caused by anesthesia, surgical trauma and some injuries
caused by medical examination may lead to the patient's
oxygen supply problems and reduce the human oxygen
saturation, resulting in some adverse reactions such as
vertigo, vomiting, weakness and other symptoms. If the
treatment measures are not taken in time, the patients' lives
will be endangered. Therefore, it is very important for doctors
to find out the problem in time by knowing the patient's
oxygen saturation in time.
◆The finger clip type pulse oximeter (hereinafter referred to
as the oximeter) has the advantages of small volume, low
power consumption, simple operation and convenient
carrying. When measuring, as long as the finger is extended
into the finger clamp photoelectric sensor, the display screen
will directly display the measured blood oxygen saturation
value.
◆The measurement principle of the oximeter is based on the
spectral characteristics of hemoglobin and oxyhemoglobin
with different absorption rates in the red and infrared light
areas, and the empirical formula of data is established by
using "lamert beer" law. The working principle of the
instrument is to use photoelectric blood oxygen detection
technology combined with volume pulse tracing technology.
Two different wavelengths of light (660nm red light and
940nm near-infrared light) are used to irradiate the fingernail
of human body through absorption perspective clip sensor,
and the measurement signal is obtained by the photosensi-
tive element. The obtained information is displayed on the
◆Measurement principle
Summary
1

display after the electronic circuit and microprocessor, which
is easy to read.
◆It is composed of double-layer components and
photoelectric detector. Bone, cell tissue, pigmentation and
vein vessels all have different absorption constants of light.
When the artery pulsates with the heart's contraction and
relaxation, the absorption of light will be different with the
increase and decrease of blood flow. The absorptivity of
different light during diastole and contraction of the heart is
transformed into the measurement result of blood oxygen
saturation. This measurement is the oxygen saturation.
1. It can not be used together with MRI or CT equipment.
2. Explosion hazard: do not use this instrument in the
environment of flammable anesthetic gas.
3. The oximeter only plays an auxiliary role in the diagnosis
of the patient. Please ask the doctor to makeadiagnosis
based on the clinical manifestations and symptoms.
4. Always check the sensor test point of the oximeter to
confirm that the patient's skin at the sensor test point is
complete and in good circulation.
5. The sensor of the finger clip pulse oximeter is not suitable
to contact the adhesive tape, which may cause the
measurement data error or mistakenly think that the tested
skin has blisters.
6. Please read this manual carefully before use.
Schematic diagram of operation
1. Infrared receiving tube
2. Infrared emitting tube
Safety warning
2

7. The oximeter does not have the function of blood oxygen
prompt, but it can not be used as a long-term continuous
monitor.
8. When using for a long time, please change the test point
periodically according to the different situation of the patient.
The test point must be replaced every2hours at most and
the skin integrity, circulation and correct adjustment of the
patient must be checked.
9. High pressure disinfection, vinyl oxide disinfectant or
immersion of the sensor in the liquid disinfectant will cause
incorrect readings.
10. Dysfunction of important indicators of hemoglobin (such
as carbohemoglobin or methemoglobin) can cause incorrect
readings.
11. Excessive staining in blood vessels, such as indocyanine
green or methylene blue, may cause incorrect readings.
12. The measurement accuracy of blood oxygen saturation
will be affected by the strong light of the surrounding
environment. (e.g. wipes or direct sunlight) add a protective
enclosure to the sensor if necessary.
13. The patient's unexpected movements may cause
incorrect readings.
14. The interference of medical high frequency signal or
defibrillator may cause wrong reading.
15. Rhythmic pulsation of the vein can cause incorrect
readings.
16. The measurement position of the sensor and the blood
pressure cuff are in the same artery or the same blood
vessel, which may cause wrong readings.
17. Patients with hypotension, severe systolic hypotension,
severe anemia or hypothermia may cause incorrect readings
18. If the patient uses a cardiotonic after the heart stops
beating, or if the patient shivers, it will cause a wrong reading
19.bright nails or nail polish coated nails cause erroneous
readings.
the recyclable accessories or components of accessories
in the equipment specified in the manual, including batteries,
shall comply with local laws and regulations.
3

◆The power consumption of the product is low, and the two
AAA batteries can be used for more than 15 hours;
◆When the battery voltage is too low which may affect
normal use, there is a low voltage warning indication;
◆When no signal, the product will enter the shutdown state
after 8 seconds;
◆Small size, light weight and very easy to carry.
This product is suitable for monitoring the pulse oxygen
saturation and pulse rate of patients.
This product is not recommended for use during exercise
and is not suitable for continuous monitoring of patients.
This product is not applicable to infants and newborns
1. Install 2pcs AAA batteries according to the positive and
negative marks of the battery compartment and directly
fasten the battery cover;
2. Insert the finger into the rubber hole (the finger should be
fully extended), with the nail facing up, and then release the
clamp;
3. Press the button in the middle of the front panel to start the
oximeter for blood oxygen detection;
4. In the use process, it is better not to shake your fingers,
and the human body should not be in motion;
5. If you want to change the display direction during use,
press the button in the middle of the front panel;
6. read the relevant data directly from the display;
7.The instrument has sleep function, and it will enter standby
sleep state if no signal is generated;
replace the battery.
4
Product features
Product scope of application
Method of use
8.When the battery power is low, the low battery power
indication will appear on the LCD, prompting the user to
interface can display four different display modes;
◆The product adopts LCD hd display, and the

Fingernails must be up when inserting fingers
Declaration: before using this product for testing, please use
medical alcohol to wipe the rubber of finger sleeve first, and
use medical alcohol to wipe the tested finger before and after
using. (the rubber material is medical rubber, which is
non-toxic, harmless and has no side effects such as allergy
to human skin.)
Key function description: in the standby mode, start the key
instrument to enter the working mode, and press the key in
the working mode to change the display mode.
2. Pulse oxygen saturation display range: 35% - 100%
3. Blood oxygen saturation: measurement accuracy: within
the range of 70% - 99% ± 2%, < 70%, no definition of
5
1. 2.
3. 4.
Report errors
Brief introduction of the whole machine
Technical specifications
◆Front and back panel
◆Power on / mode key
◆LCD display mode
1. Display mode: LCD hd display

℃~ 40
℃
106kpa
accuracy
Resolution: blood oxygen saturation ± 1%
4. Pulse rate: 25bpm ~250bpm
Measurement accuracy:±1bpm or ± 1% of the measured
value (whichever is greater)
5. Battery model: 2 AAA 1.5V alkaline batteries;
6. Automatic shutdown: when no finger is inserted, it will be
automatically shutdown after 8 seconds
7. Boundary dimension: 58mm × 32mm × 34mm
8. Operating environment: operating temperature: 5
℃
Storage temperature: - 10 ℃~ 40
Environmental humidity: 15% - 80% storage when working
Atmospheric pressure: 70kpa ~
9. Declaration: EMC of this product complies with
iec60601-1-1-2 standard.
10. Pulse wave measurement sensitivity under the condition
of weak perfusion: the necessary test equipment (bio-tek
index pulse oximeter) can measure the available pulse wave
signal whose amplitude is 6% of the amplitude of analog
pulse wave signal.
11. Anti environment light interference ability: apply
interference signal test with intex pulse oximeter
tester, the instrument can work normally
1. According to the management category of medical
devices, it is divided into: Class II equipment
2. It can be divided into: internal power supply equipment
according to the type of anti electric shock
3. According to the degree of anti electric shock: BF type
1. When the low voltage indicator is on, please replace the
battery immediately
2. Please scrub the surface of the oximeter before use
3. If you do not use the oximeter for a long time, please take
out the battery inside
6
Product classification
Maintenance and preservation

4. The product is best stored in the environment with the
ambient temperature of - 10 to 40℃(14 to 104 ) and the
humidity of 10% to 80%
5. The product must be kept dry at all times. The humid
surrounding environment may affect the service life of the
product or even damage the product
6.Please handle the used batteries according to the laws
and regulations of the local government
The oximeter is designed to be used in a specific
electromagnetic environment. The user must ensure that
the product is used in the following environment
1. 1pcs lanyard
2. 1pcs User manual
7
Spare parts
Product statement - electromagnetic radiation - for other
equipment and systems
Radiation test
RF interference
CISPR 11
RF interference
CISPR 11
Basis
Group 1
Class B
Electromagnetic
environment -
Guidance
The RF signal of the
oximeter is only
generated by the
internal function, so its
RF interference is very
low and will not interfere
with the surrounding
electrical equipment.
The oximeter is
applicable to all
institutions, including
communities and
domestic buildings
directly connected to
the public low-voltage
power supply network.

8
Problem Potential
reason
solution
1. Put the finger in
correctly and try again
2. Please try several
times. If you confirm
that the product has no
quality problem, please
go to the hospital for
diagnosis
Abnormal display
of blood oxygen or
heart rate
1.Fingers are not
put in correctly
2.The patient's
blood oxygen
value is too low to
be detected
1. Fingers putting
not deep enough
2. Fingers
shaking or human
body moving
1. The battery
may be low or
dead
2. The battery
may not be
installed correctly
3. The machine
may be damaged
1. The product is
set to stand by
after 8 seconds
when there is no
detection signal
2. Low battery
Blood oxygen or
heart rate is
unstable
Unable to power
on
The display light
goes out suddenly
1. Please put your finger
in correctly and try
again
2. Please try to keep
your body still
1. Please replace the
battery
2. Please reinstall the
battery
3. Please contact the
local customer service
center
1. Normal
2. Please replace the
battery
Possible problems and Solutions

9
Symbols and meanings
We reserve the right to change the technical appearance
of this product without notice.
Symbol Meaning
Equipment type is BF type
Please refer to the manual
before use
Perfusion Index
Blood oxygen saturation
Pulse rate
(beats per minute)
Low voltage indication
Serial number
Enterprise Name:Zhengzhou AiQURA Intelligent Technology
Co., Ltd.
Address:4th Floor, Building 26, Smart phone industrial park,
Intersection of xingang Avenue and Renmin road,
Zhengzhou airport economic comprehensive experimental
zone.