manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Intoximeters
  6. •
  7. Test Equipment
  8. •
  9. Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Other manuals for Alco-Sensor FST

5

Other Intoximeters Test Equipment manuals

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Popular Test Equipment manuals by other brands

Ambit P.F. Series instruction manual

Ambit

Ambit P.F. Series instruction manual

Megger BMM500 Series user guide

Megger

Megger BMM500 Series user guide

RJG Lynx Sensor Tester operating instructions

RJG

RJG Lynx Sensor Tester operating instructions

Mark-10 TT02 Series user guide

Mark-10

Mark-10 TT02 Series user guide

Martel 3001 Operator's manual

Martel

Martel 3001 Operator's manual

ElproSys DiagProg4 instruction manual

ElproSys

ElproSys DiagProg4 instruction manual

Redtech TRAILERteck T05 user manual

Redtech

Redtech TRAILERteck T05 user manual

Venmar AVS Constructo 1.0 HRV user guide

Venmar

Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions SafetyPAT operating manual

Test Instrument Solutions

Test Instrument Solutions SafetyPAT operating manual

Hanna Instruments HI 38078 instruction manual

Hanna Instruments

Hanna Instruments HI 38078 instruction manual

Kistler 5495C Series instruction manual

Kistler

Kistler 5495C Series instruction manual

Waygate Technologies DM5E Basic quick start guide

Waygate Technologies

Waygate Technologies DM5E Basic quick start guide

StoneL DeviceNet CK464002A manual

StoneL

StoneL DeviceNet CK464002A manual

Seica RAPID 220 Site preparation guide

Seica

Seica RAPID 220 Site preparation guide

Kingfisher KI7400 Series Training manual

Kingfisher

Kingfisher KI7400 Series Training manual

Kurth Electronic CCTS-03 operating manual

Kurth Electronic

Kurth Electronic CCTS-03 operating manual

SMART KANAAD SBT XTREME 3G Series user manual

SMART

SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies BERT Serial Getting started

Agilent Technologies

Agilent Technologies BERT Serial Getting started

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
1 A
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
1 B