Intoximeters Alco-Sensor FST User manual

Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.

(314)429‐4000
Intoximeters Inc.
1 A
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual

goto:http://training.intox.com/asfst

Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
1 B

Other manuals for Alco-Sensor FST

5

Other Intoximeters Test Equipment manuals

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Intoximeters Alco-Sensor FST User manual

Intoximeters

Intoximeters Alco-Sensor FST User manual

Popular Test Equipment manuals by other brands

Chauvin Arnoux pel 102 user manual

Chauvin Arnoux

Chauvin Arnoux pel 102 user manual

Proceq Equotip Live Series quick start guide

Proceq

Proceq Equotip Live Series quick start guide

Testboy TV 292 operating instructions

Testboy

Testboy TV 292 operating instructions

Southwire M400TP operating instructions

Southwire

Southwire M400TP operating instructions

Beta 1464FAP Operation manual and instructions

Beta

Beta 1464FAP Operation manual and instructions

Elma Instruments Elma TR1000 manual

Elma Instruments

Elma Instruments Elma TR1000 manual

KYORITSU KEW FORK instruction manual

KYORITSU

KYORITSU KEW FORK instruction manual

Agilent Technologies InfiniiVision 5000 Series Demo guide

Agilent Technologies

Agilent Technologies InfiniiVision 5000 Series Demo guide

Hioki 3030-10 user manual

Hioki

Hioki 3030-10 user manual

EXFO FTB-200 user guide

EXFO

EXFO FTB-200 user guide

GE EVT-Pro 2.0 Operation and technical manual

GE

GE EVT-Pro 2.0 Operation and technical manual

Rigol MSO2000A Series user guide

Rigol

Rigol MSO2000A Series user guide

KOSTER Moisture Tester User instructions

KOSTER

KOSTER Moisture Tester User instructions

DARKSTAR TECHNOLOGIES D220 Guide to Operation

DARKSTAR TECHNOLOGIES

DARKSTAR TECHNOLOGIES D220 Guide to Operation

Wiha Torque QuickCheck Instructions for use

Wiha

Wiha Torque QuickCheck Instructions for use

Performance Tool W89735 owner's manual

Performance Tool

Performance Tool W89735 owner's manual

Phoenix Contact PROFIPOL operating manual

Phoenix Contact

Phoenix Contact PROFIPOL operating manual

Jamara ACCU MASTER 8 BAL Instruction

Jamara

Jamara ACCU MASTER 8 BAL Instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.