Intoximeters Alco-Sensor FST User manual

Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
1 A

Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Alco-Sensor FST
Accessing On-Line Training & Manual
goto:http://training.intox.com/asfst
Thefollowingoptionsareavailablewithoutalogin:
ASFSTOperatorsCheatSheet
CalibrationLogSheet
NHTSAApprovalLetter
Creatingaloginissimpleandfree.Afterloggingin
youwillhaveaccessto:
IndividualOnlineTraining
GroupOnlineTraining
ASFSTOperatorsManual
Call(314)429‐4000for:
*TechnicalAssistance
* Supplies
*RepairInformation
Intoximeters Inc.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
Intoximeters Inc.
Alco-Sensor FST
Accessing Menu Options
1. Hold down OFF button while pressing ON
button to access MENU features.
2. Press ON button to cycle through MENU
options (listed below).
3. Press OFF button to select displayed option.
MENU Listing
(A password may be required)
rcL Recall
Display last test.
PAS Passive Test Mode
Replace mouthpiece with Passive Sampling.
Cup. Ask subject to blow at cup until instrument
samples.
rbL Roadblock Test Mode
Allows rapid repeat testing. Press ON button
after test result to begin another test.
ACC Accuracy Check
For use by Maintenance Technician.
CAL Calibration
For use by Maintenance Technician.
dSr Description
For use by Maintenance Technician.
1 B
Other manuals for Alco-Sensor FST
5
Other Intoximeters Test Equipment manuals
Popular Test Equipment manuals by other brands

KOSTER
KOSTER Moisture Tester User instructions

DARKSTAR TECHNOLOGIES
DARKSTAR TECHNOLOGIES D220 Guide to Operation

Wiha
Wiha Torque QuickCheck Instructions for use

Performance Tool
Performance Tool W89735 owner's manual

Phoenix Contact
Phoenix Contact PROFIPOL operating manual

Jamara
Jamara ACCU MASTER 8 BAL Instruction