Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Intoximeters Inc.
(314)429‐4000
www.intox.com
Alco-Sensor FST®
Basic Operation Checklist
1. Attach Mouthpiece
Insert closed end of mouthpiece into end of channel,
single hole facing up. Lever/push down to lock into
place.
2. Power Instrument ON.
Press the ON button (large button) with trigger/index
finger to power unit ON. Hold down for several
seconds if back lighting is preferred.
3. Battery condition and temperature will display.
4. When ‘blo’ appears, instruct subject on how to
properly provide a breath sample.
“Take a deep breath, hold it for a moment and blow
steadily into the mouthpiece for as long as you can”.
5. Note the subject’s result.
6. Remove & discard the mouthpiece.
7. Turn the instrument OFF.
Press the OFF button (small button under display) or
the instrument will auto-power down after fifteen
seconds.
For On Line Certified Training and a printable
Operators Manual go to:
http://training.intox.com/asfst
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000
Intoximeters Inc.
Alco-Sensor FST
Quick Reference Status Messages
E03BlankTimeout(>60seconds)
E06ExceededSampleAttemptAllotment
E07BadCalibration(InvalidSample)
E09TempTooColdforTestTypeBeing
Performed
E10TempTooHotforTestTypeBeing
Performed
E11FailedAirBlank(ReplaceMouthpiecewith
NewMouthpiece&BeginTestAgain)
E12RFIDetectionError
E21InvalidSample
E22InvalidCalibrationTarget(Targetvalue
enteredmustbebetween.015&.200)
E29FuelCellOutofRange
E30PressureSensorBaselineOutofRange
E31ResultOverRange
RefertotheOperatorsManualforanyStatus
MessagesnotshownaboveorContactIntoximeters.
(314)429‐4000