  9. IRROMETER RSU-V User manual

IRROMETER RSU-V User manual

Other IRROMETER Accessories manuals

IRROMETER RSU-C User manual

IRROMETER

IRROMETER RSU-C User manual

IRROMETER WATERMARK User manual

IRROMETER

IRROMETER WATERMARK User manual

Popular Accessories manuals by other brands

Quincy lab 140AE-1 operating manual

Quincy lab

Quincy lab 140AE-1 operating manual

HomePro ZIR000 user manual

HomePro

HomePro ZIR000 user manual

DOL 20SCR Technical user guide

DOL

DOL 20SCR Technical user guide

RAB Smart Task installation manual

RAB

RAB Smart Task installation manual

Kemppi KMS 300 quick guide

Kemppi

Kemppi KMS 300 quick guide

Stearns 57,500 Series Installation and service instructions

Stearns

Stearns 57,500 Series Installation and service instructions

iGuzzini TWILIGHT COPENAGHEN manual

iGuzzini

iGuzzini TWILIGHT COPENAGHEN manual

Adelaide Annexe & Canvas Avan Deluxe installation guide

Adelaide Annexe & Canvas

Adelaide Annexe & Canvas Avan Deluxe installation guide

OHAUS 5000 instruction manual

OHAUS

OHAUS 5000 instruction manual

HOME8 WLS1300 quick start guide

HOME8

HOME8 WLS1300 quick start guide

MAXBOTIX MaxSonar quick start guide

MAXBOTIX

MAXBOTIX MaxSonar quick start guide

ioSmart iSmart56 Installation and user guide

ioSmart

ioSmart iSmart56 Installation and user guide

Focusrite Scarlett 18i8 user guide

Focusrite

Focusrite Scarlett 18i8 user guide

Velleman ED38105 user manual

Velleman

Velleman ED38105 user manual

JumpSport SkyBounce user manual

JumpSport

JumpSport SkyBounce user manual

Prima-Temp PRIYA Instructions for use

Prima-Temp

Prima-Temp PRIYA Instructions for use

Tefcold BC85I-BC85I W user manual

Tefcold

Tefcold BC85I-BC85I W user manual

Vaisala WMS302 user guide

Vaisala

Vaisala WMS302 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Model RSU-V
INSTALLATION AND OPERATING INSTRUCTIONS
®
RSU-V – Remote Sensing Tensiometer
The IRROMETER RSU-V incorporates a pressure transducer in a potted
waterproof housing. The RSU-V can be installed in the gauge port of any
IRROMETER, thus converting the reading to a voltage signal. Excitation voltage
required is 5 VDC.
Note 1: Consult the factory if modifying an existing MLT.
Note 2: Three wire insulated cable (suitable for direct burial) should be used to connect
Note 2: RSU-V to Electronic Reading Device – #18 AWG size (standard preferred).
Calibration Information
INPUT:5 Volts
VOLTAGE RANGE:0.5 - 4.5 Volts
SPAN:4 Volt Span
TENSION RANGE:0 - 94 kPa
CALIBRATION:.0426 V/cb
2
— WARNING —
Do Not use water containing chlorine or sulfur (distilled preferred)
or cleaners containing chlorine or sulfur with the RSU-V.
When de-airing the transducer equipped IRROMETER (Model RSU-V)
care must be taken to release the vacuum slowly. A sudden release
of the vacuum could cause damage to the pressure transducer,
rendering the instrument inoperable. Transducers damaged by
sudden release of the vacuum are unrepairable and would require
complete replacement of the entire transducer assembly.
3
Reading the RSU-V with a Digital Meter
VACUUM
0kPa
50 kPa
94 kPa
RSU-V (Standard Instruments)
The DVM shows 0.5 V
The DVM shows 2.63 V
The DVM shows 4.5 V
(10/16) #729
LITHO U.S.A.
Optimizing Irrigation . . .
Maximizing Conservation . . .
Worldwide Since 1951
WARRANTY: The IRROMETER COMPANY warrants its products against defective
workmanship or materials under normal use for one year from date of purchase. Defective parts
will be replaced at no charge for either labor or parts if returned to the manufacturer during the
warranty period. The seller’s or manufacturer’s only obligation shall be to replace the defective
part and neither seller nor manufacturer shall be liable for any injury, loss or damage, direct or
consequential, arising out of the use of or inability to use the product. This warranty does not
protect against abuse, shipping damage, neglect, tampering or vandalism, freezing or other
damage whether intentionally or inadvertently caused by the user.
®
1425 Palmyrita Ave., Riverside, CA 92507
951-682-9505 • FAX 951-682-9501
techsupport@irrometer.com
www.IRROMETER.com
Remote Sensing Units – RSU-V
Multi-conductor wire can be used and twisted pairs are recommended. All
splices should be made with direct burial type waterproof splices, Spears
DBY®or similar. The following table presents recommended maximum
limits for cable extensions to RSU-V transducers. Following these
guidelines should ensure that the voltage loss created by the wire length
will not exceed 1 Centibar (kPa) of the calibrated value.
DISTANCE WIRE SIZE
Up to 200 ft. (61 m) .. . . . . . . AWG UF 18 (1.02 mm, 0.82 sq mm)
Up to 300 ft. (91 m) .. . . . . . . AWG UF 16 (1.29 mm, 1.31 sq mm)
Up to 400 ft. (121 m) .. . . . . . AWG UF 14 (1.63 mm, 2.08 sq mm)