ISOIL SBM 32 CF User manual

ISOIL IMPIANTI S.p.A. – Italy
Sede e Stabilimento
Head office and factory
24061 Albano S. Alessandro (BG)
74, via Madonna delle Rose
Tel. +39 035 4239.011
Fax +39 035 582078
E-mail: albano@isoil-impianti.it
www.isoilmeter.com
Uffici commerciali
Sales office
20092 Cinisello Balsamo (MI)
27, via F.lli Gracchi
Tel. +39 035 4239.011
Fax +39 02 66012457
E-mail: [email protected]
Sede legale
Registered office
20124 Milano
24, viale Vittorio Veneto
Capitale soc. i.v. Euro 774.00
Codice fisc. E part. IVA 13066900153
C.C.I.A.A. MI – Reg. Impr. 86449/2000
SERVICE and
OPERATOR’S MANUAL
P.D. METER
SBM 32 CF
CODE: MA/038/00/EN/03
DATED: 04/2008

Isoil impianti SpA Page 1
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03

Page 2 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
Contents
1 Introduction............................................................................................................. 4
1.1 Warranty........................................................................................................... 4
1.2 Introduction ...................................................................................................... 4
1.3 Working principles ............................................................................................ 4
2 General safety principles ........................................................................................ 5
2.1 General instructions ......................................................................................... 5
2.2 Meter operation ................................................................................................ 5
2.3 Instructions for the operator ............................................................................. 6
2.4 Servicing instructions ....................................................................................... 6
2.5 Operating precautions ...................................................................................... 6
3 Installation .............................................................................................................. 8
3.1 Precautions ...................................................................................................... 8
3.2 Start-up precautions ......................................................................................... 9
4 Maintenance ......................................................................................................... 10
4.1 Disassembly................................................................................................... 10
4.1.1 Main components disassembling........................................................... 10
4.1.2 Rotor disassembling .............................................................................. 11
4.2 Calibrating mechanism (only for mechanical counter).................................... 11
4.2.1 Calibration of flow meter ........................................................................ 12
4.2.2 Suggested lubricating oils for calibrating mechanism ............................ 12
4.3 Tests after overhaul........................................................................................ 12
4.4 Strainer/air eliminator (optional) ..................................................................... 13
4.4.1 Strainer(Fig.2)........................................................................................ 13
4.4.2 Air eliminator.......................................................................................... 13
4.5 Assembly........................................................................................................ 14
4.5.1 Measuring chamber assembly ............................................................... 14
4.5.2 Final Assembly ...................................................................................... 15
4.6 Extraordinary maintenance ............................................................................ 15
4.7 Storing............................................................................................................ 15
4.8 Spare parts..................................................................................................... 15
4.9 Maintenance schedule ................................................................................... 16
4.10 Fault diagnosis ............................................................................................. 16
5 Technical data ...................................................................................................... 18
ISOIL IMPIANTI updates own products without condition of advance notice. ISOIL IMPIANTI doesn’t take
liabilities for use the dates that are modified

Isoil impianti SpA Page 3
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
Enclosures
•SBM32 with strainer air separator .....................................................................dwg.5659/9
•SBM32 Measuring Chamber Assembly............................................................... dwg. 6262
•Strainer air eliminator...................................................................................................FIG.2
•Calibrating mechanism - exploded view ........................................................... dwg. 672/32

Page 4 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
1 Introduction
1.1 Warranty
Each device comes with a 1-year warranty, starting from the delivering date.
Such warranty of good operating of the equipments includes our efforts of repairing or
replacing, in the shortest time, the parts that fail for defective manufacturing or material
during the warranty period, without rights to any refund for damages or other expenses.
If a device is going to be transferred in our laboratory for repair, the delivery expenses is at
the customer’s expense.
For any inspection of our qualified personnel related to what stated above, the labor is at our
expenses, while the board, lodging and travel expenses are in charge at the customer.
The components furnished and installed, but not produced, by Isoil
Impianti S.p.A.are covered by the guarantee released by the respective producers.
The warranty ceases if non-original spare parts are used; the warranty ceases too for an
improper use or if the operational limits of the device are exceeded.
1.2 Introduction
ISOIL positive displacement meters are precision measuring instruments designed for use
with a variety of petrochemical products and liquids. Each meter is fully tested and calibrated
by factory before dispatch, and a regular service will maintain a high standard of
performance and accuracy.
P.D. meters must be periodically tested by a calibrating tank, Prover, or Master Meter: if out
of accuracy a service is necessary.
Experience has shown that mechanical defects are usually caused by the entry of foreign
matters into the metering compartment due to inadequate straining facilities in the pipeline.
1.3 Working principles
Liquid enters the meter through the manifold and causes the rotor to revolve by pressure on
the vanes. The proximity of the rotor to the front and rear of the casing forms an efficient seal
while the profile of the casing guides the vanes on to the measuring crescent.
The seal between vane and body is assured by the combined effects of gravity and
centrifugal forces of vanes and it is assured the self-balance of clearances generated by the
use too.
The rotor spindle extends through a pressure tight seal in the meter front cover into the
calibrating mechanism which transmits the rotor movement to the register. (fig.1)
Fig.1

Isoil impianti SpA Page 5
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
CAUTION
All parts under pressure must be released before disassembling
the meter or its accessories for adjustment, inspection, servicing or
substitution of its components.
Also make sure that all electric or electronic part, if present, are
disconnected from its power supply.
2 General safety principles
This operator’s manual contains basic safety instructions that must be followed during
system installation, operation and maintenance. Failure to comply with these instructions
may result in personal injury and can lead to personal, industrial or environmental accidents.
Some examples of possible hazards caused by non-compliance with these instructions are:
•Failure of the system and/or some components.
•Hazards to people caused by the exposition to electrical, mechanical or chemical
influences.
•Pollution of the environment through the leaking of hazardous substances
Therefore, follow the safety instructions described in this manual; in case of uncertainties,
please contact the manufacturer.
2.1 General instructions
•Read carefully the operator’s manual.
•Make sure that all the personnel assigned to the installation, operation and
maintenance is properly trained.
•Make sure that the contents of the operator’s manual are completely understood by
all personnel assigned to the operations on the system.
•Inspect parts under pressure in compliance with national regulations before the initial
operation of the system.
•Make sure that the operator’s manual is readily available to personnel on site
•Follow national safety regulation in force in the location of the plant.
•Make sure that the system operates in compliance with the relevant operational limits.
•All pressure parts must be inspected and serviced in accordance with national laws in
force.
2.2 Meter operation
•The meter must be operated only by trained and authorized personnel.
•The meter must not be operated in presence of foreign, unauthorized or not

Page 6 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
adequately trained personnel.
•The meter must be used for the purpose it is made for; the manufacturer is not
responsible for any damage deriving from uses outside of the original purposes.
•The meter must be operated inside the limits fixed by the manufacturer; the
manufacturer is not responsible for any damage deriving from uses outside of the
operational limits of the device.
2.3 Instructions for the operator
•The operator must adhere to safety and accident-prevention standards currently in
force in the country where the device is installed.
•The operator must not, by his own initiative, carry out any operation that is outside his
competence.
•The operator must carefully comply with hazard and/or prohibition instructions
contained in this manual.
•Do not use petrol, solvents or other flammable substances to clean parts. Use only
approved commercial solvents that are non-flammable and non-toxic.
2.4 Servicing instructions
•Never carry out any maintenance, servicing or regulation before having closed the
root valve, discharged the pressure from the system and disconnected the power
supply to any electrical device (if present), unless explicitly stated otherwise.
•Read carefully the rating plates on the individual equipment.
•All maintenance operations, either ordinary or extraordinary, must be done by
authorised and trained personnel.
•The maintenance operator must wear clothes adequate to the working environment
and to the situation; in particular, loose or voluminous clothes, chains, bracelets, rings,
earrings or anything that might get caught in the mechanical parts of the system
should be avoided.
•The maintenance operator must wear adequate protective devices in accordance with
safety and accident-prevention regulations.
•In explosive environments use only antisparking equipment.
•If the meter is connected to any electric or electronic equipment, disconnect all of
them from the power supply before doing any servicing or regulation operation, unless
explicitly stated otherwise in the manual.
2.5 Operating precautions
•The meter must be calibrated following the instructions reported into the operator's
manual and in what stated into the Manual of Petroleum Measurement Standards
(API) with particular attention to the following chapters:
- Chapter 4: proving systems
- Chapter 5: metering
- Chapter 6: metering assemblies
- Chapter 11 section 2.3: water calibration of volumetric provers
- Chapter 12 section 2: calculation of petroleum quantities
•The meter must always remain full of product; to achieve this, it is suggested to install
the meter so that it remains below the main line.

Isoil impianti SpA Page 7
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
•The line upstream the meter must be kept full of product to avoid that some air enters
the meter.
•Each meter must be adjusted following the instructions supplied in the operator’s
manual.
•Follow the recommendations of the manufacturer when installing pumps. Pay
particular attention to factors such as the use of foot valves, inlet pipes dimension and
conformity with NPSH when there are suction pumps. Follow the recommendations of
the manufacturer to minimize the problems due to air and vapors.
•For flashing liquids (quick gasification of the liquid) or easily vaporizing liquids at high
environmental temperatures, e.g. light hydrocarbon, it is advisable the use of
submerged aspirations and pipes larger than the nominal dimension of the pump.
•Thermal expansions that generate overpressures can easily damage the meters and
the systems in general. Put safety valves for overpressure in every section that can be
closed through regulation or isolation valves.

Page 8 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
CAUTION
All parts under pressure must be released before disassembling
the meter or its accessories for adjustment, inspection, servicing or
substitution of its components.
Also make sure that all electric or electronic part, if present, are
disconnected from its power supply.
3 Installation
3.1 Precautions
The installation of the meter counter does not require any special procedure; just pay
attention to the following points:
•The meter counter must be installed horizontally, the rotor axis must be kept
horizontal.
•All the meter’s openings are protected with covers when shipped; those must not be
removed until the meter is installed on the piping
•Before the installation of the meter, it's suggested to clean thoroughly the piping to
remove dirt, crusts and other foreign particles.
•Piping should not exercise strain on the meter. The meter is designed for overhang
and supports should be provided only on the adjacent pipes.
•Leave space enough around the meter to ease access for adjustment, servicing and
disassembly. The counter, if present, must be easily readable.
•The meter must always remain full of product; it's suggested to install the meter so
that it remains always under the main line.
•The line upstream the meter must always remain full of product to avoid that air
enters the meter's measure chamber; if the pipe arrangement allows reversal flow, a
non-return valve must be installed.
•Flow through the meter must be regular and uniform; pulsating and irregular flows
must be avoided.
•It is recommended to install flow limiting valves downstream of the meter if the flow
rate can reach values higher than the maximum allowed for the meter.
•If the line pressure can reach values higher than the maximum allowed, automatic
safety valves must be installed in adequate places.
•To protect meter from damages due to foreign particles in the liquid, a suitable
strainer with a correct mesh number (60 mesh for diesel oil, 100 mesh for gasoline)
must be installed upstream the meter.
•In case the rate of flow through the installation exceeds the meter maximum rated
capacity, it is advisable to use a flow limiting valve, which must be installed

Isoil impianti SpA Page 9
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
downstream the same meter.
•Water must not flow through the meter.
•Meters must be installed in such a way that air or vapour do not enter through the
liquid under measurement. It is always suggested to install an air separator just
upstream the meter.
•To avoid hammer shocks which may strongly damages the meter, it is not advisable
to install upstream or downstream the meter any quick closing valves.
•It is recommended to install root valves at the inlet and outlet to ease servicing
operations and isolation of the meter
3.2 Start-up precautions
•Before proceeding with the start-up make sure that:
- The meter is adequately fixed
- All the connections are tightened
- Air is bled from the pipes
•If a calibration mechanism is associated with the meter, before the start-up it must be
filled with lubricating oil
•Vent out all the air eventually present in the line
•When the meter is operated for the first time, fill it slowly with the operating fluid by
following this procedure:
- Open slowly the upstream isolation meter or fill the meter by gravity
- Open slowly the downstream isolation meter letting the flow rate rise smoothly to
the operating value
CAUTION
Be very careful when starting up the meter: if the air enters the
measuring chamber the rotor can easily reach high rotational
speeds, leading to abnormal wear of the vanes and other
components; this, in turn, will lead the meter to a major failure.

Page 10 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
CAUTION
All parts under pressure must be released before disassembling
the meter or its accessories for adjustment, inspection, servicing or
substitution of its components.
Also make sure that all electric or electronic part, if present, are
disconnected from its power supply.
4 Maintenance
Before removing the flow meter from the pipeline for repairs, it is recommended that the
possible causes and corrective actions are noted with the help of Fault Diagnosis Chart given
in this manual.
Note that certain components in the flow meter assembly are not interchangeable. Therefore,
if more that one flow meter is dismantled, it is recommended that each flow meter is
dismantled independently.
CAUTION: Be very careful when you start up the meter after maintenance: if the air enters
the measuring chamber the rotor can easily reach high rotational speeds,
leading to abnormal wear of the vanes and other components; this, in turn, will
lead the meter to a major failure. See chap.3.2 for start-up procedures.
4.1 Disassembly
The flow meter may be considered as two main assemblies: the measuring chamber and the
calibrating mechanism. To separate these two assemblies from each other, remove the
screws securing the calibrating mechanism on the flow meter body.
CAUTION: Before dismantling the meter for maintenance release pressure from the line
and drain all the fluid inside the meter through the draining hole below the
meter.
Proceed then with the maintenance of the faulty part.
4.1.1 Main components disassembling
1. Remove Spring Dowel (30) fitted across the Rotor spindle.
2. Rest the flow meter body horizontally on bench.
3. Remove 6 Nuts (5) from the Front Cover (7).
4. Remove Front Cover and collect O’Ring Seal (9) and Shim (26) fitted inside.
Note : Do not remove Spindle Seal from the Front Cover, unless it is defective. The
defective seal is indicated by product leakage between the Front Cover and
the Calibrating Mechanism. If it is necessary to remove the seal, take out the
Snap ring (1) and Backing Washer (2), which secure the Spindle Seal into the
Front Cover, and then the seal can be withdrawn.

Isoil impianti SpA Page 11
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
4.1.2 Rotor disassembling
Remove the rotor assembly in a smooth vertical lift. Collect washer (28) from the Rear cover
NOTE: Do not remove the vanes unless they are damaged; if their substitution is
necessary:
•remove the split pins (11) fitted on the opposite side of the vane adjusting nuts;
withdraw the vane blade (10) from the assembly and collect washers (13) and
springs (12); withdraw the remaining vane blades with vane rods; remove screws
(20) and collect gland clamps (19) and gland seal rings (17).
CAUTION: Don't remove or loosen the the vanes adjusting nuts(21-22), since that would
vary the clearance between the vanes and the measuring chamber, causing
loss of performance and/or meter failure.
4.2 Calibrating mechanism (only for mechanical
counter)
The calibrating mechanism comprises a train of gears which transmit movement of the rotor
to the counter. Operational failures of the mechanism are rare and they generally regard the
breaking of tension pin, due to an excessive strain.
It is recommended to repair without varying the calibrating adjustment.
CAUTION: do not remove the shimming washers between the frame and the bearing of the
mechanism box.
Fig. 1
This chapter only applies for mechanical counter; for the calibration of the meters with
electronic counter please refer to the specific manual.

Page 12 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
4.2.1 Calibration of flow meter
The calibrating mechanism comprises a train of gears which transmit movement of the
rotor to the counter. Operational failures of the mechanism are rare and they generally
regard the breaking of tension pin, due to an excessive strain.
It is recommended to repair without varying the calibrating adjustment.
Attention: do not remove the shimming washers between the frame and the bearing of the
mechanism box.
To carry out meter calibration follow next procedures:
•Break and remove seals.
•Remove the three screws (1) securing the cover (2) to the housing (3) in which
calibrating mechanism is fitted.
•Remove cover (2)
•By using square key (4 mm) turn shaft (4) till A, B, C holes placed on the bracket (5)
and on friction roller (6) will be properly aligned.
•Insert in these holes the stop pin (supplied with the meter, then using the square key
operate on the shaft (4) as follow :
- turning counter-clockwise direction, even if the quantity of fluid does not vary, on
the counter is obtained an higher volume indication;
- turning clockwise direction it is obtained a lesser indication.
NOTE: one complete turn of the screw varies the volume indicated on the meter by
approximately 0,36% (per cent).
4.2.2 Suggested lubricating oils for calibrating mechanism
Company Type Temeprature range
AGIP OTE 32
SINT 2000
-10÷+60 °C
ESSO NUTO 32 -20÷+65 °C
IP HINDRUS HI 46 -10÷+60 °C
SHELL AEROSHELL FLUID 31 -40÷+204°C
MOBIL DTE 26
MOBIL 1
-10÷+80 °C
-10÷+200 °C
NOTE: to avoid ice forming in wintertime, add two spoons of car antifreeze.
4.3 Tests after overhaul
After overhaul the p.d. meters must be tested with suitable proving systems. Error between
the value stated by the p.d.meter counter and the value stated by the proving device is
calculated as below:
Example:
Measured by the meter Measured by the proving tank Error (%)
1000 l 1003 l -0,3%
1000 l 997 l +0,3%
The formula is:

Isoil impianti SpA Page 13
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
100%
0
0⋅
−
=
V
VV
E
V = Measured by the meter
V0= Measured by the proving tank
4.4 Strainer/air eliminator (optional)
The strainer removes all solid impurities from the liquid before it reaches the meter and the
Air eliminator removes entrapped air from the liquid.
The strainer is mounted on the upstream side of the meter.
The air eliminator is mounted centrally on the top of the top of the fabricated steel body of
strainer.
4.4.1 Strainer(Fig.2)
A clean strainer is fundamental for the correct operation of the system; a dirt strainer would
create an excessive pressure drop, leading to the failure of the basket and thus letting dirt,
crusts and other harmful elements to flow into the fluid. If possible, check regularly the
pressure drop between the inlet and the outlet of the strainer; the maximum allowed pressure
drop is 130 kPa.
The basket (2) which is made of stainless steel wire mesh is positively clamped by the end
cover and nut (5,23) a drain plug is provided for draining of liquids inside.
For a correct servicing of the strainer, follow the procedure described below:
•Discharge the pressure from the system and close the valves at the inlet and outlet
•Open the drain plug
•Remove the end cover (5) with o-ring seal (4)
•Remove and clean the basket; if possible, use a water cleaning machine
•Check accurately that the basket is intact; if it is found to be damaged, replace with
another one with the same filtering grade:
- For diesel oils: 60 mesh (247 micron)
- For gasoline: 100 mesh (153 micron)
•Remove and inspect o-ring seal for visual damage, if the rubben sealing rings need
replacement, remove the damaged rings and replace with a news o-ring
•Put again the basket in place, close the cover and the drain plug.
It's suggested to replace the o-rings between the strainer body and the cover each time the
strainer is disassembled.
4.4.2 Air eliminator
The valve assembly (air vent valve) consists to stainless float connected to the main valve
via multiple linkages (11,12,13). Main valve sits against rubber sealing ring (20) and seals
the air eliminator chamber from atmosphere. Main valve slides in guide provided in bracket
(10). Bracket is clamped to the cover plate (9) by socket head screw (27). An o-ring seal(21)
is provided between body and cover plate.
Follow the instructions below for the servicing of the air eliminator:
•Discharge the pressure from the system and close the valves at the inlet and outlet
•Discharge the liquid through the drain plug
•Remove nuts (22) and lift the cover. Collect o-ring seal.
•Remove screws (27) and remove bracket from cover plate.

Page 14 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
•Remove the main valve and the rubber seal ring.
•Inspect the air vent valve, checking that it works properly and its part aren't damaged
•Check rubber seal ring. If found damage or worn, replace with new one.
•Reassembly the valve assembly
•Close the drain plug and restore the operating conditions
4.5 Assembly
4.5.1 Measuring chamber assembly
Before assembly, clean and inspect all parts for any visual damage. Ensure that all o-ring
grooves are clean and undamaged. Examine the O-Rings and ensure that they are not
damaged or swollen, replace them if necessary. Examine graphite bushes on rotor assembly
for free rotation and play, replace them if either is apparent. Examine each vane assembly
for damage or wear, replace vane(s) if either is apparent. Examine gland seal rings in rotor
assembly for damage or wear and replace if either is apparent.
The assembly procedure is the reversal of the dismantling procedure; only for installation of
the internal body special precautions should be taken to match the centering cap of the
external body's front cover.
1. Assemble Gland Seal Rings (18) and Gland Clamps (19) on Rotor.
2. Insert Vane assembly with Vane Rods (24) into the Rotor.
Note : Ensure that pocket on the Vane matches with the pocketed portion of
the Rotor slot.
3. Place Washers (13) on Vane Rods and reassemble remaining Vane on the Vane
Rod. Ensure position of slot as above.
4. While inserting Vane, reassemble Vane Rod Spring (12) in position.
5. Ensure that the Vane is seated on Washer by pressing it manually.
6. Compress Vane Rod Spring and insert Split Pin (11) in the Vane Rod.
7. Ensure that both Vane assemblies slide freely in the Rotor slots. Check by turning
the
Rotor by hand and allowing each vane assembly to fall under its own weight.
8. Clean and inspect Body and Rear Cover assembly.
9. Place Washer (28) in the Rear Cover (29).
10. Lower Rotor-Vanes assembly into the Body (18).
11. Ensure rear Ball Bearing (35) entering the Rear Cover properly. Check the rotation
of Rotor.
12. Check clearance between the vane tip and body bore. It should be within the range
0.1 mm to 0.145 mm. If clearances are not correct, make the necessary
adjustments with the help of 2 Lock Nuts (21-22) on the Vane Rod.
13. Assemble O’Ring (9) in the Body. Place Thickness (26) on the front Ball Bearing.

Isoil impianti SpA Page 15
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
14. Assemble Front Cover (7) with Dowels (16) and Nuts (6).
15. If the Spindle Seal has been removed from the Front Cover, press the seal into the
Cover with spring on the seal facing towards inside, place Backing Washer (2)
and secure the Circlip (1).
16. Assemble Spring Dowel (30) on the Rotor spindle.
17. Check free rotation of the Rotor assembly.
4.5.2 Final Assembly
For assembling Calibrating Mechanism on the measuring chamber, follow the
procedure below :
1. Align slot of the Rotor Gear to match with the Spring Dowel fitted across the
Rotor spindle.
2. Assemble Calibrating Mechanism with the help of 3 screws.
3. Check free rotation of the Rotor by rotating it. (The Rotor can be rotated by
inserting fingers through the inlet port).
4.6 Extraordinary maintenance
The user must define a maintenance scheduling table according to the fluid utilised, the
operational conditions, the estimated/real workloads and the environmental conditions.
For all extraordinary maintenance needed after a failure and/or the rising of a fault that
compromises the normal operation of the system, please contact Isoil Impianti SpA
Customer Care.
4.7 Storing
•if after a working period it is foreseen to stop the meter for a long time, drain the meter
and its accessories;
•If the meter or the equipment of the system are not immediately used, or if it is
withdrawn from service and stored, it is important to follow next instructions :
- fill the meter and its accessories with clean kerosene or lubricated oil and close
its ends with blind flanges;
- fill the carter containing the calibrating mechanism with oil till the sight glass is
reached (see chap.4.4)
•adequately protect counters against rain and dust, with damp-proof caps
4.8 Spare parts
For a correct meter maintenance use only original spare parts from Isoil Impianti S.p.A. .
Isoil Impianti S.p.A. is not responsible for any problem that can result from the use of non
original spare parts.

Page 16 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
4.9 Maintenance schedule
Task Monthly 6 Months 12 Months 24 Months
Visual inspection of manifold
and meter body; check for leaks
X
Inspection of the meter; check
for internal part status: vanes,
rotor, measure chamber, o-
rings; check for wear or
damage.
X
4.10Fault diagnosis
Fault Possible Cause Corrective Action
Liquid passing with
normal flow rate but
counter fails to
register.
1. Defective counter. Remove and check the counter
by rotating bottom coupling. In
case of any problem in the
counter, contact the factory.
2. Drive from the rotor fails to
reach the counter, due to
sheared pins in the calibrating
mechanism.
Check cross pins fitted on all
gears in the calibrating
mechanism.
No liquid passing
through.
1. Defective counter. Remove and inspect the counter
for free movement.
2. Jammed calibrating
mechanism.
Remove and inspect the
calibrating mechanism. Identify
the source of friction and replace
relevant parts.
3. Jammed rotor assembly. Dismantle meter assembly.
Inspect front & rear covers and
rotor for scoring marks. Inspect
vanes and bearings for damage.
Causes of rotor jamming :
a. Solid particles trapped on
rotor surface.
Clean the rotor surfaces.
b. Incorrect adjustment of rotor
end clearance due to loose or
defective bearing adjuster.
Check setting of bearing
adjuster. Clean and inspect the
bearing adjuster.
c. Rotor bearings jammed. Clean bearings and cover, and
inspect for any damage.
d. Rotor bearings worn out. Replace the bearings, if axial
play is observed.
e. Misalignment of front & rear Ensure that both dowel pins are

Isoil impianti SpA Page 17
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
covers due to missing dowel
pins.
used for locating the covers with
the body.
4. Clogged strainer basket. Clean the strainer regularly.
Liquid is leaking from
the joint at front cover
and calibrating
mechanism.
Spindle seal is damaged. Replace the spindle seal.
Inspect rotor spindle for wear or
scratch marks.
Liquid is leaking from
the joint at front cover
and body or rear cover
and body.
Damaged o’ring seals, unclean
o’ring grooves, cover bolts
loose.
Clean o’ring grooves, replace
o’rings and secure cover bolts
tightly.
Excess delivery
beyond 1%.
1. Vane blades damaged. Inspect and replace damaged
vane blades.
2. Jammed rotor or calibrating
mechanism.
As described above.

Page 18 Isoil impianti SpA
Code MA/038/00/EN/03 All rights reserved. Reproduction of this manual is forbidden
5 Technical data
•Maximum working pressure .................................................................................. 1000 kPa
•Working temperature..................................................................................... -10°C ÷ +50°C
•Minimum flow rate ..................................................................................................... 30 lpm
•Maximum flow rate .................................................................................................. 350 lpm
The value reproduced in label can change as regards to those shown in the manual

Isoil impianti SpA Page 19
All rights reserved. Reproduction of this manual is forbidden. Code MA038/00/EN/03
E FILTRO DEGASATORE
P.D.METER SBM32 WITH BARE SHAFT
CONTATORE SBM32 BARE SHAFT
AND STRAINER AIR SEPARATOR
5659/9
DB / GF 01/03/07
Brambilla
01
01
1:5
A 3
ISOIL Fontana
/
/
/
01/03/07
01/03/07
Table of contents
Other ISOIL Measuring Instrument manuals

ISOIL
ISOIL SBM 150 User manual

ISOIL
ISOIL MS501 User manual

ISOIL
ISOIL ISOFLUX IFX-P200 User manual

ISOIL
ISOIL MS 2500 User manual

ISOIL
ISOIL SBM 75 User manual

ISOIL
ISOIL Isoflux IFX-F100 User manual

ISOIL
ISOIL ISOMAG MV800 Operating instructions

ISOIL
ISOIL ISOMAG MV110 Operating instructions

ISOIL
ISOIL ISOMAG MS5000 User manual