itaia Photometer FM-1 User manual

Operating manual
Photometer: FM-1
Litostrojska cesta 44/d, SI-1000 Ljubljana
T +386 (1) 514 18 00, F +386 (1) 514 18 04
ELECTRONICS
RESEARCH
DEVELOPMENT


Table of contents
1. SAFETY INFORMATION AND CAUTION............................................ 1
1.
1.
G
ENERAL
.................................................................................................... 1
1.
2.
F
IRE
-
EXTINGUISHING GEAR
......................................................................... 2
1.
3.
S
AFETY PRECAUTIONS
................................................................................ 2
2. GUIDELINE FOR CONNECTION AND USE OF PHOTOMETER.... 4
2.
1.
F
IELD OF USE
............................................................................................... 4
2.
2.
M
OUNTING AND PLACEMENT
...................................................................... 4
2.
3.
C
ONNECTING
.............................................................................................. 4
2.
4.
O
NBOARD INDICATORS
............................................................................... 6
2. 4. 1. Use of 230 VAC contactor outputs........................................................ 6
2. 4. 2. Use of U/I analog outputs ...................................................................... 7
2.
5.
C
ALIBRATION
.............................................................................................. 8
3. USE OF SOFTWARE ................................................................................. 9
3.
1.
S
ETTING UP
PC
COMMUNICATION
..............................................................10
3.
2.
W
RITING DATA TO THE PHOTOMETER
........................................................10
3.
3.
R
EADING DATA FROM PHOTOMETER
..........................................................11
4. MAINTENANCE........................................................................................12
4.
1.
C
HECKING AND MAINTENANCE
..................................................................12
4.
2.
S
ERVICING
.................................................................................................12
4.
3.
T
ROUBLESHOOTING
...................................................................................13
4.
4.
S
PARE PARTS
..............................................................................................13
5. TECHNICAL DATA..................................................................................14
5.
2.
T
RANSPORT AND STOCKING
.......................................................................15
5.
3.
D
ISPOSAL
...................................................................................................15
6. CE DECLARATION OF CONFORMITY ..............................................15
7. WARRANTY OBLIGATIONS.................................................................15

Operating manual: PHOTOMETER FM
-
1
1
1. Safety information and caution
1. 1. General
The device, connected and used in accordance with following instructions, is safe to use.
All electrical parts are protected against water, dust and dirt by the relevant IP class.
Housing prevents direct contact with any parts under voltage.
! CAUTION: For any injury resulting from
improper use, the user
takes full responsibility.
!
BEFORE MOUNTING AND CONNECTING THE
DEVICE READ MANUAL CAREFULLY!
!
THE CONNECTING OF DEVICE MUST BE CARIED
OUT BY QUALIFIED PERSONNEL!
!
WHEN CONNECTING, DISCONNECT ELECTRICAL
POWER!
!
DO NOT BLOCK ACCESS TO ELECTRICAL PARTS
USED FOR EMERGENCY STOP, CLEANING AND
MAINTENANCE!
!
ALL METAL PARTS OF HOUSING MUST BE
GROUNDED!
!
MAKE SURE THAT WATER DOES NOT GET IN
TOUCH WITH PARTS UNDER VOLTAGE!
Connecting of the device must be done by qualified personnel and in accordance with
local standards. After mounting and connection control measurements must be carried
out to ensure safe and correct operation.
!

Operating manual: PHOTOMETER FM
-
1
2
1. 2. Fire-extinguishing gear
In case of fire use CO
2
based fire extinguisher and other equipment following procedures
in accordance with fire-safety regulations.
1. 3. Safety precautions
Safety precautions are meant to provide personal safety and long service life of device.
Any person take part of mounting, connecting or maintenance of the device must follow
this manual and local safety regulations. They have to be qualified for the job and use
appropriate technical equipment.
Before starting:
Disconnect circuit breakers at input and output of the device and make
sure there is no voltage present.
Make sure power is disconnected during the time of mounting and
connecting.
Check the voltage presence with an instrument.
Check grounding connection.
Provide protection against touching the parts under voltage.
Qalified personnel
Transport, mounting, connecting, commissioning and maintenance of the device can
only be preformed by personnel:
with job qualification and experience,
familiar with up to date safety standards and
with knowledge of basic working principles of the device.
Responsibility
We take no responsibility for proper operation and any consequences resulting from use
in discordance with this manual. Safety measures must be provided by user working with
device.
In case of malfunction, please, provide us following information:
type of the device,
serial number,
errors detected,
overall time of operation,
environmental conditions and
application information.

Operating manual: PHOTOMETER FM
-
1
3
Our technical support
Litostrojska cesta 44/d, SI-1000 Ljubljana
T +386 (1) 514 18 00, F +386 (1) 514 18 04
[email protected], www.itaia.si
Declaration of conformity
The device complies with relevant European directives (CE declaration of conformity):
– Electromagnetic compatibility (ULRS, no. 132/2006)
Directive 2004/108/EC and its amendments
– Low voltage Directive (ULRS, no. 27/04)
Directive 2006/95/EC and its amendments

Operating manual: PHOTOMETER FM
-
1
4
2. Guideline for connection and use of photometer
2. 1. Field of use
Photometer is designed for measuring and regulating illumination in traffic areas. It
provides adequate ilumination at any time of the day and any light conditions. Its basic
application is to regulate the illumination of tunnels and under passages based on L
20
measurement according to CIE standard. Photometer FM-1 is provided with analogue
U/I output for PLC or tunnel SCADA connection. For local illumination controll 4-channel
230VAC programmable outputs are available. Measuring range, output channel
switching and time averaging interval are PC programmable using provided software.
2. 2. Mounting and placement
Photometer is equipped with wall fixture. Pole fixture is also available (optional).
For measuring light conditions at the entrance of the tunnel the photometer is installed at
the stopping distance before entrance portal. It should be pointed directly to the entrance
of the tunnel.
When used to measure in-tunnel light conditions, the photometer should be mounted at
the height app. 4.5 metres and pointed to the spot one reach in 3 seconds at maximum
allowed speed. In this chase the driver’s view cannot be simulated very accurate due to
mounting height. Therefore, the measuring result must be adopted by multiplying with a
constant number (usually between 1.05 and 1.20).
Once mounted, undisturbed view field of the photometer must be provided. There should
be no disturbing objects in the view field of the photometer, either any light sources or
light reflecting objects in near distance. Any of these can considerably affect light
measuring accuracy.
2. 3. Connecting
Photometer is connected to the mains power supply 230 VAC and to the tunnel light
regulating system. Cables with 3-7 leads at minimum 0.75 mm
2
must be used for 230 VAC
power outputs. For analogue U/I output coaxial 2-3 lead data cable is recommended.
Maximum connection distance is up to 400 metres. Cable inlet via glands must be well
tightened after the connection. For the parametric setup connect photometer with personal
computer and run software. Connection distance can be up to 20 metres for RS-232 (200 m
for RS-485 –optional).
All necessary connectors are accessible by removing rear cover (see Pictures 1 and 2and
Table 1).
After final connection and successful commition place provided desiccant bag in the
photometer housing and push it towards front window. Take care it is not touching any
electrical part!

Operating manual: PHOTOMETER FM
-
1
5
Green/Red
Fuse F 2A
power
supply /
4Ch 230VAC
contactor
output
PC RS 232/485
connection
analog
U/I output
r.1
r.2
r.3
r.4
out 1
out 2
out 3
out 4
1
11 13
2
Yel/gn
1
7
6
5
4
7
8
10
Picture 1: Connectors behind the rear cover
pin port voltage description
1
L-
power
230 VAC power
2 ground / grounding
3
PWR
N-
power
230 VAC power
4
/
out D 230 VAC
contactor out. D
5
contactor
out C 230 VAC
contactor out. C
6
out
out B 230 VAC
contactor out. B
7 out A 230 VAC
contactor out. A
8
PC
GND O VDC
programming/controll
9
RS 232
RX +/- 12 VDC
programming/controll
10
/485
TX +/- 12 VDC
programming/controll
11
Analog
Uout O-10V DC voltage output
12
out
Iout 4-20 mA (15V) current output
13 An GND 0V DC analog ground
Table 1: Pin out and description

Operating manual: PHOTOMETER FM
-
1
6
Data aquisition I(B)
L
PE
N
TX
U I
RX
RELAYS/
CONTACTORS
230 V AC
PC
5
2
3
RS 232
DB 9
Data aquisition U(B)
R
shunt+ line
R
max 650
Ω
R
shunt+ line
R
min 1000
Ω
4-20 mA
0-10 V
U
A
B
C
D
Picture 2: Photometer connection
2. 4. Onboard indicators
Beside connectors there are LED indicators on the PCB board (seePicture 1). They indicate
current light conditions (green) and contactor output status (red).
2. 4. 1. Use of 230 VAC contactor outputs
Those outputs can be directly connected to the coil side of contactors for light-chain
switching (check contactor’s coil voltage). Use provided software to distribute switching
combinations regarding to light condition. Example is shown at Table 2.

Operating manual: PHOTOMETER FM
-
1
7
regulation light output
A
B
C
D
LED
level condition pin 7 6 5 4 green,red
1
2
3
4
0
B<L1
230 VAC
/
/
/
ON
/
/
/
1
B>L1;B<L2
/
230 VAC
/
/
/
ON
/
/
2
B>L2;B<L3 / /
230 VAC
/ / /
ON
/
3
B>L3;B<L4
/
/
/
230 VAC
/
/
/
ON
4
B>L4
230 VAC 230 VAC
/
230 VAC ON ON
/
ON
Table 2: Indicating LEDs and contactor outputs
Four green LEDs indicate current light level switch combination (no averaging delay)
regarding to light condition. For example : At first level there is LED 1 lit in second LED 2 is lit
in third LED 3, etc. Red LEDs indicate current status of 230 VAC outputs (averaging
affected).
Light levels, output combinations and averaging interval are software adjustable (referto
PC software).
2. 4. 2. Use of U/I analog outputs
Analogue outputs directly follow current light conditions (no averaging is applied). There are
two outputs for this purpose: voltage (0-10 V) and current (4-20 mA). Simultaneous use of
both is not recommended and can affect accuracy. 230 VAC contactor outputs can be used
at the same time with no affect to accuracy. (For example: tunnel lighting can be switched
on/off by using contactor outputs while light conditions are monitored via analog outputs
connected to SCADA). For distance over 100 metres use of I-current output is preferred. For
brightness calculation use following equations:
B(U)=B
range
*(U-U
dc
range
)/(10- U
dc
range
)
B(U ) = brightness at output voltage U(cd/m
2
)
B
range
= measuring range (50 / 500 / 5000 cd/m
2
) or specified
U= measuring voltage (V)
Udc
range
= offset voltage for range (U
dc50
= 0,220V / U
dc500
= 0,028V / U
dc5000
=0,0090V)
or individual calibration constants
B(I)=B
range
*(I-4-Idc
range
)/(16-Idc
range
)
B(I) = brightness at output current I(cd/m
2
)
B
range
= measuring range (50 / 500 / 5000 cd/m
2
) or specified
I= measuring current (mA)
Idc
range
= offset current derived from offset voltage for range (Udc
range
)
Idc
range
(mA) = 1,6*Udc
range
(V) ; or individual calibration constants
NOTE:

Operating manual: PHOTOMETER FM
-
1
8
Using individual calibration constants is always preferred for best accuracy.
When using analog outputs “fix-range” optionshould be chosen when programming (seePC
software) and relevant range applied in equations!
Calibration constants are placed on thecalibration sticker at the side of photometer.
2. 5. Calibration
Each photometer is individually factory pre-calibrated. Last calibration date and calibration
constants are stored in photometer’s memory and can be read using PC program. They are
also visible on the calibration sticker at the side of the photometer. Recommended
recalibration interval is 3 years.
2. 6. Your calibration data
Factory calibration:
Place of installation:___________________________________ date:_____________
Recalibrations:

Operating manual: PHOTOMETER FM
-
1
9
3. PC software
Software FM-1v1.1.exe is used to monitor photometer condition and to configure working
parameters using PC. Install the software to any folder on the computer running MS
Windows 2000 or higher. By clicking the icon FM-1 window like shown at Picture 3 appears.
Picture 3: Photometer program window
The “System settings” button is used for factory calibration and testing. It has no
meaning for the user.

Operating manual: PHOTOMETER FM
-
1
1
3. 1. Setting up PC communication
Photometer should be connected to the computer via serial COM port using RS-232 (485)
protocol. Cable length can be up to 20 metres (using optional RS-485 up to 200 m).
In the upper right corner of the screen there is a field for adjusting communication
parameters. COM1 is selected as default. If port is already occupied by other device the
message “Port closed!” appears. In this chase select another port and click “Open port”.
If port is successfully open, the message "Port OK!" appears. Communication speed is
fixed at 9600 bps. When communicating, photometer must be powered on!
In the upper left corner of the screen there is a field to select the photometer’s ID
number. This is used to identify the device when there is more than one photometer
connected on the bus (RS-485). Use last two digits of the device’s serial number (see
back side sticker) for ID.
3. 2. Writing data to the photometer
Left part of the screen contains data to be written into the photometer (black font), right
side shows data read from the photometer (red font). By clicking “Write” button all data
consisting of 16 contactor-switch combinations, range, averaging interval and other
parameters are downloaded to the photometer. Data is only applied to device with
accurate ID number. When download is done successfully red numbers as read data
appears on the right side of the screen. If there is an error at the procedure “Error!”
message appears. In this chase check ID number and try clicking “Write” button again.
Range
Use “Range” setting to define measuring range of the photometer. Always use lowest
possible range, taking care brightness never exceed the range specified. In “auto” mode
photometer switches ranges according to light conditions. Use this option when
brightness data is acquisition is done from RS232/485 or contactor outputs. When using
analog outputs, make sure to choose fixed range R1-R3 and match range in equations.
Default factory setting for range is full measuring range –R3.
Contactor out
A contactor combinations field Level 1-16 represents the upper limit of brightness (cd/m
2
)
for the level. You can use only numerical data. When the option “Expander” is selected
the software fills the contactor-output combinations like BCD code. Use this option when
optional Expander unit is used to controll more than 4 light chains. In this chase
contactor-combinations are programmed at Expander unit.
The contactor-output combinations can be set for each level independently when option
“Custom” is set. When using less than 16 brightness level combinations, fill the rest of
the Level data with zero, and Level 16 with combination used when brightness at any
level specified are exceeded. (see: Picture 4, Table 2).

Operating manual: PHOTOMETER FM
-
1
11
Picture 4: Contactor output example
Averaging
Averaging interval represents the time in minutes for brightness-averaging before
contactor outputs are switched. This eliminates contactor switching due to short-time
light changes (approaching car, ect.) and provides hysteresis between the brightness
levels. The numerical data between 0 and 12 minutes can be selected.
3. 3. Reading data from photometer
By clicking “Read” button data is read from photometer. Data is only read from device
corresponding set ID. After reading is done successfully red numbers as read data
appear on the right side of the screen. Brightness data is not affected with averaging
interval.

Operating manual: PHOTOMETER FM
-
1
12
4. Maintenance
4. 1. Checking and maintenance
It is recommended to:
clean front screen of the photometer (minimum 2 months interval),
clean the housing (1 year interval) and
recalibrate (3 year interval).
Note!
The device requires regular checking and maintenance. Keep your maintenance and
service log up to date. Shorten check intervals if required.
•Maintenance personnel must be job qualified;
•Maintenance contractor company must provide education and training;
•Regular testing and checking must be exercised as described in this manual
and local standards;
•Any maintenance activity must be noted in the log.
4. 2. Servicing
•Servicing can only be done by authorised qualified personnel.
•Keep servicing log.
Servicing for the time of warranty is provided by manufacturer exclusively, after this time it
can be done by approved maintenance contractor.

Operating manual: PHOTOMETER FM
-
1
13
4. 3. Troubleshooting
Error
Possible cause
Action
No LED is lit. - no power supply
- burned fuse
- light conditions are out of
set range
- check power
- change fuse
- fix photometer setup
No contactor output is
active. - burned fuse
- light conditions are out of
set range
- to long averaging interval
- change fuse
- fix photometer setup
- shorten averaging interval
No analog output is active. - burned fuse
- no connection or short circuit - change fuse
- check connection
Note: “Port closed!” - note appears at start of the
program if another device is
occupying the port
- close other applications
- check serial port
Note: “No communication!” - note appears by unsuccessful
communication with
photometer
- check ID number of the device
- check power supply and connection to
the photometer
Note: “Communication
error!” - note appears by unsuccessful
communication with
photometer
- try data transmission again
- check serial COM port
Note: "... is not a valid
integer value." - note appears when other than
integer value is being entered - entered data must be numerical integer
Note: “Maximum value is
12 minutes!”
- note appears if averaging interval
exceeds 12 minutes - entered data must be between 0 and 12
4. 4. Spare parts
Parts that need to be changed must be replaced with original or functional equivalent
parts approved by manufacturer.
User serviceable parts:
- fuse F 2 A 5x20 mm

Operating manual: PHOTOMETER FM
-
1
14
5. Technical data
Basic technical data is shown on the back sidesticker.
Sensor silicon photodiode, Vλfilter
Spectral range human eye like
Spectral error < 5 % to CIE
Full viewing angle 20 °(as standard)
Measuring ranges 0-65cd/m
2
;0-650cd/m
2
;0-6500cd/m
2
(see calibration)
Accuracy > 3 %
Temperature control heater 25 W
Sensor-room temperature stability ± 1 °C
Temperature range -40 °C to +50 °C
Power supply 230 V AC ±15 %
Frequency 50-60 Hz
Max. control power 450 W
Power consumption < 30 W
Fuse F 2 A (5x20 mm)
Casing IP 65 ; Aluminium / Stainless steel / ABS
Size 395 x 120 x 180 (220) mm
Weight 3,9 kg
Fixture wall, X - Y adjustable fixture / pole -optional
Communication port RS-232/485 -optional / 9600 bps
Analog outputs:
0 - 65/650/6500 cd/m
2
0–10 V ; min. loop resistance: 1000 Ω
4–20 mA ; max. loop resistance: 650 Ω
Contactor 230VAC outputs:
Number of channels 4
Output voltage 230 V AC (common fuse 2 A)
Adjustable ranges 0–6500 cd/m
2
Averaging interval 0–720 sec.
5. 1. Environmental conditions
Maximum environmental temperature at use is +50 °C; 24 hour average temperature
must not exceed +35 °C. Lowest environmental temperature at use is –40 °C. Relative
air humidity must not exceed 50 % at maximum temperature +50 °C. Higher humidity is
allowed at lower temperature (like: 95 % at +20 °C).
Sea-level elevation must not exceed 2000 metres. When using above 1000 metres
consider lower air dielectricity and different cooling conditions.

Operating manual: PHOTOMETER FM
-
1
15
5. 2. Transport and stocking
Transport the device in it’s original packing. If it is not available use protection air bubble
foil and protect exposed areas with styropor. If needed place it on wooden pallets.
Temperature while transporting and stocking should wary between –25 °C and +55 °C.
Avoid long time stocking. After transport and mounting, allow at least 1 hour time to
settle down before commission. In case that device is not to be used for longer period,
store it in dry, well ventilated place at room temperature!
5. 3. Disposal
Lifetime of the device significantly depends on the way of use, installation, maintenance and
the working conditions. Production year is visible on the back side sticker. Deteriorated or
ruined device dispose to the place for industry electronic-components disposal according to
local regulation.
Components particularly danger for environment (batteries, chemicals, etc.) dispose to
special places.
6. CE declaration of conformity
CE declaration of conformity ensures that device is safe for use and it has been checked
and testedto the specifiedstandards. CE statement is enclosed.
7. Warranty obligations
Manufacturer of the device ensures that the product will work in accordance to specification
for the time of limited warranty of 1 year. For this time we take full responsibility and costs
(transport not included) for removal of any malfunction due to materials or workmanship.
Warranty does not cover any mechanical damage or damage being result of natural cause
or unprofessional use.

ES - IZJAVA O SKLADNOSTI
CE DECLARATION OF CONFORMITY
Proizvajalec (naziv, naslov):
Manufacturer (name, address):
ITAIA d.o.o.
Litostrojska cesta 44/d
1000 Ljubljana
s polno odgovornostjo izjavljamo, da
We declare under our sole responsibility that
(tip, model):
(type, model):
ustreza zahtevam naslednjih predpisov:
is in conformity with the provisions of the following regulations:
-Low voltage Directive (Uradni list RS, no. 27/04)
Low voltage directive 2006/95/EC and its amendments
- Electromagnetic compatibility (Uradni list RS, no. 132/2006)
EMC directive 2004/108/EC and its amendments
in izpolnjuje zahteve naslednjih standardov:
and complies with the requirements of the following standards
SIST EN 61010-1:2010
Safety requirements for electrical equipment for measurement,control and
laboratory use-Part1:General requirements
SIST EN 61000-3-2:2006
EMC-Part 3-2: Limits for harmonic current emissions (equipment input
current <= 16A per phase)
SIST EN 60529:1997
+
A1:2000
Degrees of protection provided by enclosures (IP Code)
SIST EN 61000-3-3:1997 +A1:2009
EMC-Part 3-3: Limitation of voltage changes,voltage fluctuations and
flicker in public low-voltage supply systems, for equipment with rated
current <= 16A per phase and not subject to conditional connection
SIST EN 61000-6-1:2007
EMC-Part 6-1: Generic standards-Immunity for residential,commercial and
light-industrial environments
SIST EN 61000-6-3:2007
EMC- Part 6-3: Generic standards-Emission standard for residential,
commercial and light-industrial environments
Leto, ko je bil CE znak nameščen na proizvod : 07
Year of affixing of CE mark:
FM
1/8
_xx
Peter Strmčnik, direktor
Ime in priimek, funkcija
Name, surname, position

Limited warranty conditions
Our products are workmanship and materials fault free and will work according to
specification for minimum time of warranty -1 year. We provide spare parts and servicing
for at least 5 additional years. We will respond to failure notice immediately, at maximum
7 days and fix it in maximum time of 48 working days.
For warranty period we take full responsibility and costs (transport not included) for
removal of any malfunction due to materials or workmanship. After this period we charge
servicing according to used time and replaced parts.
Warranty related obligations come in power with full payment.
Warranty obligations cease to exist:
•in case of improper use or use in discordance to this manual;
•in case of mechanical damage;
•in case of consequential damage due to failure or damage of other devices,
•in case of natural or environmental cause (e.g. lightning discharge, flooding, fire,
etc.),
•if the device has been repaired by unauthorized personnel,
•if there are unoriginal spare parts built in the device,
•if there are changes made to the device or their parts without written approval of the
manufacturer.
Servicing obligations cease to exist:
•if the device has been ruined,
•if the device has been repaired by unauthorized personnel,
•if there are unoriginal spare parts built in the device,
•if there are changes made to the device or their parts without written approval of the
manufacturer,
•for the time of overdue payment for previous services.
Sale Data: ___________ Serial number: _________________
Other manuals for Photometer FM-1
1
Table of contents
Other itaia Measuring Instrument manuals