English
8
2.1.1 Potential improper use
The device is not suitable for post-curing procedures outside of the dental field.
The following is considered as misuse of the system:
•Non-compliance with the specified intended use, intended operator specifications or
intended environment
•Modifications, maintenance and repairs without prior authorization by Ivoclar Vivadent
•Operation with inappropriate parameters
Improper use of the devices can lead to the following:
•Danger for patients and operating personnel
•Impairment of the operability of the devices
Ivoclar Vivadent assumes no liability for damage resulting from improper use.
Typical misapplications of the device to be avoided:
Improper use Consequences
Post-curing of non-approved materials Insufficient material properties in terms of
mechanics, colour, precision, biocompat-
ibility
Post-curing of printed objects / objects in
an articulator
•Articulator does not fit into the chamber,
door cannot be closed
•Device cannot be used
•Damage to the device
Post-curing of non-dental objects Insufficient material properties in terms of
mechanics, colour, precision, biocompat-
ibility due to different volume, weight, layer
thickness, materials, etc. of the objects
Incorrect cleaning and disposal •Skin irritations
•Environmental damage
•Damage to the device
Switching off the device by the user during
operation or opening the door whilst a
program is running
•Interruption or cancellation of the curing
process/program
•Objects not completely cured
•Insufficient material properties in terms
of mechanics, colour, precision,
biocompatibility
Note:
If a curing job in progress is cancelled, it
cannot be continued.
It is not possible to perform a post-curing
procedure after a program cancellation as
this would lead to an undesirable change in
the material properties.