ixtur MRP-45 User manual

23.7.2018
IXTUR OY | VARESPELLONTIE 6, FI-21500 KAARINA FINLAND, WWW.IXTUR.COM
Technical Specifications and
Instructions
PNEUMATIC MAGNET IXTUR MRP-45
Original User Manual

Ixtur Ltd / MRP-45
1
Content
Introduction and Warnings ............................................................................................................................ 2
Technical Specifications ................................................................................................................................ 3
Holding Force .................................................................................................................................................. 4
Installing the Device ....................................................................................................................................... 7
MRP-45 Lifting Adapter .................................................................................................................................. 9
Using the Ixtur MRP-45 Pneumatic Magnet ............................................................................................... 9
Maintenance and Inspections ..................................................................................................................... 10
Warranty ......................................................................................................................................................... 10
Patent Pending .............................................................................................................................................. 10
Declaration of Conformity ............................................................................................................................ 11

Ixtur Ltd / MRP-45
2
Introduction and Warnings
MRP-45 is a pneumatic magnet meant for attaching ferromagnetic parts in lifting, work holding or
automation applications. It is suitable for plates and round objects. The magnet’s state, attached or
detached, is changed pneumatically. MRP-45 pneumatic magnet doesn’t need compressed air to
keep its state. Air pressure is needed only to change the state from attached to detached or vice
versa.
Please read this manual carefully before deployment to ensure safe use of the device.
This device includes a source of magnetic field. A fluctuating magnetic field is produced
during the use in the proximity of the device.
This device produces a magnetic field that may interfere with medical devices such as
metallic implants or cardiac pacemakers. Do not use this device and keep out of close
proximity unless you have approval from the manufacturer of the medical device or from
your physician.
Do not disassemble or replace the parts of the device. Only pneumatic connectors are user
changeable.
The device must not be used for lifting or transporting people.
The permitted holding force must not be exceeded.
Always pay special attention to the factors that may decrease the holding force of the
magnet.
Do not operate a damaged or malfunctioning device.
The designer and manufacturer of the system, which this device will be integrated into,
should always ensure that the complete system meets the local and national safety
regulations.

Ixtur Ltd / MRP-45
3
Technical Specifications
Dimensions:
- Length: 80.0 mm
- Width 54.8 mm
- Height: 80.0 mm (height with lifting adapter 100.0 mm)
- Weight: 1790 g (weight with lifting adapter 2000 g)
Capacity:
- Rated holding force: 1324 N (S355 S ≥ 12 mm plate)
- Rated holding force: 883 N (S355 cylinder ø ≥ 120 mm)
- Rated lifting capacity for plate with safety factor 3: 45 kg (S355 S ≥ 12 mm)
- Rated lifting capacity for solid round object with safety factor 3: 30 kg (ø ≥ 120 mm)
- Residual holding force flat: maximum 1200 g
- Residual holding force for other shapes:
MRP-45’s magnetic structure (similar to standard lever magnets) makes residual grip-
ping capacity sensitive to material and geometry of the lifted part. In extreme case, a
part up to 4 kg may stay connected to the magnet. This may happen just after the mag-
net has been turned OFF. When this behavior is critical to the application, the actual
parts need to be pretested with the MRP-45 magnet.
Environmental conditions:
- Operating temperature: 0 °C … +50 °C
- Storage temperature: -20 °C … +50 °C
- Humidity 0 % … 90 %
- IP67 (The device is protected from dust and temporary immersion up to 1 m)
Requirements for compressed air:
- Recommended pressure: 6.0 bar
- Functional range: 5.0 … 8.0 bar
- Water separation
- Particle filter ≤ 5µm
Information related to medical implants
- Safety distance for a person with active implanted device is 50 cm of air or other non-
magnetic material. The distance is measured from the magnet or magnetic material
attached to it.
o Maximum static magnetic field of MRP-45 is 200 mT on the gripping surface.
o Interference with active implanted devices, e.g. cardiac pacemakers – Action level for expo-
sure to static magnetic fields is 0,5 mT [Directive 2013/35/EC].

Ixtur Ltd / MRP-45
4
Holding Force
Holding force and rated lifting capacity of MRP-45 can only be applied when the forces are per-
pendicular to the magnet's gripping surface.
The following factors affect magnet’s holding force:
- Material properties
- Thickness of the load
- Distance between gripping surface and attached load (airgap)
- Object diameter when lifting round part
Holding force and lifting capacity of MRP-45 pneumatic magnet are rated with steel EN S355. Most
other steel materials produce lower holding force. Also hardened steel produces lower holding
force.
Material’s capacity to conduct magnetic field depends on its thickness. A thinner material has
weaker capacity compared to a thicker material. Increasing material thickness increases the mate-
rial’s capability to conduct magnetic field until the magnet’s specific maximum value is reached.
See figure 2: MRP-45 holding force in proportion to plate thickness. Figure 2 is made for plate, but
same principle is valid also for pipe wall thickness when round parts are lifted.
Any material that separates the magnet's gripping surface from the surface of the load reduces the
total magnetic holding force. Reduction can be caused by non-magnetic materials or by materials
with weak magnetic qualities, such as, zinc or paint or similar coating, rust, frost or air. The dis-
tance between the surfaces caused by materials above is commonly called as airgap.
Surfaces must be clean and the airgap zero to achieve rated holding force. The load surface must
also be flat and it must occupy the gripping surface of the device completely. See figure 3: MRP-45
holding force in proportion to airgap.
In case of round parts, the magnet’s holding force depends also on part’s diameter. Rated holding
force can be achieved with ø ≥ 120 mm. See figure 4: MRP-45 holding force in proportion to round
part diameter. Round part is assumed to be a solid material.
Figure 1. The gripping surface and magnetic poles of MRP-45.

Ixtur Ltd / MRP-45
5
Figure 2. MRP-45 holding force in proportion to plate thickness
Figure 3. MRP-45 holding force in proportion to airgap
0
200
400
600
800
1000
1200
1400
1600
0 2 4 6 8 10 12 14 16 18 20
Holding force [N]
Plate thickness [mm]
0,0
10,0
20,0
30,0
40,0
50,0
60,0
70,0
80,0
90,0
100,0
0 0,1 0,2 0,3 0,4 0,5
Holding force %
Airgap [mm]

Ixtur Ltd / MRP-45
6
Figure 4. MRP-45 holding force in proportion to diameter of a solid round object
The holding forces specified in this document are valid for steel EN S355. Most other
magnetic materials result in lower holding force.
The rated lifting capacities are valid for solid cylinders Ø ≥ 120 mm or plates of 12 mm
thick. Pipes or thinner plates lead to lower lifting capacity.
When lifting a piece from a stack, also the piece below the lifted piece can be attached
to the magnet.
The forces caused by acceleration must be considered. For example, sudden moves
cause stronger forces than slow and smooth moves. Sudden moves may drop the load.
Before magnet attachment, make sure that the magnet's gripping surface and the load
material surface are clean of foreign materials to minimize the airgap that reduces the hold-
ing force.
Due to residual magnetism, the magnet's lifting surface may contain small magnetic
particles or pieces which reduce the holding force.
The load material surface may bend, which produces an airgap and reduces the hold-
ing force. Especially thin materials may cause this effect.
Do not lift the load higher than necessary. Unnecessary elevation of the load during
lifting increases the risk of damage or injury.
0
100
200
300
400
500
600
700
800
900
1000
0 20 40 60 80 100 120 140 160 180 200
Holding force [N]
Diameter [mm]

Ixtur Ltd / MRP-45
7
Installing the Device
MRP-45 pneumatic magnet can be mounted either from side(s) or from top. Mounting is done us-
ing M6 screws (4 pcs.) according to Figure 4. The pneumatic connections are made with 4.0 mm
hose.

Ixtur Ltd / MRP-45
8
Figure 5. Mounting dimensions of MRP-45.
When the MRP-45 magnet is turned ON, a magnetic field is generated between the steel plates of
the magnet.
The installation construction around the magnet must not short-circuit the magnetic field. The parts
of the installation that connect the MRP-45 steel plates together must be made of non-magnetic
material, e.g. aluminium or stainless steel.
Figure 6. Magnetic field of MRP-45.
When combining multiple MRP-45 magnets in an installation, all magnets have to be positioned in
the same direction.
Figure 7. Installation of multiple MRP-45 magnets together.

Ixtur Ltd / MRP-45
9
MRP-45 Lifting Adapter
A lifting adapter is available for MRP-45 magnet. The adapter can be used for example when
MRP-45 is connected to a crane hook. The lifting adapter is made of aluminium.
Figure 6. MRP-45 with the lifting adapter
Using the Ixtur MRP-45 Pneumatic Magnet
The front (logo side) and back steel plates of MRP-45 have a pneumatic connector. When the air
pressure in the back side pneumatic connection is 6.0 bars higher than in the front side connection,
the magnet is set ON (attach the metal). When the air pressure in the front side pneumatic connec-
tion is 6.0 bars higher than in the back side connection, the magnet is set OFF (detach the metal).
The same pressure in both air hose contacts keeps the magnet’s current state. Even if the pres-
sure is lost, the magnet keeps the current state. For more accurate air pressure information, look at
the Technical Specifications of this user’s manual.

Ixtur Ltd / MRP-45
10
Figure 6. MRP-45 Pneumatic connectors
The magnetic holding force of MRP-45 magnet weakens, if magnetic field is short circuited outside
gripping surface. This may happen if the steel plates are connected to each other with unwanted
magnetic material.
If there is a chance that the magnetic field is short circuited by external magnetic part (besides the
load) during the gripping of the load, the MRP-45 magnet’s steel plates should be shielded by non-
magnetic material from such direction(s).
Short circuiting the magnetic field between the steel plates of MRP-45 reduces the holding
force and may cause the lifted object to drop. The load must be the only magnetic material in con-
tact with both plates.
Maintenance and Inspections
MRP-45 pneumatic magnet is maintenance free.
If MRP-45 pneumatic magnet has been dropped or it has been exposed to severe impact, the
magnet must be inspected for any damages.
Do not use the device if you notice any damage or malfunction.
Warranty
The MRP-45 Magnet has a 12 months warranty from the date of delivery. The warranty covers
material and manufacturing defects. Warranty does not cover defects caused by a misuse of the
device against the regulations and instructions in this manual. The warranty does not cover normal
wear and tear.
Patent Pending
Several patent applications are pending related to MRP-45 and Ixtur magnet technology.

Ixtur Ltd / MRP-45
11
Declaration of Conformity
Declaration of incorporation of partly completed machinery
Directive 2006/42/EC
We hereby declare that this partly completed machinery, Pneumatic Magnet MRP-45, complies
with the essential requirements of the Directive 2006/42/EC on Machinery and the following har-
monized standards have been applied and fulfilled:
- EN ISO 12100:2010
- ISO/TR 14121-2:2007
- EN 13155:2003+A2:2009
The technical documents of partly completed machinery is compiled in accordance with section B
of Annex VII of Directive 2006/42/EC, that can be transmit, in response to a reasoned request by
the national authorities.
Partly completed machinery must not be put into service until the final machinery into which it is to
be incorporated has been declared in conformity with the provisions of the Directive 2006/42/CE on
Machinery.
This manual suits for next models
2
Table of contents
Other ixtur Industrial Equipment manuals

ixtur
ixtur MAP-40 Instruction sheet

ixtur
ixtur MRP-31FK User manual

ixtur
ixtur MRP-20NK User manual

ixtur
ixtur MRP-28NK User manual

ixtur
ixtur MRP-31F User manual

ixtur
ixtur MRP-130F User manual

ixtur
ixtur MAP-6 Instruction sheet

ixtur
ixtur MRP-100NK User manual

ixtur
ixtur MAP-180 Instruction sheet

ixtur
ixtur MRP-42K User manual