manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. J. MITRA & CO. PVT. LTD.
  6. •
  7. Medical Equipment
  8. •
  9. J. MITRA & CO. PVT. LTD. DENGUE NS1 Ag Parts list manual

J. MITRA & CO. PVT. LTD. DENGUE NS1 Ag Parts list manual

WARNING FOR USERS
CAUTION: ALL THE SAMPLES TO BE TESTED SHOULD BE HANDLED AS THOUGH
CAPABLE OF TRANSMITTING INFECTION. NO TEST METHOD CAN OFFER
COMPLETE ASSURANCE THAT HUMAN BLOOD PRODUCTS WILL NOT TRANSMIT
INFECTION.
1. The use of disposable gloves and proper biohazardous clothing is STRONGLY
RECOMMENDED while running the test.
2. In case there is a cut or wound in hand, DO NOT PERFORM THE TEST.
3. Do not smoke, drink or eat in areas where specimens or kit reagents are being
handled.
4. Tests are for in vitro diagnostic use only and should be run by competent person only.
5. Do not pipette by mouth.
6. All materials used in the assay and samples should be decontaminated before
disposal or by autoclaving at 121ºC at 15psi for 60 min. They should be disposed off
in accordance with established safety procedures and standard biosafety guidelines
for handling & disposal of potentially infective material.
7. Wash hands thoroughly with soap or any suitable detergent, after the use of the kit.
Consult a physician immediately in case of accident or contact with eyes, in the event
that contaminated material are ingested or come in contact with skin puncture or
wounds.
8. Spills should be decontaminated promptly with Sodium Hypochlorite or any other
suitable disinfectant.
9. Assay Buffer contains Sodium Azide as a preservative. If these material are to be
disposed off through a sink or other common plumbing systems, flush with generous
amounts of water to prevent accumulation of potentially explosive compounds. In
addition, consult the manual guideline "Safety Management No. CDC-22",
Decontamination of Laboratory Sink Drains to remove Azide salts" (Centre for
Disease Control, Atlanta, Georgia, April 30, 1976).
PRECAUTIONS
In order to obtain reproducible results, the following instructions must be followed:
1. Do not use the kit beyond the expiry date.
2. Do not mix reagents from different batches.
3. Do not open the foil pouch until it attains room temperature.
4. Do not re-use the test cartridge.
5. Follow the given test procedure and storage instructions strictly.
6. Do not paste any sticker or write anything on the QR-Code as this will lead to
erroneous result.
7. Do not temper the QR-Code as this will lead to erroneous result.
8. Do not touch the membrane with the pipette tip.
Important Note: Dengue NS1 Ag Quanti Card is only operational in conjection with
iQuant Analyzer.
SAMPLE / SPECIMEN COLLECTION AND STORAGE
1. Serum / plasma samples may be used with this test.
2. If serum / plasma specimens cannot be tested immediately, they should be
refrigerated at 2-8ºC. For storage for more than 3 days, freeze the specimen at -20ºC
or below.
3. Repeated freezing and thawing of the specimen should be avoided.
4. Specimens containing precipitate or particulate matter may yield inconsistent test
results. Such specimens must be centrifuged at 10,000 rpm for 15 minutes and the
clear supernatant should only be used for testing.
5. The use of hemolytic, lipaemic, icteric or bacterially contaminated specimens should
be avoided as it may lead to erroneous results.
BEFORE YOU START
The Assay Buffer Solution provided in the kit has closed nozzle and screw cap with pin
(outside). Before using Assay Buffer, keep the vial vertically straight and tap down gently on
the working platform, so that Assay Buffer comes down at the bottom of the vial. To
orifice/puncture the closed nozzle, follow the instruction as illustrated below:
INTRODUCTION
Dengue virus is a flavivirus found largely in areas of the tropic and sub-tropics. There are
four distinct but antigenically related serotypes of dengue viruses, and transmission is by
mosquito, prinicipally Aedes aegypti and Aedes albopictus.
The mosquito-borne dengue viruses (serotype 1-4) cause dengue fever, a severe flu-like
illness. The disease is prevalent in third world tropical regions and spreading to sub-tropical
developed countries - including the United States. WHO estimates that 50-80 million cases
of dengue fever occur worldwide each year, including a potentially deadly form of the
disease called dengue haemorrhagic fever (DHF) and dengue shock syndrome (DSS).
Primary infection with dengue virus results in a self-limiting disease characterized by mild to
high fever lasting 3 to 7 days, severe headache with pain behind the eyes, muscle and joint
pain, rash and vomiting. Secondry infection is the more common form of the disease in
many parts of Southeast Asia and South America. IgM antibodies are not detectable until 5-
10 days in case of primary dengue infection and until 4-5 days in secondary infection after
the onset of illness. IgG appear after 14 days and persist for life in case of primary infection
and rise within 1-2 days after the onset of symtoms in secondary infection. This form of the
disease is more serious and can result in DHF and DSS. The major clinical symptoms can
include high fever, haemorrhagic fever, and circulatory failure, and the fatality rate can be as
high as 40%. Early diagnosis of DSS is particularly important, as patients may die within 12
to 24 hours if appropriate treatment is not administered.
Primary dengue virus infection is characterized by elevations in specific NS1 antigen levels
0 to 9 days after the onset of symptoms; this generally persists upto 15 days. Earlier
diagnosis of Dengue reduces risk of complication such as DHF or DSS, especially in
countries where dengue is endemic.
INTENDED USE
Dengue NS1 Ag Quanti Card is a sensitive immuno- chromatographic test for the qualitative
detection of Dengue NS1 Antigen in human Serum/ Plasma with iQuant Analyzer . This test
is for in vitro diagnostic use only and is intended as an aid in the earlier diagnosis of dengue
infection.
PRINCIPLE (ANTIGEN-ANTIBODY REACTION)
Dengue NS1 Ag Quanti Card is a fluorescence immunoassay based on sandwich principle.
The test area is coated with specific anti-dengue NS1 Ag antibodies. When a sample is
added to the cartridge, Dengue NS1 antigen if present will form a complex with another
specific NS1 antibodies conjugate to fluorochorme. On addition of Assay Buffer, this
complex migrates along the nitrocellulose membrane to the test region and forms an
antibody-antigen-antibody fluorescence immunocomplex. The result will be displayed by i-
Quant Analyzer.
MATERIALS PROVIDED
Dengue NS1 Ag Quanti Card kit contains following components to perform the assay:
a) Dengue NS1 Ag Quanti Card Device b) Assay Buffer
c) Instruction Manual
KIT PRESENTATION
24 Test Pack 48 Test Pack 96 Test Pack
STORAGE AND STABILITY
The kit should be stored at 2-8ºC in the coolest and driest area available. Expiry date on the
kit indicates the date beyond which kit and its components should not be used. Dengue NS1
Ag Quanti Card should not be frozen and must be protected from exposure to humidity.
DESCRIPTION OF SYMBOLS USED
The following are graphical symbols used in or found on J. Mitra diagnostic products
and packing. These symbols are the most common ones appearing on medical devices and
their packing. They are explained in more detail in the British and European Standard
EN ISO 15223-1:2016.
Manufactured By In vitro diagnostic
medical device
No. of tests See Instruction for use
Lot Number Temperature
Batch Number Limitation
Manufacturing Date Caution, see instruction for use
Expiry Date Catalogue Number
Do not use if package Keep away from sunlight
is damaged
Single use only
Fluorescence immunoassay (antigen-antibody) for
qualitative detection of Dengue NS1 Antigen in Human Serum/ Plasma
DENGUE NS1 Ag
QUANTI CARD
2ºC
8ºC
TEST PROCEDURE
1. Bring the complete kit and specimen to be tested to room temperature
prior to testing.
2. Remove the test cartridge from the foil pouch prior to use and place it on a flat and dry
surface. The test should be performed
immediately after removing the test card from
the foil pouch.
3. Label the test cartridge with patient's name or
identification number. Do not write on QR code.
4. Add 50µl Serum/ Plasma sample using
micropipette onto the sample pad in the sample
well 'S'. Care should be taken to avoid any
spillage on the QR Code.
NOTE : Make sure that the sample from the
micropippete has been completely transferred to
the sample pad.
5. Add 2 drops of the Assay Buffer in the buffer
well 'B'.
6. Allow the reaction to occur for 30 minutes. In
the meantime enter the patient's details in the
iQuant analyzer testing window and select the
NS1 test from the pop down menu in the
testing window of the iQuant analyzer.
7. Insert the test cartridge into the i-Quant Analyzer with arrow ( ) marked side on the
top of cartridge facing towards the analyzer and press RUN icon. Note down the value
displayed on the screen of i-Quant Analyzer & interpret the result as mentioned below.
8. Discard the Dengue NS1 Ag Quanti Card immediately after reading results at 30
minutes considering it to be potentially infectious.
Important Note: Do not read results after 30 minutes.
INTERPRETATION OF RESULTS
The iQuant Analyzer will display results as Reactive, Equivocal or Non-Reactive as follows:
NS1 Antigen Reactive: >1.1 U and above: interpret the result as Dengue NS1 Antigen
reactive.
Equivocal: >0.9 to ≤1.1 U: Interpret the result as equivocal. Repeat the test after
centrifuging the sample at 5000 rpm for 20 minutes. Even after repeating the test, if result
comes equivocal, further test the sample with alternative method or collect another sample.
Non-Reactive: Below ≤0.9 U: Interpret the result as Dengue NS1 Antigen non-reactive.
LIMITATIONS AND INTERFERENCES
1. The test is for in vitro diagnostic use only.
2. This test detects the presence of Dengue NS1 antigen in the specimen and should not
be used as the sole criteria for the diagnosis of Dengue virus infection.
3. Serological cross-reactivity across the Flavivirus group (Dengue virus, St. Louis
encephalitis, Japanese encephalitis, West Nile and yellow fever virus) is common.
4. As with all diagnostic tests, all results must be corelated with other clinical findings. If
the test result is negative and clinical symptoms persist, additional follow-up testing
using other clinical methods is recommended. A negative result at any time does not
preclude the possibility of an early infection of Dengue virus.
5. This is only a screening test. Therefore, isolation of virus, antigen detection in fixed
tissues, RT-PCR and more specific alternative diagnosis method must be used in
order to obtain a confirmation of dengue virus infection.
PERFORMANCE CHARACTERISTICS OF DENGUE NS1 Ag QUANTI CARD TEST
The kit has been evaluated in-house with the known panel of fresh as well as frozen Dengue
NS1 antigen positive and Negative samples. The performance of the test kit was evaluated
and compared with the a license commercially available ELISA test kit. The samples
included cross-reacting samples; Epstein-Barr virus, Malaria, Rheumatoid factor,
Leptospirosis, Japanese encephalitis, yellow fever and West Nile viruses. Following is the
in-house evaluation.
Sample Type No. of Result of Dengue Ns1
Samples tested licensed test Ag Quanti Card
Dengue NS1 Antigen 2070 2070 2069
Negative
Dengue Antigen Positive 255 255 255
Sensitivity : Specificity : 100% 99.95%
Precision:
Intra assay precision (Reproducibility)
Within run (Intra assay) precision have been determined by testing 10 replicates of two
negative and four dengue NS1 antigen positive samples; 3 weak and 1 medium. The C.V.
(%) of all the samples were within 10% of the unit.
Inter assay precision (Reproducibility)
Between run (Inter assay) precision have been determined by testing 10 replicates of two
negative and four dengue NS1 antigen positive samples; 3 weak and 1 medium in 10
sequential days. The C.V. (%) of all the samples were within 10% of the unit.
LIMITED EXPRESSED WARRANTY DISCLAIMER
The manufacturer limits the warranty to the test kit, as much as that the test kit will function
as an in vitro diagnostic assay within the limitations and specifications as described in the
product instruction-manual, when used strictly in accordance with the instructions
contained therein. The manufacturer disclaims any warranty expressed or implied including
such expressed or implied warranty with respect to merchantability, fitness for use or
implied utility for any purpose. The manufacturer’s liability is limited to either replacement of
the product or refund of the purchase price of the product and in no case liable to claim of
any kind for an amount greater than the purchase price of the goods in respect of which
damages are likely to be claimed. The manufacturer shall not be liable to the purchaser or
third parties for any injury, damage or economic loss, howsoever caused by the product in
the use or in the application there of.
TROUBLE SHOOTING
PROBLEM POSSIBLE CAUSE SOLUTION
1. False negative a) Low sample volume Repeat the test with proper
along with less volume volume of samples and/or
of Assay Buffer used. Assay Buffer.
2. False positive a) Less volume of Assay Repeat the test with proper
Buffer added/used. volume of Assay Buffer
BIBLIOGRAPHY OF SUGGESTED READING
1. Dengue haemorrhagic fever: Diagnosis, Treatment, Prevention and Control. WHO
2009 edition.
2. Evaluation of diagnostic test: Dengue, Rosanna W. Peeling, Harrey Artsob etal.
(2010). Nature reviews.
3. Use of dengue Ns1 antigen for early diagnosis of dengue virus infection. Kassim FM;
Izate MN, etal. (2011) Southeast Asian J. Trop. Med. Public Health. May; 42(3); 562-
9.
4. Dengue NS1 antigen detection: A useful tool in early diagnosis of dengue virus
infection. S. Datta, C Wattal (2010). Indian Journal of Medical Microbiology, vol 28,
No. 2; 107-110.
S
B
Fig. (a)
Fig. (e)
S
B
MN/DAQ/057 VER-01 R- 02
Rev. Date: Jan.-18
J. MITRA & CO. PVT. LTD.
A 180-181, Okhla Indl. Area, Phase-1, New Delhi-110 020, INDIA
Ph: +91-11-47130300, 47130500, 26818971-73
e-mail: [email protected] Internet: www.jmitra.co.in
For in-vitro diagnostic use only, not for medicinal use
Fig. i
close vial provided
in the kit
Fig. v
screw the cap
tightly after
use
Fig. iii
invert the cap &
press on the closed
nozzle
Fig. iv
give a half
turn twist to
puncture the nozzle
Fig. ii
open the cap
Fig. (f)
S
B

Popular Medical Equipment manuals by other brands

Storz 110 Series manual

Storz

Storz 110 Series manual

Quart Healthcare Elite Turn Setup Notes

Quart Healthcare

Quart Healthcare Elite Turn Setup Notes

GE IV-908 instructions

GE

GE IV-908 instructions

ChoiceMMed MD300CF3 user manual

ChoiceMMed

ChoiceMMed MD300CF3 user manual

Allergan TrueTear Patient guide

Allergan

Allergan TrueTear Patient guide

broda 100-PS operating manual

broda

broda 100-PS operating manual

Dornoch UL-QD2800 Technical manual

Dornoch

Dornoch UL-QD2800 Technical manual

Konica Minolta SRX-101A Operation manual

Konica Minolta

Konica Minolta SRX-101A Operation manual

Cyclomedica Technegas TechnegasPlus user manual

Cyclomedica

Cyclomedica Technegas TechnegasPlus user manual

Sutter CURIS flow Instructions for use

Sutter

Sutter CURIS flow Instructions for use

iFog COVIX user manual

iFog

iFog COVIX user manual

tecnodent SERENITY NEXT operating instructions

tecnodent

tecnodent SERENITY NEXT operating instructions

iMerciv BuzzClip quick start guide

iMerciv

iMerciv BuzzClip quick start guide

SMARTfit Mini user manual

SMARTfit

SMARTfit Mini user manual

Gate Bure Standard GAS 56-311 Instructions for use

Gate

Gate Bure Standard GAS 56-311 Instructions for use

Otto Bock 50K4-1 Agilium Freestep Instructions for use

Otto Bock

Otto Bock 50K4-1 Agilium Freestep Instructions for use

Tristel Stella System A user guide

Tristel

Tristel Stella System A user guide

OFA LION X30A Instructions for use

OFA

OFA LION X30A Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.