Jamar Digital Pinch Gauge User manual

Digital Pinch Gauge
Instruction Booklet
1`

11 12
A
B
C
LBS KG
Z E R O
1
2
3
4
5
678
9
10
13
Appearance
2
11 12
A
B
C
LBS KG
Z E R O
1
2
3
4
5
678
9
10
13
Appearance
2

1.Function buttons
(1) Power button:Press one time to turn on.
Press again to turn off.
(2) Up button: Press one time to save the
▲
current record and jump to the last record
and display the value of that record at time of
testing.
When pressed down for 5-6 seconds, it will
enter the state of unit conversion; the figure of
KG or LBS will flash in turn when pressed once
again, and will confirm the unit mode when the
power button is pressed again .It will display your
unit when turned on again.
(3) ▼ Down button: Press one time to save the
value of this record, skip to the next record and
display the value of the record at time of testing,
and display the next record in case of browse-up.
(4) ZERO Clear button: Press one time to set the
current record to the value of 0.0; continue to
hold down until the emergence of the figure
AL.CL,and then all the records in the instrument
will be reset to 0.0, and it will automatically go
back to the first record number.
2. Display screen
(6) Record number: "1" represent for the first record,
"99" for the 99 records, "0" for the first 100 records
(7) Reading: shows the data of the measurement.
(10) Unit: KG represents for kilogram, LBS for pound.
3. Testing keys
3
1.Function buttons
(1) Power button:Press one time to turn on.
Press again to turn off.
(2) Up button: Press one time to save the
▲
current record and jump to the last record
and display the value of that record at time of
testing.
When pressed down for 5-6 seconds, it will
enter the state of unit conversion; the figure of
KG or LBS will flash in turn when pressed once
again, and will confirm the unit mode when the
power button is pressed again .It will display your
unit when turned on again.
(3) ▼ Down button: Press one time to save the
value of this record, skip to the next record and
display the value of the record at time of testing,
and display the next record in case of browse-up.
(4) ZERO Clear button: Press one time to set the
current record to the value of 0.0; continue to
hold down until the emergence of the figure
AL.CL,and then all the records in the instrument
will be reset to 0.0, and it will automatically go
back to the first record number.
2. Display screen
(6) Record number: "1" represent for the first record,
"99" for the 99 records, "0" for the first 100 records
(7) Reading: shows the data of the measurement.
(10) Unit: KG represents for kilogram, LBS for pound.
3. Testing keys
3

Specifications
1). Capacity: 60 LBS
2). Division: 0.1Lb/0.1KG
3). Voltage: 3 V
4). Battery: 1 x CR2032 lithium cell
5). Number of records: 100 PCS
6). Temperature: 0-40 C
7). Measurement: 103 x 54 x 18 mm
8). Weight: 145 g
Operations
At time of purchase, the battery has been
already installed.
1. Testing operation:
(8) Top metal key
(11) Bottom metal key
4. Battery
(12) Battery cover
(13) Battery: 1 X CR2032 lithium cell
5. Others
(5) Hole for the wrist band to pass through
(9) Logo
4
Specifications
1). Capacity: 60 LBS
2). Division: 0.1Lb/0.1KG
3). Voltage: 3 V
4). Battery: 1 x CR2032 lithium cell
5). Number of records: 100 PCS
6). Temperature: 0-40 C
7). Measurement: 103 x 54 x 18 mm
8). Weight: 145 g
Operations
At time of purchase, the battery has been
already installed.
1. Testing operation:
(8) Top metal key
(11) Bottom metal key
4. Battery
(12) Battery cover
(13) Battery: 1 X CR2032 lithium cell
5. Others
(5) Hole for the wrist band to pass through
(9) Logo
4

55
1).Press the power button to turn it on. The first
display is "HELO", and then the unit displays the
record number "1" and reading "0.0", if not 0.0, pre
-ss ZERO to set it to 0.0. After re-powering the instru
-ment, it will show the record and the corresponding
data that was saved before it was shut off the last
time.
2).Squeeze the metal keys with your fingers. The
d a t a w i l l g o u p w i t h y o u r f i n g e r p i n c h
increase, and the unit will show your maximum finger
pinch.
3). Press down button ▼ to save the current record and
skip to the next record.
Press up button ▲ to save the current record and
skip to the last record.
4). Press clear button ZERO to set the current record
to 0.0.
5). Press power button to shut off the unit.
LBS
e.g
LBS
e.g
LBS
e.g
LBS
e.g
LBS
e.g

2. Query operation
(1) Press ▲ up button to browse the last record.
(2) Press ▼ down button to browse the next record.
3. Unit conversion
Battery Replacement
When the screen displays "LO", this indicates
the device has a low battery. Do not replace the
battery while the unit is on. Battery replacement steps:
1). Make sure unit is off. Press down the battery
cover top slightly and pull back, the cover will
slide open. See below:
2). Push the golden metal chip outside to make the
cell bounce up; with small tweezers, remove
the exhausted cell. See below:
It will enter the state of unit conversion when
pressed down for 5-6 seconds, the figure of KG
or LBS will flash in turn when pressed once again, co
-nfirm the unit mode when press the power
button .It will display your unit when turned on again.
6
2. Query operation
(1) Press ▲ up button to browse the last record.
(2) Press ▼ down button to browse the next record.
3. Unit conversion
Battery Replacement
When the screen displays "LO", this indicates
the device has a low battery. Do not replace the
battery while the unit is on. Battery replacement steps:
1). Make sure unit is off. Press down the battery
cover top slightly and pull back, the cover will
slide open. See below:
2). Push the golden metal chip outside to make the
cell bounce up; with small tweezers, remove
the exhausted cell. See below:
It will enter the state of unit conversion when
pressed down for 5-6 seconds, the figure of KG
or LBS will flash in turn when pressed once again, co
-nfirm the unit mode when press the power
button .It will display your unit when turned on again.
6

4). Slide the cover back in locked position.
See below:
Caution
This instrument should be kept away from all
liquids. Avoid dropping or throwing at all times.
Golden metal chip
1
2
cell
golden metal chip
3). Insert the new cell below the golden metal chip
with the + marking up, and then press it down to
fasten.See below:
7
4). Slide the cover back in locked position.
See below:
Caution
This instrument should be kept away from all
liquids. Avoid dropping or throwing at all times.
Golden metal chip
1
2
cell
golden metal chip
3). Insert the new cell below the golden metal chip
with the + marking up, and then press it down to
fasten.See below:
7
D
istributed by:
Performance
H
ealth Supply Inc.
28100 Torch Parkway, Ste. 700
Warrenville, IL 60
555
, USA
Phone: 800-323-
55
47
Fax: 800-
5
47-4333

88
D
istributed by:
Performance
H
ealth Supply Inc.
28100 Torch Parkway, Ste. 700
Warrenville, IL 60
555
, USA
Phone: 800-323-
55
47
Fax: 800-
5
47-4333
CEpartner4U
Esdoornlaan 13
3951 DB Maarn
The Netherlands
www.cepartner4u.eu

Character display signification Operation
Clear all test data
to 0.0
Remove your finger from the Zero
button immediately,wait till 0.0 occurs
Initial data are missed
Low battery Replace a new lithium cell
Over load
Turn off the gauge and turn on again
Run error
Notice
remove the lithium cell
and then re-assemble it
LBS KG OZ
remove the lithium cell
and then re-assemble it
Character display signification Operation
Clear all test data
to 0.0
Remove your finger from the Zero
button immediately,wait till 0.0 occurs
Initial data are missed
Low battery Replace a new lithium cell
Over load
Turn off the gauge and turn on again
Run error
Notice
remove the lithium cell
and then re-assemble it
LBS KG OZ
remove the lithium cell
and then re-assemble it
Character display signification Operation
Clear all test data
to 0.0
Remove your finger from the Zero
button immediately,wait till 0.0 occurs
Initial data are missed
Low battery Replace a new lithium cell
Over load
Turn off the gauge and turn on again
Run error
Notice
remove the lithium cell
and then re-assemble it
LBS KG OZ
remove the lithium cell
and then re-assemble it

2). Push the golden metal chip outside to make the
cell bounce up; with small tweezers, remove
the exhausted cell. See below:
See below:
Caution
This instrument should be kept away from all
liquids. Avoid dropping or throwing at all times.
Golden metal chip
2). Push the golden metal chip outside to make the
cell bounce up; with small tweezers, remove
the exhausted cell. See below:
See below:
Caution
This instrument should be kept away from all
liquids. Avoid dropping or throwing at all times.
Golden metal chip
Battery Replacement
N o t i c e w h e n t h e s c r e e n d i s p l a y s " L O " , t h i s
indicates the device has a low battery. Replace it.Do not
replace the battery while the unit is on.
1). Make sure unit is off. Press down the battery
cover top slightly and pull back, the cover will
slide open. See below:
4). Slide the cover back in locked position.
3). Insert the new cell below the golden metal chip
with the + marking up, and then press it down to
fasten.See below:
Battery replacement steps:
WARNNING No touching any buttons while replacing a cell.
1
golden metal chip
2
cell
Battery Replacement
N o t i c e w h e n t h e s c r e e n d i s p l a y s " L O " , t h i s
indicates the device has a low battery. Replace it.Do not
replace the battery while the unit is on.
1). Make sure unit is off. Press down the battery
cover top slightly and pull back, the cover will
slide open. See below:
4). Slide the cover back in locked position.
3). Insert the new cell below the golden metal chip
with the + marking up, and then press it down to
fasten.See below:
Battery replacement steps:
WARNNING No touching any buttons while replacing a cell.
1
golden metal chip
2
cell
Table of contents
Popular Medical Equipment manuals by other brands

COOK Medical
COOK Medical I-IGTCFS-426-08 Instructions for use

Medela
Medela Invia Series Instructions for use

Synthes
Synthes CODMAN NEURO CODMAN CERTAS Plus manual

Shenzhen Mindray Bio-Medical Electronics
Shenzhen Mindray Bio-Medical Electronics BeneHeart D6 Operator's manual

Leica BIOSYSTEMS
Leica BIOSYSTEMS CM3600 XP Instructions for use

Lightmed
Lightmed LightLas 532/670 Operator's manual

Bio Compression Systems
Bio Compression Systems SC-2004-OC operating instructions

Nellcor
Nellcor NPB-190 Service manual

Fukuda Denshi
Fukuda Denshi LX-8000 Series Operation manual

FHSurgical
FHSurgical C-MAX Operator's manual

I-Tech
I-Tech Power Q-6000 Plus user manual

Clarity
Clarity Platinum Operator's manual