manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. janitza
  6. •
  7. Measuring Instrument
  8. •
  9. janitza UMG 605US User manual

janitza UMG 605US User manual

Doc No. 1.043.002.1.b
Power Quality Analyser
UMG 605US
Operating manual and
technical data
Product No. 33.03.129
Janitza electronics GmbH
Vor dem Polstück 1
D-35633 Lahnau
Support Tel. 0049 6441 9642-22
Fax 0049 6441 9642-30
e-mail: [email protected]
Internet: http://www.janitza.com
www.janitza.com
Profibus
Ethernet
Power supply voltage
Temperature
measurement
input
4 x current
measurement
4 x voltage
measurement
RS232 RS485 2 digital inputs
2 digital outputs
A B
2
UMG 605US
General Information 4
Copyright 4
Protected trademarks 4
Liability disclaimer 4
Comments on the manual 4
Meaning of the symbols used 5
Inspection on receipt 6
Scope of supply 7
Available accessories 7
Notes on Use 8
Product Description 10
Intended use 10
UMG605US features 11
Methods of measurement 12
Operating concept 12
GridVis programming software 13
Three-phase 4-wire systems 14
Three-phase 3-wire systems 16
Use 18
Key functions 18
Concealed key (service) 18
Display mode 19
Programming mode 20
Display password 21
Homepage password 21
Installation 22
Installed position 22
Power supply voltage 23
Current measurement 24
Ammeter 26
Direct measurement 28
Voltage measurement 29
Interfaces 31
RS485 33
RS485 profibus DP V0 slave 35
Digital inputs and outputs 37
Temperature measurement input 41
Putting into Service 42
Applying the power supply voltage 42
Frequency measurement 42
Applying the measuring-circuit voltage 43
Phase sequence 44
Applying the measuring-circuit current 44
Checking the energy measurement 45
Configuration 46
Current transformer ratio 46
Current measurement connection options47
Voltage transformer ratio 48
Voltage measurement connection options49
Interfaces 50
RS232 50
RS485 50
Ethernet 51
Profibus 52
Recordings 55
3
UMG 605US
System information 56
Overrange 56
Serial number 57
Date 57
Firmware release 57
Time 57
Service and maintenance 58
Repair and calibration 58
Front film 58
Battery 58
Firmware update 58
Service 58
Trouble shooting 59
Technical specifications 62
General information 62
Ambient conditions during operation 62
Transport and storage 62
Power supply voltage 63
Protection class 64
Inputs and outputs 65
Temperature measurement input 66
Interfaces 67
Function parameters 68
Specifications for
IEC 61000-4-30 Class S 71
Measuring inputs 72
Parameter list 74
Measured value displays 80
Declaration of conformity 81
Dimensioned drawings 82
UMG605US connection example 83
Quick Reference Instructions 84
4
UMG 605US
General Information
Copyright
This manual is subject to the statutory copy-
right protection provisions and may not be
photocopied, reprinted, reproduced as a
whole or in parts, by mechanical or electronic
means, nor copied or further published in any
other way whatsoever without the legally bind-
ing, written consent of
Janitza electronics GmbH,
Vor dem Polstück 1,
D 35633 Lahnau,
Germany.
Protected trademarks
All trademarks and their resulting rights belong
to the respective holders of these rights.
Liability disclaimer
Janitza electronics GmbH does not accept
any responsibility whatsoever for errors or
deficiencies within this manual and does not
undertake any obligation to keep the content
of this manual up to date.
Comments on the manual
We welcome your comments. If anything in
this manual seems unclear to you, please let
us know by sending an eMail to:
[email protected]
5
UMG 605US
m
c
Meaning of the symbols used
The following pictograms are used in this
manual:
Dangerous voltage!
Danger or risk of severe injury.
Disconnect the system and device
from the power supply before starting
the work.
PE (protective earth) terminal.
Important!
Please note and follow the
documentation. This symbol is
intended to warn you of possi-
ble hazards that can occur during
installation, starting up and use.
6
UMG 605US
m
m
m
Inspection on receipt
Fault free and safe use of this device requires
appropriate transport, proper storage, erection
and assembly as well as careful operation and
maintenance. If it can be assumed that safe
operation is no longer possible, the device
must be immediately taken out of service and
secured against being accidentally started
up.
The device must be unpacked and packed
with the usual care, without the use of force
and only using suitable tools. The devices
must be visually inspected for perfect
mechanical condition. Please also note and
follow the installation instructions enclosed
with the device.
It can be assumed that safe operation is no
longer possible if the device, e.g.
• has visible damage,
• no longer works, despite intact mains power
supply,
• has been exposed to unfavourable conditions
(e.g. storage outside the permissible climatic
limits without adjustment to the ambient
climate, condensation, or similar) for a lengthy
period or was exposed to unfavourable effects
or loads during transport (e.g. fall from a large
height even if there is no visible external
damage, or similar).
Please check the scope of supply for
completeness before you start installing the
device.
All supplied options and design
versions are described on the delivery
note.
The installation and start-up instruc-
tions also describe options which do
not belong to the scope of supply.
All screw-type terminals belonging to
the scope of supply are plugged into
the device.
7
UMG 605US
Product No. Name
21 01 102 Battery, lithium CR2450N, 3 V/540 mAh
08 02 427 RS232, Connection cable (UMG 605US – PC), 2m, 5 poles
1) Refer to delivery note for product number.
2) Design version.
Scope of supply
Number Product No. Name
1 52 16 xxx1) UMG605US XX2)
1 33 03 0xx Installation and start-up instructions
1 33 03 084 Quick guide (supplement to the operating instructions)
1 51 00 116 CD with the following content:
- “GridVis” programming software
- Functional descriptions, GridVis, UMG605US
- GSD file “u605US0c2d.GSD” for profibus DP V0
1 10 01 807 Screw-type terminal, plug-in, 2 pin
1 10 01 808 Screw-type terminal, plug-in, 3 pin
1 10 01 809 Screw-type terminal, plug-in, 5 pin
1 10 01 810 Screw-type terminal, plug-in, 6 pin
1 89 10 051 Slot-head screwdriver (0.40x2 mm), ESD
1 08 01 505 Patch cable, 2m, twisted, grey (UMG605US - PC/switch connection)
1 52 00 008 RS485, external terminating resistor, 120 ohm
Available accessories
8
UMG 605US
Notes on Use
Please read these operating instructions and
all other publications which have to be used
to work with this product (in particular for
installation, operation or maintenance).
Note and follow all safety instructions as well
as any warnings. If you do not follow the
instructions, personal injuries and/or damage
to the product could be the result.
Any unauthorised change or use of this device
which extends beyond the given mechanical,
electrical or other operating limits can cause
personal injuries and/or damage to the
product.
Any such unauthorised change is “misuse”
and/or “negligence” under the product's
warranty and therefore excludes the warranty
for cover of possible resulting injuries or
damage.
This device may be solely operated and
maintained by skilled persons.
Skilled persons are people who, on the basis
of their relevant training and experience, are
capable of identifying risks and avoiding
possible hazards which operation or
maintenance of the device can cause.
When using the device, any additional legal and
safety regulations required for the respective
use must be observed.
9
UMG 605US
m
c
m
Important!
If the device is not operated according
to the instruction manual, protection
is no longer ensured and the device
can cause hazards.
Conductors made of individual wires
must be fitted with wire end ferrules.
Only pluggable screw terminals with
the same number of poles (pins) and
of the same type may be plugged
together.
10
UMG 605US
Product Description
Intended use
The UMG605US is intended to be used for
the measurement and calculation of electrical
variables such as voltage, current, energy,
work, harmonic components, etc. in building
installations, at distribution boards, circuit-
breakers and busbar trunking systems.
Measuring-circuit voltages and currents must
originate from the same system.
The UMG605US is permanently installed in
control cabinets or small distribution boards.
It can be installed in any position.
The measurement results can be displayed,
stored and read out via serial interfaces and
further processed.
The voltage measurement inputs are designed
for measurements in low-voltage systems in
which rated voltages up to 300 V conductor
to earth and surge voltages of overvoltage
category III can occur.
The current measurement inputs of the
UMG605US are connected via external ../1A
or ../5A current transformers.
Mains failure detection
The mains failure detection takes place via the
voltage measurement inputs. The selection of
voltage measurement inputs can be configured
using the GridVis software.
Mains failure stored energy time
The UMG605US bridges the following mains
failures at the auxiliary voltage input:
Mains voltage Stored energy time
230V AC max 80ms
Measurement in medium and high-voltage
systems takes place with current and voltage
transformers. Special safety requirements
must be complied, which are not dealt with in
any greater detail here.
The UMG605US fulfils the test requirements
for use in industrial areas.
11
UMG 605US
UMG605US features
- Measurement in IT, TT and TN systems,
- 4 voltage measurement inputs, - 4 current measurement inputs,
- Continuous scanning of the voltage and current measurement inputs with 20kHz,
- Measurement of power quality according to DIN EN61000-4-30:2009 class S,
- Flicker meter according to DIN EN61000-4-15:2011 class F3,
- Analysis and evaulation acoording to DIN EN50160 with software GridVis, which belongs to the
contents of delivery,
- Measurement of harmonics and interharmonics (Uln, Ull, I) according to DIN EN61000-4-7,
- Measurement of audio remote frequence (U, I, P, Q),
- Real energy, measurement uncertainty class 0.5 for ../5A current transformers,
- Real energy, measurement uncertainty class 1 for ../1A current transformers,
- Fourier analysis 1st to 63th harmonic component for U, I, P (cons./supply) and Q (ind./cap.),
- Detection of events such us overvoltage, voltage drop, power failure and overcurrent,
- Datalogger / event memory (128MB Flashdisk),
- Registers more than 2400 measured values (readings) pro Messzyklus (200ms),
- Registering and storage of transients (>50 µs) mit bis zu 16.000 Abtastpunkten,
- 2 digital inputs, 2 digital outputs, temperature measurement input,
- LC display, 2 keys,
- Profibus DPV0,
- RS485 (modbus RTU, modbus master, BACnet (optional)),
- RS232,
- Ethernet; web server, EMAIL, BACnet (option),
- Programming own applications in Jasic,
- Operating temperature range -10°C .. +55°C,
- Mounting on top hat rails 35 mm.
12
UMG 605US
m
Operating concept
You can program the UMG605US and call up
measured values in several ways.
• Directly at the device using 2 keys and the
display. You can change the values in the
parameter list (see Appendix) and call up
the measured values from the measured
value displays.
• Via the GridVis programming software.
• In devices with an ethernet interface, via
the homepage of the UMG605US.
• Via the RS485 with the modbus protocol.
You can change and call up data with the
help of the modbus address list (is filed on
the enclosed data carrier).
Operation of the UMG605US via the integrated
display and the two keys only is described in
these operating instructions.
The GridVis programming software and the
homepage have their own “online help”.
Methods of measurement
The UMG605US measures continuously and
calculates all effective values over a 200 ms
interval.
Use the parameter list in the
appendix to these instructions for
programming at the UMG605US
and the modbus address list on the
data carrier included in the scope of
supply for programming via a serial
interface.
13
UMG 605US
PC UMG
605US
Ethernet
PC UMG
605US
RS232
RS485
PC UMG
605US
RS232
GridVis programming software
The UMG605US can be programmed and read
out using the GridVis®programming software
included in the scope of supply. This requires
a PC to be connected to the UMG605US via a
serial interface/ethernet.
GridVis features
• Programming the UMG605US
• Configuring recordings
• Reading out recordings.
• Storing data in a database
• Graphic display of measured values
• Programming customer-specific applica-
tions.
Fig. 13.2 Connection of a UMG605US to a PC
via an interface converter.
Interface
converter
Fig.13.3 Connection of a UMG605US to a PC
via the ethernet.
Fig. 13.1 Connection of a UMG605US to a PC
via an RS232 cable.
14
UMG 605US
Three-phase 4-wire systems
The UMG605US can be used in three-phase
4 conductor systems (TN, TT system) (50 Hz,
60 Hz) with earthed PEN conductor. The
bodies of the electrical system are earthed.
The conductor to neutral conductor voltage
may not exceed 300 V AC.
The UMG605US is only suitable for
environments in which the impulse voltage
withstand level of 4 kV (overvoltage category
III) is not exceeded.
Fig. 14.1 Table of rated voltages of the grid
suitable for the voltage inputs.
UL-N / UL-L
66 V / 115 V
120 V / 208 V
127 V / 220 V
220 V / 380 V
230 V / 400 V
240 V / 415 V
260 V / 440 V
277 V / 480 V
Maximum rated voltage
of the grid
15
UMG 605US
Fig.15.1 Block diagram, UMG605US in TN
system.
UMG605US
Earthing the
system DC
AC/DC
PE
230/400V 50/60Hz
L2
L3
N
L1
Auxiliary power
Voltage measurement
4M
4M
4M
4M
L1 L3L2 N
4M
L4
N
L1
240V
50/60Hz
16
UMG 605US
Three-phase 3-wire systems
The UMG605US can be used in unearthed
three-phase 3 wire systems (IT system). The
conductor to conductor voltage may not
exceed 480V AC (50 Hz, 60 Hz).
The UMG605US is only suitable for
environments in which the impulse voltage
withstand level of 4 kV (overvoltage category
III) is not exceeded.
In the IT system the neutral point (star point)
of the voltage generator is not earthed. The
bodies of the electrical system are earthed.
Earthing via high-resistance impedance is
allowed.
IT systems are only allowed in certain systems
with their own transformer or generator.
UL-L
66 V
115 V
120 V
127 V
200 V
230 V
240 V
260 V
277 V
347 V
280 V
400 V
415 V
440 V
480 V
Fig. 16.1 Table of rated voltages of the grid
suitable for the voltage inputs.
Maximum rated voltage
of the grid
17
UMG 605US
Fig.17.1 Block diagram, UMG605US in IT
system without N.
Fig.17.2 Block diagram, UMG605US in IT
system with N.
400V 50/60Hz
DC
AC/DC
L2
L3
Auxiliary power
Voltage measurement
4M
4M
4M
4M
L1 L3L2
4M
L4
Earthing the
system
Impedance
L1
UMG605US
N
230/400V 50/60Hz
UMG605US
Earthing the
system
DC
AC/DC
L2
L3
N
L1
Auxiliary power
4M
4M
4M
4M
L1 L3L2 N
4M
L4
Impedance
Voltage measurement
18
UMG 605US
12
Use
The UMG605US has a display, keys 1 and 2
and the Service key to make it easier to install
and start up the UMG605US without a PC.
Important parameters such as current
transformers and device address are included
in the parameter list (see Appendix) and can
be directly programmed at the device.
A differentiation is made between operation
with the
• display mode and
• Programming mode.
Key functions
Press the key “briefly”:
• page forwards
• Digit/value +1
Press the key for “long time”:
• page backwards
• Digit/value -1
Simultaneously press both keys for around
1 second and keep them pressed:
• Switch between display mode and
programming mode.
Display
Key 1
Key 2
Concealed key
The UMG605US is
operated using keys 1
and 2.
Keys 1 and 2
Concealed key (service)
The Service key is intended for use by
instructed service employees only.
19
UMG 605US
1
2
N L1
RxD TxD
V
L1 L2 L3 L4
Input Output
RxD TxD L1 L2 L3 L4
Input Output
Hz
mImportant!
The user can use the GridVis/Jasic to
reconfigure the function of the keys
and selection of the values to be
displayed.
Display mode
After the power supply is resumed the device
is in Display mode.
In Display mode you can use Keys 1 and 2 to
page between the measured value displays.
Press Key 2 to page between the
measured values for current, voltage,
power output, etc.
Use Key 1 to select the phase for the
measured values.
The factory default setting for the measured
value displays is shown in the “measured
value displays” in the Appendix.
Fig.19.1 “Display Mode” display example.
Displayed measured value: UL1-N = 230.0 V.
Fig.19.2 Display example for rotating field and
frequency.
20
UMG 605US
Address Content
PRG
Programming mode
The most important settings required for
operation of the UMG605US can be displayed
and changed in programming mode.
The parameter list in the Appendix contains
the addresses for the most important settings.
You can make further settings using the GridVis
software included in the scope of supply.
If you simultaneously press Keys 1 and 2 for
around 1 second, programming mode opens
via a password query. If a display password has
not been programmed, the first programming
menu opens directly.
Programming mode is denoted in the display
by the text “PRG”. The digits of the address
flash.
If you are in programming mode and have not
pressed a key for approximately 60 seconds or
simultaneously press Keys 1 and 2 for around
1 second, the device returns to display mode.
Fig. 20.1 “Programming Mode” display example,
address 000 with content 5,000.

Other janitza Measuring Instrument manuals

janitza UMG 96 RM Instruction manual

janitza

janitza UMG 96 RM Instruction manual

janitza UMG 96-S2 User manual

janitza

janitza UMG 96-S2 User manual

janitza UMG 605 PRO Instruction manual

janitza

janitza UMG 605 PRO Instruction manual

janitza 800-CT8-A User manual

janitza

janitza 800-CT8-A User manual

janitza UMG 96S User manual

janitza

janitza UMG 96S User manual

janitza UMG 96-S2 Instruction manual

janitza

janitza UMG 96-S2 Instruction manual

janitza UMG 512-Pro Instruction manual

janitza

janitza UMG 512-Pro Instruction manual

janitza optec UMG-96 User manual

janitza

janitza optec UMG-96 User manual

janitza 800-CT8-A User manual

janitza

janitza 800-CT8-A User manual

janitza UMG 96-PA Instruction manual

janitza

janitza UMG 96-PA Instruction manual

janitza UMG 96 RM-EL User manual

janitza

janitza UMG 96 RM-EL User manual

janitza UMG 103-CBM User manual

janitza

janitza UMG 103-CBM User manual

janitza UMG 96-PA User manual

janitza

janitza UMG 96-PA User manual

janitza UMG 500 A User manual

janitza

janitza UMG 500 A User manual

janitza UMG 96S 24V User manual

janitza

janitza UMG 96S 24V User manual

janitza UMG 801 User manual

janitza

janitza UMG 801 User manual

janitza UMG 605-PRO User manual

janitza

janitza UMG 605-PRO User manual

janitza UMG 96 RM-PN User manual

janitza

janitza UMG 96 RM-PN User manual

janitza UMG 96-PQ-L Instruction manual

janitza

janitza UMG 96-PQ-L Instruction manual

janitza UMG 96 RM-P User manual

janitza

janitza UMG 96 RM-P User manual

janitza UMG 508 User manual

janitza

janitza UMG 508 User manual

janitza UMG 804 Instruction manual

janitza

janitza UMG 804 Instruction manual

janitza UMG507 Manual

janitza

janitza UMG507 Manual

janitza UMG 511 Guide

janitza

janitza UMG 511 Guide

Popular Measuring Instrument manuals by other brands

Gigahertz Solutions HFW59D instruction manual

Gigahertz Solutions

Gigahertz Solutions HFW59D instruction manual

Endress+Hauser Memograph M RSG40 technical information

Endress+Hauser

Endress+Hauser Memograph M RSG40 technical information

Hanna Instruments GROCHEK COMBO instruction manual

Hanna Instruments

Hanna Instruments GROCHEK COMBO instruction manual

Gima WT1 quick start guide

Gima

Gima WT1 quick start guide

Anritsu Network Master Series Operation manual

Anritsu

Anritsu Network Master Series Operation manual

Extech Instruments EN510 user manual

Extech Instruments

Extech Instruments EN510 user manual

ICP DAS USA iWSN-9601 Series quick start

ICP DAS USA

ICP DAS USA iWSN-9601 Series quick start

Dwyer Instruments RPME Series Specifications-installation and operating instructions

Dwyer Instruments

Dwyer Instruments RPME Series Specifications-installation and operating instructions

Rinstrum M4201 quick start guide

Rinstrum

Rinstrum M4201 quick start guide

Agilent Technologies X Series manual

Agilent Technologies

Agilent Technologies X Series manual

SkyGolf SkyPro quick start guide

SkyGolf

SkyGolf SkyPro quick start guide

Bosch LR 24 Professional Original instructions

Bosch

Bosch LR 24 Professional Original instructions

YATO YT-7307 quick start guide

YATO

YATO YT-7307 quick start guide

Mastech MS8238G user manual

Mastech

Mastech MS8238G user manual

Dytran 3224C operating guide

Dytran

Dytran 3224C operating guide

Anton Paar DMA 35 Standard operating procedure

Anton Paar

Anton Paar DMA 35 Standard operating procedure

Union INCA1021 Translation of the original operating instructions

Union

Union INCA1021 Translation of the original operating instructions

Nasaco NGSLR Operation manual

Nasaco

Nasaco NGSLR Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.