4
GENERAL INFORMATION
PRECAUTIONS
WARNINGS
The Jorvet IV Fluid and Blood Transfusion Warmer is a dry-heat device. This unit is designed to warm
IV uids safely through the heating plate. The Warmer does not control the uid ow rate. Fluids can be
warmed to temperatures between 95°-107.6° F (35°C-42°C) at ow rates 25ml/hr to l000ml/hr.
uIMPORTANT
Before using the Jorvet IV Fluid and Blood Transfusion Warmer, please read and understand this
User's Guide and the SAFETY PRECAUTIONS prior to use. If you have any questions, please
contact Jorgensen Laboratories at 1.800.525.5614.
uCarry out the inspections and maintenance periodically according to the manual to avoid patient or
user injury.
uThis operation manual is only used by professional trained Veterinary Personnel.
uThis device is used for heating the infusion, blood transfusion and can be used only by the professional
trained Veterinary Personnel.
uSecurely attach the warmer to an IV Stand to prevent damage to the unit by sliding down or dropping
to the oor.
uThe unit should only be used with the power cord that came with the unit and is not shared during
the use of the warmer.
uIt is not allowed to use voltage other than that specied on the product label, or it could cause
damage or a re.
uInspect the warmer and all accessories completely before use to verify it is working in a safe manner.
uDO NOT only rely on the warmer's alarm to notify of a problem, the warmer should be monitored
throughout the infusion.
uStop the use of the warmer during an alarm
uDO NOT connect other brand heating devices or other brand heating accessories to this warmer as
it could lead to wrong heating temperature or dangerous infusion.
uKeep the infusion line straight and unfolded to prevent a long-term occlusion and heating to uids/
blood for long period of time.
uDO NOT use the warmer where there is strong sunlight, cold & hot, wind or a dusty environment.
uMicrowave will inuence the function of the warmer. All other devices next to the warmer must meet
the EMC standard. Mobile phone, X-Ray or MRI will be the cause of a malfunction due to its high
frequency radiation.