31WarrantyThe Blood Pressure Monitor is guaranteed for 2-year from the date of purchase. If theBlood Pressure Monitor does not function properly due to defective components or poor workmanship, we will repair or replace it freely. The warranty does not cover damages to your Blood Pressure Monitor due to improper handling. Please contact local retailer for details.
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.Table 1Table 2
Guidance and declaration of manufacturer-electromagnetic immunityThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environm-ent.Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.Electromagnetic environment -guidanceCompliancelevelIEC 60601test levelIMMUNITY test± 8 kV contact±2 kV,±4 kV, ±8 kV, ±15 kV airElectrostaticdischarge (ESD)IEC 61000-4-2
Guidance and declaration of manufacturer-electromagnetic emissionsThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.Electromagnetic environment -guidanceComplianceEmissions testThe device uses RF energy only forits internal function. Therefore, itsemissions are very low and are notlikely to cause any interference innearby electronic equipment.Group 1,class B.Radiated emission CISPR 11The device is suitable for use in allestablishments, including domesticestablishments and those directlyconnected to the public low-voltagepower supply network that supplies buildings used for domestic purposes.Conducted emission CISPR 11Harmonic emissionsIEC 61000-3-2Class AVoltage fluctuations/ flicker emissions IEC 61000-3-3Complies
Electrostatictransient/burstIEC 61000-4-4SurgeIEC 61000-4-5±0.5kV, ±1kV (differential mode)
Voltage dips, short interrupti-ons and voltagevariations on p-ower supply in-put linesIEC 61000-4-11 0 % UT; 0,5 cycleAt 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°0 % UT; 1 cycleand70 % UT; 25/30 cyclesSingle phase: at 0°0 % UT; 250/300 cycle
Power frequency(50/60 Hz)magnetic fieldIEC 61000-4-8 30 A/m; 50Hz or 60Hz Power frequency magnetic fields should be at levels charactertic of a typical location in a typical comme-rcial or hospital environment.30 A/m; 50Hz or 60Hz± 8 kV contact±2 kV,±4 kV, ±8 kV, ±15 kV air
Electromagnetic Compatibility Information
3233
± 2 kV ,100kHz, for AC power port
0 % UT; 0,5 cycleAt 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°0 % UT; 1 cycleand70 % UT; 25/30 cyclesSingle phase: at 0°0 % UT; 250/300 cycle
±0.5kV, ±1kV (differential mode)± 2 kV ,100kHz, for AC power portMains power quality should be thatof a typical commercial or hospitalenvironment.Mains power quality should be thatof a typical commercial or hospitalenvironment.Mains power quality should be thatof a typical commercial or hospitalenvironment.
Group 1,class B.Table 2(continued)
Guidance and declaration of manufacturer-electromagnetic immunityThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environm-ent.Electromagnetic environment -guidanceCompliancelevelIEC 60601test levelIMMUNITY test
Electromagnetic Compatibility Information
34Radiated RF EM fieldsIEC 61000-4-3
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Conducted disturbances Induced by RF fieldsIEC 61000-4-63 V in 0.15 MHz- 80 MHz 6 V in ISM and/or amateur radio bands between 0.15 MHz and 80 MHz 80 % AM at 1kHz
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended seperation distance 80 MHz to 800 MHz 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transm- itter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
3 V in 0.15 MHz- 80 MHz 6 V in ISM and/or amateur radio bands between 0.15 MHz and 80 MHz 80 % AM at 1kHz10 V/m80MHz-2.7Ghz 80%AM at 1kHz10 V/m80MHz-2.7Ghz 80%AM at 1kHz