In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "2124"; press the "M" key to confirm, Display "S.MOdE",
press "M" key, display "last set value", use "up" key and "down" key to set, after
adjustment is completed, press "M" key to confirm and save.
4. Alarm interval time setting
Opcode: 2125 units (minutes)
In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "2125"; press the "M" key to confirm, Display "S_ErT", press
"M" key, display "00000" (the last set value,), set the value (range 1~09999) by "up"
key and "down" key, press "M" after adjustment is completed key to confirm saving.
5. Liquid level/level alarm upper limit setting
Opcode: 2311
In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "2311"; press the "M" key to confirm, Display "Ad1H", press
"M" key to set the upper limit alarm value, set the value by "up" key and "down" key,
"Ad1F" alarm upper limit hysteresis value, "Ad1d" alarm upper limit delay, all
adjustment Press the "M" key to confirm saving when finished.
6. Liquid level/level alarm lower limit setting
Opcode: 2312
In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "2312"; press the "M" key to confirm, Displays "Ad2H", press
the "M" key to set the upper limit alarm value, use the "Up" and "Down" keys to set
the value, "Ad2F" alarm upper limit hysteresis value, "Ad2d" alarm upper limit delay,
all adjustments Press the "M" key to confirm saving when finished.
7. Instrument zero calibration
Opcode: 1000
In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "1000"; press the "M" key to confirm, Display "bd_0" Press
the "up and down" keys to select "NO (exit without calibration) or YES (calibration
zero)". Press the "M" key to confirm saving and exit.
8. Instrument zero point removal
Opcode: 1511
In the measurement state, press the "M" key, it will display "-COd-"; press the "M"
key to confirm, it will display "00000", the units digit flashes, press the "up and
down" key to adjust to "1511"; press the "M" key to confirm, Display "CUTO", press
the "M" key, it will display "00000" (the last set value), use the "up" and "down" keys
to set the value (range 1~29999), press "M" after the adjustment is completed key
to confirm saving.
9. Meter offset setting