- 6 -
2.2 Parameter Settings
Before starting a measurement, the parameter settings must be configured.
Prior to the parameter settings, please read the software operation manual.
Before setting the parameters, connect the 3718-A/3719-A, a power supply for
3718-A/3719-A, the USB-RS485 Converter (converter, connecting cable) and a PC
(that has the software installed).
Sampling time settings: The settable range is 1-60 seconds, at a rate of 1.0 second.
All data compiled within the configured sampling time will be
transmitted via the analog output (4-20mA).
This sampling time is shared between 2 channels.
* The factory setting is 1 second.
Output range settings: Configure the range of the particle count and concentration.
The analog output will be 4 to 20mA within the configured
range. The following figure shows this setting range, which is
settable for each channel.
* The factory setting is 01 (0-10CNT).
Alarm Settings: Set the measurement range (0-100%) in 1% increments.
An alarm setting can be configured per 2 channels. If the detected
particle counts exceeds the configured alarm level of concentration,
the alarm signal will be sent. For more details on the alarm settings,
refer to 2.4 Connecting the Analog Output Terminal.
*The factory setting is 0 (for both CH1 and CH2).
Address Settings: Set the address of this instrument. The settable range is 0-512.
The addresses are used to identify each instrument when using
multiple instruments.
*The factory setting is 0.
Sensor Communication Settings: Configure the communication setting of the
instrument.
Configure the communication settings between this
instrument and upper-layer PC/CRMS distributor.