Service manual
Counting scales
KERN CXB
Version 1.3
4/2009
GB
CXB-SH-e-0913

GB KERN CXB
Version 1.3 4/2009
Service manual
Counting scales
Table of Contents
1Basic information _______________________________________________ 3
2Introduction____________________________________________________ 4
3General Information _____________________________________________ 4
4Error messages_________________________________________________ 5
5Explanation of display ___________________________________________ 6
5.1 Symbol indicator _________________________________________________ 7
6Explanation of keyboard _________________________________________ 8
7External calibration______________________________________________ 9
7.1 Calibration of non-approval models (CXB) ____________________________ 9
7.2 Back to the original value of calibration_______________________________ 9
7.3 Calibration of approval models (CXB-M) _____________________________ 10
8Service mode access ___________________________________________ 11
9Linearity calibration ____________________________________________ 12
10 Spec. calibration _______________________________________________ 15
10.1 Reference values for capacity setting _______________________________ 17
11 Temperature calibration_________________________________________ 18
11.1 Indoor air (normal) temperature calibration___________________________ 18
11.2 High temperature calibration_______________________________________ 19
11.3 Low temperature calibration _______________________________________ 20
11.4 Check / Delete temperature calibration data __________________________ 20
12 Spare parts ___________________________________________________ 21
2 CXB-SH-e-0913

1 Basic information
Grundlegende Hinweise
The device must be repaired only by trained specialist staff or personnel with
professional formation (such as a repair-specialist accredited by law
concerning verification).
The service manual is obligatory for repair work.
After repair, original conditions of the device have to be restored.
Only original spare parts should be used.
Instructions about conformity-evaluated scales:
Repair must be carried only at 100% compliance with the type approval.
A violation of this specification will result in a loss of the type approval!
After successful repair the balance will have to be reverified before it can be
used again in a statutorily regulated field.
Das Gerät darf nur von geschultem oder beruflich ausgebildetem
Fachpersonal (z. B. eichrechtlich anerkannter Instandsetzer) repariert werden.
Die Serviceanleitung ist bindend für Reparaturen.
Das Gerät muss nach erfolgter Reparatur wieder in den Originalzustand
zurückversetzt werden.
Es dürfen nur Originalersatzteile verwendet werden.
Hinweis zu konformitätsbewerteten Waagen:
Reparatur darf nur in 100% -iger Übereinstimmung mit der
Bauartzulassung erfolgen. Ein Verstoß gegen diese Vorgabe führt zum Erlöschen
der Bauartzulassung!
Nach erfolgreicher Reparatur muss eine Nacheichung erfolgen, um die Waage
wieder im gesetzlich geregelten Bereich verwenden zu können.
CXB-SH-e-0913 3

2 Introduction
1. There are 3 ways to rectify the linearity for setting the temperature calibration,
which can make the scale more accurate.
nInside air temperature (normal temperature)
oHigh temperature
pLow temperature
♦It is adequate to only do the inside air temperature calibration. Upon the
customer’s request, we also offer the high temperature and low temperature
calibration as well.
2. Sampling speed is also depended on the user’s need (01~05).
3. Linearity calibration, capacity setting and temperature calibration can be set
separately.
4. Weighing unit: g or kg (depending on model)
5. Tare range is full range tare.
3 General Information
1. The newly manufactured scale without passing the calibration and setting process
will automatically enter the linearity calibration mode. After the linearity calibration
procedure is done, the program will automatically enter the Specification Setting
Mode.
2. Zero Range
The internal value should be in between 120,000~250,000 with the platter on
and without any load (check Æsee chapter 10 “Spec. calibration”)
3. Maximum Capacity
The internal value should be over 300,000 when the scale is fully loaded.
4. When rubbing the tetragon, the inaccuracy value should be smaller than 10
divisions.
5. Tare range is full range tare.
4 CXB-SH-e-0913

4 Error messages
Non-approval models:
E1 ⇒ZERO over 300,000 internal value.
E2 ⇒ZERO below 10,000 internal value.
Approval models:
E1 ⇒ZERO over the zero position in weight calibration + 10% of full scale.
E2 ⇒ZERO over the zero position in weight calibration - 10% of full scale.
All models:
E4 ⇒ZERO is unstable.
E5 ⇒Turning on machine core value is under 0.
E6 ⇒Core value is over 250,000.
E7 ⇒Core value is below 120,000.
- - OL - ⇒The weighing range is exceeded.
NOTES:
•Non-approval models (CXB): After E1, E2 or E4 is displayed press
to continue.
•Approval models (CXB-M): After E1, E2 or E4 appear, set the jumper SWA1
on main board to “ADJ” position and press to return the counting
mode.
Then press + + at the same time followed by pressing
to access calibration (see chap. 7.3).
CXB-SH-e-0913 5

5 Explanation of display
Weight Reference weight
Loading status display Pieces
6 CXB-SH-e-0913

5.1 Symbol indicator “3”
Display weight
Here, the weight of your goods is displayed.
Overlay Windicates:
Æ0ÅZeroing display
PRE-
TARE Tare in memory
Battery very low
Display reference weight
The reference weight of a sample is shown here. This value is either entered by user
of calculated by balance.
Overlay Windicates:
Placed number of pieces insufficient for reference calculation
Placed reference weight insufficient for reference calculation
Display quantity
Here, all the parts placed on balance are immediately displayed by number.
Overlay Windicates:
M+ Data in summation memory
Stability display
Battery charge status display
red Battery is almost discharged
green Battery is completely discharged
CXB-SH-e-0913 7

6 Explanation of keyboard
Choice Function
- •Number keys
•Delete key
•Call counting with tolerance control
•Store reference weights in memory
•Call stored reference weights
•Addition in total memory
•Call up total memory
•Delete summation memory
•Enter reference weight through weighing
•Display reference weight stored last
•Enter target number of pieces
•Numeric entry reference weight
•Display reference weight stored last
•Enter target weight
•Zeroing key
•Back to weighing mode
•Taring key
•Enter numerical tare
8 CXB-SH-e-0913

7 External calibration
7.1 Calibration of non-approval models (CXB)
♦Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8)
♦Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
0
W
a
PRE-
TARE
Press the key and keep it pressed down whilst simultaneously operating the
key
Weight value of calibration weight is flashing on display. However, you may enter a
value of your choice via the number keys. *
3000
W
a
PRE-
TARE
CAL
Carefully place calibration weight in the centre of the weighing plate. Calibration
will be carried out automatically after dead stop control.
Remove calibration weight during count down to zero.
The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration
weight be incorrect. Turn balance off, then restart it and repeat the calibration
process.
* The calibration should be made with the recommended calibration weight (see
chap. 1 "Technical data" in the operating manual).
Note
Use the key to exit calibration mode. The balance returns to weighing mode.
7.2 Back to the original value of calibration
(only non-approval models)
nPress and hold key followed by pressing key until the scale is counting
backward to zero, then release it.
oThe scale will show the original value, which is the value setting at the factory.
CXB-SH-e-0913 9

7.3 Calibration of approval models (CXB-M)
♦Before calibrating, adjust the mini jumper SWA1 to “ADJ” position (see chap. 8)
♦Observe stable environmental conditions. A warming up time of 30 minutes is
required for stabilization.
Operation
Ensure that there are no objects on the weighing plate.
ÖPress + 1+ 5keys
ÖPress key
The display shows:
0.000
M+
0
(Example)
ÖPress key
The display shows:
15.000
(Example)
M+
0
Weight value of calibration weight is flashing on display.
ÖCarefully place calibration weight in the centre of the weighing plate.
ÖPress key
Calibration will be carried out automatically after dead stop control.
ÖRemove calibration weight during count down to zero.
The balance returns automatically into weighing mode. An error message will
appear on the display should a calibration error occur or should the calibration
weight be incorrect. Turn balance off, then restart it and repeat the calibration
process.
194452
194452
* The calibration should be made with the recommended calibration weight (see
chap. 1 "Technical data" in the operating manual).
10 CXB-SH-e-0913

8 Service mode access
♦Access to the service mode is controlled by jumper SWA1 on the main PCB. The
jumper has two positions LOCK, which prevents access to the service mode and
ADJ, which allows access to the calibration and configuration routines.
♦SWA 1 jumper is located adjacent to the RF screening cover of the main PCB.
♦Place your balance upside down and remove seal/cap. For calibration and
configuration routines the jumper “SWA1” must be set to position “ADJ”.
Sealing mark / unlocking switch
Position of unlocking switch Status
To the right Unlock the balance for calibration process (ADJ)
To the left Verification position - calibration locked (LOCK)
CXB-SH-e-0913 11

9 Linearity calibration
♦Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position.
♦After finish the calibration, please re-adjust the mini jumper SWA1 to
“LOCK” position.
~Divide the capacity into 5 sections before linearity calibration.
E.g. CXB 30K2 (capacity = 30kg)
Divide it into 5 phases: 3kg, 6kg, 12kg, 12kg, 3kg
(The total calibrated weight should be between 1.1~1.25 times of the capacity
weight. E.g.: 3kg + 6kg + 12kg + 12kg + 3kg = 36kg are 1.2 times of 30kg)
nTurn on the machine, after the scale resets back to zero,
then press key + 1key + 6key (at the same time) to enter the linearity
calibration mode. (Please press 9key to pass over this linearity calibration
mode.)
Weight Piece Weight Pcs
0 0 L. XXXXXX
key in *n value core value
display the number of times of calibration (zero internal value)
♦*n = the calibrated value of phase X / the calibrated value of 1st phase
[*n = multiplier (2,3,4…) of 1st calibration weight / 1st phase]
♦*n ⇒must be an integral
oPress . key
Weight Piece Weight Pcs
0 1 1 XXXXXX
internal value of the phase
12 CXB-SH-e-0913

pPut the 1st weight mass (e.g. weight mass=3kg, *n=1), then press . key
Weight Piece Weight Pcs
0 2 1 XXXXXX
input *n value internal value of the phase
qAdd the 2nd weight mass (e.g. 6kg, in sum 9kg) and input *n value (e.g. *n=2),
then press . key
Weight Piece Weight Pcs
0 3 1 XXXXXX
input *n value, e.g. n=2 internal value of the phase
rAdd the 3rd weight mass (e.g. 12kg, in sum 21kg) and input *n value
(e.g. *n=4), then press . key
Weight Piece Weight Pcs
0 4 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
sAdd the 4th weight mass (e.g. 12kg, in sum 33kg) and input *n value
(e.g. *n=4), then press . key
Weight Piece Weight Pcs
0 5 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
CXB-SH-e-0913 13

tAdd the 5th weight mass (e.g. 3kg, in sum 36kg) and input *n value (e.g. *n=1),
then press . key
Weight Piece Weight Pcs
0 6 1 XXXXXX
input *n value, e.g. n=1 internal value of the phase
uPress key (“CONFIRM” key), the buzzer beeps once
Weight Piece Weight Pcs
S Y S U P
vRemove all the weight mass, the scale will reset back to zero automatically,
then the procedures finished.
14 CXB-SH-e-0913

10 Spec. calibration
♦Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position.
♦After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK”
position.
nTurn on the machine, after the scale resets back to zero,
then press key + 1key + 7 key (at the same time) to enter the spec.
calibration mode.
Weight Piece Weight Pcs
0 3 0 0 1 0 2 1 X X 2 0 2 0 5 3
program version keyboard default value core value
(to check keyboard function) (internal value of zero)
see chapter 3
“General Information“
oPress key
Weight Piece Weight Pcs
Ñ0Ò0 0 0 0 0 0 0 0 0 0 0 0 0 0 1
full capacity setting preset calibrated value refer to “Approx. 1”
key in “ full scale+9d ” suggest key in “full scale”
(overload function max.+ 9d)
Approx. 1: 0 0 0 0 0 1
a b c d e
a⇒input the division 1 = e.g. 1g / 10g 2 = e.g. 0,2g / 2g 5 = e.g. 0,5g / 5g
b⇒input the decimal point 0 = e.g. 1g 1 = e.g. 0,2g
2 = e.g. 0,01kg 3 = e.g. 0,001kg
c⇒input the unit of the unit weight 0, 1, 2, 8, 9 (full capacity are 10nof division)
d⇒define the keyboard function of ;kg/lb ; PST CE keys (use always 1 )
0 = PST CE, unit weight in kg
1 = 10 preset, unit weight in kg (OIML approval model)
2 = kg/lb (1st unit = kg, 2nd unit = lb) 3 = 10 preset, kg/lb (1st unit kg, 2nd unit lb)
4 = PST CE, lb unit only 5 = 10 preset, lb unit only
6 = lb/kg (1st unit lb, 2nd unit kg) 7 = 10 preset, lb/kg (1st unit lb, 2nd unit kg)
0 = 100 (x 1) » CXB 15K1 / CXB 30K2
1 = 101 (x 10)
2 = 102 (x 100)
8 = 10-2 (x 0.01)
9 = 10-1 (x 0.1) » CXB 3K0.2 / CXB 6K0.5
Example: CXB 3K0.2 a= 2.
value of ax 10n = Readout 2 x 10-1 = 0,2g c= 9
CXB-SH-e-0913 15

e ⇒different versions of models setting
0 = Non-approval model (External calibration is applicable)
1 = OIML approval model (The display resolution should be set at 1/3000,
External calibration is NOT applicable)
4 = Non-approval model (External calibration is NOT applicable)
pAfter input spec., press key (not necessary to put the weight mass on the
platter, the calibrated weight value is the reference for the temperature
calibration procedure).
♦ The scale will be self-tested and reset back to zero.
Press key+1key+5key, then press key to enter the temperature
calibration mode (see chapter 11).
NOTES:
(1) The last number of full capacity value must be variable number, can be fixed “0”.
(2) The digit and location of preset calibrated value, refer to the location of capacity
digit.
(3) Input capacity + 9d in order to run 9 times overload function.
(4) If max. number of digit is “0”, please input “0”.
(5) Reference values for capacity setting see chapter 10.1.
16 CXB-SH-e-0913

10.1Reference values for capacity setting
Weighing unit – g–
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 1 9 1 0 (1)
Weight Piece Weight Pcs
Ñ6Ò0 0 4 5 6 0 0 0 0 5 1 9 1 0 (1)
Weight Piece Weight Pcs
Ñ1Ò5 0 0 9 1 5 0 0 0 1 0 0 1 0 (1)
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 0 0 1 0 (1)
CXB 3K0.2
CXB 6K0.5
CXB 15K1
CXB 30K2
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 1 0 0 1 1 (1)
Weight Piece Weight Pcs
Ñ6Ò0 0 4 5 6 0 0 0 0 2 0 0 1 1 (1)
CXB 3K1M
CXB 6K2M
Weighing unit – kg –
Weight Piece Weight Pcs
Ñ1Ò5 0 0 9 1 5 0 0 0 1 3 0 1 0 (1)
Weight Piece Weight Pcs
Ñ3Ò0 0 1 8 3 0 0 0 0 2 3 0 1 0 (1)
CXB 15K1
CXB 30K2
Weight Piece Weight Pcs
Ñ0Ò3 0 0 9 0 3 0 0 0 1 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ0Ò6 0 1 8 0 6 0 0 0 2 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ1Ò5 0 4 5 1 5 0 0 0 5 3 0 1 1 (1)
Weight Piece Weight Pcs
Ñ0Ò3 0 0 9 0 3 0 0 0 1 2 0 1 1 (1)
CXB 3K1M
CXB 6K2M
CXB 15K5M
CXB 30K10M
CXB-SH-e-0913 17

11 Temperature calibration
♦Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position.
♦After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK”
position.
11.1Indoor air (normal) temperature calibration
n* The scale will be self-tested and reset back to zero.
Press key+1key+5key, then press key to enter the temperature
calibration mode.
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
indoor temperature(°C) zero internal value
oNo weight on the platter, press .key to calibrate zero.
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value indoor temperature(°C) internal value of full capacity
pPut the full capacity weight on the platter, press .key to calibrate full capacity
value.
qAfter scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures
finished. Put the weight mass on the platter again to check if the display
shows the accurate weight. If it’s correct, turn off the power, then re-adjust the
mini jumper SWA1 to “LOCK” position.
NOTES:
(1) The procedures are for the scale without the temperature IC or for the scale in
the normal temperature environment.
(2) To achieve the accuracy of the scale, please run the high temperature and low
temperature calibration procedures.
(3) The procedures for high temperature and low temperature are totally reversed.
18 CXB-SH-e-0913

11.2High temperature calibration
⇒scales must be in the temperature lab chamber
nPress key+1key+5key, then press key to enter the temperature
calibration mode
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
high temperature(°C) zero internal value
oNo weight on the platter, press .key to calibrate zero
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value high temperature(°C) internal value of full capacity
pPut the full capacity weight on the platter, press .key to calibrate full capacity
value.
qAfter scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures
finished. Put the weight mass on the platter again to check if the display
showing the accurate weight. If it’s correct, turn off the power, then re-adjust
the mini jumper SWA1 to “LOCK” position.
CXB-SH-e-0913 19

11.3Low temperature calibration
⇒scales must be in the control room of low temperature
nPress key+1key+5key, then press key to enter the temperature
calibration mode
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
low temperature(°C) zero internal value
oNo weight on the platter, press .key to calibrate zero
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value low temperature(°C) internal value of full capacity
pPut the full capacity weight on the platter, press .key to calibrate full capacity
value.
qAfter scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures
finished. Put the weight mass on the platter again to check if the display
showing the accurate weight. If it’s correct, turn off the power, then re-adjust
the mini jumper SWA1 to “LOCK” position.
11.4Check / Delete temperature calibration data
nPress key+1key+5key
Weight Piece Weight Pcs
Sho t 0 0 XXXXXX
show temperature temperature (°C) internal value
oPress 2or 8key to check the temperature calibration data.
pPress 2or 8key to select the desirable data, and double-press .key to
delete it.
20 CXB-SH-e-0913