INSTALLATION INSTRUCTIONS
DWG. NO.
REV.
SHEET OF
DRAWN BY
DATE
APPROVED
KING-GAGE
®
MODEL 5993 INDICATOR
PNEUMATIC COLUMN DISPLAY
1 2
D. Kennedy
K-1000-51
B 8/09 Revised top connection block
A 5/97 Released
1/22/10
C
C 1/10 Caution! Contains Liquid
Indicator Identication
(Model & Serial No.)
The label on the front of the cover has the model number which
species case material, range and type of indicating liquid (uid
ll) used. See Detail 1.
If scales have been installed, tank designation appears on top
or side of scale. This should correspond to the applicable tank
installation. See Detail 2.
5993S - xx x
Indicating Liquid:
M - Mercury
R - No. 294 Red
P - No. 175 Purple
Range in Inches
600
700
800
60
70
WATER
K-xxx-1
Tank
Designation
Serial
Number
Detail 2 - Indicator Scale
Detail 1 - Model Number Designation
Dimensions by Indicator Range:
Range Dim. "A" Dim. "B"
20 ..................26-1/4 ...............25-3/8
35 ..................42-1/4 ...............41-3/8
43 ..................50-1/4 ...............49-3/8
51 ..................58-1/4 ...............57-3/8
66 ..................74-1/4 ...............73-3/8
81 ..................90-1/4 ...............89-3/8
Overow Check Valve
This check valve will prevent uid ll from escaping if
the indicator is overpressured.
B-1. Thread the vent plug into the top of the
check valve
B-2. Install this assembly into the front center
connection (Detail 4).
4
20
40
50
EACH MARK = 1 GALLONK-123-1
60
70
80
90
100
110
GALLONS
100
200
300
400
500
EACH MARK = 10 GALLONSK-123-1
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
GALLONS
AB
3-5/16 in.
84 mm
2-7/8 in.
73 mm
4-1/2 in.
108 mm
®
®
KING-GAGE
LIQUID LEVEL INDICATING SYSTEM
Overflow
Check Valve
7/16 in.
11 mm
5/16 in. DIA
(2 pls)
ADJUST AIR
INSIDE BARS
1-1/16 in.
27 mm
1/4 NPT
Detail 3 - Removing Sealing (Shipping) Plugs
9/16 in. Hex
Sealing Plug
(discard)
Swagelok fitting
Swagelok fitting
Vent Plug
Well Tube Plug
(discard)
Vent Plug
Overflow
Check Valve
Vent Plug
9/16 in. Hex
Sealing Plug
(discard)
Swagelok fitting
Swagelok fitting
Vent Plug
Well Tube Plug
(discard)
Vent Plug
Overflow
Check Valve
Vent Plug
Detail 4 - Overow Check Valve and
Fitting Installation
Installation
This KING-GAGE column indicator has a uid ll which is
sealed during shipment and installation. Do not remove seal
plugs until the indicator has been physically mounted as out-
lined below.
Note: Trapped air may prevent uid ll from showing in the
glass tube until seal plugs have been removed.
A-1. Mount indicator in a vertical position to a wall or other
rigid structural member. Indicator should be located to
ensure visibility of the display with adequate overall
clearance for connections at the top and bottom of the
housing. Refer to mounting dimensions shown.
A-2. Remove front cover and discard any packing material.
A-3. Remove and discard metal seal plug from center connec-
tion. Remove white vent plug from left hand connection
and save for nal installation. Remove and discard well
tube plug (Detail 3).
A-4. Parts bag includes tube tting and overow check valve to
complete the installation (Detail 4).
CAUTION! Contains Liquid
This item contains a uid ll as the
indicating medium. DO NOT REMOVE
seal plugs at the top of the case until
the indicator has been mounted in a
vertical position.
¡CUIDADO! Contiene Líquido
INSTALLATION INSTRUCTIONS
DWG. NO.
REV.
SHEET OF
DRAWN BY
DATE
APPROVED
KING-GAGE
®
MODEL 5993 INDICATOR
PNEUMATIC COLUMN DISPLAY
K-1000-51
2 2
D. Kennedy
B 8/09 Revised top connection block
A 5/97 Released
1/22/10
C
C 1/10 Caution! Contains Liquid
Indicator Scales
Installation: Scales are furnished in pairs, one for each side
of the indicating tube, with matching serial numbers (Detail 2).
Install or align scales only after the indicator has been physi-
cally mounted.
Alignment Check: If scales have been factory installed, refer
to step C-3 below to check alignment. Disconnect signal tubing
or air supply line before making any adjustment to scale align-
ment.
C-1. Remove front cover and remove any packing material (felt
blocks, etc.)
C-2. Scales are fastened to the support brackets using small
screws and washers supplied. See Detail 5.
C-3. Align the lowest mark on the scale to coincide with level of
liquid in the glass indicating tube. Tighten screws to
secure scales once properly aligned. Any misalignment
will cause reading errors.
C-4. Replace front cover and connect signal tubing and/or air
supply line.
100
200
300
400
500
4
20
40
50
600
60
A - 00123 - B
A - 00123 - B
Washer & Screw
(loosen to adjust
scale)
Lowest Mark on Scale
(aligns with top of
liquid)
Detail 5 - Indicator Scale and Adjustment
Signal Tube Installation
The tubing between the tank downpipe and indicator. Tubing
may be copper, nylon or polyo of the appropriate diameter
as recommended to convey the air ow to the downpipe and
resulting pressure back to the indicator.
D-1. Route tubing to avoid sharp bends and minimize exposure
to damage or high trafc areas. Recommended minimum
tubing diameters –
1/4” O.D. for distances up to 125 feet/38 m
3/8” O.D. for distances up to 550 feet/170 m
7/16” O.D. for distances greater than above
D-2. Install tube tting and connect signal tubing at indicator.
Tubing must be leak-tight for its entire run. Do not tee into
the signal tube for any equipement that consumes air.
Compressed Air Supply
Compressed air supply must be clean, dry and oil-free to en-
sure trouble free operation. Under normal operating conditions,
supply pressure should be 55-60 psig (380-420 kPa).
E-1. Install supply line at the Compressed Air Inlet connection
of the indicator.
E-2. Adjust the rotameter control so that the small internal oat
rests between the horizontal marks.
Pressure Signal
Tank Signal
Line Connection
Compressed
Air Supply
Rotameter
Tank Top
Fitting
and
Downpipe
4
20
40
50
EACH MARK = 1 GALLONK-123-1
60
70
80
90
100
110
GALLONS
100
200
300
400
500
EACH MARK = 10 GALLONSK-123-1
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
GALLONS
®
®
KING-GAGE
LIQUID LEVEL INDICATING SYSTEM
ADJUST AIR
INSIDE BARS

Popular Touch Panel manuals by other brands

Vega VEGADIS 62 operating instructions

Vega

Vega VEGADIS 62 operating instructions

Advantech PPC-3150SW user manual

Advantech

Advantech PPC-3150SW user manual

Fluidwell F112-P manual

Fluidwell

Fluidwell F112-P manual

EVOC W21 Series user manual

EVOC

EVOC W21 Series user manual

i3-TECHNOLOGIES i3TOUCH PXr Series manual

i3-TECHNOLOGIES

i3-TECHNOLOGIES i3TOUCH PXr Series manual

Advantech TPC-1070 user manual

Advantech

Advantech TPC-1070 user manual

Winmate W07IB3S-EHT1 quick start guide

Winmate

Winmate W07IB3S-EHT1 quick start guide

Moxa Technologies MPC-122-K Quick installation guide

Moxa Technologies

Moxa Technologies MPC-122-K Quick installation guide

Advantech PPC-315W TGL user manual

Advantech

Advantech PPC-315W TGL user manual

DFI TPC170 Quick manual

DFI

DFI TPC170 Quick manual

Seneca VISUAL 7N Installation instruction

Seneca

Seneca VISUAL 7N Installation instruction

Bristan WavePlus Dual Outlet Touchpad Installation instructions and user guide

Bristan

Bristan WavePlus Dual Outlet Touchpad Installation instructions and user guide

Graff Aqua-Sense Manual instruction

Graff

Graff Aqua-Sense Manual instruction

AMX Modero NXD-1500VG Dimension Guide

AMX

AMX Modero NXD-1500VG Dimension Guide

Atlona Velocity AT-VSP-550-BL installation guide

Atlona

Atlona Velocity AT-VSP-550-BL installation guide

Cermate Technologies PV035-TST installation manual

Cermate Technologies

Cermate Technologies PV035-TST installation manual

Eaton HMI04BU quick start guide

Eaton

Eaton HMI04BU quick start guide

DT Research 524T Basic operation guide

DT Research

DT Research 524T Basic operation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.