Kobold DAG-A3V User manual

Operating Instructions
for
Digital Indicating Unit
for Panel Mounting
Model: DAG-A3V

DAG-A3V
page 2 DAG-A3V K06/1222
We don’t accept warranty and liability claims neither upon this publication nor in
case of improper treatment of the described products.
The document may contain technical inaccuracies and typographical errors. The
content will be revised on a regular basis. These changes will be implemented in
later versions. The described products can be improved and changed at any time
without prior notice.
© Copyright
All rights reserved.
1. Contents
1.Contents ........................................................................................................ 2
2.Note .............................................................................................................. 3
3.Instrument Inspection .................................................................................... 3
4.Regulation Use ............................................................................................. 3
5.Assembly ...................................................................................................... 4
6.Electrical Connection .................................................................................... 5
7.Function description and operation ............................................................... 6
8.Setting up the device ..................................................................................... 7
8.1Switching on ........................................................................................ 7
8.2Standard parameterization: .................................................................. 7
8.3Programming interlock RUN ................................................................ 8
8.4Extended parameterization .................................................................. 8
9.Reset to default values ................................................................................ 12
10.Functional principle of the switching points ................................................. 13
11.Technical Information .................................................................................. 14
12.Order Codes ............................................................................................... 14
13.Safety advice .............................................................................................. 15
14.Error elimination .......................................................................................... 17
15.Disposal ...................................................................................................... 18
16.EU Declaration of Conformance ................................................................. 19
17.UK Declaration of Conformity ...................................................................... 20
Manufactured and sold by:
Kobold Messring GmbH
Nordring 22-24
D-65719 Hofheim
Tel.: +49(0)6192-2990
Fax: +49(0)6192-23398
E-Mail: [email protected]
Internet: www.kobold.com

DAG-A3V
DAG-A3V K06/1222 page 3
2. Note
Please read these operating instructions before unpacking and putting the unit into
operation. Follow the instructions precisely as described herein.
The instruction manuals on our website www.kobold.com are always for currently
manufactured version of our products. Due to technical changes, the instruction
manuals available online may not always correspond to the product version you
have purchased. If you need an instruction manual that corresponds to the
purchased product version, you can request it from us free of charge by email
([email protected]) in PDF format, specifying the relevant invoice number and
serial number. If you wish, the operating instructions can also be sent to you by
post in paper form against an applicable postage fee.
Operating instructions, data sheet, approvals and further information via the QR
code on the device or via www.kobold.com
The devices are only to be used, maintained and serviced by persons familiar with
these operating instructions and in accordance with local regulations applying to
Health & Safety and prevention of accidents.
When used in machines, the measuring unit should be used only when the
machines fulfil the EC-machine guidelines.
3. Instrument Inspection
Instruments are inspected before shipping and sent out in perfect condition.
Should damage to a device be visible, we recommend a thorough inspection of the
delivery packaging. In case of damage, please inform your parcel service /
forwarding agent immediately, since they are responsible for damages during
transit.
Scope of delivery:
The standard delivery includes:
Digital Indicating Unit for Panel Mounting model: DAG-A3V
4. Regulation Use
Any use of the device, which exceeds the manufacturer’s specification, may
invalidate its warranty. Therefore, any resulting damage is not the responsibility of
the manufacturer. The user assumes all risk for such usage.

DAG-A3V
page 4 DAG-A3V K06/1222
5. Assembly
Please read the Safety instructions on page 13 before installation and keep this
user manual for future reference.
1.
After removing the fixing elements, insert the device.
2.
Check the seal to make sure it fits securely.
3.
Click the fixing elements back into place and tighten the
clamping screws by hand. Then use a screwdriver to
tighten them another half a turn.
CAUTION! The torque should not exceed 0.1 Nm!

DAG-A3V
DAG-A3V K06/1222 page 5
6. Electrical Connection
Type DAG-A3V0 Type DAG-A3V3
supply 230 VAC supply 24 VDC galv. isolated
Connection examples:
Below you find some connection examples, which demonstrate some practical
applications:

DAG-A3V
page 6 DAG-A3V K06/1222
7. Function description and operation
Operation
The operation is divided into two different levels.
Menu Level
Here it is possible to navigate between the individual menu items.
Parameterization level:
The parameters stored in the menu item can be parameterized here.
Functions that can be adjusted or changed are always indicated with a
flashing of the display. Adjustments made at the parameterization level
should be always confirmed by pressing the [P] key to save them.
However, the display automatically saves all adjustments and then switches
to operation mode if no further keys are pressed within 10 seconds.

DAG-A3V
DAG-A3V K06/1222 page 7
8. Setting up the device
8.1 Switching on
Once the installation is complete, you can start the device by applying the current
loop. Check beforehand once again that all the electrical connections are correct.
Starting sequence
For 1 second during the switching-on process, the segment test (8 8 8 8 8) is
displayed, followed by an indication of the software type and, after that, also for 1
second, the software version. After the start-up sequence, the device switches to
operation/display mode.
8.2 Standard parameterization:
To be able to parameterize the display, press the [P] key in operating mode for 1
second. The display then changes to the menu level with the first menu item
TYPE.

DAG-A3V
page 8 DAG-A3V K06/1222
8.3 Programming interlock RUN
8.4 Extended parameterization
By pressing the [▲] & [▼] buttons during standard parameterization for one
second, the display switches to the extended parameterization mode.
Operation is the same as in standard parameterization.

DAG-A3V
DAG-A3V K06/1222 page 9

DAG-A3V
page 10 DAG-A3V K06/1222

DAG-A3V
DAG-A3V K06/1222 page 11

DAG-A3V
page 12 DAG-A3V K06/1222
9. Reset to default values
To return the unit to a defined basic state, a reset can be carried out to the
default values.
The following procedure should be used:
• Switch off the power supply
• Press button [P]
• Switch on voltage supply and press [P]-button until
„--- -“
is shown in the display.
With reset, the default values of the program table are loaded and used for
subsequent operation. This puts the unit back to the state in which it was
supplied.
Caution! All application-related data are lost

DAG-A3V
DAG-A3V K06/1222 page 13
10. Functional principle of the switching points
Limit value exceedance
“
High
”
The setpoint S1-S2 is “off”
below the threshold and “on”
on reaching the threshold.
Limit value undercut
“low”
The setpoint S1-S2 is
“on” below the
threshold and switched
“off” on reaching the
threshold.
Alarms / optical switching point display
An activated switching point can be optically indicated by flashing of the 7-segment
display.

DAG-A3V
page 14 DAG-A3V K06/1222
11. Technical Information
Operating instructions, data sheet, approvals and further information via the QR
code on the device or via www.kobold.com
12. Order Codes
Operating instructions, data sheet, approvals and further information via the QR
code on the device or via www.kobold.com

DAG-A3V
DAG-A3V K06/1222 page 15
13. Safety advice
Please read the following safety advice and the assembly chapter 2 before
installation and keep it for future reference.
Proper use
The DAG-A3V-device is designed for the evaluation and display of sensor signals.
Danger! Careless use or improper operation can
result in personal injury and/or damage to the
equipment.
Control of the device
The panel meters are checked before dispatch and sent out in perfect condition.
Should there be any visible damage, we recommend close examination of the
packaging. Please inform the supplier immediately of any damage.
Installation
The DAG-A3V-device must be installed by a suitably qualified specialist (e.g. with
a qualification in industrial electronics).
Notes on installation
• There must be no magnetic or electric fields in the vicinity of the device, e.g. due
to transformers, mobile phones or electrostatic discharge.
• The fuse rating of the supply voltage should not exceed a value of 6A N.B. fuse.
• Do not install inductive consumers (relays, solenoid valves etc.) near the device
and suppress any interference with the aid of RC spark extinguishing combinations
or free-wheeling diodes.
• Keep input, output and supply lines separate from each other and do not lay them
parallel with each other. Position “go” and “return lines” next to one another. Where
possible use twisted pair. So, the best measuring results can be received.
• Screen off and twist sensor lines. Do not lay current-carrying lines in the vicinity.
Connect the screening on one side on a suitable potential equaliser (normally
signal ground).
• The device is not suitable for installation in areas where there is a risk of
explosion.

DAG-A3V
page 16 DAG-A3V K06/1222
• Any electrical connection deviating from the connection diagram can endanger
human life and/or can destroy the equipment.
• The terminal area of the devices is part of the service. Here electrostatic
discharge needs to be avoided. Attention! High voltages can cause dangerous
body currents.
• Galvanic insulated potentials within one complex need to be placed on a
appropriate point (normally earth or machines ground). So, a lower disturbance
sensibility against impacted energy can be reached and dangerous potentials, that
can occur on long lines or due to faulty wiring, can be avoided

DAG-A3V
DAG-A3V K06/1222 page 17
14. Error elimination
Error description Measures
1. The unit permanently indicates overflow. • The input has a very high measurement, check the
measuring circuit.
• With a selected input with a low voltage signal, it is only
connected on one side or the input is open.
• Not all of the activated switching points are
parameterised. Check if the relevant parameters are
adjusted correctly.
2. The unit permanently shows underflow. • The input has a very low measurement, check the
measuring circuit .
• With a selected input with a low voltage signal, it is only
connected on one side or the input is open.
• Not all of the activated switching points are
parameterised. Check if the relevant parameters are
adjusted correctly.
3. The word "HELP " lights up in the
7-segment display.
• The unit has found an error in the configuration memory.
Perform a reset to the default values and re-configure
the unit according to your application.
4. Program numbers for parameterising of
the input are not accessible.
• Programming lock is activated
• Enter correct code
5. "Err1" lights up in the 7-segment display • Please contact the manufacturer if errors of this kind
occur.
6. The device does not react as expected. • If you are not sure if the device has been parameterised
before, then follow the steps as written in chapter 6 and
set it back to its delivery status.

DAG-A3V
page 18 DAG-A3V K06/1222
15. Disposal
Note!
Avoid environmental damage caused by media-contaminated parts
Dispose of the device and packaging in an environmentally friendly manner
Comply with applicable national and international disposal regulations and
environmental regulations.
Batteries
Batteries containing pollutants are marked with a sign consisting of a crossed-out
garbage can and the chemical symbol (Cd, Hg, Li or Pb) of the heavy metal that
is decisive for the classification as containing pollutants:
Cd1 Hg2 Pb3 Li4
1. „Cd" stands for cadmium
2. „Hg" stands for mercury
3. „Pb" stands for lead
4. „Li" stands for lithium
Electrical and electronic equipment

DAG-A3V
DAG-A3V K06/1222 page 19
16. EU Declaration of Conformance
We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole
responsibility that the product:
Digital Indicating Unit for Panel Mounting Model: DAG-A3V
to which this declaration relates is in conformity with the standards noted below:
EN 61010-1:2010 + A1:2019 + A1:2019/AC:2019 Safety requirements for
electrical equipment for measurement, control and laboratory use - Part 1:
General requirements
EN 61326-1:2013 Electrical equipment for measurement, control and
laboratory use - EMC requirements - Part 1: General requirements
EN IEC 63000:2018 Technical documentation for the assessment of electrical
and electronic products with respect to the restriction of hazardous substances
Also, the following EC guidelines are fulfilled:
2014/35/EU Low Voltage Directive
2014/30/EU EMC Directive
2011/65/EU RoHS (category 9)
2015/863/EU Delegated Directive (RoHS III)
Hofheim, 16 Jan. 2023
H. Volz M. Wenzel
General Manager Proxy Holder

DAG-A3V
page 20 DAG-A3V K06/1222
17. UK Declaration of Conformity
We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole
responsibility that the product:
Digital Indicating Unit for Panel Mounting Model: DAG-A3V
to which this declaration relates is in conformity with the standards noted below:
BS EN 61010-1:2010+A1:2019
Safety requirements for electrical equipment for measurement, control, and
laboratory use. General requirements
BS EN 61326-1:2013
Electrical equipment for measurement, control and laboratory use. EMC
requirements. General requirements
BS EN IEC 63000:2018
Technical documentation for the assessment of electrical and electronic products
with respect to the restriction of hazardous substances.
Also, the following UK guidelines are fulfilled:
S.I. 2016/1091 Electromagnetic Compatibility Regulations 2016
S.I. 2016/1101 Electrical Equipment (Safety) Regulations 2016
S.I. 2012/3032 The Restriction of the Use of Certain Hazardous Substances
in Electrical and Electronic Equipment Regulations 2012
Hofheim, 28 April 2023
H. Volz M. Wenzel
General Manager Proxy Holder
Table of contents
Other Kobold Measuring Instrument manuals

Kobold
Kobold AFH-G User manual

Kobold
Kobold OME Series User manual

Kobold
Kobold ADI-1F Series User manual

Kobold
Kobold TIN-SS User manual

Kobold
Kobold MIM User manual

Kobold
Kobold SZM User manual

Kobold
Kobold KSV User manual

Kobold
Kobold DAG-A44 Series User manual

Kobold
Kobold DOG-4 User manual

Kobold
Kobold KAL-E EX Series User manual

Kobold
Kobold MAS Series User manual

Kobold
Kobold NUS-7 Series User manual

Kobold
Kobold DSS Series User manual

Kobold
Kobold NBK-16 User manual

Kobold
Kobold NBK-19 User manual

Kobold
Kobold DSV Series User manual

Kobold
Kobold DRS Series User manual

Kobold
Kobold MAK User manual

Kobold
Kobold KME User manual

Kobold
Kobold DRF34 Series Installation instructions