HND-P215
HND-P215 K06/0422 page 9
(*) This menu can only be invoked if the logger memory contains no data! If parameter
should be changed the logger memory has to be cleared before! (key 6, p.r.t. 7.8
Operation Of Logger)
Menu Parameter Values Meaning
KEY
Menu
KEY
KEY
or
SEt
ConF
Set Configuration: Generic Configurations
Adr. 01,11..91 Base address of interface (if Out=Ser)
dAC. CH1, CH2,
or CH DIF Choice of the input to be the source for the analogue output
(if Out=dAC)
dAC.0 eg.
-5.00..5.00 mbar Enter desired value at which the analogue output potential
should be 0 V (if Out=dAC)
dAC.1 eg.
-5.00…5.00 mbar Enter desired value at which the analogue output potential
should be 1 V (if Out=dAC)
SEt
CAL
Set Calibration: Adjustment of Sensors
OFS.1 Sensordep., e.g.
-5.00…5.00 mbar The offset of sensor 1 will be displaced by this value to
compensate for deviations in the probe or in the measuring
device.
OFF Zero displacement is inactive (=0.0°)
SCL.1 -2.000…2.000 The measuring scale of sensor 1 will be changed by this
factor [%] to compensate deviations of temperature probe or
measuring device.
OFF Scale correction factor inactive (=0.000)
OFS.2 Sensordep., e.g.
-5.00…5.00 mbar The offset of sensor 2 will be displaced by this value to
compensate for deviations in the probe or in the measuring
device.
oFF Zero displacement inactive (=0.0°)
SCL.2 -2.000…2.000 The measuring scale of sensor 2 will be changed by this
factor [%] to compensate deviations of temperature probe or
measuring device.
oFF Scale correction factor inactive (=0.000)
SEt
AL.
Set Alarm: Settings Of Alarm Function
AL. 1 On
larm sensor 1 on, with buzzer sound
no.So Alarm sensor 1 on, without buzzer sound
oFF no alarm function for sensor 1
AL.Lo/AL.1 Sensor1-Min...
AL.1-Hi Min alarm rail Sensor 1 (not when AL.1 oFF)
Sensor1-Min is the lower display range of sensor 1
AL.Hi/AL.1 AL.1-Lo...
Sensor1-Max Max alarm rail Sensor 1 (not when AL.1 oFF)
Sensor1-Max is the upper display range of sensor 1
AL. 2 On Alarm sensor 2 on, with buzzer sound
no.So Alarm sensor 2 on, without buzzer sound
oFF no alarm function for sensor 2
AL.Lo/AL.2 Sensor2-
Min...AL.2-Hi Min alarm rail Sensor 2 (not when AL.2 oFF)
Sensor2-Min is the lower display range of sensor 2
AL.Hi/AL.2 AL.2-Lo...
Sensor2-Max Max alarm rail Sensor 2 (not when AL.2 oFF)
Sensor2-Max is the upper display range of sensor 2
AL.DIF
On Alarm sensor difference on, with buzzer sound
no.So Alarm sensor difference on, without buzzer sound
oFF no alarm function for sensor difference
AL.Lo DIF -19999..AL.DIF-
Hi Min alarm rail of difference (not when AL.DIF oFF)
AL.Hi DIF AL.DIF-Lo..19999 Max alarm rail of difference (not when AL.DIF oFF)
SEt
LoGG
Set Logger: Configuration Of Logger Function *
Func
CYCL Cyclic: logger function ‚cyclic logger‘ *
Stor Store: logger function ‚individual value logger‘ *
oFF no logger function *
CYCL 1..3600 Cycle time of cyclic logger [seconds] *
Lo.Po on/oFF Low-power logger with very low power consumption
(only for cyclic logger and slow measuring rate)
*
SEt
CLOC
Set Clock: Setting Of Real Time Clock
CLOC HH:MM Clock: Setting of time hours:minutes
dAtE TT.MM Date: day.month
YEAr YYYY Yea