KS RnD KS200-DM User manual

KS200-DM
OPACITY SMOKE METER
MANUAL
(Ver. 0.4.0)

contents
1. Safety guideline ……………………………………………………………………………… 3
1-1 Purpo e of u e …………………………………………………………………………… 3
1-2 Term of u e …………………………………………………………………………… 3
2. Product descriptions……………………………………………………………………… 4
2-1 Outlook …………………………………………………………………………………… 4
2-2 Front Part …………………………………………………………………………………… 5
2-3 Back Part …………………………………………………………………………………… 5
2-4 Specification …………………………………………………………………………………… 6
3. Installation and precautions ……………………………………………………………… 7
3-1 How to in tall ……………………………………………………………………………… 7
3-2 precaution ………………………………………………………………………………… 7
4. Measurement Mode ……………………………………………………………………… 7
4-1 Warm up…………………………………………………………………………………………… 7
4-2 Di play……………………………………………………………………………………………… 8
4-3 Mea urement(Peak hold) ………………………………… ………………………………… 9
4-4 Mea urement …………………………………………………………………………………… 9
. Select Mode …………………………………………………………………………………… 10
5-1 Key & Function …………………………………………………………………………… 10
5-2 Set Zero & Span …………………………………………………………………………… 11
5-3 Set Date ………………………………………………………………………………………… 11
5-4 Set HOLD Mode ……………………………………………………………………………… 11
5-5 Set RPM ………………………………………………………………………………………… 11
6. Remote Controller …………………………………………………………………………… 12
6-1 Connet remote controller ……………………………………………………………………… 12
6-2 How to u ing remote controller ……………………………………………………………… 12
7. Others ……………………………………………………………………………………… 12
7-1 Len …………………………………………………………………………………………… 13
7-2 Probe …………………………………………………………………………………………… 13
7-3 Error ………………………………………………………………………………………… 13
7-4 Acce orie ……………………… ……………………………………………………………… 14

1. Safety guideline
1-1. Purpo e to u e
KS200-DM mea ure the concentration of automobile exhau t ga to diagno e and maintain the
condition of the vehicle. .
1-2. Term of u e
① U e the rated voltage.
② An altitude of le than 1000m and a humidity of le than 85%.
③ In tall in a place where there i no direct unlight, vibration, or udden temperature change.
④ In tall it in a well-ventilated place.
⑤ In tall at lea t 25cm above the ground.
*Notice
If anything unusual occurs, please stop using it immediately and contact the place
of purchase or headquarters for inspection.
Waring
① Plea e check power before u ing.
② When probe in and out watch out for burn ..
③ Plea e in tall the probe in a location that i not affected by the wind. .
④ Don’t tay longer when the exhau t ga i generated.
* As the exhaust gas contains carbon monoxide (CO) and nitrogen oxide (NOx) that cause fatal
damage to the human body, use it in a ventilable place at all times.

KS200-DM Manual
2. Specification
2-1. Product
① Full ② Front
③ Side ④ Back
4/15

KS200-DM Manual
2-2. Di play
2-3. Back and Side
① ② ③ ④ ⑤ ⑥ ⑦
NO. NAME Description
① Fan Refre h the cell with out ide air
② INLET Ga inlet
③ Power witch 220V (1A Fu e in it)
④ RS232 9Pin connector(communication)
⑤ RPM RPM connector
⑥ Remote control Remote control connector
⑦ OIL T Oil temperature en or connector
5/15

KS200-DM Manual
2-4. Specification
NAME Description
Model Name KS200-DM
Measurement
items Smoke opacity , RPM, ℃
Measurement
principle Partial flow opacimeter
Measuring range
opci
ty
0.0~100.0%
K
0.00~21.42m-1
R
P
M
0~8000rpm Oil
Te
mp.
0~150℃
Resolution
function 0.1% 0.01m-1 1rpm 1℃
Repeatability Under 1%
Response time 0.5 ec (90%)
Preheating time Within 10 minute
Light source Green LED (565nm)
Detector Photodiode
Optical path length 430mm
Printer Built-in thermal printer
Power 220V/AC 60Hz (100W)
Size 415(W) * 130(D) * 220(H) excluding leg , handle
Weight 5.8 kg
Probe 1.5m
Communication RS232(4 time / 1 ec)
Option Printer, RPM, Oil Temperature , Standard Filter , Remote Controller
6/15

KS200-DM Manual
3. Installation and precautions
3-1. How to In tall
follow the picture, lay down the machine and tight the bolt
(proceed both ide)
Stand the main body of the machine correctly,
level it up, and tighten it trongly o that it doe not hake.
① In ert one ho e into the flexable probe tube and the other into the inlet of the mea uring
in trument and rotate the butterfly clip to fix it.
② Connect the power
4. Measurement Mode
4-1. Warm up
① Power [ON] and tart initialization.
Ver ion → Time → Warming up(within 10 min) → Zero → Span → Cal
② Warming up i carried out tep by tep, and the initial count value may vary depending on the
urrounding temperature or the tate of u e of the device.
※ Please place the probe end in a clean atmosphere.
7/15

KS200-DM Manual
4-2. Di play
① Mea ure mode tap [Di play] key move on next tep, The corre ponding LED i on.
Opacity Meter
0.0
●
%
○
m-1
○
rpm
○
℃
Smoke (%) → Light ab orption coefficient (m-1) → RPM (rpm) → Oil temperature (°C)
Opacity Dust Meter
2021/1/1 AM 09:00:00
Opacity 35 0%
k 1 42
RPM 800
Oil Temp 25℃
(print ex.)
4-3. Mea urement peak hold (Mea urement Free acceleration mode)
① Check whether it i elected a a peak hold.
② Pre [HOLD] key.
③ In ert the probe of the mea uring in trument. Step on the accelerator pedal.
④ The mea urement creen i held at the highe t value, and pre the [PRINT] key to print the held
mea urement value.
⑤ Remove the probe. (※ watch out for burn .)
Opacity Dust Meter
2021/1/1 AM 09:00:00
Opacity Peak 35 0%
k Peak 1 42
RPM 2560
Oil Temp 25℃
(ex print)
8/15

KS200-DM Manual
4-4. Mea urement (Free acceleration mode)
① Before tating mea urement, check whether the election mode i elected a the peak hold
mode.
② In ert the probe of the mea uring in trument. Step on the accelerator pedal.
③ Pre [Accel] key and [Set] key choo e mea urement.
Opacity Meter Opacity Meter
ACCL ▶▶
L-40
○
%
○
m-1
○
rpm
○
℃
○
%
○
m-1
○
rpm
○
℃
④ Choo e permi ion tandard. Pre [▲], [▼] Increa e or decrea e by 5% .
pre [SAVE] key choo e etting value.
⑤ [AC-1] di played on the creen, pre the [SET] key to tart the fir t mea urement.
Opacity Meter Opacity Meter
AC-1 ▶▶
0.0
○
%
○
m-1
○
rpm
○
℃
◉
%
◉
m-1
◉
rpm
◉
℃
⑥ LED i on and ound i ringing.
⑦ The maximum value i held and di played.
⑧ Step on the accelerator pedal.
⑨ Pre the [SET] key to di play [AC-2] on the creen, and pre the [SET] key to tart the econd
mea urement.
Opacity Meter Opacity Meter
AC-1 ▶▶
0.0
○
rpm
○
℃
◉
%
◉
m-1
◉
rpm
◉
℃
⑩ Repeat the proce ⑥~⑨.
⑪ the mea urement i complete, pre the [PRINT] key to print the recorded value.
⑫ Remove the probe from the exhau t pipe. (※ the probe tip i hot, o be careful of burn .)
9/15

KS200-DM Manual
Opacity Dust Meter
2021/1/1 AM 09:00:00
Free Acceleration Test
Opacity(%) RPM
1th 5 2 3750
2th 4 8 3550
3th 4 5 3300
AVR 4 8 (30 0)
Diff 0 7 ( 5 0)
-----------------------------
Judgment PASS
-----------------------------
(ex Print)
. MODE
5-1. Key & Function
NO. NAME Description
①SELECT U e it when choo ing a menu or additional function.
②UP Increa e Value
③Down Decrea e Value
④ ESC U e when ending the election mode and witching to the
mea urement mode.
⑤SAVE Enter mea urement mode and Save the data
*Each time you pre SET key, the mode change in the order below.
10/15

KS200-DM Manual
NO.
NAME
Description
①CAL Set Zero, Span
②DATE Set Date
③HOLD HOLD Mode
④VER Di play ver ion
⑤TEST Di play AD value
⑥ PC Di play value at PC
⑦ RPM Choo e RPM (Factory mode)
5-2. Set Zero, Span
① Pre the SET key in the mea urement mode to move to the "CAL".
②Pre the SELECT key to perform the ZERO function and then the SPAN function.
5-3. Set DATE
① Pre the SET key in the mea urement mode to move to the "DATE".
②Pre the SELECT key to enter the etting mode, u ing ▼ , ▲ key et up and move on next.
Set up Year → Date → Time .
5-4. Set Hold mode
① Pre the SET key in the mea urement mode to move to the "HOLD".
② Pre the SELECT key to enter the etting mode, u ing ▼ , ▲ key et up and pre SET key
again di play “SAVE” and move on mea urement creen .
Peak HOLD ←→ HOLD
* Peak Hold - Keep updating in real time during mea urement.
* Hold –Fix the mea ured value when you pre the [HOLD].
11/15

KS200-DM Manual
6. Remote Controller (Options)
Remote control can operate the mea uring in trument from
a long di tance. It help omeone who operate the mea uring
in trument in the driver' eat.
6-1. Connet remote controller
Connect the remote control to the remote control terminal
by referring to the chapter 2 configuration diagram.
6-2. How to u ing remote controller
① Pre the remote control on/off key for about 3 econd while the
mea uring in trument i turned on.
② The power i turned on with the ound.
③ The mea urement value are di played on the remote control and can be operated in the ame
way a the operation key of the mea uring in trument.
④ Pre the remote control on/off key for about 3 econd to turn off the power.
7. Others
7-1. Len
If the len get contaminated, it can cau e an error in mea urement, o with a oft and clean
cloth,
Clean both len e regularly.
12/15

KS200-DM Manual
7-2. Probe
① Be careful not to touch the end of the probe tip.
② After u e, the probe tip i hot, o be careful of burn .
7-3. About error Code
NO. Error Check point
①Zr Er
Sp Er
It’s Zero error.
Make sure the probe end is placed in a contaminated place. Make sure the
measuring cell is contaminated and wipe it clean.
Check and setting Zero and Span again
② T Er Temperature sensor error.
③ R Er RPM sensor is not connected
13/15

KS200-DM Manual
7-4. Accessories
⑨ ① ② ⑤ ⑥
③ ④ ⑩ ⑧ ⑦
NO. Part No. NAME De cription
① DP-001 Power cable Power cable
② DP-002 Probe Smoke probe
③ DP-003 Probe handle Smoke probe handle
④ DP-004 Ho e band Probe ho e band
⑤ DP-005 Probe ho e Probe ho e-1.5M
⑥ DP-006 Remote Control Remote Controller-10M (Option)
⑦ DP-007 Printer Printer (Option)
⑧ DP-008 Printer roll Printer roll - 5pc
⑨ DP-009 RS 232 RS232 -10M (Option)
⑩ DP-010 Standard Filter Standard Filter- 40,60,80% (Option)
14/15

KS200-DM Manual
15/15
Table of contents
Other KS RnD Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Gossen MetraWatt
Gossen MetraWatt mavolog pro operating instructions

KROHNE
KROHNE ALTOSONIC V installation manual

Agilent Technologies
Agilent Technologies B1500A Training manual

MRC
MRC PH-230SD Operation manual

Sper scientific
Sper scientific 850026 manual

Cooper Instruments & Systems
Cooper Instruments & Systems DPG 100 user guide