Labom BE Series User manual

Pressure gauges with switch function
Type series BE, BR, BF, BG, BN, BP
LABOM Mess- und Regeltechnik GmbH Im Gewerbepark 13 27798 Hude Germany
BA_037_2019-05_10.00
Page 1/5
Operating Instructions
1General Information.......................................................................................................2
1.1 Intended Use...........................................................................................................2
1.2 General Safety Notes..............................................................................................2
1.3 Conformity with EU Regulations..............................................................................2
1.4 ATEX Approval........................................................................................................2
2Transportation and Storage..........................................................................................2
3Installation and Commissioning ..................................................................................2
3.1 Mechanical Installation............................................................................................3
3.2 Electrical Connection ..............................................................................................3
3.3 Adjusting of contacts...............................................................................................4
3.4 Additional requirements relating to type BN4xxx pressure switches and type
BP4xxx pressure limiters.........................................................................................4
4Operation........................................................................................................................5
4.1 Zero-point correction...............................................................................................5
4.2 Maintenance / Service.............................................................................................5
5Disassembly...................................................................................................................5

BA_037_2019-05_10.00 Pressure gauges with switch function
Page 2/5
1 General Information
This document contains necessary information for the proper installation and use of this
device. In addition to this instruction, be sure to observe all statutory requirements, appli-
cable standards, the additional technical specifications on the accompanying data sheet
(see www.labom.com) as well as the specifications indicated on the type plate.
1.1 Intended Use
The device is intended to measure pressure of gases, vapors and liquids as specified in
the data sheet.
1.2 General Safety Notes
The installation, set up, service or disassembly of this device must only be done by
trained, qualified personnel using suitable equipment and authorized to do so.
Warning
Media can escape if unsuitable devices are used or if the installation is not
correct.
Danger of severe injury or damage
Ensure that the device is suitable for the process and undamaged.
1.3 Conformity with EU Regulations
The CE-marking on the device certifies its compliance with the applicable EU Directives for
placing products on the market within the European Union.
You find the complete EU Declaration of Conformity (document no. KE_001 (für BE4xxx,
BR42xx, BG3xxx), KE_012 (für BP4xx0), KE_014 (für BN4xx0) or KE_039 (für BE2xxx,
BF2xxx, BG2xxx)) at www.labom.com.
1.4 ATEX Approval
If you purchased a device with ATEX approval, please refer to the accompanying docu-
ment XA_013, XA_014 or XA_021 for ATEX-relevant information.
2 Transportation and Storage
Store and transport the device only under clean and dry conditions preferably in the origi-
nal packaging. Avoid exposure to shocks and excessive vibrations.
Permissible storage temperature: -40…70 °C
S3 models with case filling
Permissible storage temperature: -20…60 °C
3 Installation and Commissioning
Ensure that the device is suitable for the intended application with respect to pressure
range, overpressure limit, media compatibility, temperature range and process connection.

BA_037_2019-05_10.00 Pressure gauges with switch function
Page 3/5
3.1 Mechanical Installation
Use gaskets, if required, that are suitable for the process connection and resistant to the
media.
Before starting operation, check the process connection carefully for leaks under pressure.
Install safety pressure gauges in the line with DIN EN 837-1 S1 and S3 so that the exhaust
equipment can freely discharge to the rear in the event of a fault or malfunction. Provide
for corresponding openings if mounting the unit on the wall.
Use the spanner flats to screw in the device.
Measuring devices where oil or grease residues in the measuring element are not permit-
ted are marked on the scale with the oil can icon.
Measuring devices for use with oxygen are marked on the scale with the oil can icon
and the word “oxygen” directly below this.
Seal cylindrical screw threads by fitting a sealing washer to the front face of the gasket.
Seal taper screw threads by screwing tight; typically a gasket material is applied to the
male thread. Make sure that the set of gasket you use are correct and in perfect condition
when mounting the instrument; non-matching gasket can cause faults.
Vent the pressure gauge after installation using the vent
valve (Figure 1).
The devices are supplied with a closed vent valve
(CLOSED position). Moving the red lever to the OPEN
position equalises the pressure level inside and outside
the case (Figure 2).
Figure 1: Vent valve
Figure 2: Move valve
3.2 Electrical Connection
Complete the mechanical installation before you connect the device electrically.
Set up all electrical connections while the voltage supply is switched off.

BA_037_2019-05_10.00 Pressure gauges with switch function
Page 4/5
3.3 Adjusting of contacts
Setting the contact can be done by using the adjustment lock in the centre of the sight
glass. Just press down the separate or permanently mounted key (Figure 3) until the
pointer of the lock catches the adjusting pin of the contact (Figure 4).
Figure 3: Key
Figure 4: Setting the contact
The set pointer can be adjusted over the entire scale range. It must be ensured that the
contact will be set in a clockwise direction only. If the contact is accidentally set behind the
desired value, turn back the contact at least 5% below the desired value and make the
setting again in a clockwise direction.
The functionality and thus the switching reliability of the built-in switch contacts is guaran-
teed within the entire nominal range. Following valid standards though the accuracy stated
in the data sheet is assured for switching points that are set between 10% and 90% of the
nominal range only.
Operating the contact outside of the scale range is not permitted.
You can find additional information on contacts in the document TA_039, available on our
website at www.labom.com.
3.4 Additional requirements relating to type BN4xxx
pressure switches and type BP4xxx pressure limiters
The evaluation unit and downstream control current circuit are not part of our scope of
supply.The following evaluation units conform to the requirements of the Association of
Technical Inspection Agencies (VdTÜV) Fact Sheet No. 100.
Pepperl+Fuchs, switching amplifier:
Type KHA6-SH-Ex1, PTB 00 ATEX 2043
Type KFD2-SH-Ex1, PTB 00 ATEX 2042
The use of alternative evaluation units is within the responsibility of the operator. The op-
erator is also responsible for the control current circuit and the locking switch required with
pressure limiters, that comply with the Association of Technical Inspection Agencies
(VdTÜV) Fact Sheet No. 100.

BA_037_2019-05_10.00 Pressure gauges with switch function
Page 5/5
4 Operation
During operation, take care that the device remains within its intended pressure and tem-
perature ranges. No other monitoring is necessary.
The permissible ambient and media temperature depends on the type of device and its
design. This information can be found in the relevant data sheet.
Restricted ambient conditions may apply in explosion protected environments (see
XA_013, XA_014 or XA_021).
4.1 Zero-point correction
Small measuring errors or deviations
caused by difference in level between pres-
sure gauge and point of measurement can
be corrected on measuring devices with
micro control position pointers. To do so
turn the adjusting screw on the pointer hub
(see figure 5).
Similarly, a displacement of the zero point
caused by use and long service life can be
corrected if necessary.
Figure 5: Zero-point correction
4.2 Maintenance / Service
When properly installed in accordance with applicable specifications, this device is
maintenance-free. However, we recommend an annual recalibration of the device.
5 Disassembly
When measuring hot media, make sure that the device has cooled down prior to any dis-
mounting or wear appropriate protective clothing to avoid burns.
Switch off the power supply to the device before disconnecting the electrical connections.
Once this is done, the device may be mechanically removed.
Warning
Opening pressurized lines might cause severe injuries.
Danger of severe injuries or damage
Relieve the process pressure before attempting to remove the device.
Shut off the pressure supply for all feed lines to the device and relieve the
pressure in them.
Warning
Hazardous deposits and residues might remain on opened process connec-
tions and removed devices.
Danger of injury
After the device has been removed, seal off the measuring point and
mark the open process connection accordingly. Consider a possible dan-
ger due to residues when handling the removed device.
This manual suits for next models
5
Table of contents
Other Labom Measuring Instrument manuals