Labom GV4610 Series User manual

Temperature transmitter clamp-on
for temperature measurement on pipes,
Type series GV4610
LABOM Mess- und Regeltechnik GmbH Im Gewerbepark 13 27798 Hude Germany
BA_084_2019-05_10.00
Page 1/20
Operating Instructions
1General Information.......................................................................................................2
1.1 General Safety Notes..............................................................................................2
1.2 Intended Use...........................................................................................................2
1.3 Conformity with EU Regulations..............................................................................2
2Transportation and Storage..........................................................................................2
3Installation and Commissioning ..................................................................................2
3.1 Mechanical Installation............................................................................................2
3.2 Mount / Dismount Display .......................................................................................4
3.3 Electrical Connection ..............................................................................................5
3.4 Device orientation ...................................................................................................5
3.5 Adjustment of the Display Contrast.........................................................................5
4Operation........................................................................................................................6
4.1 Calibration...............................................................................................................6
4.2 Maintenance / Service.............................................................................................6
5Disassembly...................................................................................................................6
6User Manual...................................................................................................................7
6.1 Basics of the Operating Concept.............................................................................7
6.2 Display Mode / Measured Value Display.................................................................8
6.3 Menue Mode / Operating Menue...........................................................................11
6.4 The Menue Tree....................................................................................................13

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 2/20
1 General Information
This document contains necessary information for the proper installation and use of this
device. In addition to this instruction, be sure to observe all statutory requirements, appli-
cable standards, the additional technical specifications on the accompanying data sheet
(see www.labom.com) as well as the specifications indicated on the type plate.
1.1 General Safety Notes
The installation, set up, service or disassembly of this device must only be done by
trained, qualified personnel using suitable equipment and authorized to do so.
Ensure that the device is suitable for the process and undamaged.
1.2 Intended Use
The device is intended to measure surface temperatures as specified in the data sheet.
1.3 Conformity with EU Regulations
The CE-marking on the device certifies its compliance with the applicable EU Directives for
placing products on the market within the European Union.
You find the complete EU Declaration of Conformity (document no. KE_052) at
www.labom.com.
2 Transportation and Storage
Store and transport the device only under clean and dry conditions preferably in the origi-
nal packaging. Avoid exposure to shocks and excessive vibrations.
Permissible storage temperature: -40…80 °C
3 Installation and Commissioning
The device is adapted to a certain pipe diameter as specified in the order. Before installing
the device, be sure that the device is suitable for the intended process application with re-
spect to pipe diameter, ambient and medium temperature.
3.1 Mechanical Installation
Mount the device on a clean and straight pipe section free of imperfections like nicks or
burrs.
Mount the measuring insert at the bottom side of the pipe, if it is not completely filled with
medium.
First mount the clamping element, then install the measuring insert and finally make the
electrical connections.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 3/20
3.1.1 Mounting the clamping element
There are two options how to connect the device to the process:
Figure 1: mounting options: a) clamping block, b) clamping shoe
3.1.1.1 Mounting the clamping block
Required tools:
Hexagon socket wrench 3 mm (mounting screws)
Hexagon socket wrench 2 mm (vibration protection)
Turn back the vibration protection, so that it
does not exceed the semi-cycle groove for the
pipe (see picture to the right).
Mount the clamping block with the included
screws at the intended position on the pipe. Do
not exceed the allowed tightening torque of 2
Nm for pipe diameters up to 17,2 mm and 4
Nm for bigger pipes.
Hand-tighten the vibration protection. Do not
tighten the vibration protection screw too hard.
You might damage the clamping block or the
pipe.
Figure 2: Clamping block from below
3.1.1.2 Mounting the clamping shoe
Required tools:
depends on the used hose clamps
Mount the clamping shoe with two appropriate hose clamps or tightening straps.
The maximum width for a safe fit is 5 mm.
Tighten the clamps so that the clamping shoe cannot be moved by hand applying a
usual hand force.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 4/20
3.1.2 Mounting the measuring insert
The mounting procedure for the measuring insert is identical for all clamping elements.
First apply heat conductive paste to the tip of the measuring insert. Heat conductive paste
is mandatory for this type series for a good measuring accuracy.
Now insert the measuring insert into the bore in the clamping element. You can position
the insert in two ways. Choose the position that simplifies the later cable routing.
Finally tighten the connection by turning the screw connection (see figure 3). No tools are
required for this step. Never try to turn the housing, if you have a device with field housing
or integrated transmitter. You might destroy the device by this.
Do not turn or move the clamping element with a mounted measuring insert. You might
damage or destroy the insert.
Figure 3: Mounting steps for the measuring insert
3.2 Mount / Dismount Display
In order to dismount the display, turn off the device cover and pull out the display. The
three small studs at the outer diameter of the display (all 120°) facilitate removal.
The display cable can be disconnected either at the display or at the clamp module. When
reconnecting the cable, please take polarity (red lead) into account. Otherwise the cable
will not fit into the plug.
When mounting the display, you can choose its orientation. Complete the process by firmly
reattaching the cover.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 5/20
3.3 Electrical Connection
Complete the mechanical installation before you connect the device electrically.
Set up all electrical connections while the voltage supply is switched off.
Output (2-wire) 4...20 mA (20...4 mA)
Permissible supply voltage UV= 12…30 VDC
Permissible load RB≤ (UV- 12 V) / 22 mA
Circular connector M12
Cable gland
Figure 4: Options for the electrical connection
Pay attention to the following points regarding electrical connections via cable gland:
The display can be dismounted as described in Chapter 3.2. Use a small screwdriv-
er to press the spring of the terminal block completely downwards before inserting
the cable. Otherwise, it is not possible to ensure a safe electrical connection.
Do not forget to tighten the cable gland after the electrical connection is finalised.
3.4 Device orientation
The case can be turned in relation to the process connection as well as the display in rela-
tion to the case.
Turning the case in relation to the process connection (±170°) allows to choose the posi-
tion of the electrical connection. Turning the display in relation to the case allows to
choose the viewing angle independently of the position of the electrical connection.
3.5 Adjustment of the Display Contrast
The display contrast can be adjusted.
Press and hold the ESC key and use the arrow keys to adapt the contrast.
The setting is saved and the contrast screen closes as soon as the ESC key is released.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 6/20
4 Operation
During device operation, take care that the device remains within its intended temperature
range. No other monitoring is necessary.
Permissible media temperature: -40…150 °C
Permissible ambient temperature: -20…80 °C
4.1 Calibration
We recommend an annual recalibration.
Uninstall only the measuring insert. This way the measuring position remains unchanged
as it is defined by the clamping element.
Apply new heat conductive paste when you reinstall the measuring insert after the calibra-
tion. This ensures a good and constant thermal coupling between insert and pipe.
For further details please consult the separate calibration guideline for the Clamp-on family
(TA_008).
4.2 Maintenance / Service
When properly installed in accordance with applicable specifications, this device is
maintenance-free. However, we recommend an annual recalibration of the device.
In case of damage or defects, the customer can replace the following elements:
Display module
Cable glands (if applicable)
Clamping element
For defects to other components, repairs must be performed in the factory.
5 Disassembly
When measuring hot media, make sure that the device has cooled down prior to any dis-
mounting or wear appropriate protective clothing to avoid burns.
Switch off the power supply to the device before disconnecting the electrical connections.
Once this is done, the device may be mechanically removed.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 7/20
6 User Manual
The device can be configured via the display module as well as the HART protocol. The
following pages describe operation and configuration of the device using the display mod-
ule.
An overview of the menue structure can be found on the last page of this document.
6.1 Basics of the Operating Concept
The display module consists of a dot-matrix
display with 80x128 pixels as well as a
4-button control panel.
The four buttons below the display allow an
intuitive operation of the device. The gen-
eral functionality of the buttons is identical in
all operating modes.
Figure 5: Control elements
If you press and hold the up or down button, it will automatically be triggered multiple
times. This allows to easily navigate through longer selection lists. If you press and hold
the ESC button, you always return to the measured value display.
Button
Function
/
Select function, increase/decrease value
OK
Confirm selected function or selected value
ESC
Cancel action
ESC long
Return to measured value display
Table 1: General button functions
The structure of the display is the same in every operating mode. The display area is di-
vided into three zones:
Header
Icon indicating device status
Data area
Figure 6: Display structure

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 8/20
The icon for the device status (see 6.2.3) is displayed in each operating mode. The con-
tents of the header and the data area depend on the operating mode:
Display of measured value
Header: Icon description, if applicable. Otherwise "Value"
Data area: Measured values and parameters according to the selected screen
mode (see 6.2.4)
Display of device data (see 6.2.1)
Header: Title for the displayed device data
Data area: Device data
Operating menue
Header: Selected menue item
Data area: Submenue or dialogue during setting procedure
6.2 Display Mode / Measured Value Display
After applying the supply voltage and completion of the initialisation, the device switches to
the display mode. In the display, the currently measured value is shown.
If the measured value is larger than the displayable number due to setting a fixed decimal
point or an inappropriate unit, the largest displayable number is shown flashing. In this
case please select a different setting of the decimal point or a different unit.
In display mode you can review the device data, enter the menue or lock/unlock the
menue.
Button
Function
/
Scroll through the pages with device data
ESC
Return to measured value display
OK
Go to the operating menue
ESC+OK long
Activate / deactivate menue lock (see 6.3.2)
Table 2: Button functions in display mode
6.2.1 Quick Access to Device Data
You can access a number of device parameters directly from the measured value display
using the / -buttons. This allows a quick overview of the device configuration.
With the -button you can display variable data (min/max values), with the -button
static information such as configuration data.
From any screen of the device data, you can enter the operating menue with OK and go
back to the measured value display by pressing ESC.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 9/20
The sequence of the screens with device data is as follows:
Min/Max values (measurement)
----- Measured value display (starting point) ---
Temperature measurement (senor type)
Current output (characteristic curve, limits, measuring range)
HART data (address, tag, descriptor)
Device identification (device ID, order number, serial number)
Module information (hardware and software versions, serial numbers)
6.2.2 Locking the Menue
You can lock the menue with a key combination at the device. Press and hold the ESC-
key and press then OK to activate the menue lock. Hold down both keys until the display
confirms that the menue lock has been activated. You can still access the device data but
the menue is now locked.
An active menue lock is indicated by the header text „Write protection“ in combination with
the lock icon.
Use the same key combination to unlock the menue again.
6.2.3 Icons indicating device status
In the top right corner of the display, depending on the device status an NE107-conform
icon is displayed. The following icons are defined:
Error/failure
Critical error, alarm current is activated.
The error description is indicated in the display. The device
restarts after confirming the error.
Warning
Temperature limit is exceeded.
Limit reached
The output current reached the pre-set upper or lower cur-
rent limit.
Function control
The output current does not correspond to the measured
value due to an active simulation.
Write protection
The write protection of the device was activated via HART
or with the key combination in the display.
Table 3: Icons for device state
The list of icons is sorted in descending priority. Only the icon with the highest priority is
shown. The two most important icons for fault and/or warning are flashing when displayed.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 10/20
6.2.4 Display layouts
You can configure the layout of the measured value display as well as the displayed infor-
mation individually. There are four different layouts available:
Designation
Layout
Description
Example
Four values
Under the main value, three additional
values are shown. The 4th value can
use the entire display width.
Three val-
ues
Under the main value, two additional
values are shown. Both of them can
use the entire display width.
Two values
Two values are shown in the same
size, one under the other.
Large dis-
play
The main value is shown at the maxi-
mum size (best readability from a dis-
tance). One additional value can also
be shown.
Table 4: Display layouts
You can define what information shall be shown in the layout’s placeholders.
You can select the 1st value (main value) from the following data:
1. Temperature with the selected unit
2. Pressure in per cent of the measuring range
3. Pressure in milliampere
For all further values, you can additionally choose from the following data:
Bar graph
Device ID (see 6.4.9.1)
HART tag
HART descriptor
When information (such as the device ID) cannot be displayed in a short layout placehold-
er, "###" appears on the screen. Then select another layout or assign the value to a longer
layout placeholder.
You can configure the screen mode in the menue "Display" submenue "Screen mode"
(see 6.4.4.2).

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 11/20
6.3 Menue Mode / Operating Menue
Press OK in the measured value display to go to the operating menue. Then the main
menue appears in the display.
In the operating menue you can navigate in the menues by using the arrow buttons. The
selected menue item is indicated by triangles on the left and right. The OK button selects
the menue item or switches to the corresponding submenue. You can return to the higher-
level menue by pressing ESC. From the main menue, pressing ESC returns to the meas-
ured value display.
From every position in the operating menue, you can return to the measured value display
by pressing and holding the ESC button (for at least one second).
Figure 7: Navigating the operating menue
In long menues and selection lists, a scroll bar on the left side shows the position of the
currently selected item.
Menue items that open a setting dialogue differ from submenues by three trailing points,
e.g. "Lower range …" (calling a setting dialogue) and "System" (calling a submenue).
If no button is pressed in menue mode for five minutes, the device automatically returns to
the display mode without saving any values.
Button
Function
Scroll up in the menue, increase value/position in list
Scroll down in the menue, decrease value/position in list
OK
Select menue, confirm value/list entry
ESC
Cancel the data entry or menue selection, return to the next higher
menue
ESC long
Cancel menue mode, return to display mode
Table 5: Button functions in the operating menue

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 12/20
6.3.1 Displaying and entering parameters
When entering parameters, either numerical inputs or a selection lists with fixed options is
available.
In general, the actual selection will be displayed first (view mode). Press OK to switch to
edit mode to change the parameter. After this is done, the display will then switch back to
view mode so that you can check the new setting.
6.3.1.1 Selection of a value from a selection list
Up to four options are shown at the same time in one selection list. When the selection list
is longer, a scroll bar on the left side of the display indicates the position of the selected
item in the selection list.
When you press the ESC button, you cancel the entry and go back to view mode. The se-
lected value is not saved.
With the OK button, the selected value is saved. Saving the value is confirmed with an in-
formation window and the new value is shown in view mode.
With ESC, you leave the view mode and return to the menue.
Figure 8: Procedure for value selection (e.g. changing language from english to german)
6.3.2 Setting a numeric value
When setting numeric values, the screen shows the following elements (from top to bot-
tom):
Designation of the parameter that can be set (in header)
Help text (if applicable)
Numeric value and unit
Function of buttons (if applicable)
After selecting a menue item for numerical entry (e.g. Upper range value), at first the value
is displayed only. The numeric value itself is shown in square brackets and its unit behind
it or in the lower right area. Unused leading digits are marked with bottom lines. You need
to press OK again to enter the edit mode.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 13/20
Numeric values are entered digit by digit. First, always the leftmost digit is selected (visible
with a triangle above and below the number). By pressing OK, you go to the next digit.
Figure 9: Elements when setting a numeric value
You change the selected digit by pressing the or button. The higher value digit is
also increased or decreased when passing zero. For instance, you can easily go from 100
to 90 without having to edit two digits.
Lower value digits are not influenced, unless the parameter limit is reached. The value is
then set to the parameter limit.
You can set negative numbers by decreasing the value below zero.
When you press the ESC button, you can cancel the entry at any time and return to the
display of the set value. Any change of the value is not saved.
With the OK button you confirm the set digit value and jump to the next digit.
When you confirm the rightmost digit, the entire value is saved. You can save the new val-
ue from any digit by pressing and holding the OK button.
Figure 10: Procedure for setting a numeric value (e.g. from 1.0 to 0.9)
6.4 The Menue Tree
In the following, the display and adjustment options are described by their position in the
menue tree. An overview of the menue tree can be found on the last page of this docu-
ment.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 14/20
6.4.1 Main menue
The main menue has the following entries:
Menue entry
Description
Quick setup
Selection of the most important settings
Adjustment
Adjustment functions for temperature measurement and current
output
Display
Functions for configuring the display
Measurem./Output
Configuration of the temperature measurement and current out-
put
Diagnosis
Diagnostic information such as min/max values
Simulation
Simulation of temperature and current for function check of the
measurement chain
Communication
Information and settings regarding HART communication
System
Device data and function such as factory reset
Tabelle 6: The main menue
6.4.2 "Quick Setup" menue
In Quick Setup, basic configuration options are combined to make it possible to quickly
configure key functions. All functions of the quick setup can also be found at another posi-
tion in the menue tree.
The following functions are available in the Quick Setup menue:
Menue entry
Description
Language
Select menue language
Temperature unit
Select unit for temperature (see 6.4.4.1)
Lower range val.
Setting of the temperature value that should correspond to 4 mA
(start of range) (see 6.4.5.2)
Upper range val.
Setting of the temperature value that should correspond to 20 mA
(end of range) (see 6.4.5.2)
Device ID
Setting the device ID (see 6.4.9.1)
Table 7: "Quick Setup" menue
6.4.3 "Adjustment" menue
The following functions are available for the temperature adjustment:
Menue entry
Description
Lower adjust.
Offset correction with applied reference temperature
Upper adjust.
Span correction with applied reference temperature
Table 8: "Adjustment" menue

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 15/20
6.4.3.1 Upper and lower adjustment
The lower adjustment results in an offset of the characteristic curve. It thus affects the
measuring range.
The upper adjustment changes the slope of the characteristic curve by correcting the span
of the measuring range.
Execute the lower adjustment prior to the upper adjustment for a correct full adjustment.
Figure 11: Effect of the upper and lower adjustment on the characteristic curve
You can perform the upper and lower adjustment at any reference temperature level. For
instance, you can perform the lower adjustment of a -40…150 °C device at 0 °C. The ref-
erence temperature level for the upper offset can also be freely chosen. To ensure that the
adjustment is as precise as possible, the upper and lower adjustment should be performed
as close as possible to the lower range value and upper range value of the measurement
ranges.
6.4.3.2 Current adjustment
You can use the current adjustment, if the reading at the end of the measurement chain
(re-converted current value) does not correspond to the measured pressure. With this
function you can compensate deviations in the output stage as well as the downstream
measurement chain.
Proceed as follows (example for 4 mA):
Select function "Current adjust." –> "Adjust. 4 mA"
Use "OK" to activate constant current mode (4 mA). CAUTION! The output current
value now no longer corresponds with the measured value! This is indicated by the
icon "Function check" (see 6.2.3).
Read the displayed current value at the end of the measurement chain.
Enter this current value (e.g. 3.996) at the device. The device now corrects the cur-
rent output so that 4 mA are shown at the end of the measurement chain.
When leaving the function, the constant current mode is disabled and the current value
corresponds again to the measured value.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 16/20
6.4.4 "Display" menue
In the "Display" menue, you find all the settings that affect the display on the screen.
Menue entry
Description
Language
Select menue language
Units
Setting units for the different measured and displayed values
Screen mode
Configuration of the screen layout and content (see 6.2.4)
Decimal point
Select setting of decimal places of the 1st value by specifying the
decimal point
Backlight
Switch backlight on/off
Table 9: "Display" menue
6.4.4.1 Submenue "Units"
You can select the unit of every value shown on the display. These settings do not affect
the internal calculations of the device or HART communication.
The shown conversions are only meant for your orientation. The device uses conversion
values with ten decimal places.
Temperature unit
The unit in which the temperature is to be shown can be selected from the following list:
Unit
Description
°C
Degree Celsius
°F
Degree Fahrenheit ( TFahrenheit = TCelsius*1,8 + 32 )
°R
Degree Rankine ( TRankine = TKelvin*1,8 )
K
Kelvin ( TKelvin = TCelsius + 273,15 )
Table 10: Possible units for temperature
6.4.4.2 Submenue "Screen mode"
In the "Screen mode" submenue, you configure the representation of the measured values
and additional information on the display.
With the menue item “Screen layout” you configure the information that is displayed and its
layout. Up to four values can be displayed at the same time. In the additional menue en-
tries "1st value" to "4th value" you define the contents of the placeholders in the layout.
The various layouts as well as the possible content of the values are described in Chapter
6.2.4.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 17/20
6.4.5 Menue "Measurement/Output"
Configure the measured value as well as the current output in the "Measurement/output"
menue.
Menue entry
Description
Output function
Setting the output function (linear, inverse)
Lower range val.
Setting the temperature value that shall correspond to 4 mA (start
of range)
Upper range val.
Setting the temperature value that shall correspond to 20 mA
(end of range)
Alarm current
Select the alarm current: High (>21 mA) or low (<3.6 mA)
Minimum current
Limit of the lower output current (3.8…4.0 mA)
Maximum current
Limit of the upper output current (20…21 mA)
Table 11: "Measurement/Output" menue
Independent from the setting of the minimum and maximum current limit, the set measur-
ing range always corresponds to a current range of 4…20 mA.
6.4.5.1 Output function
You can set the output function proportionally rising to the measured value (selection "lin-
ear") or proportionally falling (selection "inverse").
6.4.5.2 Setting the measurement range
You can freely set the lower and upper range value within the nominal range of the device.
Please note that when the turndown is increased, the measurement error also increases. If
you use a very small range you must therefore accept a significant measurement error.
We recommended a maximum turndown of 10:1.
The lower range value must be below the upper range value. For an inverse characteristic
curve, select the output function setting “inverse”.
6.4.5.3 Minimum and maximum current
In the standard setting, the current output is limited at 3.8 and 20.5 mA, meaning a further
drop or rise in the measured variable does not change the current. You can freely select
these current limits for the lower limit between 3.8 and 4 mA and for the upper limit be-
tween 20 and 21 mA.
6.4.6 Menue "Diagnosis"
In this menue you can view and configure various diagnostic information. The following
diagnostic functions are available:
Menue entry
Description
Operating hours
Display of operating hour counter
Min/Max values
Displaying and resetting the min/max values
Last error
Display and reset of the last critical error
Table 12: "Diagnosis" menue

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 18/20
6.4.7 Menue "Simulation"
In the "Simulation" menue, you can simulate the temperature as well as the current to test
the subsequent measuring chain.
Menue entry
Description
Current sim.
Setting a fixed current value
Temperature sim.
Setting a fixed temperature value
Table 13: "Simulation" menue
6.4.8 Menue "Communication"
In the "Communication" menue the settings for the HART communication are summarised.
Menue entry
Description
HART address
Setting the HART address for device identification in multi-drop
mode
Current mode
Setting the current mode (proportional/constant)
HART data
Display of HART information (HART-Tag, HART-Descriptor etc.)
Send-preambles
Setting of the number of send-preambles for
HART communication
Table 14: "Communication" menue
6.4.8.1 HART address
This address corresponds with the short address which is used for the HART-
Communication. It can be set within a range of 0 to 63. Please note, that setting the short
address between 1 and 63 will not automatically activate the constant current mode. This
must be done under the menue “Current mode” (see 6.4.8.2).
6.4.8.2 Current mode
The current mode determines whether the output current of the device is set to respond
proportionally to the measured value (selection “proportional”) or whether it should remain
constant at 4 mA (selection “constant”). When the current mode “constant” is in use, the
measurement value can only be read using HART-commands (e.g. for HART-multidrop-
operation).
6.4.9 Menue "System"
In the "System" menue, device-relevant functions are summarised.
Menue entry
Description
Device ID
Setting the device ID (e.g. to display a free-text in the display)
Device data
Display of device data, such as from measured value display (see
6.2.1)
Factory reset
Reset to factory settings
Device reset
Restarting the device (such as after a power outage)
Table 15: "System" menue

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 19/20
6.4.9.1 Device ID
Using the device ID, you can show a custom text in the display if you configure the screen
mode accordingly (see 6.2.4). For instance, you can show the tag number continuously in
the display.
The device ID can be up to 16 characters long and consist of numbers, empty spaces,
capital letters and special characters.
For the selection, the characters are arranged in the order shown below. From the end of
the list (special character "@"), you are automatically guided back to the start (number
"0").
Character set:
0123456789 ABCDEFGHIJKLMNOPQRSTUVWXYZ[\]^_!“#$%&‘()*+,-./:;<=>?@
6.4.9.2 Factory reset
When carrying out a factory reset, all parameters are set to the state at the time of deliv-
ery. This also includes the temperature and current adjustment. Exceptions are the follow-
ing operational parameters: “Min/max values”, “HART change counter” and “operation
hours”.

BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 20/20
6.4.10 Overview with menue tree and device functions
Table of contents
Other Labom Measuring Instrument manuals