Page: 83-530-000 Rev B
12
1. Functions: 1. Voltage/current setting
2. OVP/UVL setting
3. Actual voltage/current readback
4. Voltage/current setting readback
5. OVP/UVL setting readback
6. Output On/Off
7. Foldback protection setting
8. Status readback (cc/cv , ovp , otp , foldback, ac fault)
9. Model identification readback
11. Calibration (not user access.) (DAC-Zero and max points , Display-Zero and max)
12. Last setting recall (at power off).
13. Re-start mode setting
14. Front Panel Lock/Unlock
2. Application software: 1.Drivers to LabView and LabWindows
2. Simple operating program
9. DIGITAL PROGRAMMING AND READBACK
1. Vout programming accuracy ---
2. Iout programing accuracy ---
3. Vout programming resolution --- 0.02% of full scale (To be confirmed)
4. Iout programming resolution --- 0.04% of full scale (To be confirmed)
5. Vout readback accuracy ---
6. Iout readback accuracy ---
7. Vout readback resolution ---
8. Iout readback resolution ---
OV Response time mS
Other Functions
10. MECHANICAL CONSTRUCTION
V 7.5 10 12.5 20 25 30 40 50 60 80 100 125 150 200 250 300 400 500 600
1. Cooling ---
2. Weight Kg/Lb
3. Dimensions ---
4. Types of connectors ---
5.Front panel material & finish ---
6. Mounting method --- Sandard 19" Rack Mount, provision for standard slides. Side/Rear Support is required; do not mount by F/P only.
7. Mounting positions --- Horizontal.
8. Output ground connection ---- M5 Stud
9. Color and finish ---
10. Output Terminal covers ---
11. ENVIRONMENTAL CONDITIONS
1. Operating temp C
0~50 C, 100% load. No operation > 50C
2. Start up temp. C 0
3. Storage temp C
4. Operating humidity % 20~80% RH Non-condensing
5. Storage humidity %
6. Altitude --- Operating:
7. Vibration & Shock (208/400VAC) G
8. Audible Noise db
RAL7035 Textured finish.
50º C up to 7500 ft., (2500m) 45º C from 7501 to 10,000ft (2501m - 3000m)
Non Operating 40,000 ft (12,000m)
2) Output: Iout =>50A: bus-bars, Iout<37.5 a threaded stud terminal.
6) IEMD Option, refer to IEMD Option Spec 08-030-100.
-20C to +70C
10~90% RH Non-condensing
4) RS232/RS485, STD Phone Jack connector.
Standard for all models with bus-bars. Closed connection for models with output connector.
5) Isolated analog option: refer to Isolated Analog option spec. 08-030-200
connector 747321. Standard 25 pin D connector.
Chassis Top/Sides/Bottom Yellow iridite. Front Panel RAL 7035 painted.
ASTM D4169, Standard Practice for Performance Testing of Shipping Containers and Systems, Shipping Unit: Single Package
Assurance Level: Level II; Acceptance Criteria: Criterion 1 - No product damage Criterion 2 - Package is intact Distribution Cycle: Cy
Air (intercity) and motor freight (local, single package up to 100 lbs.) (Note: for packages over 100 lbs, distribution cycle 12 - Air
(intercity) and motor freight (local), unitized is used)
+/-0.5% of rated output current for units with Io <187.5 +/-0.7% of rated output current for Io <187.5
43/97
0.1%+0.4% of rated output current Accuracy same as IEEE Programming Accuracy
0.02% of full scale
Fan driven, Airflow from Front to Rear. Supplemental vents on side that shall not be blocked. EIA Rack mounting,
0.1%+0.2% of rated output voltage Accuracy same as IEEE Programming Accuracy
10. Self test.
8. SOFTWARE SPEC.
+/-0.5% of rated output voltage
stackable. "Zero Stackable" top and bottom. Slides or suitable rear support required.
0.02% of full scale
20 mS maximum between output V exceeding IEEE limit and supply inhibit turning on.
65dBA at Full Load, measured 1m from Front Panel
Set Over-Voltage Limit, Set Local/Remote
Operating parameters and status available for display via host computer.
H:3U ,W: 19" Rack, See 02-530-000 for outline drawing.
1) Input: Threaded Studs and terminal cover. Strain relief optional.
3) Analog programming: DB25, plastic connector, AMP, 747461-5, Female on Power Supply, Male on Mating