Declaration of CE Compliance
This product conforms to the standards set forth in the following directives,
89/392/CEE, low voltage 73/23/CEE, electromagnetic compatibility 89/336/CEE
plus modifications. In accordance with annex V, additional documentation is
available upon request.
COD Heater Block
Code 5-0069 (110-120 V / 60-50 Hz)
Code 5-0070 (220-240 V / 50-60 Hz)
IMPORTANT: READ ALL INSTRUCTIONS IN THIS MANUAL BEFORE USE OR
MAINTENANCE OF THIS UNIT. THE MANUFACTURER IS NOT RESPONSIBLE
FOR DAMAGE RESULTING FROM IMPROPER USE OF THIS EQUIPMENT
LaMOTTE COMPANY
Helping People Solve Analytical ChallengesSM
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
5-0069 ·6/02
Warning and Instruction labels are affixed to the unit with information regarding the
potential hazards of use and maintenance. Please do not remove these labels.
Replace them if they become illegible.
WARNING – USE WITH CAUTION!
READ ALL INSTRUCTIONS BEFORE USE OR MAINTENANCE OF THE
UNIT.
SAFETY RULES
1) The heating plate may reach temperatures of up to 150°C. The green LED
indicator will be lit while the unit is heating. The unit will remain hot for a period of
time after use. Allow adequate time for the unit to cool before handling.
2) Containers, tubes and materials used with the COD Heater Block must be
compatible with the temperature ranges of this unit.
CLEANING
Always unplug the unit before cleaning. The heating plate must be cool. Use a
damp, nonflammable cloth and mild detergent.
PERSONAL PROTECTION EQUIPMENT
Use protective safety equipment appropriate for the temperature settings and
materials.
MAINTENANCE
All repairs to the unit must be made by authorized service centers. Contact your
local dealer for a list of service centers. Repairs by non-authorized service centers
will void the terms of the warranty.
WARRANTY
The warranty period begins from the date of the purchase receipt. Each unit is
registered by the serial number.
The manufacturer is engaged in the continuous improvement of this product, and
reserves the right to make modifications without notice.
SAFETY DEVICES
The heating element is equipped with a temperature-sensitive safety switch to
prevent damage to the unit from overheating. The unit will automatically shut off
when the temperature exceeds 170°C. Once the safety switch has been triggered,
the unit will no longer operate, and must be checked and reset by an authorized
service center before further use.
Electric scheme 230V/50Hz (115V/60Hz)
1 5x20mm retarded fuse 3.15A (5A)
2 Bipolar lighted switch
3 Pt100 Temperature probe
4 Resistor
5 Electronic control card
6 Safety thermostat
Instructions for use:
1. Connect the instrument to power source 230V/ 50Hz (115V/ 60Hz) by the
cable and two pin plus ground.
2. Turn unit on by pressing the main switch (A) located on the front panel. The unit
is preset to begin a heat cycle at a temperature of 148°C and a heating time
setting of 120 minutes. The temperature
(B) and time settings (C) are shown on
the LEDs on the front panel. The
heating time begins when the reactor
reaches the selected temperature,
indicated by the lighting of the green
LED (D), located on the front panel.
The unit will beep and automatically
shut off when the end of the heating
period is reached. The unit must be
turned off at the main switch (A) before
beginning a new heating cycle.
3. To adjust the temperature and heating
time, turn on the unit by the main
switch (A), and press the temperature
(B) and heating time (C) buttons to
select the required setting.
Settings:
Temperature: 100°C or 148°C
Time:
30 minutes, 60 minutes, 120
minutes, or continuous operation at
the selected temperature.
The settings for temperature or heating time may be changed at any point of
the heating cycle.
4. The heating block is equipped with two fuses, which are enclosed in the safety
plug located on rear panel. A blown fuse can be changed by removing the rear
connection socket and opening the small cover using the tip of a screwdriver.
5. This unit is designed and assembled according to safety standard IEC 1010
(1990) and Modification 1 (1991).
Power 350 W
Weight 2,15 kg
Supply 230V/50 Hz code
10.0127
Supply 115V/60 Hz code
10.1127
Verifying the temperature of heating block
The heating element in this unit is manufactured to hold the selected temperature
while minimizing temperature variations from the set point. The Pt100 temperature
probe maintains close alignment of set temperature and actual temperature of the
heating block. However, if verification is required, a thermometer may be inserted
in the opening provided in the heating block.
A
B
C
D

This manual suits for next models

1

Other LaMotte Laboratory Equipment manuals

LaMotte ColorQ User manual

LaMotte

LaMotte ColorQ User manual

LaMotte DPD 1A P-6740-G User manual

LaMotte

LaMotte DPD 1A P-6740-G User manual

LaMotte ColorQ Pro-7 User manual

LaMotte

LaMotte ColorQ Pro-7 User manual

Popular Laboratory Equipment manuals by other brands

diagenode Megaruptor 3 instructions

diagenode

diagenode Megaruptor 3 instructions

The Snell Group FLIR E30 quick start

The Snell Group

The Snell Group FLIR E30 quick start

Wavelength Electronics LD2TC5 LAB user guide

Wavelength Electronics

Wavelength Electronics LD2TC5 LAB user guide

Heathrow Scientific Mini Vortexer instruction manual

Heathrow Scientific

Heathrow Scientific Mini Vortexer instruction manual

VWR 613-2891 instruction manual

VWR

VWR 613-2891 instruction manual

Agilent Technologies PR02 Operation manual

Agilent Technologies

Agilent Technologies PR02 Operation manual

Hanil M 13 Operation manual

Hanil

Hanil M 13 Operation manual

Hettich ROTOFIX 46 operating instructions

Hettich

Hettich ROTOFIX 46 operating instructions

Flexicon FF20 Instruction handbook

Flexicon

Flexicon FF20 Instruction handbook

Cole Parmer HO-400 Series user manual

Cole Parmer

Cole Parmer HO-400 Series user manual

Curiox Laminar Wash MINI 1000 System user manual

Curiox

Curiox Laminar Wash MINI 1000 System user manual

Scientifica MTP Manual Adjustment

Scientifica

Scientifica MTP Manual Adjustment

Endress+Hauser analytikjena Biometra ThermoShaker TSC Short manual

Endress+Hauser

Endress+Hauser analytikjena Biometra ThermoShaker TSC Short manual

Getinge WD15 CLARO installation manual

Getinge

Getinge WD15 CLARO installation manual

POL-EKO SMART ST 1 instruction manual

POL-EKO

POL-EKO SMART ST 1 instruction manual

Sanyo MDF-U730 Service manual

Sanyo

Sanyo MDF-U730 Service manual

Tuttnauer Valueklave 1730 Operation & maintenance manual

Tuttnauer

Tuttnauer Valueklave 1730 Operation & maintenance manual

Thermo Scientific MaxQ 2000 CO2 Plus Operation manual

Thermo Scientific

Thermo Scientific MaxQ 2000 CO2 Plus Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.