UH50 Ultrasound Heatmeter Configuration Instructions UH 106-101c Page 9 / 13
Tariff control (optional)
The following options are available for tariff control.
Note: The summation of the quantity of heat and
volume in the standard registers is performed
independently of the tariff situation!
Threshold value tariff (tariffs T2, 3, 4, 5, 6)
Tariff registers 1, 2 and 3 can be controlled via up to
3 threshold values.
Either the quantity of heat or the volume can be
summated in the tariff registers.
The threshold values can be derived from the
flowrate (tariff T2), the heat power (tariff T3), the
return temperature (tariff T4), the flow temperature
(tariff T5) or the temperature difference (tariff T6).
Supplied quantity of heat (tariff T7)
In tariff register 1, a quantity of heat is summated
that is calculated from the flow temperature (instead
of from the temperature difference).
Returned quantity of heat (tariff T8)
In tariff register 1, a quantity of heat is summated
that is calculated from the return temperature
(instead of from the temperature difference).
Heat/cold meter (tariff T9)
In tariff register 1, the measured quantity of cold; in
tariff register 2, the measured quantity of heat is
summated. In both cases a threshold can be defined
via the flow temperature (“cold threshold”, “heat
threshold”).
Temperature above “heat threshold” and
temperature difference > +0.2 K --> quantity of heat
is acquired
Temperature below “cold threshold” and
temperature difference < -0.2 K --> quantity of cold
is acquired
Tariff control via timer switch (tariff T10)
For tariff control, one switch-off time and one switch-
on time per day can be defined. At the switch-on
time, summation of the quantity of heat or volume is
started in tariff register 1; at the switch-off time, it is
ended.
Tariff control via M-bus (tariff T11)
In tariff registers 1, 2 and 3, either the quantity of
heat or the volume can be summated. With the
relevant M-bus command, one of the 3 tariffs can be
activated or all tariffs can be deactivated.
Surcharge quantity tariff by means of return
temperature (tariff T12)
The quantity of heat is summated depending on the
return temperature in tariff registers 1 or 2.
The summated quantity of heat is calculated from
the difference of the return temperature from the
defined return temperature threshold (instead of
from the temperature difference).
Above return threshold: T1 is summated
Below return threshold: T2 is summated
Display of the tariff situation on the LCD
The current tariff status is shown in the user loop
together with the quantity of heat or the volume.
No tariff status is shown for tariffs T7 and T8.
for tariffs T2, T3, T4, T5, T6, T10, T11 and T12
no tariff register active
tariff register 1 active
tariff register 2 active
tariff register 3 active
for tariff T9 (cold/heat meter)
no tariff register active
tariff register 1 active
tariff register 2 active
The type of tariff and the associated parameters are
displayed in service loop LOOP 4.
for T2, T3, T4, T5, T6
at intervals of 2s with threshold value 1/2/3
for T7
for T8
for T9;
at intervals of 2s
for T10;
switching times at intervals of 2s
for T11
for T12
The contents of the tariff registers are displayed in
the user loop after the quantity of heat.
for tariffs T2, T3, T4, T5, T6, T10, T11 and T12
tariff register 1
tariff register 2
tariff register 3 (not for T12)
for tariff T7
for tariff T8
for tariff T9,
at intervals of 2s
Error messages
The heat meter constantly performs self-diagnostics
and can display various error messages.
Error Error / action to be taken:
code:
F0 No flow;
air in measuring unit / pipe, vent pipe
F1 Interruption of flow sensor
F2 Interruption of return sensor
F3 Electronic for temperature evaluation defective
F4 Battery empty; replace!
F5 Short-circuit flow sensor
F6 Short-circuit return sensor
F7 Fault in the internal memory
F8 F1, F2, F3, F5 or F6 pending for longer than 8 hours.
No more measurements are performed.
F9 Error in the electronics
Message F8 has to be reset in parameter setting
mode (manually, PappaWin). All other error
messages are cleared automatically once the error
has been corrected.