4.Button and Operation Instructions
1. Power on: Turn on the power, and the controller will power on. The temperature display window will display
"HY-2," the speed display window will display "8888," and the time display window will display "8888." All the
indicators will light up, and after about 3 seconds, it will enter the normal display status.
2. Temperature: Press the "Temperature" button, and the number on the right side of the temperature display
area will flash. Use the "◀" shift button, "▼" decrease button, and "▲" increase button to set the desired
temperature. Then press the "Temperature" button again to exit the temperature setting state. The set
temperature value will be automatically saved, and the instrument will operate according to this temperature
value.
3. Speed: Press the "Speed" button, and the number on the right side of the speed display area will flash. Use
the "◀" shift button, "▼" decrease button, and "▲" increase button to set the required speed. Then press the
"Speed" button again to exit the speed setting state. The set speed value will be automatically saved, and the
instrument will run at this speed value.
4. Time: If the total timing function is set to "0", it means that there is no timing function and the equipment runs
continuously. The timing function is set by referring to the description in item 8 above.
5. Sterilization: When "St=0", press and hold the "Sterilization" button for 3 seconds, the identifier "STERILIZE"
under the interface lights up, the UV lamp is turned on, and the device enters sterilization operation. Press the
"Sterilization" button again within 3 seconds, the identifier "STERILIZE" goes out and sterilization stops. The
sterilization lamp off delay setting is set by referring to the description in "Temperature - Internal Parameter
Table (1)".
6. Operation control: Press the "Run/Stop" button, the identifier "RUN" at the lower right of the interface lights
up and the equipment starts to rotate; press the "Run/Stop" button again, the identifier "STOP" at the bottom
right of the interface lights up, and the equipment stops rotating.
7. Gating: During operation, if the door is opened, the interface identifier "OPEN" lights up, the equipment stops
rotating, and when the door is closed, the identifier "OPEN" goes out and the equipment continues to rotate.
8. Lighting: When "Lt=0", click the "Lighting" button to turn the light on and off. The illumination off delay setting
is set by referring to the description in "Temperature - Internal Parameter Table (1)”.
5.Function Setting Description
1. Timing Function
When the total timing time is set to "0", it means that there is no timing function, and the device runs continuously.
When the total timing time is not "0", the timer starts according to the selected timing mode. The letter "m" (for
minutes) or "h" (for hours) will flash, and the motor will stop when the timer ends. The time display area will show
"End," and the buzzer will sound. Click the "Start/Stop" button to restart the operation.
Note:
1) If the set value of the timing time is changed during operation, the controller will restart the timer according to
the new set time.
2) The timing mode can be selected by modifying the "ndt" value in the "Time - Internal Parameter Table".
3) The timing mode can also be selected by modifying the "Hn" value in the "Time - Internal Parameter Table".
2. Refrigeration Function
The working mode of the compressor can be selected (intermittent, balanced, or disconnected), and the
compressor can be started manually or automatically during intermittent operation. For specific settings, please
refer to parameters such as "ndc," "CP," "Htd," "Crc," and "Cnp" in "Temperature - Internal Parameter Table
(2)".