Leica TP1020 User manual

Instructions for Use
Leica TP1020
Tissue Processor
Instructions for Use
Leica TP1020 V 3.0 RevG, English – 07/2016
Order No. 14 0422 80101 RevG
Always keep this manual with the instrument!
Read carefully before working with the instrument.


3
Leica TP1020
The information, numerical data, notes and value
judgments contained in this manual represent
the current state of scientific knowledge and
state-of-the-art technology as we understand it
following thorough investigation in this field.
We are under no obligation to update the present
manual periodically and on an ongoing basis ac-
cording to the latest technical developments, nor
to provide our customers with additional copies,
updates etc. of this manual.
To the extent permitted in accordance with
the national legal system as applicable in each
individual case, we shall not be held liable for
erroneous statements, drawings, technical illus-
trations etc. contained in this manual. In particu-
lar, no liability whatsoever is accepted for any
financial loss or consequential damage caused
by or related to compliance with statements or
other information in this manual.
Statements, drawings, illustrations and other
information as regards contents or technical
details of the present Instructions for Use are not
to be considered as warranted characteristics of
our products.
These are determined only by the contract
provisions agreed between ourselves and our
customers.
Leica reserves the right to change technical
specifications as well as manufacturing process-
es without prior notice. Only in this way is it pos-
sible to continuously improve the technology and
manufacturing techniques used in our products.
This document is protected under copyright laws.
Any copyrights of this document are retained by
Leica Biosystems Nussloch GmbH.
Any reproduction of text and illustrations (or
of any parts thereof) by means of print, photo-
copy, microfiche, web cam or other methods
– including any electronic systems and media
– requires express prior permission in writing by
Leica Biosystems Nussloch GmbH.
For the instrument serial number and year of
manufacture, please refer to the type plate at the
rear side of the instrument.
© Leica Biosystems Nussloch GmbH
NOTE
Leica Biosystems Nussloch GmbH
Heidelberger Str. 17 - 19
D-69226 Nussloch
Germany
Phone: +49 (0)6224 143-0
Fax: +49 (0)6224 143-268
Internet: http://www.LeicaBiosystems.com
Assembly contracted to Leica Microsystems Ltd. Shanghai.

4Instructions for Use V 3.0 RevG, English – 07/2016
Table of contents
1. Important Information ............................................................................................................................. 7
2. Safety ......................................................................................................................................................... 9
2.1 Safety notes .................................................................................................................................................. 9
2.2 Warnings ....................................................................................................................................................... 9
3. Instrument Components and Specifications..................................................................................... 13
3.1 Technical data ............................................................................................................................................ 13
3.2 Description of the instrument................................................................................................................... 15
3.3 Standard delivery—packing list .............................................................................................................. 16
3.4 Overview of the Leica TP1020 - with additional wax bath (optional accessory) ............................ 19
3.5 General overview Leica TP1020 - with fume control system.............................................................. 21
3.6 Accessories supplied together with the instrument ............................................................................ 23
3.6.1 Standard tissue basket.............................................................................................................................. 23
3.6.2 Glass beaker with holder .......................................................................................................................... 23
3.6.3 Aluminum container with holder (instruments with vacuum function only or
as an optional accessory)......................................................................................................................... 24
3.6.4 Wax bath...................................................................................................................................................... 25
4. Initial Operation ..................................................................................................................................... 26
4.1 Unpacking instructions ............................................................................................................................. 26
4.2 Installation site requirements .................................................................................................................. 29
4.3 Electrical connection................................................................................................................................. 30
4.3.1 Checking the voltage selector setting .................................................................................................... 30
4.3.2 Adjusting the voltage selector ................................................................................................................. 31
4.3.3 Connecting the power cable .................................................................................................................... 32
4.4 Switching on the instrument .................................................................................................................... 33
4.5 Setting the time .......................................................................................................................................... 33
4.6 Installing the accessories......................................................................................................................... 34
4.6.1 Installing the wax baths ............................................................................................................................ 34
4.6.2 Connecting the wax baths ........................................................................................................................ 35
4.6.3 Fastening the wax bath connection cables (instruments with fume control system only) ........... 36
4.6.4 Installing the reagent stations ................................................................................................................. 36
5. The User Interface ................................................................................................................................. 37
6. Programming the Instrument ............................................................................................................... 43
6.1 General description ................................................................................................................................... 43
6.1.1 Special characteristics of two-basket programs ................................................................................. 44
6.2 Setting-up / editing programs .................................................................................................................. 45
6.2.1 Selecting programming mode.................................................................................................................. 45
6.2.2 Selecting the program............................................................................................................................... 45
6.2.3 Selecting the number of baskets............................................................................................................. 45
6.2.4 Selecting the starting station ................................................................................................................... 46
6.2.5 Activating the vacuum function (instruments with vacuum function only)...................................... 46

5
Leica TP1020
Table of contents
6.2.6 Selecting infiltration times per station.................................................................................................... 47
6.2.7 Displaying total program duration........................................................................................................... 48
6.2.8 Leaving the programming mode .............................................................................................................. 48
7. Preparations for Tissue Processing................................................................................................... 49
7.1 Filling the reagent stations ....................................................................................................................... 49
7.2 Filling the wax baths .................................................................................................................................. 50
7.3 Inserting the tissue basket ....................................................................................................................... 52
8. Operating the Instrument in the Manual Processing Mode .......................................................... 54
8.1 Lifting and lowering the tissue basket.................................................................................................... 54
8.2 Moving the tissue basket to the next station......................................................................................... 55
8.3 Activating the vacuum function (instruments with vacuum function only)...................................... 56
9. Operating the Instrument in the Automatic Processing Mode...................................................... 57
9.1 Starting a program ..................................................................................................................................... 57
9.1.1 Immediate start........................................................................................................................................... 57
9.1.2 Delayed start............................................................................................................................................... 59
9.2 Editing and modifying programs while a processing cycle is in progress ....................................... 63
9.3 Display end of processing time................................................................................................................ 63
9.4 Pausing a process...................................................................................................................................... 64
9.5 Resuming a paused process .................................................................................................................... 65
9.6 Stopping or terminating a process.......................................................................................................... 65
9.7 End of an automatic process.................................................................................................................... 66
9.8 Remove the specimens ............................................................................................................................. 67
9.9 Finishing your daily work .......................................................................................................................... 67
10. Warning and Error Codes – Troubleshooting ................................................................................... 68
10.1 Warning codes ........................................................................................................................................... 68
10.2 Error codes.................................................................................................................................................. 70
10.3 "POWER FAILURE" and "WRONG STATION" messages ....................................................................... 72
10.4 Measures to remove specimen material in case of prolonged power failure................................. 73
10.5 "ABORT" message ...................................................................................................................................... 75
11. Cleaning................................................................................................................................................... 76
12. Maintenance........................................................................................................................................... 78
12.1 General maintenance instructions.......................................................................................................... 78
12.2 Wax bath...................................................................................................................................................... 79
12.2.1 Altering the standard working temperature .......................................................................................... 79
12.2.2 Reset after excess temperature shutdown ........................................................................................... 80
12.3 Replacing the fuses ................................................................................................................................... 81

6Instructions for Use V 3.0 RevG, English – 07/2016
Table of contents
13. Optional accessories ............................................................................................................................ 82
13.1 Three-level tissue basket.......................................................................................................................... 82
13.2 Basket removal device.............................................................................................................................. 83
13.3 Basket holder for second tissue basket................................................................................................. 84
13.4 Connecting the fume outlet tube (optional accessory for instruments
with fume control system) ........................................................................................................................ 87
13.5 Inserting the activated carbon filters (optional on instruments with fume control system) ......... 88
14. Ordering Information............................................................................................................................. 89
14.1 Accessories ................................................................................................................................................ 89
15. Appendix ................................................................................................................................................. 90
15.1 Modifying the instrument configuration................................................................................................. 90
15.2 Reagents appropriate for use with the instrument............................................................................... 93
15.3 Factory-set programs ................................................................................................................................ 94
16. Warranty and Service ........................................................................................................................... 98
17. Decontamination Certificate .............................................................................................................. 99

7
Leica TP1020
1. Important Information
(5)
ENTER
Symbols used in the text and their meaning
Instrument type
All information provided in these Instructions for
Use applies only to the instrument type indicated
on the title page.
A nameplate indicating the instrument serial
number is attached to the rear side of the instru-
ment (Fig. 1 is provided as an example only).
Dangers, warnings and cautions
appear in a gray box and are marked by
a warning triangle .
Notes,
i.e. important information for the user,
appear in a gray box and are marked
with the symbol .
Numbers in parentheses refer to item
numbers in illustrations.
Function keys to be pressed on the in-
strument touch screen are written in
bold-print capital letters.
Fig. 1
Manufacturer
Observe the instructions for use!
In vitro diagnostics (IVD) medical device
This product fulfills the requirements
of the Council's Directive 98/79/EC
concerning in vitro diagnostics (IVD)
medical devices.
Date of Manufacture
Order No.
Serial number
Symbol for labeling electrical and elec-
tronic equipment in accordance with
Section 7 of the German Electrical and
Electronic Equipment Act (ElektroG).
ElektroG is the law regarding the sale,
return and environmentally sound dis-
posal of electrical and electronic equip-
ment.
Environmental protection symbol of the
China RoHS directive. The number in
the symbol indicates the "Environment-
friendly Use Period" of the product. The
symbol is used if a substance restricted
in China is used in excess of the maxi-
mum permitted limit.

8Instructions for Use V 3.0 RevG, English – 07/2016
Qualification of personnel
• TheLeicaTP1020maybeoperatedonlybytrained
laboratory personnel.
• Alllaboratorypersonneldesignatedtooperatethe
Leica instrument must read these Instructions for
Use carefully and must be familiar with all techni-
cal features of the instrument before attempting to
operate it.
Designated use
The instrument has been designed so that it is safe to
use by the operator as well as for processing speci-
mens—provided that it is operated according to these
Instructions for Use.
The Leica TP1020 is a modular automated tissue
processor designed for the following laboratory
applications:
• Fixation
• Dehydration
• Paraffininfiltration
of histological tissue specimens.
The Leica TP1020 must be operated exclusively with
the reagents listed in chapter 15.
Any other use of the instrument is con-
sidered improper! Failure to adhere to
these instructions may result in an ac-
cident, personal injury, damage to the
instrument or accessory equipment.
1. Important Information

9
Leica TP1020
2. Safety
2.1 Safety notes
These Instructions for Use includes important
information related to the operating safety and
maintenance of the instrument.
The Instructions for Use are an important part of
the product, and must be read carefully prior to
startup and use and must always be kept near
the instrument.
2.2 Warnings
The safety devices installed in this instrument by
the manufacturer only constitute the basis for
accident prevention. Primarily responsible for
accident-free operation is above all the institu-
tion which owns the instrument and, in addition,
This instrument has been built and tested in ac-
cordance with the following safety regulations
on electrical measuring, control, regulating and
laboratory devices.
To maintain this condition and ensure safe
operation, the user must observe all notes and
warnings contained in these Instructions for Use.
the designated personnel who operates, services
or repairs the instrument.
To ensure trouble-free operation of the instru-
ment, make sure to comply with the following
instructions and warnings.
Make sure to comply with the safety instructions and warnings in this chapter.
Make sure to read these instructions, even if you are already familiar with the operation and
use of other Leica products.
For current information about appli-
cable standards, please refer to the CE
declaration for the instrument and to
our Internet site:
www.LeicaBiosystems.com
The protective devices on both instrument and accessories may neither be removed nor modi-
fied. Only authorized and qualified service personnel may repair the instrument and access the
instrument's internal components.
If additional requirements for accident
prevention and environmental protec-
tion exceeding the scope of these In-
structions for Use are imposed by laws/
regulations of the country of operation,
these Instructions for Use must be
supplemented by appropriate instruc-
tions to ensure compliance with such
requirements.

10 Instructions for Use V 3.0 RevG, English – 07/2016
2. Safety
Working with the instrument
Transport and installation
• Keepinmind Chapter 3, "Technical data"!
• Duringtransport,keeptheinstrumentupright!
• Whentransportingtheinstrument,donotliftitupbythecarouselcover!
• Donotoperatetheinstrumentinroomswithexplosionhazard.
• Caution!Thevoltageselectorisfactorypreset.Beforeconnectingtheinstrumenttothepower
supply, make sure this setting is correct for your laboratory's electrical connection values.
The socket for connecting the power cable is sealed with an adhesive tape indicating the
factory preset voltage for your particular instrument.
• Connectingtheinstrument tothe powerwith thevoltageselectorsetatawrongvalue can
cause severe damage to the instrument!
• Whenadjustingthevoltageselectorsetting,theinstrumentmustbedisconnectedfromthe
power supply.
• Theinstrumentmayonlybeconnectedtothepowerwiththecablesuppliedtogetherwith
the instrument and it may only be connected to grounded sockets.
• Thecarouselmaynotberotatedmanually!Severedamagewillresultfromdoingso!
• Caution when lowering the carousel! Keep your fingers out of the space between the
container lid and the upper rim of the container!
• Caution!Incaseofapowerfailure,thecarouselautomaticallydescendsintoastation.
• Instruments equipped with vacuum function may only be operated with the aluminum
containers supplied together with the instrument. Glass beakers used in combination with
the vacuum accessories have a high hazard potential when damaged!
• While operating the instrument, no liquid may enter in contact with any of the electrical
connections or the interior of the instrument.
• Makesuretoobservethelevelindicatorsonthereagentcontainersandwaxbaths.
• Warning!Usecautionwhenhandlingsolvents!Makesurethepremisesareadequatelyventi-
lated! Explosion hazard!
• Spilled reagents have to be wiped away immediately. In case of long-term exposure, the
instrument surfaces are only conditionally resistant to solvents.

11
Leica TP1020
Cleaning
2. Safety
Working with the instrument (continued)
• Alwaysobserveworker'sprotectionrulesanduseadequateprotectivegear(gloves,labora-
tory coats).
• Theheatedwaxbathsmayonlybeusedwithparaffin.Undernocircumstancesmaytheybe
filled with solvents. When solvents heat, a highly explosive mixture builds up!
• Caution!Theinteriorcontainersofthewaxbathsbecomeveryhotwhentheheatingfunction
is activated! Do not touch the gray upper rim of the containers with your hands! Burn hazard!
• Cautionwhenhandlinghotparaffin!Burnhazard!
• Turnoffthepowerswitchbeforecleaningtheinstrument.
• Caution!Theinteriorcontainersofthewaxbathsbecomeveryhotwhentheheatingfunction
is activated! Do not touch the gray upper rim of the containers with your hands! Burn hazard!
• Cautionwhenhandlinghotparaffin!Burnhazard!
• Whencleaningtheinstrument,noliquidmayenterincontactwithanyoftheelectricalcon-
nections or the interior of the instrument.
• Spilled reagents have to be wiped away immediately. In case of long-term exposure, the
instrument surfaces are only conditionally resistant to solvents.
• Tocleanthepaintedsurfaces,thecontainerplatformandthecontrolpanel,donotusesol-
vents containing acetone or xylene; neither use abrasive cleaning powders! Only mild
household detergents may be used! The lacquered surfaces and the control panel are not
resistant to xylene or acetone!

12 Instructions for Use V 3.0 RevG, English – 07/2016
Wax baths
Maintenance
Activated carbon filter (optional accessory)
2. Safety
• Onlyauthorizedandqualifiedservicepersonnelmayaccesstheinternalcomponentsofthe
instrument for service and repair!
• Beforereplacingthefuses,turnoffthepowerswitchandunplugtheinstrument!
• Burnt-outfusesmayonlybereplacedbyfusesofthesametypeandspecification.Forappro-
priate brands and specifications, see Chapter 3 "Technical Data".
• Iftheexcesstemperaturerepeatedly-pleasecallLeicaTechnicalService.DONOTcontinue
to use the wax bath.
• Checkifthewaxbathworkstrouble-free.Useofdefectivewaxbathshastobediscontinued
for safety reasons!
• Emptythewaxbathscompletelybeforetheparaffincools!Whencoolingdown,theparaffin
contracts, which can cause damage to the wax bath.
• Warning!Firehazard!Ifafilterbecomessaturatedwithsolvent,thereisapotentialfirerisk!
It is important that the activated carbon filters are changed at factory recommended inter-
vals.

13
Leica TP1020
3.1 Technical data
Type TP1020
Approvals ..........................................................................................................................................UL / cUL / VDE
Electrical data
Nominal voltage ...................................................................................................100 / 120 / 230 /240 V AC ±10 %
Rate frequency..........................................................................................................................................50 - 60 Hz
Power fuses (Type MDA; Bussmann) .................................................................. 2x T 10.0 A L250 V, UL listed
Nominal capacity...........................................................................................................................................700 VA
Classifications according to IEC-1010/EN61010-1:
Protective class.........................................................................................................................................................I
Pollution degree.......................................................................................................................................................2
Overvoltage category............................................................................................................................................. II
Heat emission (max.).................................................................................................................................... 700 J/s
Operating temperature range..............................................................................................................5 °C – 40 °C
Relative humidity................................................................................................................. 80 % non-condensing
Dimensions
Carousel lid ................................................................................................................................................820 mm Ø
Height....................................................................................................................................................595 - 780 mm
Pitch diameter of rollers ...................................................................................................................... 610 mm Ø
Weight
Net weight (including accessories)...............................................................................................................60 kg
Net weight, including packaging material.............................................................................................. 116 kg
Wax baths
Quantity:.............................................................................................................................................2 (optionally 3)
Capacity:...............................................................................................................................................................1.8 l
Nominal voltage / rated frequency: .......................................................................................230 V AC, 50-60 Hz
Nominal capacity per station:......................................................................................................................150 VA
Temperature setting range ....................................................................................................45 °C – 65 °C ±3 K
Excess temperature shutdown ...........................................................................................................85 °C±5 K
3. Instrument Components and Specifications

14 Instructions for Use V 3.0 RevG, English – 07/2016
3. Instrument Components and Specifications
Technical data (continued)
Reagent stations
Number................................................................................................... 10 (9 when 3rd wax bath is installed)
Capacity............................................................................................................................................................ 1.8 l
Tissue basket
Quantity:......................................................................................................................................... 1 (optionally 2)
Capacity:.............................................................................................................100 cassettes (optionally: 200)
Programs
Quantity:................................................................................................................................. 9, freely selectable
Programmable time per station.................................................................................................. 99 hrs. 59 min.
Delayed start time of processing: ..................................................................................................up to 9 days
Dripping time:...................................................................................................................................... 60 seconds
Vacuum function (types 3, 5, 7, 8 and 11 only, see Chapter 3.3)
Pressure differential..............................................................................max. 500 hPa ±10 % (approx. 0.5 bar)

15
Leica TP1020
3.2 Description of the instrument
The Leica TP1020 is an automatic tissue processor designed for laboratory
applications. It is used for the fixation, dehydration and infiltration of his-
tological tissue samples with fixatives, alcohol, solvents and paraffin wax.
The reagent stations numbered 1 - 10 are used to contain reagents. Station
10 may be replaced with an optional third wax bath. Stations 11, 12, and if
used 10, are heated, temperature controlled wax baths that can be filled
with either wax pellets or molten paraffin wax.
Embedding cassettes used to hold the tissue samples, are placed into the
tissue basket. The basic instrument is designed for a single tissue basket.
An optional second basket can be added. The basket, or baskets, are moved
clockwise from station to station.
To ensure thorough infiltration the basket containing the tissue samples
is agitated, by moving up and down, at each station. This function can be
switched off at any time.
During processing as the tissue basket moves from station to station there
is a delay period of sixty (60) seconds during which time the basket is sus-
pended above the station. Excess liquid can drip down during this process.
This ensures that there is minimal reagent carryover from station to station.
All instrument functions are activated through the control panel. Real time
is displayed via LCD. The instrument can be operated in manual and auto-
matic processing mode. Automatic processing is controlled via 9 different
programs which can be individually set up, altered and edited.
If a power failure occurs, the specimens are protected from drying out
- even when overnight processing has been selected, since in case of a
power failure the tissue basket will always be immersed into a station. Once
the power is restored, processing will be resumed where it had been inter-
rupted. After a long-term power failure, critical excess immersion time in a
station will be visually displayed.
3. Instrument Components and Specifications

16 Instructions for Use V 3.0 RevG, English – 07/2016
3. Instrument Components and Specifications
3.3 Standard delivery—packing list
The Leica TP1020 is available in eleven different configurations, which are described below. All vari-
ants also include accessories as described in the table on page 17.
Variant 1 Order No.
One basket operation / standard tissue basket.
Without vacuum / without fume control.
1 Basic instrument 14 0422 30543
12 Glass beakers with handle 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
Variant 2
Two basket operation (double-loading)/standard tissue baskets.
Without vacuum / without fume control.
1 Basic instrument 14 0422 31418
11 Glass beakers with handle 14 0422 42545
3 Wax baths, (1.8 liters) 14 0422 30665
2 Standard tissue baskets 14 0422 30585
Variant 3
One basket operation / standard tissue basket.
With vacuum / without fume control.
1 Basic instrument 14 0422 30536
10 Aluminum reagent containers for vacuum 14 0422 42545
10 Reagent container carriers 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
Variant 4
One basket operation / standard tissue basket.
Without vacuum / with fume control system.
1 Basic instrument 14 0422 30537
12 Glass beakers with handle 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
1 Tubing band clamp 14 0422 31973
If the vacuum function or extraction is required for the intended processes to be carried out
using this instrument, this information must be included in the original order.
It is not possible to retrofit any instrument variant for use with vacuum function or extraction!

17
Leica TP1020
3. Instrument Components and Specifications
Standard delivery continued
Variant 5 Order No.
One basket operation/standard tissue basket. With vacuum/with fume control system.
1 Basic instrument 14 0422 30535
10 Aluminum reagent containers for vacuum 14 0422 42545
10 Reagent container carriers 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
1 Tubing band clamp 14 0422 31973
Variant 6
Two basket operation (double-loading)/standard tissue baskets. Without vacuum/with fume control system.
1 Basic instrument 14 0422 31416
11 Glass beakers with handle 14 0422 42545
3 Wax baths, (1.8 liters) 14 0422 30665
2 Standard tissue baskets 14 0422 30585
1 Tubing band clamp 14 0422 31973
Variant 7
Two basket operation (double-loading)/standard tissue baskets. With vacuum/with fume control system.
1 Basic instrument 14 0422 31412
9 Aluminum reagent containers for vacuum 14 0422 42545
9 Reagent container carriers 14 0422 42545
3 Wax baths, (1.8 liters) 14 0422 30665
2 Standard tissue baskets 14 0422 30585
1 Tubing band clamp 14 0422 31973
Variant 8
Two basket operation (double-loading)/standard tissue baskets. With vacuum/without fume control system.
Connection for optional third wax bath includes second tissue basket carrier.
1 Basic instrument 14 0422 31414
9 Aluminum reagent containers for vacuum 14 0422 42545
9 Reagent container carriers 14 0422 42545
3 Wax baths, (1.8 liters) 14 0422 30665
2 Standard tissue baskets 14 0422 30585
Variant 9
Basic Instrument for processing with chloroform as clearing reagent
1 Basic instrument 14 0422 42238
12 Glass beakers with handle 14 0422 42545
2 Wax baths for chloroform 14 0422 32001
1 Standard tissue basket 14 0422 30585

18 Instructions for Use V 3.0 RevG, English – 07/2016
All instrument variants of the Leica TP1020 come with the following accessories:
4 Replacement fuses T 10.0 A L250 V 14 0600 00759
1 Screwdriver 5.5 x 200 14 0170 10702
1 Crank handle, size 4, mounted 14 0222 30663
1 Instructions for Use printed (German/English,
with Language CD 14 0422 80200) 14 0422 80001
3. Instrument Components and Specifications
Standard delivery continued
Variant 10
One basket operation/standard tissue basket. With fume control system and accessories.
1 Basic instrument 14 91020 US01
12 Glass beakers with handle 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
1 Tubing band clamp 14 0422 31973
1 Activated carbon filter - Standard 14 0422 30673
1 Activated carbon filter - Xylene 14 0422 30674
1 Exhaust hose, diameter 50 mm, 4 m long 14 0422 31975
Variant 11
One basket operation/standard tissue basket. With vacuum, fume control system and accessories.
1 Basic instrument 14 91020 US02
10 Aluminum reagent containers for vacuum 14 0422 42545
10 Reagent container carriers 14 0422 42545
2 Wax baths, (1.8 liters) 14 0422 30665
1 Standard tissue basket 14 0422 30585
1 Tubing band clamp 14 0422 31973
1 Activated carbon filter - Standard 14 0422 30673
1 Activated carbon filter for formaldehyde 14 0422 30674
1 Exhaust hose, diameter 50 mm, 4 m long 14 0422 31975
You will find the standard accessories and - in case you ordered them - further parts in the
upper portion of the packing box.
Compare the delivered parts with the packing list and your actual order. If you find any discrep-
ancies, please contact your local Leica sales organization immediately.
The country specific power cord needs to be ordered separately. Please find a list of all power cords available
for your device on our website www.LeicaBiosystems.com within the product section.

19
Leica TP1020
3. Instrument Components and Specifications
3.4 Overview of the Leica TP1020 - with additional wax bath
(optional accessory)
Fig. 2
Container station
platform Control panel
Wax bath
Notch for wax
bath connection
cable
Glass container
with holder
Beaker carrier with
fill level marks
Standard
tissue basket
Reagent station lid
Carousel cover
Tissue basket
holder

20 Instructions for Use V 3.0 RevG, English – 07/2016
Fig. 3 - TP1020 - rear side
3. Instrument Components and Specifications
Overview of the Leica TP1020 - with additional wax bath (optional accessory, continued)
Connections
for wax baths
(connection
socket for station
10 with cap (not
pictured) closed)
11
12
10
Socket for power cable
Mains switch
Voltage selector Wax bath plug
Continued