level 1 HOTLINE HL-90 User manual

=
OPERATING
ERES
we
ADD
WATER
m
OVER
TEMP

CORE
TREE

smiths
Smiths
Medical
ASD,
inc,
Anesthesia
and
Safety
Devices
Division
October
1,
2004
160
Weymouth
Street
.
Rockland
MA
02370
USA
Tel:
41
781
878
8011
Fax:
+1
781
878
8201
wwwsmiths-medical.
com
Dear
Valued
Customer,
"(thas
only
been
one
year
since
the
introduction
of
our
6-month
reservoir
mainteriance
protocol
for
the
HOTLINE’
Blood
and
Fluid
Warmer.
We
have
been
working
to
extend
this
protocol
to
reduce
HOTLINE”
reservoir
maintenance
from
twice
a
year
7
46
ohte
a
year,
and
to
reduce
the
maintenance
for
the
L
evel
T
Fast
Flow
Fluid
Warmer,
from
12
times
a
year
to
once
a
year”.
Smiths
Medical
ASD,
Inc.
is
pleased
to
arinounce
the
new
ONE-YEAR
MAINTENANCE
PROTOCOL
for
HOTLINE’
Blood
and
Fluid
Warmer
and
L
evel
1
Fast
Flow.
Fluid
Warmer
Benefits
for
you:
<
Maintain
a
microbe-free
environment
We
use
a
disinfecting
procedure,
unlike
hotplate
warmers
which
use
superf
ictal
cleaning
that
does
not
disinfect
or
sterilize,
.…
Maintenance
time
is
reduced
for
the
HOTLINE”
Warmer
from
twice?
a
year
to
once
a
year
and
for
the
L
evel
1
Fast
Flow
Fluid
Warmer
from-12
times
a
year
to
once
a
year
Using
hydrogen
peroxide
mixing
bottles.
will
make
for
easier
mixing,
pouring
and
storage
of
the
0.3%
hydrogen
peroxide/distilled
water
solution,
To
find
out
mioré
about
the
mixirig
bottles
please
contact
your
local
sales
representative
or
Customer
Service
at
800-258-5361,
e
Yearly
maintenance
costs
are
reduced
The
HOTLINE”
Warmer
and
the
Level
Y
Fast
Flow
Fluid
Warmer
require
cleaning
maintenance
costs
only
once
a
year,
unlike
hotplate
warmers
which
may
require
cleaning
maintenance
costs
up
to
12
times
a
year.
If
you
would
like
additional
information
regarding
the
new
one-year
maintenance
protocol
for
the
HOTLINE"
Blood
and
_
Fluid
Warmer
and/or
the
Level
1°
Fast
Flow
Fluid
Warmer,
please
do
not
hesitate
to
contact
us
at
the
below
numbers.
Technical
Support
"1-800-553-8351,
prompt
5
TECHSERV
@
smiths-medical.com
~
Jill
Goudreau
.
1-800-553-8351
х7893
jill.
goudreau@
smiths-medical.com
Poonam
Virdi,
Product
Manager
1-800-553-8351
х7851
poonam.virdi@
smiths-medical.com
Sincerely,
Ji
C,
Goudreau
Associate
Praduct
Manager,
Fluld
Warming
Data
on
file
Portex’
Bivona’
Level
7
ОНО
PIN
45-21-10
Rev,
A

48
al,
smiths
ONE-YEAR
RESERVOIR
MAINTENANCE
PROTOCOL
_The
one-year
maintenance
solution
İs
a
0,3%
hydrogen
peroxide/distilled
water
solution.
The
below
protocol
should
be
followed
annually
for
the
Level
1°
Fast
Flow
Fluid
Warmer
(H-
-1200,
H-1025,
H-1000,
H-1100),
as
wéll
as
for
the
initial
set
up
in
your
facility.
Level
1°
Fast
Flow
Fluid
Warmer
To
prepare
0.3%
hydrogen
peroxide/distilled
water
solution
for
the
reservoir,
mix
140
mi.
of
3%
hydrogen
peroxide
solution
and 1,260
ml.
of
distilled
water,
1.
¡ana
pop
Empty
reservoir
of
current
solution.
.
Place
a
contalner
under
the
drain
valve
of
the
Level
1*
Fast
Flow
Fluid
Warmer.
Drain
the
re-circulating
solution
by
turning
the
drain valve
clockwise
90
degrees.
É
When
all
of
the
re-circulatinig
solution
has
drained
from
the
unit,
close
the
drain
valve.
Refill
the
unit
with
1.4
liters
of
fresh
0.3%
hydrogen
peroxide/distilled
water
solution.
Insert
a
D-series
disposable
set
(D-50/DI-50,
D-60HL/DI-60HL,
D-100/DI-100,
D-300/DI-300)
into
the
Level
1°
Fast
Flow
Fluid
Warmer.
1f
you
use
a
Level
1°
Fast
Flow
Fluid
Warmer
for
urological
and/or
gynecological
procedures,
Insert
an
IR
-series
disposable
set
(IR-500,
IR-600AR1-600AR1-600B,
[R-700,
1R-40).
.
7.
Turn
the
unit
on
and
run
for
a
30-minute
disinfection
period.
During
the
12-month
period,
any
solution
last
from
the
reservoir
through
the
disposable
must
be
topped
off
with
the
0.3%
hydrogen
peraxide/distilled
water
solution.
Switch
the
unit
off,
discard
the
disposable according
to
established
hospital
procedures
and
empty
the
reservolr
of
the
re-circulating
solution
by
turning
the
drain
valve
clockwise
90
degrees.
When
afl
of
the
re-circulating
solution
has
been
drained
from
the
unit,
close
the
drain
valve,
Refill
the
reservoir
with
1.4
liters
of
fresh
0.3%
hydrogen
peroxide/distilled
water
solution.
The
Level
1º
Fast
Flow
Fluid
Warmer
is
now
ready
for
use.
Smiths
Medteal
ASD;
Ine,
160
Weymouth
Street
Rockland,
MA
02370
USA
Tel:
+4
784
878
8011
Fax:
+1
7818788201
www,smiths-madical.oom
PIN
46-21-1094
Rev.
A

sea
smiths
ONE-YEAR
RESERVOIR
MAINTENANCE
PROTOCOL
.
The
one-year
maintenance
solution
is
a
0.3%
hydrogen
peroxide/distilled
water
solution.
The
below
protocol
should
be
followed
annually
for
the
HOTLINE"
Blood
and
Fluid
Warmer
{HL-90),
as
well
as
for
the
initial
set
up
in
your
facility,
HOTLINE
®*
Blood
and
Fluid
Warmer
To
prepare
0.3%
hydrogen
péeroxide/distilled
water
solution
for
the
reservoir,
mix
140
mL
of
3%
hydrogen
peraxide
solution
and
1,260
mL.
of
distilted
water.
-
4
2
3,
4
5,
6,
“solution.
The
HOTLINE*
Blood
and
Fluid
Warmer
is
now
ready
for
use.
Empty
solution
currently
in
HOTLINE*
Warmer
raservolr.
Fill
the
reservoir
with
1.4
İlters
of
fresh
0.3%
hydrogen
peraxida/distilled
water
solution.
Insert
a
disposable
set
(L-70
or
L-70
Ni)
into
the
HOTLINE®
Warmer.
Turn
the unit
on
and
fet
the
solution
circulate
for
a
30-minute
disinfection
period.
Switch
the
unit
off
and
empty
the
reservoir.
Discard
the
disposable
according
to
established
hospital
procedures.
Refill the
reservoir
with
1.4
liters
of
fresh
0.3%
hydrogen
peroxide/distilled
water
During
the
12-month
period,
any
solution
lost
from
the
reservoir
through
the
disposable
must
be
topped
off
with
the
0.3%
hydrogen
percxide/distilled
water
solution.
Smiths
Medical
ASD,
Inc.
160
Woyinouth
Street
Rookdand,
MA
02870
USA
Tel:
+4
781
878
8014
Fax:
+4
784
478
8201
wewsmithe-modloutonin
.
BIN
46-21-198
Rev,
A

smiths
Smiths
Medical
ASD,
inc.
Anesthesia
and
Safety
Devices
Division
160
Weymouth
Street
Rockland
MA
02370
USA
Tel:
781
878
8011
Fax:
761
878
8201
www.enilths-medical.com
RECOMMENDED
ONE-YEAR
MAINTENANCE
CHECKLIST
Model:
Level
1°
Fast
Flow
Fluid
Warmer
Ref.
H-1000,
H-1100,
H-1025,
H-1200
‘
One-Year
Reservoir
Maintenance
Protocol
for
the
Level
1
Fast
Flow
Fluid
Warmer*
=
이
A
new
one-year
reservoir
maintenance
protocol
was
launched
in
2004
for
Level.1
Fast
Flow
Fluid
Warmer.
This
maintenance
protocol
involves
using
a
0.3%
hydrogen
peroxide/distilied
water
solution
that
maintains
a
microbe-free
environment
for
one
year.
For
details
on
this
upgraded
version
of
our
reservoir
maintenance
protocol,
please
contact
your
local
Smiths
Medical
ASD,
Inc.
representative
or
contact
our
Technical
Support
Department
at
1-800-553-8351.
.
One-Year
Maintenance
Tasks
All
testing
and
maintenance
should
be
o
performed
by
qualified
personnel
only.
Date
of
purchase:
Serial
number:
Task
Date
Date
‘Date
..
Date
Change
0.3%
Hydrogen
.
Peroxide/Distilled
Water
Solution
Replace
O-Rings
Clean
Fan
Filter
Test
Over-Temp
Alarm
Test
Float
Switch
Test
Disposable
Interlocks
Verify
Temperature
Calibration
*The
one-year
maintenance
protocol
is
in
addition
to
the
maintenance
protocols
listed
in
the
Fast
Flow
Fluid:
Warmer
operator’s
manual.
P/N
73-10-129
Rev.
A
m
Bivona’
aval

Smiths
160
Weymouth
Street
Rookland
MA
02370
USA
Tel:
781
878
8011
Fax:
781
878
8201
www.smiths-medical.com
RECOMMENDED
ONE-YEAR
MAINTENANCE
CHECKLIST
Model:
HOTLINE“
Blood
and
Fluid
Warmer
”Ref
HL-90
HOTLINE
Blood
and
Fluid
Warmer
One-Year
Reservoir
Maintenance
Protocol*
A
new
one-year
reservoir
maintenance
protocol
was
launched
in
2004
for
the
HOTLINE
Warmer
reservoir.
This
maintenance
protocol
involves
using
a
0.3%
hydrogen
peroxide/distilled
water
solution
that
maintains
a
microbe-free
environment
for
one
year.
For
details
on
this
upgraded
version
of
our
reservoir
maintenance
protocol,
please
contact
your
local
Smiths
Medical
ASD,
Inc.
representative
or
contact
our
Technical
Support
Department
at
1-800-553-8351.
One-Year
Maintenance
Tasks
All
testing
and
maintenance
should
be
performed
by
qualified
personnel
only.
Date
of
purchase:
Serial
number:
Task
Date
Date
Date
Date
Change
0.3%
Hydrogen
Peroxide/Distitled
Water
Solution
Replace
O-Ring
and
Seals
Lubricate
O-Rings
Test
Alarm
Switch
Test
Float
Switch
Test
Disposable
Switch
Test
Over-Temp
Alarm
Verify
Temperature
Calibration
*The
one-year
protocol
is
an
addition
to
the
maintenance
protocols
listed
in
our
HOTLINE®
Warmer
operators
manual.
“The
one-year
reservoir
maintenance
protocol
supercedes
tite
6-Month
maintenance
protocol.
| |
P/N
73-10-128
Rev.
A
=
Bivona’
|
af

~

INSTRUCTIONS
FOR
USE
HOTLINE®
WARMER
Table
of
Contents
ABOUT
THIS
MANUAL
K
K
tte
3
REFERENCES
eus
sucer
3
MESSAGES
...............
rr
итак
3
INDICATIONS
2...
K
erent
es
3
PRINCIPLES
OF
OPERATION
2.0...
cece
cee
eee
teen
ro
3
INFUSATE
DELIVERY
TEMPERATURES
,,,
aa
3
DESCRIPTION
.........
eee
4
HOTLINE
WARMER
............
K
arara
4
DISPLAY
PANEL
し
にし
に
に
に
に
に に
に
に に に
に に
に
に
レト
トレ
トト
ト
トト
トト
し
トト
て て
に
し
て し て て
ここ
て
て
て
て
し
て て
て
4
HOTLINE
WARMINGSET
..........
ee
eee
eee
eee
4
SAFETY
ee
K
4
CONTRAINDICAMONS
..............
1...
ear
4
WARNINGS
keen
eee
ere
skeerne
ssnne
5
CAUTIONS
..............
Ene
EEK
E
ee
5
ELECTRICALSAFETY
..............
ete
te
ee
raras
arara
arena
5
POWER
REQUIREMENT
に
に
に
に
rene
5
ELECTRICAL
SAFETY
TESTING
.................
RP
5
STORAGE
..........
e
6
PREPARATION
AND
SET-UP
..........
ete
etree
renee
eee
een
eal
6
INSTRUCTIONS
FOR
USE
........................
本
STEP
1:
HOTLINE
WARMER
SET-UP
νεο
νε
ων
ων
κενο
ον νε
χ
νε ων ον
εως
εν
7
STEP
2:
SET-UP
THE
HOTLINE
WARMING
SET
ων
ον
7
STEP
3:
CONNECT
THE
INTRAVENOUS
ADMINISTRATION
SET
.............
7
STEP
和
USE
ses
ere
resserre
7
STEP
5:
AFTER
USE
.....,.
чеки
0
κ
χε
εν
7
TROUBLESHOOTING
.............,....,..,
4444444
rr
rr
enon
eens
8
MAINTENANCE
K
K
enene
8
TESTING
K
renee
10
TEMPERATURE
VERIFICATION
..
0.0.0.0
erer
11
RECOMMENDED
MAİNTENANCE
CHECKLİST
...........
rro
12
LIMITED
WARRANTY
SECTION
..............
0mm
13
SERVİCE
...............
n
13
SPECIFICATIONS
............
L
K
nar
14
PHYSICAL
SPECİFİCATIONS
..........
0.
eee
ners
14
ELECTRICAL
SPECIFICATIONS
...........
ea
14
OPERATING
SPECIFICATIONS
.....0ooooocoocoroncor
rr
rr
14
ENVIRONMENTAL
SPECIFICATIONS
.
00...
ea
ee
14
ο
ο
αμ
rv
renee
14
ELECTRICAL
SCHEMATICS
AND
PARTS
LIST
...............,,...,.,....,,,,...,
15
SYMBOLS
...........
0...
learn
a
anana
нь
16


ABOUT
THIS
MANUAL
REFERENCES:
The HL-90
HOTLINE?
Warmer
will
be
referred
to
as
HOTLINE?
The
L
Series
Fluid
Warming
Set
will
be
referred
to
as
the
HOTLINE?
Warming
Set.
The
distilled
circulating
water
will
be
referred
to
as
the
Circ.
H,0.
MESSAGES:
Messages
that
are
headed
by
"NOTE:"
indicate
information
or
procedures
that
if
not
followed
correctly
can
cause
improper
results.
Messages
that
are
headed
by
"CAUTION:”
indicate
information
or
procedures
that
if
not
followed
correctly
can
cause
improper
results
and
damage
to
the
equipment.
Messages
that
are
headed
by
"WARNING:"
indicate
information
or
procedures
that
if
not
followed
correctly
can
cause
improper
results,
damage
to
the
eguipment,
and serious
patient
harm.
INDICATIONS
The
HOTLINE®
Warmer
is
designed
for
use
with
the
HOTLINE®
Warming
Set
to
warm
blood
and
I.V.
fluids
and
deliver
them
to
the
patient's
intravenous
access
site
at
normothermic
temperatures
under
gravity
flow
conditions.
PRINCIPLES
OF
OPERATION
The
HOTLINE®
delivers
fluids
and
blood
at
normothermic
temperatures
at
routine,
gravity
flow
rates.
All
other
fluid
warming
systems
suffer
from
cool-down
between
the
warmer
and
the
patient
connection.
HOTLINE®
overcomes
this
problem
by
providing
active
warming
of
the
patient
line
all
the
way
to
the
patient
connection,
protecting
the
patient
fine
against
exposure
to
cold
and
eliminating
patient
line
cool-down.
Active
warming
is
achieved
by
jacketing
the
sterile
patient
|.V.
line
with
a
layer
of
precisely
temperature
controlled
circulating
warm
water.
An
8'
(2.4m)
HOTLINE®
Warming
Set
allows
blood
and
intravenous
fluid
to
be
delivered.to
the
patient
at
normothermic
temperature
at
gravity
flow
rates
to
4,500
ml.
per
hour
(83mL
per
minute).
The
HOTLINE?
Warmer
and
Warming
Set
are
an
integrated,
highly
effective
design
with
unparalleled
ease
of
use.
Properly
trained
users
can
confidently
and
safely
have
the
HOTLINE®
set-up
and
operating
within
seconds.
NOTE:
The
HOTLINE’
Warmer
must
be
used
only
with
HOTLINE
Fluid
Warming
Sets.
Hotline
(HL90)}
IV.
Fluid
Warmer
Typical
Infusate
Delivery
Temperatures
at
Patient
End
of
L-70
41.9°C
water
bath
setpoint
temperature
44,0
42,0
40,0
38,0
36,0
34,0
32,0
30,0
28,0
26,0
24,0
22,0
20,0
Temperature
(*
C)
——
20"C
Infusate
一
本
一
10"Ciniusats
50
250
500 750
1000
1250
1500
2000
2500
3000
3500 4000 4500
5000
Flow
Rate
(mL/hr)
Ambient
Temperature
Range:
19.5°C
—
20.5°C

DESCRIPTION
HOTLINE®
WARMER:
An
on
board water
supply
is
heated
to
42°C
and
circulated
through
the
HOTLINE®
Warming
Set.
Electronic
circuitry
continuously
monitors
the
water
temperature.
Circulating
water
temperature
and
visual
alarms
are
indicated
on
the
Display
Panel
on
the
front
of
the
unit.
A
green
“OPERATING”
or
“WARMING*"
light
illuminates
on
this
panel
when
the
unit
is
set-up
and
operating
correctly.
DISPLAY
PANEL:
Five
displays
provide
information
for
safe,
convenient
use
of
the
device.
SYMBOL
DISPLAY
DESCRIPTION
|
Displays
temperature
of
circulating
water
in
degrees
Celsius.
“OPERATING”
|
This
green
indicator
illuminates
when
the
power
switch
is
or
“WARMING”
|
“ON”
and
the
HOTLINE
Warming
System
has
been
properly
installed.
indicates
fluid
warming
is
taking
place.
“DISPOSABLE”
|
This
red
indicator
and
pulsed
audible
alarm
indicate
the
HOTLINE
Warming
Set
is
not
properly
installed.
Fluid
warming
is
not
taking
place.
“ADD
WATER?”
|
This
red
indicator
and
pulsed
audible
alarm
indicate
additional
distilled
water
must
be
added
to
the
reservoir.
Fluid
warming
is
not
taking
place.
=
This
red
indicator
and
pulsed
audible
alarm
indicate
a
TEMPERATURE"
|
malfunction
of
the
primary
temperature
control
system.
Fluid
warming
is
not
taking
place.
HOTLINE
should
be
removed
from
service.
er
[HO
HOTLINE?
WARMING
SET:
Single
Use
HOTLINE®
Warming
Sets
are
provided
in
single
unit
packaging
with Sterile Fiuid
Path.
The
priming
volume
of
the
warming
set
is
17.5
mL.
A
plug-in
fitting
on
the
HOTLINE®
Warming
Set
engages
a
socket
in
the
HOTLINE®
Warmer.
This
is
the
only
connection
necessary
to
provide
the
warming
function.
The
HOTLINE®
Warming
Set
is
easily
unplugged
from
the
HOTLINE®
Warmer
and
discarded,
or
it
may
be
used
to
continue
subsequent
unwarmed
infusion.
SAFETY
The
HOTLINE®
employs
a
safe,
circulating
water
heating
system,
inherently
free
of
“hot
spots”
to
actively
warm
the
patient
line.
The
primary
temperature
contro!
circuit
limits
the
circulating
water
to
42°C
maximum.
In
the
unlikely
event
of
a
malfunction
of
this
circuit,
a
second
“watch-dog”
circuit
will
visually
and
audibly
alarm
and
stop
the
circulating
water
pump
if
the
temperature reaches
43°C.
Fluid
in
the
HOTLINE®
Warming
Set
is
never
exposed
fo
any
damaging
or
dangerous
temperatures
while
the
unit
is
operating.
The
HOTLINE®
is
manufactured
to
be
in
compliance
with
UL
2601-1 (HL-90
115V)
and
EN
60601-1
(HL-90
230V).
CONTRAINDICATIONS:
+
Not
for
use
in
warming
cryo-precipitates
or
granulocyte
suspensions.

WARNINGS:
Failure
to
follow
warnings
may
result
in
patient
injury
or
death.
The
HOTLINE?
Warming
Set
is
a
Single
Use
device
and
is
not
intended
for
resterilization.
Sterile
Fluid
Path.
if
caps
are
not
in
place,
then
sterility
will
be
compromised.
Set-up, priming
and
use require
aseptic
technique.
Follow
applicable
hospital
policies
and
procedures.
Prime
the
circulating
water
path
before
connecting
to
the
intravenous
administration
set.
This
is
to
confirm
that
there
is
not
a
breach
between
the
circulating
water
path
and
IV
path.
If
fluid
exits the
patient
end
before
connecting
to
intravenous
administration
set,
remove
and
replace
administration
set.
*
All
air
must
be
removed
from
the
intravenous
fluid
lines
before
connecting
to
the
patient.
Failure
to
do
so
may
result
in
introduction
of
air
to
the
patient,
which
may
contribute
to
serious
patient
injury
or
death.
Monitor
intravenous
fluid
fine
to
make
sure
it
is
air-free.
*
Do
not
stick
the
HOTLINE®
Warming
Set
tubing
with
needles,
which
will
compromise
integrity
of
the
patient
intravenous
line.
If
a
disposable
with
a
breached
circulating
water
path/IV
path
is
used,
then
patient
illness
may
occur
because
of
the
fluid
warmer's
circulating
water
entering
the
patient's
blood
stream.
ㆍ.
Do
not
use
HOTLINE?
in
the
presence
of
flammable
anesthetics
because
of
the
risk of
fire
or
possibility
of
explosion.
«
©
Do
not
use
HOTLINE"
in
high
energy
fields
such
as:
MRI,
X-Ray,
etc.
HOTLINE?
may
act
as
a
projectile
in
a
strong
magnetic
field,
cause image
artifacts
or
not
function
as
intended.
CAUTIONS:
」
・
Federal
Law
(USA)
restricts
sale
of
this
device
to
or
on
the
order
of
a
physician.
*
©
Do
not
mount
the
HOTLINE?
more
than
42"
(107cm)
above
the
floor.
For
convenience
42"
(107
cm)
is
indicated
on
the
HOTLINE"
line
power
cord
by
a
black
mark.
Mounting
the
HOTLINE“
above
42"
(107cm)
may
result
in
instability
of
the
pole
and
tipping.
«
Do
not
autoclave
or
immerse
any
part
of
the
HOTLINE?
in
liquids,
which
may
cause
damage
and
improper
functioning.
¢
—
Insure
that
the
HOTLINE®
pole
clamp
is
screwed
tightly
onto
the
IV
pole.
Failure
to
securely
mount
the
HOTLINE?
onto
the
IV
pole
may
cause
the
HOTLINE?
to
slide
down
the
IV
pole.
+
Not
for
use
with
pressure
devices
generating
over
300
mmHg.
Pressure
greater
than
300
mmHg
may
compromise
the
integrity
of
the
HOTLINE®
Warming
Set.
NOTE:
Hand
pumps
can
generate
over
1000
mmHg.
o
de
o.
ELECTRICAL
SAFETY:
POWER
REQUIREMENT
Most
of
the
current
drawn
by
the
HOTLINES
is
for
the
300-Watt
immersion
heater.
When
the
device
is
first
turned
on
and
the
digital
display
shows
rapidly
rising
temperatures
below
41°C,
the
300-Watt
heater
is
in
a
full
ON
condition.
Beyond
approximately
40°C,
the
proportional
controller
cycles
the
heater
ON/OFF
with
proportionally
shorter
ON
times
as
the
circulating
water
nears
the
41.5°C
set
point.
ELECTRICAL
SAFETY
TESTING:
All
testing
and
maintenance
should
be
performed
by
qualified
personnel.
Safety
testing
is
to
be
conducted
on an
annual
basis.
This
involves
earth
leakage
and
ground
continuity,
which
must
be
tested
according
to
UL 2601-1
and
EN
60601-1.
NOTE:
This
device
is
equipped
with
disposable
sensing
interlocks.
A
HOTLINE®
Warming
Set
is
required
to
correctly
operate
the
device
and
perform
leakage
current
testing.
Do
not
defeat
the
disposable
sensing
interlocks
or
try
to
operate
the
HOTLINE
without
a
Warming
Set
in
place.
Earth
teakage
must
be
tested
according
to
UL
2601-1
and
EN
60601-1.
Earth
leakage
current
test
should
be
performed
with the
immersion
heater
circuit
in
the
full
ON
condition.
For
this
reason,
leakage
current
test
should
be
performed
on
a
HOTLINE“
that
has
room
temperature
water
in
the
water
tank.
5

Ground
continuity
must
be
tested
according
to
UL
2601-1
and
EN
60601-1.
HOTLINE®
is
certified
that
it
is
in
compliance
with
the
European
Communities
Council
Directive
relating
to
Electromagnetic
Compatibility
(EMC):
(89/336/EEC).
Test
methods
and
acceptance
criteria
as
specified
in
EN
60601-1-2
demonstrate
conformance.
STORAGE
Store
in
a
cool
dry
place.
Do
not
expose
to
extreme
temperature.
See
Environmental
Specifications
for
more
details.
PREPARATION
AND
SET-UP
HARDWARE:
WARNING:
This
unit
should
be
tested
by
hospital
biomedical
personnel
prior
to
placing
it
in
service.
All
testing
and
maintenance
should
be
performed
by
qualified
personnel.
Step
1:
CLAMP
THE
HOTLINE?
SECURELY
TO
AN
I.V.
POLE.
CAUTION:
Do
not
mount
the
HOTLINE®
more
than
42"
(107cm)
above
the
floor.
For
convenience,
42"
(107cm)
is
indicated
on
the
HOTLINE®
line
cord
by
a
black
mark.
Mounting
the
HOTLINE®
above
42"
(107cm)
may
result
in
instability
of
the
pole
‘and
tipping.
CAUTION:
This
unit
is
cooled
by
convection.
Be
sure
the
air
vents
on
the
bottom
and
the
back
of
the
unit
are
kept
clear.
CAUTION:
Insure
that
HOTLINE?
pole
clamp
is
screwed
tightly
onto
the
IV
pole.
Failure
to
securely
mount
the
HOTLINE?
onto
the
IV
pole
may
cause
the
HOTLINE?
to
slide
down
the
pole.
Step
2:
FILL
THE
WATER
TANK
4.
Unscrew
the
fillport
plug
on
the
water
tank.
2.
Fill
the
water
tank
with
1.4
liters
of
either:
*
Distilled
water
Or
if
being
used
as
a
maintenance
solution,
disinfect
per
Instructions
on
page
9
before
filling
with
either:
«
A0.3%
Hydrogen
Peroxide
/
distilled
water
solution:
mix
140
mL
of
3%
Hydrogen
Peroxide
and
1,260
mL
of
distilled
water
(this
will
equal
1.4L
of fluid)
OR
ㆍ
A35%
Isopropyl
alcohol
/
distilled
water
solution:
mix
700
mL
of
70%
Isopropyl
alcohol
and
700
mL.
of
distilled
water
(this
will
equal
1.4L
of fluid)
These suggested
instructions
are
designed
to
be
used
in
conjunction
with
established
hospital
procedures.
NOTE:
Use
Distilled
Water
only.
Failure
to
do so
may
cause
a
build-up
of
mineral
deposits
in
the
circulating
water
path
which
may
impair
heater
performance.
CAUTION:
Do
not
fill
the
HOTLINE’
water
tank
with
a
HOTLINE’
Warming
Set
in
place.
Failure
to
remove
the
Warming
Set
before
the
fill
procedure
may
result
in
an
air
lock
in
the
HOTLINE’
Warmer.
3.
Replace
the
fillport
plug.
4.
Plug
the
HOTLINE®
into
the
applicable
outlet
(115V~
or
230V~).

INSTRUCTIONS
FOR
USE
STEP
1:
HOTLINE®
WARMER
SET-UP
1.
Check
that
the
water
level
is
above
the
minimum
level
mark
on
the
tank.
Add
distilled
water
to
the
tank
through
the
fillport
if
required.
2.
Plug
into
power
outlet.
STEP
2:
SET-UP
THE
HOTLINE®
WARMING
SET
To
set
up
the
HOTLINE®
Warming
Set
you
will
need
the
following:
+
HOTLINE®
Warmer
HOTLINE?
Warming
Set
(L
-
Series)
Intravenous
fluid
or
blood
Intravenous
administration
set
Extension
Set,
8"
(20cm)
or
less
in
length
(optional)
1.
Plug
the
Twin-Tube
Connector
into
the
socket.on
the
right
side
of
the
HOTLINE®
Warmer.
2.
-Activate
the
Power
Switch
on
the
left
side
of
the
Warmer.
The
green
light
on
the
display
panel
will
:
iluminate
and
the
circulating
water
bath
temperature
display
will
begin
to
increase.
-3.
The
circulating
water
path
will
automatically
prime
when
the
unit
is
turned
on.
inspect
the
patient
end
of
the
tubing
for
leaks
to
confirm
the
integrity
of
the
intravenous
pathway.
STEP
3:
CONNECT
THE
INTRAVENOUS
ADMINISTRATION
SET
1.
Connect
the
IV.
fluid
and
the
intravenous
administration
set
to
the
HOTLINE?
Warming
Set.
2.
Fully
prime
the
intravenous
administration
set,
the
HOTLINE®
Warming
Set,
and
patient
extension
set
(if
used).
:
3.
Connect
to
the
patient's
intravenous
access
site
without
entrapping
air.
STEP
4:
USE
1.
The
circulating
water
bath
temperature
display
will
reach
37°C
in
about
4
minutes.
“2.
Adjust
the
rate
of
flow
using
the
clamp
on the
intravenous
administration
set.
NOTE:
Do
not
kink
the
disposable
set
tubing.
Do
not
restrict
the
water
circulation
through
the
tubing.
STEP
5:
AFTER
USE
1.
Tum
off
the
Power
Switch
on
the
left
side
of
the
Warmer.
2.
Remove
the
HOTLINE®
Warming
Set.
3.
After
use,
this
product
may
be
a
potential
biohazard.
Handie
and
dispose
of
in
accordance
with
accepted
medical
practice
and
applicable
local,
state
and
federal
regulations.
4.
Wipe
down
the
extemal
surfaces
of
the
HOTLINE®
Warmer
with
mild
detergent,
water,
and
a
soft
cloth.
For
external
disinfection,
a
10%
bleach/distilled
water
solution
may
be
used.

TROUBLESHOOTING
Problem:
Check:
No
power
+
Unit
plugged
in?
+
Power
switch
on?
ㆍ
NOTE:
/f
plugged
in
and
power
switch
is
“ON”,
the
green
or
red
LEDs
will
glow.
DISPOSABLE
alarm
・
Disposable
properly
installed?
Push
Twin-tube
Connector
firmly
into
HOTLINE®
Warmer
Socket.
ADD
WATER
alarm
+”
Add
distilled
water
to
the
water
tank
to
the
MAX
fill
line.
ㆍ
CAUTION:
Do
not
fill
the
HOTLINE?
water
tank
with
a
HOTLINE®
Warming
Set
in
place.
CIRCULATING
WATER
e
Check
Warming
Set
for
tubing
kinks
or
other
restrictions.
BATH
OVER
ㆍ
©
Check
for
air
lock:
Tum
Power
Switch
“OFF”,
remove
the
Warming
TEMPERATURE
alarm
Set,
gently
shake
HOTLINE®
to
dislodge
air,
plug
in
Warming
Set
and
tum
Power
Switch
“ON”.
ㆍ.
there
are
no
restrictions
or
air
present,
remove
the
HOTLINE?
from
service
and
return
it
for
repair
or
replacement.
Hot
Cabinet
・
Blocked
air
vents
on
the
bottom
or
the
back
of
the
unit?
ㆍ
NOTE:
Room
temperature
above
42°C
may
cause
the
HOTLINE®
to
shut
down
and
the
over
temperature
alarm
to
activate.
In
this
unusual
situation,
turn the
power
switch
off
and
allow
the
unit
to
cool
down
before
returning
it
to
service.
Hard
to
install
the
Warming
|
+
Grease
O-rings
in
the
socket
as
per
the
instructions
in
this
manual.
Set
See
“Maintenance”.
Water
leaks
at
the
socket
|・
Replace
O-rings.
Order
P/N
80-04-001
where
the
Warming
Set
plugs
into
the
unit.
MAINTENANCE
Alt
testing
and
maintenance
should
be
performed
by
qualified
personnel.
EXTERNAL
CLEANING:
EVERY
USE
For
routine
cleaning
use
only
mild
detergents,
water
and
soft
cloth.
For
external
disinfecting,
a
10%
bleach/distilled
H,O
solution
may
be
used.
CAUTION:
Do
not
autoclave.
Do
not
use
alcohol,
abrasive
cleaning
agents,
solvents
or
cold
sterilants.
These
agents
will
cause
the
unit
to
crack.
Do
not
immerse
any
part
of
the
HOTLINE?
in
liquids.

GENERAL
INSPECTION:
EVERY
USE
Visually
check
the
condition
of
the
unit.
Remove
from
service
any
unit
which
shows
physical
damage
or
one
in
which
the
HOTLINE®
Warming
Set
does
not
install
easily.
CHANGE
DISTILLED
WATER
SOLUTION:
EVERY
30
DAYS
1.
Unplug
the
unit
before
servicing.
2.
Remove
the
fillport
plug
from
the
water
tank.
3.
Invert
the
unit
over
a
sink
or
suitable
container
to
remove
the
water
from
the
unit.
Rinse
tank
with
distilled
H,O
twice
before
returning
to
use.
4.
Re-attach
to
1.V.
pole.
5.
Refill
the
unit
with
1.4
Liters
of
distilled
water
solution.
Refer
to
Set-up
and/or
Disinfection instructions
for
filling
options.
OTE:
Use
Distilled
Water
only.
Failure
to
do so
may
cause
a
build-up
of
mineral
deposits
in
the
circulating
water
path
which
may
impair
heater
performance.
CAUTION:
Do
not
fill
the
HOTLINE?
water
tank
with
a
HOTLINE®
Warming
Set
or
a
TEMP
CHECK
in
place.
Failure
to
remove
the
Warming
Set
before
the
fill
procedure
may
result
in
an
air
lock
in
the
HOTLINE’
Warmer.
6.
Replace
the
fillport
piug.
LUBRICATE
O-RING
SEALS:
EVERY
30
DAYS
Using
a
cotton
swab
apply
a
small
amount
of
silicone
grease
to
the
O-rings.
Silicone
grease
is
available
from
LEVEL
1:
Silicone
Grease
P/N
80-04-002
REPLACE
O-RING
SEALS:
EVERY
12
MONTHS
Remove
the
socket
head
screws
with
a
1/8"
(.31cm) hex
wrench.
Remove
the
disposable
interface
block,
being
careful
not
to
damage
the
micro-switch
lever.
Remove
the
oid
O-rings
and
clean
the
sockets
with
a
cotton
swab.
Apply
silicone
grease
to
2
new
O-rings
and
install
into
the
sockets.
Reassemble
in
reverse
order
being
careful
not
to
damage
the
micro-switch
lever.
O-Ring
Kit:
P/N
80-04-0071
DISINFECTION
To
disinfect
the
circulating
water
bath,
the
following
procedure
may
be
employed:
1.
Unplug
the
unit
before
servicing.
2.
Remove
the
fillport
cap.
3.
Empty
all
water
from
the
reservoir
tank.
4.
Fill
the
water
tank
with
1.4
liters
of
either:
DALI
・
A0.3%
Hydrogen
Peroxide
/
distilled
water
solution:
mix
140
mL
of
3%
Hydrogen
Peroxide
and
1,260
mL
of
distilled
water
(this
will
equal
1.4L
of
fluid)
OR
+
A35%
Isopropyl
alcohol
/
distilled
water
solution:
mix
700
mL
of
70%
Isopropyl
alcohol
and
700
mL
of
distilled
water
(this
will
equal
1.4L
of
fluid)
These
suggested
instructions
are
designed
to
be
used
in
conjunction
with
established
hospital
procedures.
NOTE:
Use
Distilled
Water
only.
Failure
to
do
so
may
cause
a
build-up
of
mineral
deposits
in
the
circulating
water
path
which
may
impair
heater
performance.
CAUTION:
Do
not
fill
the
HOTLINE®
water
tank
with
a
HOTLINE?
Warming
Set
in
place.
Failure
to
remove
the
Warming
Set
before
the
fill
procedure
may
result
in
an
air
lock
in
the
HOTLINE®
Warmer.
5.
Insert
a
Warming
Set
into
the
machine.
6.
Turn
the
unit
on
and
let
the
solution
circulate
for
30
minutes.
7.
Switch
the
unit
off
and
empty.
9

8.
Thoroughly
rinse the
unit
with
distilled
water
or
equivalent.
9.
Refill
the
reservoir
tank
with
1.4
liters
of
distilled
water
solution.
Refer
to
Set-up
instructions
for
filling
options.
10.
These
suggested
instructions
are
designed
to
be
used
in
conjunction
with
established
hospital
procedures.
TESTING
All
testing
and
maintenance
should
be
performed
by
qualified
personnel.
CAUTION:
if
any
alarm
conditions
are
not
met
remove
the
unit
from
service.
NOTE:
Alarm
testing
requires
a
HOTLINE®
Warming
Set
to
be
installed
and
the
unit
operating.
Alarm
Test
Switch:
The
HOTLINE®
is
equipped
with
two
membrane
switches
on
the
side
label.
Pressing
the
Alarm
Test
Switch
activates
all
alarm
conditions.
Pressing
and holding
the
Over
Temp
Test
Switch
activates
the
over
temp
alarm
condition.
To
Test
the
HOTLINE®
Alarm
Circuitry:
Press
the
Alarm
Test
Switch.
Observe
that
the
green
LED
goes
out.
The
three
red
LED's
will
light.
The
audible
alarm
will
sound.
Water
will
stop
circulating.
(This
can
be
checked
by
observing
the
return
port
inside
the
water
tank.)
APRON
>
T
©
test
the
Circulating
Water
Bath
Over
Temperature
Alarm:
The
unit
should
be
running
at
an
operating
temperature
of
approximately
41°C
to
42°C.
Press
and
hold
the
Over
Temp
Test
Switch.
The
Circulating
Water
Bath
Over
Temperature
Alarm
will
activate
at
43°C.
Observe
that
the
green
LED
goes
out.
The
Circulating
Water
Bath
red
LED
will
light.
The
audible
alarm
will
sound.
Water
will
stop
circulating.
(This
can
bé
checked
by
observing
the
return
port
inside
the
water
tank.)
NOOR
WN
=
Add
Water
Alarm:
The
HOTLINE?
is
equipped
with
a
float
switch,
which
senses
the
water
level
in
the
water
tank.
When
the
waiter
is
too
low,
the
ADD
WATER
Alarm
will
activate.
To
test
the
Add
Water
Alarm
circuit:
1.
Unscrew
the
fillport
plug
on
the
water
tank.
2.
Gently
depress
the
float
switch.
(This
action
will
simulate
the
low
water
condition.)
NOTE:
Use
a
non-metal
tool
to
depress
the
float
switch
because
the
float
switch
contains
a
magnet.
Observe
that
the
green
LED
goes
out.
The
red
LED
will
light.
The
audible
alarm
will
sound.
Water.
will
stop
circulating.
(This
can
be
checked
by
observing
the
return
port
inside
the
water
tank.)
gapw
Disposable
Alarm:
An
interlock
switch
located
in
the
Disposable
Interface
Block
senses
a
properly
installed
Warming
Set.
When
the
switch
does
not
sense
a
Warming
Set
the
DISPOSABLE
alarm
will
activate.
To
test
the
Disposable
Alarm
circuit:
Slowly
remove
the
HOTLINE®
Warming
Set
from
the
HOTLINE?,
Observe
that
the
green
LED
goes
out.
The
red
LED
will
light.
The
audible
alarm
will
sound.
Water
will
stop
circulating.
(This
can
be
checked
by
observing
the
retum
port
inside
the
water
tank.)
の
た
の
ON
ュー
In
any
alarm
condition
the
pump
should
not
be
running.
A
small
amount
of
water
dripping
from
the
disconnection
is
normal
and
should
stop
in
a
few
seconds.
10

TEMPERATURE
VERIFICATION
All
testing
and
maintenance
should
be
performed
by
qualified
personnel.
Displayed
Circulating
Water
Temperature:
Should
be verified
using
a
Level
1
TEMP
CHECK
Thermometer
(HLTA-40).
Other
methods
of
temperature
verification
may
be
inaccurate.
Units
can
be
returned
for
temperature
verification.
Principles
of
Operation:
The
Circ.
H,O
is
pumped
from
the
water
tank
through
the
heater.
The
Circ.
H,O
then
passes
the
HOTLINE®
thermistor,
assembled
within
the
heater
on
the
output
side,
and
then
through
the
TEMP
CHECK.
After
this
point the
Circ.
H¿O temperature
begins
to
drop
due
to
the
effect
of
ambient
temperature
on
the
disposable
set.
The
temperature
of
the
water
tank
is
typically
0.5°C
to
2.0°C
lower
than
the
temperature
from
the
heater.
Verifying
the
operating
temperature
of
the
Circ.
HO
by
temperature
readings
in
the
water
tank
will
be
inaccurate.
The
TEMP
CHECK
senses
the
highest
temperature
of
the
Circ.
H,O
bath.
TEMP
CHECK
REFER
TO
MANLIAL FOR
CALIBRATION
INSTRUCTIONS,
CAUTION:
Refer
to
the
TEMP
CHECK
HLTA-40
Thermometer
Operator's
Manual
for
complete
Temperature
Verification
and Calibration
instructions.
SET-UP:
To set
up
the
TEMP
CHECK
you
will
need
the
following:
*.
TEMP
CHECK
(HLTA-40)
*
HOTLINE?
Warmer
ㆍ
HOTLINE?
Warming
Set
1.
Plug
the
HOTLINE®
Warmer
into
power
outlet.
2.
Place
the
TEMP
CHECK
onto
the top
right
corner
of
the
HOTLINE®
Warmer
and
plug
it
into
the
socket
on
the
right
side
of
HOTLINE®.
3.
Plug
the
Twin-Tube
Connector
on
the
HOTLINE®
Warming
Set
onto
the
socket
on
the
right
side
of
the
TEMP
CHECK,
4.
Remove
the
black
label
from
the
auxiliary
outlet
on
the
back
of
the
HOTLINE?
and
plug
in
the
TEMP
CHECK
power
cord.
CAUTION:
The
auxiliary
outlet
is
for
use
only
with
LEVEL
1
accessories.
5.
Activate
the
Power
Switch
on
the
left
side
of
HOTLINE®.
Allow
15
minutes
for
the
temperature
to
stabilize.
6.
if
the
TEMP
CHECK
display
indicates
a
temperature
between
41°C
and
42ºC,
and
the
HOTLINE“
display
equals
the
TEMP
CHECK
display,
water
bath
verification
is
complete.
Refer
to
the
TEMP
CHECK
Manual
for
OVERTEMP
ALARM
verification.
7.
If
the
TEMP
CHECK
display
does
not
indicate
a
temperature
between
41°C
and
42°C,
refer
to
the
TEMP
CHECK
Manual
for
calibration
instructions.
11

RECOMMENDED
MAINTENANCE
CHECK
LIST
MODEL:
HOTLINE®
FLUID
WARMER
REF:
HL-90
All
testing
and
maintenance
should
be
performed
by
qualified
personnel.
Write
your
date
of
purchase
here
Write
your
serial
number
here
30-day
maintenance
tasks
Change
Distilled
Water
Solution
Grease
O-ring
Seals
Test
Aiarm
Switch
12
months
maintenance
tasks
Test
Alarm
Switch
Test
Float
Switch
Test
Disposable
Switch
Test
Over
Temperature
Alarm
Replace
O-ring
seals
Verify
Temperature/
Calibrate
You
may
photo
copy
this
form
for
your
record
keeping.
Notes:
12
Table of contents
Other level 1 Food Warmer manuals