Steps 1–4 should be performed with the
device turned OFF. Once the unit has been
turned ON, do not remove electrodes
without rst turning OFF the device.
IMPORTANT!
PAGE 2 QUICK REFERENCE GUIDE
2CONNECT ELECTRODES
TO LEAD WIRES
Firmly insert
the red or black
end of the lead
wire pin (it
doesn’t matter
which one) into
the electrode
connectors.
Make sure no bare metal of the pins are exposed.
3CONNECT LEAD WIRES
TO TOP OF DEVICE
This unit operates
with one or two
sets of electrodes.
When using one set,
plug lead wire into
Channel 1 on top
of the unit. Use
Channel 2 to operate
a second set if
directed by your
healthcare provider.
4PLACE ELECTRODES
ON SKIN 6SELECT THE
THERAPEUTIC MODE
Remove electrodes from
protective backing. Keep
backing to properly store
electrodes after use.
Apply electrodes to the
exact site indicated by
your doctor or therapist.
Make sure the skin at the
application site is clean
and dry.
There are 4 therapeutic modes available: TENS, EMS,
IF and MIC. Press the Mbutton and continue to
press until you have advanced to the appropriate
therapeutic mode.
Consult your physician for your suitable
therapeutic mode.
Press the Sbutton to enter and cycle through
to each setting state.
SET THERAPEUTIC PROGRAM
There are 4 programs available in theTENS therapeutic mode: B -Burst, N -Normal,
M -Pulse Width Modulation, and M1 -Pulse Rate Modulation. The therapeutic
program can be selected by pressing the sand tbuttons.When you choose to
Mprogram, program Moutside of the box will be ashing.
SET CYCLE TIME (OPTIONAL)
Cycle time is adjustable from 5to 30 seconds. Only modulation has this parameter
setting. Press Sbutton to enter this menu, and then press the sand tbuttons to
adjusting the setting.
SET TIMER
Press the Sbutton to enter this setting. The treatment time is adjustable from
1 to 60 minutes or Continuous. Press the sand tbuttons to adjust setting.You
can set the timer to“Continuous”mode by pressing the scontrol when it shows
60 minutes.The output will be shut o when time is up. If in continuous mode, you
need to manually turn o device.
SET PULSE WIDTH
Pulse Width is adjustable from 50 uS to 300 uS. Press Sbutton to enter this menu,
then press the sand tbuttons to adjust the setting.
5TURN DEVICE ON
See Electrode Placement Guide for more information.
1INSERT BATTERY
Press arrow and slide battery compartment
cover open. Insert 9V battery, matching the
+ and – ends to those in unit.
The included AC adapter may
also be used as a power source.
TENS SETTING
2
1
3
4
SET PULSE RATE
Pulse rate is adjustable from 1Hz to 150 Hz (0.5 Hz to 5 Hz for Burst). Press the S
button after scrolling through and completing all settings to start the therapy program.
The sand tbuttons can be used to increase sor decrease tthe intensity level.
5
1
2
3
4
3
5
Check that electrodes are securely placed on
the skin before activating device. Press and release the
button in the center of the unit to turn it ON.
LG QUAD Combo™
QUICK REFERENCE GUIDE PAGE 3
Press the Sbutton to enter and cycle through
to each setting state.
SET THERAPEUTIC PROGRAM
There are 3 programs available in EMS therapeutic mode: S -Synchronous,
A -Alternate and D -Delay. The therapeutic program can be selected by pressing the
sand tbuttons.When you choose to Dprogram, program Doutside of the box
will be ashing.
SET TIMER
Press the Sbutton to enter this setting. The treatment time is adjustable from
1 to 60 minutes or Continuous. Press the sand tbuttons to adjust setting.You
can set the timer to“Continuous”mode by pressing the scontrol when it shows
60 minutes.The output will be shut o when time is up. If in continuous mode, you
need to manually turn o device.
SET PULSE WIDTH
The pulse width determines the length of time. Each electrical signal is applied
through the skin, which controls the strength and sensation of the stimulation.
Press the Sbutton to enter this setting. The pulse width is adjustable from
50 to 300 uS. Press the sand tbuttons to adjust the setting.
SET PULSE RATE
The pulse rate determines how many electrical impulses are applied through the skin
each second. Press the Sbutton to enter this menu. Adjust the setting by pressing the
sand tbuttons.The pulse rate is adjustable from 1Hz to 150 Hz.
SET DELAY TIME (OPTIONAL)
Delay time is adjustable form 1to 10 seconds. Only the“Delay”therapeutic program
has this parameter setting. Press the Sbutton to enter this menu, and then press the
sand tbuttons to adjust the setting.
SET RAMP TIME
The ramp time controls the time of output current that increase from 0to the setting
level, and from the setting value to 0. When the ramp time is set, each contraction
may be ramped up and down in order that the signals come on and come o gradually
and smoothly.The ramp time is adjustable from 1to 6seconds.
SET CONTRACT TIME
The contract time controls the time of stimulation and can be adjusted. Press the S
button to enter this menu, and then press the sand tbuttons to adjust the setting.
Both channels’stimulation is cycled on and o by the contraction and relaxation
settings. The range is adjustable from 1to 60 seconds.
EMS SETTING
2
1
3
4
1
2
3
2
4
5
Caution: Contract time does not include the ramp up and ramp down time,
ON time = Ramp up + Contract time + Ramp down.
SET RELAXATION (OFF) TIME
The o time controls the time of relaxation and can be adjusted. Press the Sbutton to
enter this menu, and then press the sand tbuttons to adjusting the setting. Both
channels’stimulation is cycled on and o by the contraction and relaxation settings.
The range is adjustable from 0to 60 seconds.
Press the Sbutton after scrolling through and completing all settings to start the
therapy program. The sand tbuttons can be used to increase sor decrease t
the intensity level.
6
7
5
8
6
7
8
IMPORTANT TIPS AND SAFETY FEATURES
The Safety Lock Feature automatically activates after there is no operation in
the panel for 30 seconds by locking out the ability to press the buttons.This is a safety
feature to prevent accidental changes to your settings and to prevent accidental
increases to the intensity levels.You can press either one of the tbuttons to unlock
the device.
Once you have activated the treatment timer, you can press the Mbutton or the
t button to control stop the treatment.
Turn o the device by pressing and holding the power on/o button for
3 seconds. If in“lock”mode, press the tbutton to unlock the device and then press
and hold the power on/o button for 3 seconds to turn o.
The device will automatically power o after 2 minutes of inactivity in the panel.
Low battery indicator signals when battery is low. Replace battery when
stimulation feels weaker.
Live Pain Free, Live Good Today!
TOPVIE W