manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Lika
  6. •
  7. Monitor
  8. •
  9. Lika LD120-M7 User manual

Lika LD120-M7 User manual

LD120
User manual
LD120-M7
SM5
Description
This manual describes the LD120 display and the
sensors of the SM5 series. The purpose of this
system is to display displacements on industrial
machines and automation systems. The
measurement system includes a LED display,
ma netic tape and a ma netic sensor. As the sensor
is moved alon the ma netic tape, it detects the
displacement which is shown on the display. The
flexibility of the tape allows it to be used for both
linear and an ular applications.
Chapters
1 Safety summary
2 Identification
3 Installation
4 Mountin steps
5 Electrical connections
6 Setup
7 RS485 serial interface
8 Dimensional drawin and cut-out
1 - Safety summary
We stron ly recommend carefully readin this user
manual and followin the installation uidelines:
•Measurement system (sensor) should be installed
as close as possible to the display.
•Always use shielded and twisted cables if
possible.
•Avoid runnin the sensor cable near hi h
volta e power cables (e. . drive cables).
•Install EMC filters on sensor power supply if
needed.
•Avoid mountin sensor near capacitive or
inductive noise sources such as relays, motors,
and switchin power supplies.
Connect accordin to the chapter 5: “Electrical
connections”.
2 - Identification
The display and sensor can be identified by the
label's data (orderin code, serial number). This
information is listed in the delivery document. For
technical features of the product to make reference
at the technical catalo ue.
3 - Installation
Install the product accordin to the protection level
provided. Protect the system a ainst knocks,
friction, solvents, and respect the environmental
characteristics of the device.
- Mounting steps
.1 Display
Push the display into the cut-out (approx. 68 x 33
mm2) without panel clips.
Install panel clips on the display's housin and screw
until fixed.
.2 Magnetic tape
See manual supply with the ma netic tape.
.3 Sensor mounting
.3.1 Sensor type SM5-R (rectangular)
Sensor can be fixed by means of two M3 screws
over the buttonholes. Make sure that the ap
between sensor and tape is in respect with (fi . 1)
alon the total measurin len th. Avoid contact
between the parts. You can check planarity and
parallelism between sensor and ma netic tape usin
a feeler au e.
MAN LD120 I_E 1.5 Pa . 9 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
fi ure 1
D = 0,1 mm - 2,1 mm
.3.2 Sensor type SM5-C (circular)
The sensor can be fixed in a correspondin
mountin hole by means of the two nuts. Make sure
that the ap between sensor and tape is in respect
with (fi . 2) alon the total measurin len th.
Observe the correct ali nment of the marker on the
tape. Avoid contact between the parts. You can
check planarity and parallelism between sensor and
ma netic tape usin a feeler au e.
fi ure 2
D = 0,1 mm - 2,1 mm
5 - Electrical connections
5.1 SM5 sensor
Plu in the sensor's Mini-DIN connector (circular) on
backside of the display.
5.2 Display
Make the clamp connections as follow:
Serial interface is option.
The ACCU input can supply the device in stand-by
mode with a low consumption. In this way the LED’s
are deactivate while the electronics count and the
sensor remain active.
NOTE:
This device is to be supplied by a Class 2 Circuit or
Low-Volta e Limited Ener y or Ener y Source not
exceedin 30Vdc.
MAN LD120 I_E 1.5 Pa . 10 www.lika.it
www.lika.biz
D
<
1
°
<
3
°
<
3
°
<3°
DMarker
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
6 - Setup
6.1 Key's function
 : UP (select value)
 : Shift links (select di it)
* : Save (save data)
P : Pro ram (pro rammin /chan e parameter)
6.2 Key combinations / Quick functions
6.2.1 Set datum (reference)
Push * key for 3 s to set actual value to datum
value.
Datum value is = rEF + OFS1 + OFSx (where OFSx is
the actually set Offset value).
This function is enabled only if “F_rSt” parameter is
set “on”.
6.2.2 Incremental measurement
Push P and * key simultaneously to switch from
absolute measurement to incremental.
In incremental mode the decimal point flashes.
Zero settin in incremental mode (see 6.2.1) does
not chan e absolute value in the back round.
6.2.3 Mm/inch display modes
Mm/ich display modus can be chan ed by pushin
 key for 3 s.
6.2. Offset value modification
Push P and  keys simultaneously to display Offset
1 value (OFS1). Use  and  keys to chan e value
and save with * key. Further Offset values OFS2 and
OFS3 can be chan ed only in setup menu.
This function is enable only if “F_oFS” parameter is
set “on”.
 key allows to scroll OFS1, OFS2 and OFS3 values.
OFS1 = actual value + OFS1 + rEF
OFS2 = actual value + OFS1 + OFS2 + rEF
OFS3 = actual value + OFS1 + OFS3 + rEF
6.2.5 Datum modification
Push P and  keys simultaneously to display datum
value rEF. Use  and  keys to chan e value and
save with * key.
This function is enabled only if “F_rEF” parameter is
set “on”.
6.3 Setup / Parameter setting
Push P key for 3 s to enter setup and “SEtUP” is
displayed.
Push  key to enter Menu 1 (parameters)
Push * key to enter Menu 2 (RS485 serial I/F)
Push P key to access the next Parameter and
Parameter settin .
Push P key for 3 s to exit the setup at any point.
6.3.1 Default parameters (factory settin s)
Default parameter values are written in BOLD
characters. The unit can be reset to default values by
pushin * key while switchin on.
6.3.2 Menu 1: Parameters
rES Resolution [10, 50, 100, 1000, InCH, FrEE]
Linear resolution value expressed in µm (microns).
10 = 0,01 mm
50 = 0,05 mm
100 = 0,1 mm
1000 = 1 mm
InCH = sets display mode and measurement unit to
inch (with 3 decimals, e. . 1.000).
FrEE = allows to input a free resolution factor e. .
for an le displays (basis of calculation is max.
resolution 0,01 mm).
* = save, P = next parameter, P for 3 s. = exit
FrEE [0.0001, 1.0000]
Only if rES = FrEE.
E. . an le display with ran e from 0° to 90° and 0,1°
resolution on a rotatin table with circumference of
785,4 mm. Total workin ran e is 785,4 mm : 4 =
196.35 mm.
FrEE = 900 : 19635 = 0,0458
Default value = 0.0001
* = save, P = next parameter, P for 3 s. = exit
dEC Decimal point [0, 1, 2, 3]
Only if rES = FrEE.
Modification of decimal point position.
0 = no decimal point
1 = one decimal (e. . 1,0)
2 = two decimals (e. . 1,00)
3 = three decimals (e. . 1,000)
* = save, P = next parameter, P for 3 s. = exit
MAN LD120 I_E 1.5 Pa . 11 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
dir Countin direction [uP, dn]
uP = up (standard direction)
dn = down (inverted direction)
* = save, P = next parameter, P for 3 s. = exit
F_rEL Incremental measurement function [on, oFF]
Enables incremental measurement function (by
pushin P and * keys).
on = function on
oFF = function off
* = save, P = next parameter, P for 3 s. = exit
F_rSt Datum function [on, oFF]
Enables datum function (by pushin * key).
on = function on
oFF = function off
* = save, P = next parameter, P for 3 s. = exit
F_oFS Offset function [on, oFF]
Enables offset function (by pushin P and  keys).
on = function on
oFF = function off
* = save, P = next parameter, P for 3 s. = exit
rEF Datum value [-99999, 99999]
Absolute reference value for the measurin system.
This value is displayed by pushin * key for 3 s.
(displayed value includes previously set offset
values).
Default value = 0
* = save, P = next parameter, P for 3 s. = exit
OFS1 Offset1 value (-99999, 99999)
First offset value (e. . tool correction). This value is
added to actual value (see 6.2.3.)
Default value = 0
* = save, P = next parameter, P for 3 s. = exit
OFS2 Offset2 value [-99999, 99999]
Second Offset value. This value is added to actual
value and OFS1.
Default value = 0
* = save, P = next parameter, P for 3 s. = exit
OFS3 Offset3 value [-99999, 99999]
Third Offset value. This value is added to actual
value and OFS1.
Default value = 0
* = save, P = next parameter, P for 3 s. = exit
NOTE:
To set ne ative values, select si n di it with  key
and chan e with  key. If “-” si n flashes ne ative
value is selected, otherwise positive value is
selected. Push * key to save.
F_SAP Save actual value [on, oFF]
Saves the last displayed value after power off.
on = last value memory on
oFF = last value memory off
* = save, P = next parameter, P for 3 s. = exit
When the setup is completed the display shows
"rESEt".
•Push P key to display rEF+OFS1+OFSx
•Push * key to reset the display.
•Push P key twice to quit the setup.
6.3.3 Menu 2: RS 85
Ad xx Device address [00, 31]
Settin of device address (only if ordered with serial
interface (option I4).
Default value = 0
* = save, P = next parameter
H_cnt Hourmeter (1/10 h)
Elapsed time indication (display connected to power
supply). Resolution is 1/10 hour (6 minutes).
 = hourmeter reset, P = exit
MAN LD120 I_E 1.5 Pa . 12 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
7 - RS 85 serial interface (option I )
If the display is provided with RS485 serial interface,
the followin commands can be used.
7.1 RS 85 parameters
Baud rate = 9600
Data bits = 8
Parity bit = No
Stop bit = 1
Flow control = Xon/Xoff
7.2 Serial commands
Serial commands must have the followin structure:
| A D C M N D = X
where :
| : PC keyboard symbol
AD : device address (00 to 31) 2 di it
CMND : command (see command list)
X : value ran e (see command list)
Upon receipt of a wron command the display will
answer with the same command + ? and checksum
(e. . sent command: |02azs → answer |02azs?EF)
Any common terminal pro ram can be used for
communication with LD120 (e. . Hyperterminal).
Command will be send after confirmation by ENTER
key (carria e return).
Answers are structured as follow:
ADCMND:SXXXXXCK
where:
AD : device address
CMND : command
XXXXX : value
CK : checksum
The checksum is equal to the least si nificant byte
of the summin the hex values of all characters
transmitted.
Example:
The displayed position is 8,29. The position of device
with address 01 is read by means of the |01TPOS
command.
The answer is: 01TPOS:+008290F
The sum of hex values of all characters is the
followin : 30+31+54+50+4F+53+3A+2B+30+30+
38+ 32+39 = 30F
The least si nificant byte of 30F is 0F which is the
checksum.
7.2.1 Command list
(below the device address is indicated with AD)
Zeroing of device address
|00RSET
Device address is set to value 0.
Device address [1, 31]
|00INIT=X
Sets device address to value X.
Display device address
|00DADR
Displays device address until P key is pushed.
Change device address [1, 31]
|ADRADR=X
Chan es actual device address AD to X.
Answer: ADRADR:+XCHKS (CHKS is checksum and X
is value).
Read actual position
|ADTPOS
Reads actual position of device AD (resolution is
0,01mm).
Change counting direction [0, 1]
|ADRDIR=X
Sets countin direction.
X=0 → uP = standard direction
X=1→ dn = inverted direction
Answer: ADRDIR:+00000XCHKS
Read counting direction
|ADTDIR
Reads the actual countin direction.
X=0→uP , X=1→dn
Answer: ADTDIR:+0000XCHKS
MAN LD120 I_E 1.5 Pa . 13 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
Decimal point [0, 3]
|ADRDEC=X
Modification of decimal point position.
0 = no decimal point
1 = one decimal (e. . 1,0)
2 = two decimals (e. . 1,00)
3 = three decimals (e. . 1,000)
Answer: ADRDEC:+0000XCHKS
ATTENTION: decimal point settin has no influence
on mm/inch display mode (see command |
ADRMMI=0).
Read decimal point position
|ADTDEC
Reads value of actual decimal point position.
X=0→ 0, X=1→ 1, X=2→ 2, X=3→ 3
Answer: ADTDEC:+0000XCHKS
Display mm/inch [0, 1]
|ADRMMI=X
Sets mm or inch display mode.
X=0→ mm
X=1→ inch
Answer: ADRMMI:+0000XCHKS
Read mm/inch display mode
|ADTMMI
Reads value of actual display mode.
X=0→ mm, X=1→ inch
Answer: ADTMMI:+0000XCHKS
Incremental measurement function [0, 1]
Enables incremental measurement function (key
combination P and *).
|ADRRLA=X
X=0→ oFF (function off)
X=1→ on (function on)
Answer: ADRRAE:+0000XCHKS
Read incremental measurement
|ADTRAE
Reads status of incremental measurement function.
X=0→ oFF, X=1→ on
Answer: ADTRAE:+0000XCHKS
Incremental measurement [0, 1]
Sets from absolute display mode to incremental
(relative).
|ADRRLA=X
X=0→ oFF
X=1→ on
Answer: ADRRAE:+0000XCHKS
Read incremental measurement
|ADTRLA
Reads status of absolute/incremental display mode.
X=0→ oFF, X=1→ on
Answer: ADTRLA:+0000XCHKS
Datum function [0, 1]
|ADRRSE=X
Enables datum function (by pushin * key).
X=0→ oFF (function off)
X=1→ on (function on)
Answer: ADRRSE:+0000XCHKS
Read datum function
|ADTRSE
Reads status of datum function.
X=0→ oFF, X=1→ on
Answer: ADTRSE:+0000XCHKS
Datum value modification [0, 1]
|ADRRFE=X
Enables datum value modification (by key
combination P and ).
X=0→ oFF (not enables)
X=1→ on (enabled)
Answer: ADRRFE:+0000XCHKS
Read datum value modification
|ADTRFE
Reads status of datum value modification.
X=0→ oFF, X=1→ on
Answer: ADTRFE:+0000XCHKS
Offset function [0, 1]
|ADROFE=X
Enables offset function (by key combination P and
)
X=0→ oFF (function off)
X=1→ on (function on)
Answer: ADROFE:+0000XCHKS
MAN LD120 I_E 1.5 Pa . 14 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
Read offset function
|ADTOFE
Reads status of offset function.
X=0→ oFF, X=1→ on
Answer: ADTOFE:+0000XCHKS
Resolution [10, 50, 100, 1000]
|ADRRES=X
Sets linear resolution.
X=10→ 0,01 mm
X=50→ 0,05 mm
X=100→ 0,1 mm
X=1000→ 1 mm
Answer: ADRRES:+XCHKS
Read resolution
|ADTRES
Reads value of actual resolution.
X=10→ 0,01 mm, X=50→ 0,05 mm,
X=100→ 0,1 mm, X=1000→ 1 mm
Answer: ADTRES:+XCHKS
Free resolution factor FrEE [0.0001, 1.0000]
|ADRFRE=X
Sets free resolution factor FrEE (see chapter 6.3.2).
Answer: ADRFRE:+XCHKS
Read FrEE factor
|ADTFRE
Reads value of actual FrEE factor.
Answer: ADTFRE:+X.XXXXCHKS
Datum value [-99999, 99999]
|ADRREF=X
Absolute reference value for the measurement
system. Resolution is 0,01.
Answer: ADRREF:XCHKS
Read datum value
|ADTREF
Reads actual datum value.
Answer: ADTREF:XCHKS
Offset1 value [-99999, 99999]
|ADROF1=X
Sets Offset1 (OFSt1) value (resolution 0,01mm).
Answer: ADROF1:XCHKS
Read Offset1 value
|ADTOF1
Reads actual Offset1 value.
Answer: ADTOF1:XCHKS
Offset2 value [-99999, 99999]
|ADROF2=X
Sets Offset2 (OFSt2) value (resolution 0,01mm).
Answer: ADROF2:XCHKS
Read Offset2 value
|ADTOF2
Reads actual Offset2 value.
Answer: ADTOF2:XCHKS
Offset3 value [-99999, 99999]
|ADROF3=X
Sets Offset3 (OFSt3) value (resolution 0,01mm).
Answer: ADROF3:XCHKS
Read Offset3 value
|ADTOF3
Reads actual Offset3 value.
Answer: ADTOF3:XCHKS
Save actual value function [0, 1]
|ADRSPE=X
Saves the last displayed value after power off.
X=0→ oFF
X=1→ on
Answer: ADRSPE:+0000000XCHKS
Read save actual value function
|ADTSPE
Reads status of save actual value function.
X=0→ oFF, X=1→ on
Answer: ADTSPE:+0000000XCHKS
8 - Dimensional drawing and cut-out
8.1 Display Cut-out
Provide a 68 x 33 mm2 (w x h) cut-out.
8.2 SM5 sensor
Check details on product catalo ue.
MAN LD120 I_E 1.5 Pa . 15 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]
LD120
Rev. Man.Vers. Description
0 1.0 1^ issue
1.- SW + Manual update
5 1.5 Chap.6 correction
Lika Electronic
Via S. Lorenzo, 25 - 36010 Carrè (VI) - Italy
Tel. +39 0445 382814
Fax +39 0445 382797
Italy: eMail [email protected] - www.lika.it
World: eMail [email protected] - www.lika.biz
MAN LD120 I_E 1.5 Pa . 16 www.lika.it
www.lika.biz
PH SYSTEMS - www.phsystems.be - [email protected]

Other Lika Monitor manuals

Lika LD210 User manual

Lika

Lika LD210 User manual

Lika LD350 User manual

Lika

Lika LD350 User manual

Lika Posicontrol LD220 Series User manual

Lika

Lika Posicontrol LD220 Series User manual

Lika LD355 User manual

Lika

Lika LD355 User manual

Lika LD200 User manual

Lika

Lika LD200 User manual

Lika SM5 Series User manual

Lika

Lika SM5 Series User manual

Lika LD200 User manual

Lika

Lika LD200 User manual

Lika POSICONTROL LD111 Series User manual

Lika

Lika POSICONTROL LD111 Series User manual

Lika MC150 series User manual

Lika

Lika MC150 series User manual

Lika MC111 User manual

Lika

Lika MC111 User manual

Lika IFS-10-PM User manual

Lika

Lika IFS-10-PM User manual

Lika LD14 Series User manual

Lika

Lika LD14 Series User manual

Lika LD14 Series User manual

Lika

Lika LD14 Series User manual

Lika LD140 Series User manual

Lika

Lika LD140 Series User manual

Popular Monitor manuals by other brands

Kenwood SM-230 manual

Kenwood

Kenwood SM-230 manual

Axis T8412 user manual

Axis

Axis T8412 user manual

LG W1953SE user guide

LG

LG W1953SE user guide

Mermaid ventura 150 TFT x4 user manual

Mermaid

Mermaid ventura 150 TFT x4 user manual

ViewSonic VSD220 user guide

ViewSonic

ViewSonic VSD220 user guide

Panasonic TH-42PF30U operating instructions

Panasonic

Panasonic TH-42PF30U operating instructions

Samsung S27E332H user manual

Samsung

Samsung S27E332H user manual

SMART SBID-7075 Installation and maintenance guide

SMART

SMART SBID-7075 Installation and maintenance guide

D-Boss LT-26HMF owner's manual

D-Boss

D-Boss LT-26HMF owner's manual

Acer V206HQL Guide

Acer

Acer V206HQL Guide

Elo TouchSystems M467 User manual & installation guide

Elo TouchSystems

Elo TouchSystems M467 User manual & installation guide

Maxell MA-6510WH user manual

Maxell

Maxell MA-6510WH user manual

Philips 200P4MG/00 Specifications

Philips

Philips 200P4MG/00 Specifications

HP M2350A Service manual

HP

HP M2350A Service manual

Eneo VM-UHD32M Quick installation guide

Eneo

Eneo VM-UHD32M Quick installation guide

Sharp LL-S242A-W Setup manual

Sharp

Sharp LL-S242A-W Setup manual

Eizo FlexScan M1900 Specifications

Eizo

Eizo FlexScan M1900 Specifications

LG 34BL85C owner's manual

LG

LG 34BL85C owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.