LINET SafeSense 3 Parts list manual

SafeSense® 3
Monitoring System
D9U001SFS-0101
Version: 03
Publication Date: 2021-09
Instructions for Use
and Technical Description
Stand-alone Version, Version for Image 3 and
Version for Eleganza 2

D9U001SFS-0101_032
Manufacturer:
L I N E T spol. s r.o.
Želevčice 5
274 01 Slaný
Tel.: +420 312 576 111
Fax: +420 312 522 668
E-mail: [email protected]
http://www.linet.com
Service department: [email protected]
SafeSense® 3
Monitoring System
Author: LINET, s.r.o.
Related links: www.linet.com
D9U001SFS-0101
Version: 03
Publication Date: 2021-09
Copyright © L I N E T, s.r.o., 2021
Translation © L I N E T, s.r.o., 2021
All rights reserved.
All trademarks and brands are the property of the appropriate owners. The manufacturer reserves the right to changes in the
contents of the instructions for use that relate to the product´s technical regulations. It is for this reason that the contents of the in-
structions for use may indicate dierences from the current manufacture of the product. Reproduction, also excerpts, only with prior
permission of the publisher. Subject to changes due to technical developments. All technical data are rated data and are subject to
construction and manufacturing tolerances.

D9U001SFS-0101_03
3
Table of Contents
1 Symbols and Denitions .................................................................................................................................................4
1.1 Warning Notices..............................................................................................................................................................4
1.1.1 Types of Warning Notices ............................................................................................................................................4
1.1.2 Structure of Warning Notices .......................................................................................................................................4
1.2 Instructions......................................................................................................................................................................4
1.3 Lists.................................................................................................................................................................................4
1.4 Symbols on the Package ................................................................................................................................................5
1.5 Symbols and Labels on the Product ...............................................................................................................................6
1.6 Product Labels ................................................................................................................................................................9
1.7 Visual signalisation........................................................................................................................................................11
1.7.1 LEDS ON THE HUB CONTROL PANEL....................................................................................................................11
1.8 Denitions .....................................................................................................................................................................11
1.9 Abbreviations ............................................................................................................................................................... 12
2 Safety Instructions........................................................................................................................................................ 13
3 Intended Use (Stand-alone Version, Version for Image 3 and Version for Eleganza 2) ......................................... 15
3.1 User population............................................................................................................................................................ 15
3.2 Contraindications ......................................................................................................................................................... 15
3.3 Operator....................................................................................................................................................................... 15
4 Mattresses compatible with SafeSense® 3 ................................................................................................................ 15
5 Product Description...................................................................................................................................................... 16
5.1 System Desciption ....................................................................................................................................................... 16
5.1.1 Stand-alone Version.................................................................................................................................................. 16
5.1.2 Version for Image 3 (Image 3 bed with SafeSense® 3)............................................................................................ 17
5.1.3 Version for Eleganza 2 (Eleganza 2 bed with SafeSense® 3).................................................................................. 18
5.2 Hub .............................................................................................................................................................................. 19
5.3 Air Pad ......................................................................................................................................................................... 20
5.4 Wetness Pad................................................................................................................................................................ 21
6 Technical Specication ................................................................................................................................................ 22
6.1 Mechanical Specications............................................................................................................................................ 22
6.2 Technical Requirements on the Part of the Customer.................................................................................................. 22
6.3 Environment conditions................................................................................................................................................ 23
6.3.1 Use Conditions.......................................................................................................................................................... 23
6.3.2 Storage and Transport Conditions ............................................................................................................................ 23
6.4 Electrical Specications ............................................................................................................................................... 23
6.5 Electromagnetic Compatibility...................................................................................................................................... 24
6.5.1 Manufacturer instructions - electromagnetic emissions ............................................................................................ 24
6.5.2 Manufacturer instructions - electromagnetic susceptibility........................................................................................ 25
7 Use and Storage Conditions........................................................................................................................................ 26
8 Scope of Delivery and Product Variants ..................................................................................................................... 26
8.1 Delivery ........................................................................................................................................................................ 26
8.2 Scope of Delivery......................................................................................................................................................... 26
8.3 SafeSense® 3 Variants................................................................................................................................................ 27
8.3.1 Stand-alone Version.................................................................................................................................................. 27
8.3.2 Version for Image 3................................................................................................................................................... 27
8.3.3 Version for Eleganza 2.............................................................................................................................................. 27
9 Putting into Service ...................................................................................................................................................... 28
9.1 Stand-alone SafeSense® 3 ......................................................................................................................................... 28
9.2 Image 3 with SafeSense® 3 ........................................................................................................................................ 33
9.2.1 Instructions for putting into service ........................................................................................................................... 34
9.3 Eleganza 2 with SafeSense® 3 ................................................................................................................................... 44
9.3.1 Instructions for putting into service ........................................................................................................................... 45
10AC/DC Adaptor (only for Stand-alone Version) ........................................................................................................ 56
11 Manipulation ................................................................................................................................................................ 57
11.1 Bed Exit Monitoring .................................................................................................................................................... 58
11.2 Motion Monitoring....................................................................................................................................................... 58
11.3 Decubitus/Movement Notication............................................................................................................................... 58
11.4 Wetness Monitoring.................................................................................................................................................... 58
11.5 Control Elements........................................................................................................................................................ 59
11.5.1 Hub Control Panel................................................................................................................................................... 59
11.5.2 CareMonitor............................................................................................................................................................. 60
11.5.3 Attendant Control Panel with SafeSense® 3 Control (only for Image 3 and Eleganza 2)....................................... 82
12 Cleaning/Disinfection ................................................................................................................................................. 87
13 Troubleshooting .......................................................................................................................................................... 88
13.1 Error Codes................................................................................................................................................................ 89
14 Maintenance ................................................................................................................................................................ 90
14.1 Regular maintenance................................................................................................................................................. 90

D9U001SFS-0101_034
1 Symbols and Denitions
1.1 Warning Notices
1.1.1 Types of Warning Notices
Warning notices are dierentiated by the type of danger using the following key words:
► CAUTION warns about the risk of material damage.
► WARNING warns about the risk of physical injury.
► DANGER warns about the risk of fatal injury.
1.1.2 Structure of Warning Notices
SIGNAL WORDS!
Type and source of danger!
► Measures to avoid the danger.
1.2 Instructions
Structure of instructions:
► Perform this step.
Results, if necessary.
1.3 Lists
Structure of bulleted lists:
■ List level 1
□ List level 2
● List level 3
14.2 Spare Parts ................................................................................................................................................................ 90
14.3 Safety Technical Checks............................................................................................................................................ 90
15 Disposal ....................................................................................................................................................................... 91
15.1 Environment Protection.............................................................................................................................................. 91
15.2 Disposal ..................................................................................................................................................................... 91
15.2.1 Within Europe ......................................................................................................................................................... 91
15.2.2 Outside Europe ....................................................................................................................................................... 91
16 Warranty....................................................................................................................................................................... 92
17 Standards and Regulations ....................................................................................................................................... 92

D9U001SFS-0101_03
5
1.4 Symbols on the Package
FRAGILE, HANDLE WITH CARE
THIS WAY UP
DO NOT POLLUTE THE ENVIRONMENT
PAPER RECYCLING SYMBOL

D9U001SFS-0101_036
1.5 Symbols and Labels on the Product
READ INSTRUCTIONS FOR USE
ACTIVATION BUTTON (TO ACTIVATE SafeSense® 3)
PLAY OR STOP BUTTON (TO ACTIVATE OR DEACTIVATE SELECTED MONITORING)
SELECT AND TIME BUTTON (SELECT TYPE OF MONITORING OR SET PERIOD OF
SELECTED MONITORING)
WEEE SYMBOL (RECYCLE AS ELECTRONIC WASTE,
DO NOT PUT INTO THE HOUSEHOLD WASTE)
MANUFACTURER
MANUFACTURING DATE
ON
OFF
WARNING
ONLY SUITABLE FOR INDOOR USE

D9U001SFS-0101_03
7
APPLIED PARTS TYPE B
REFERENCE NUMBER (PRODUCT TYPE DEPENDING ON CONFIGURATION)
SERIAL NUMBER
BATTERY (HUB)
PLAY ICON (ON)
STOP ICON (OFF)
DECUBITUS/MOVEMENT NOTIFICATION
MOTION MONITORING
BED EXIT MONITORING
WETNESS MONITORING

D9U001SFS-0101_038
CHECKED BY AUTHORIZED SERVICE DEPARTMENT
DO NOT WASH (AIR PAD)
DO NOT BLEACH (AIR PAD)
DO NOT TUMBLE DRY (AIR PAD)
DO NOT WRING (AIR PAD)
DO NOT IRON (AIR PAD)
PROFESSIONAL CHEMICAL CLEANING - MODERATE PROCEDURE (AIR PAD)
DO NOT USE PHENOL (AIR PAD)
DRY NATURALLY (AIR PAD)
RINSE WITH WATER (AIR PAD)
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition

D9U001SFS-0101_03
9
DISINFECT USING SOLUTION CONTAINING LESS THAN 1000 ppm OF CHLORINE
(REFER TO INSTRUCTIONS FOR USE) (AIR PAD)
HANDWASH WITH DETERGENT (INITIAL TEMPERATURE OF HOT WATER SHOULD
NOT EXCEED 50°C) (AIR PAD)
COVER MATERIALS ARE FIRE RESISTANT
TO BS7175, SOURCE 0, 1 AND 5 (AIR PAD)
CE MARKING
MEDICAL DEVICE (COMPATIBLE WITH MEDICAL DEVICE REGULATION)
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition
1.6 Product Labels
Serial label of the SafeSense 3 Monitoring System is placed on the Hub bottom side. The serial label contains information about
manufacturer‘s address, manufacturing date, product reference number and product serial number. Hub serial label is also placed
on the Hub bottom side. The Hub serial label contains information about manufacturer‘s address, manufacturing date, component
reference number, component serial number, symbols and electrical specication.

D9U001SFS-0101_0310
IPXX
20 -XX XX
S6018957
REFREF
VeVersion: HW: FW: XX.XX XX.XX
bar code
XXXXXXXXXXXXXXXX
Power In: 24-40 VDC, max. 0,5 A
1: Li-ion 7,3 V / 2,6 Ah
2: CR1220
HUB - SafeSense
bar codebar code
L I N E T spol. s r.o.
Želevčice 5, 274 01 Slaný
Czech Republic
Made in Czech Republic
Fig. SafeSense® 3 Hub Serial Label (example)
Fig. SafeSense® 3 Air Pad
Type Label
Fig. SafeSense® 3 Air Pad
Wash Label
LINET spol. s r. o.,
Želevčice 5, 274 01 Slaný
The Czech Republic
COVER TEXTILE(S) PASS
THE REQUIREMENTS
OF CTB 117-2013 SECTION 1
(CLASS 1 NORMAL FLAMMABILITY)
and 16 CFR Part 1632.6-2013 Edition

D9U001SFS-0101_03
11
1.7 Visual signalisation
SIGNALISATION MEANING
ACTIVATION LED GREEN SafeSense® 3 ON - battery charged to more than 75%
LIGHT BLUE SafeSense® 3 ON - battery charged to 25 - 75%
RED SafeSense® 3 ON - battery fault
LED OFF SafeSense® 3 OFF
PLAY/STOP LED GREEN monitoring ON
BLUE monitoring OFF
ERROR LED RED internal error
LED OFF no error detected
1.7.1 LEDS ON THE HUB CONTROL PANEL
1.8 Denitions
Adult Patient having a physical size equal to or more than 146 cm, a
mass equal to or more than 40 kg and a body mass index (BMI)
equal to or more than 17 (according to IEC 60601-2-52).
Alert Information signal (other than alarm signal)
MQTT Broker MQTT Broker is a software part of the SafeSense® 3 system.
MQTT (Message Queuing Telemetry Transport) represents a light-
weight communication protocol for the internet of things according
to the standard ISO/IEC 20922. The MQTT Broker consists of
software components running on the server allowing to share mes-
sages among connected devices (clients with SafeSense Applica-
tion and Hubs). The MQTT Broker is congured to use encrypted
communication in a network to eliminate vulnerable and insecure
client connections.

D9U001SFS-0101_0312
AC ( ~ ) Alternating Current
ACP Attendant Control Panel
CE European Conformity
CPR Cardiopulmonary Resuscitation
dBA Sound Intensity Unit
DC ( ) Direct Current
EMC Electromagnetic Compatibility
HF High Frequency
HW Hardware
ICU Intensive Care Unit
IP Ingress Protection
IV Intravenous
LED Light Emitting Diodes
ME Medical Electrical (Equipment)
ON Activation
OFF Deactivation
ppm parts per million, millionth (1000 ppm = 0,1%)
REF Reference Number (product type depending on conguration)
SN Serial Number
SW Software
SWL Safe Working Load
USB Universal Serial Bus
WEEE Waste Electrical and Electronic Equipment
1.9 Abbreviations

D9U001SFS-0101_03
13
WARNING!
An additional multiple socket-outlet or extension cord shall not be connected to the medical electrical system.
2 Safety Instructions
WARNING!
This medical device is portable medical electrical equipment!
WARNING!
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or patient is established!
WARNING!
No modication of this equipment is allowed.
WARNING!
Inappropriate handling of the power supply cord, e. g. by kinking, shearing or other mechanical damages is
hazardous!
WARNING!
When routing cables from SafeSense® 3 in a medical bed avoid squeezing those between parts of the medical
bed!
WARNING!
Incompatible mattresses can put system functions out of order.
WARNING!
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
WARNING!
Only authorised and trained person using the tool is allowed to change Hub battery!
WARNING!
This medical device is not intended for oxygen enriched environment!
WARNING!
This medical device is not intended for use with ammable substances!
WARNING!
You cannot rely unconditionally on alert notications! This medical device is not intended to be used as an
alarm system! SafeSense® 3 was not developed or tested as a diagnostic tool and must not be considered as
such or as another instrument of that type.

D9U001SFS-0101_0314
Additional instructions for correct use:
► Follow the instructions for use carefully.
► Ensure any user has read and understood the instructions for use completely before operating the product.
► Use the system exclusively if it is in perfect working order.
► If necessary, check the system functions daily or at each shift change.
► Use the system exclusively with the correct mains supply.
► Ensure that the system is operated exclusively by qualied personnel who have been trained according to the instructions
for use.
► Ensure that the patient (health permitting) has been informed about the operation of the system and all applicable safety
instructions.
► Contact service department of the manufacturer immediately to let it replace any damaged parts by the original spare
parts.
► Ensure that maintenance and installation are performed exclusively by qualied personnel who have been trained by the
manufacturer.
► To avoid injury or crushing, take extra caution when operating any moving parts of the bed.
► Never use the system in areas where there is a hazard of explosion.
► Never handle the mains plug with wet hands.
► Disconnect the product from the mains exclusively by pulling the mains plug.
► When pulling the mains plug, always hold the plug, not the cable.
► Position the mains cable so that there are no loops or kinks in the cable; protect the cable from mechanical wear and tear.
► Improper handling of mains cable can cause an electric shock hazard and other serious injuries or damages.
► To prevent failures, use exclusively the manufacturer’s original accessories and mattresses in perfect working order.
► Do not hang anything on any cable.
► Select a suitable location for the placement of bed accessories and other objects to prevent involuntary activation of
buttons or controls which may result in the adjustment of bed positioning.
► Replace the Air Pad once every 2 years to ensure its full functionality!

D9U001SFS-0101_03
15
3 Intended Use (Stand-alone Version, Version for Image 3 and
Version for Eleganza 2)
The SafeSense® 3 Monitoring System is intended to use for the automated, contactless and unobtrusive continuous
mobility detection of patients in bed, bed exit and continuous wetness detection in the context of healthcare and nursing
facilities.
3.1 User population
► Not specic. Indication is limited by limitation of hospital/nursing bed patients (weight ≥ 40 kg and ≤ 180kg , height >= 146
cm, BMI >= 17)
► Caregivers (nurses, doctors, technical personnel, transport personnel, cleaning personnel)
3.2 Contraindications
► Patients in whom proper positioning cannot be achieved or maintained.
► Patients who do not meet the weight limits tested or specied.
► An explosive atmosphere or in the presence of ammable anesthetics or gases.
3.3 Operator
► Health care professional (nurses, doctors)
4 Mattresses compatible with SafeSense® 3
Compatible Mattresses:
■ EectaCare 10
■ EectaCare 20
■ PrimaCare 10
■ PrimaCare 20
■ CliniCare 10
■ CliniCare 20
■ CliniCare 30
■ Universal mattress Wul
■ ViskoMatt Comfort
■ MicroMatt 7
■ HeavyMatt 270
■ ViskoMatt 2
■ ViskoMatt 3
■ ViskoMatt 5
■ ViskoMatt 3 bariatric
WARNING!
Following mattresses are manufactured in versions with dierent dimensions. Use only mattresses whose
dimensions are compatible with corresponding medical bed!

D9U001SFS-0101_0316
5 Product Description
Nurse Call System
LAN/WiFi
Air Pad
Air
Hose
Wetness Pad
Hub
5.1 System Desciption
Based on the functional principle of ballistocardiography, patients are monitored by a bed-side hardware unit of the SafeSense®
3 Monitoring System. This system consists of an electronic-free pad (SafeSense® 3 Air Pad) located under the mattress, which
is connected to a hardware unit (SafeSense® 3 Hub) through an air hose. From there, the data is sent to remote devices via (W)
LAN. The data is visualized using an SafeSense Application. Targeted and ecient patient care is also supported by an integrated
notication function.
5.1.1 Stand-alone Version
Fig. Diagram of the SafeSense® 3 Stand-alone system

D9U001SFS-0101_03
17
Nurse Call System
LAN/WiFi
Image 3 Control Unit
Air Pad
Air
Hose
315
Hirschmann
7 pin
connector
Inputs
Wetness Pad
ACP with
SafeSense 3
Control
Connectivity
Box
Hub
5.1.2 Version for Image 3 (Image 3 bed with SafeSense® 3)
Fig. Diagram of the SafeSense 3 system with Image 3 bed
WARNING!
SafeSense® 3 Monitoring System in Version for Image 3 is compatible only with Image 3 medical bed and de-
pends on the specic conguration of the Image 3 bed because Image 3 bed must be SafeSense® 3 ready for
use with SafeSense® 3!

D9U001SFS-0101_0318
Nurse Call System
LAN/WiFi
Eleganza 2 Control Unit
Air Pad
Air
Hose
315
Hirschmann
7 pin
connector
Inputs
Wetness Pad
ACP with
SafeSense 3
Control
Connectivity
Box
Hub
5.1.3 Version for Eleganza 2 (Eleganza 2 bed with SafeSense® 3)
Fig. Diagram of the SafeSense 3 system with Eleganza 2 bed
WARNING!
SafeSense® 3 Monitoring System in Version for Eleganza 2 is compatible only with Eleganza 2 medical bed
and depends on the specic conguration of the Eleganza 2 bed because Eleganza 2 bed must be SafeSense®
3 ready for use with SafeSense® 3!

D9U001SFS-0101_03
19
5.2 Hub
1 2 3
1. Hub Control Panel
2. Air Hose
3. Attendant Control Panel Connector with plug (only for Image 3 and Eleganza 2 medical bed)
Hub interconnects parts of the SafeSense® 3 system and enables communication between the SafeSense® 3 system installed on
a medical bed and CareMonitor user interface.

D9U001SFS-0101_0320
5.3 Air Pad
Fig. Air Pad with Air Hose
Air Pad connected to the Hub via Air Hose enables Bed Exit Monitoring and Motion Monitoring.
Table of contents
Other LINET Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Teledyne
Teledyne T3VNA use & care

Apogee
Apogee MP-100 owner's manual

Spectris
Spectris PARTICLE MEASURING SYSTEMS MiniCapt Operation manual

PCB Piezotronics
PCB Piezotronics 3710 Series Installation and operating manual

Geokon
Geokon 4410 instruction manual

Summit Technology
Summit Technology PowerSight PS4500 user manual