8
The bed may not be used if faults were detected which may damage the patient, staff or a third
person, the bed or the surrounding interior
Before using the bed, the backup batterymust be charged first (if the bed is equipped with a
battery, it is alreadyconnected to the source when delivered); some bed adjustments cannot be
carried out without a connected battery
Onlypersons mayoperate the bed who have the knowledge or practical experience that
guarantees problem-free handling of the bed
The operating staff is obliged to acquaint the bedridden patient with the control functions that are
designed for the patient; this above all involves handling the hand control; the patient must also be
acquainted with the safety instructions (depending on the patient’s condition)
The user (operator) is obliged, before use, to make sure the bed is functional and fault-free
The bed may be used only on even and solid floors and under the conditions stated in paragraph
“Conditions of Use “
The bed may not be overloaded even for a short time, however only if this is truly essential in
specific situations (such as during reanimation), the individual parts of the mattress platform must
be in the lowest position
If a patient is on the bed, the brakes on the castors must be locked (except when travelling with the
bed) –if the bed brakes are unlocked, there is the danger that the patient, when leaving the bed or
leaning against it, may suffer an accident.
During the transport of the bed with the patient, the side-rails must be always locked in upper
position.
The mattress platform height must be adjusted to the height and condition of the patient
No other persons may sit on the bed
When handling the mobile parts of the bed, care must be taken that the patient, other person or
objects are not trapped
If a lifting pole or infusion stand is positioned on the bed, greater attention must be paid when
travelling, lifting or tipping to the space around the lifting pole and infusion stand
During the adjustment to Trendelenburg position, staff must give attention to the distance of lifting
pole (if it is used on the bed) and a wall, objects placed on the wall or other ambient objects. There
is a potential risk of collision and damage of ambient objects
Before cleaning or maintaining the bed, a feeding cable must always be disconnected from the
mains.
The bed may not be used in an environment where there is danger of an explosion, or the
presenceof inflammable anaesthetics
When making repairs to the bed, onlyoriginal materials and components maybe used; when
replacing damaged parts with non-original or unsuitable materials, the manufacturer shall not be
liable for any damage that may arise
Whenmaintaining and cleaning the bed in the space between the undercarriage and mattress
platform the adjustment functions must be blocked on the blocking panel so no accident occurs
due to the unwanted switching on of the foot switches and subsequent movement of the mattress
platform in a downward direction
Only Linet mattresses and accessories may be used on the bed; if incompatible mattresses are
used this may result in the bad function of the siderails and endanger the safetyof patients
When transporting the bed, the mains cable must be suspended on a plastic hook on the bed end
or other safe place to prevent the cable for being run over and damaged
The mains cable must be led in such a manner so it does not get wound round mobile parts or
constricted between them; a serious danger of an electrical current accident may arise if the mains
cable is damaged
The movable parts must be checked for wear and tear at least once a month.
In case of power failure or insufficient electric tension it is possible to adjust safety positions of
mattress platform thanks to the backup battery (if the bed is equipped with backup battery) and
with the help of mechanical unblocking of backrest.