Capintec CAPTUS 600 User manual

CAPTUS®600
THYROID UPTAKE SYSTEM
OWNER’S MANUAL
Capintec, Inc.
Sales and Marketing
and Customer Support
6 Arrow Road
Ramsey, NJ 07446
Phone (800) ASK-4CRC
Fax (201) 825-1336
Capintec, Inc.
Service Center
620 Alpha Drive
Pittsburgh, PA 15238
Phone (800) 227-6832
Fax (412) 963-0610
Copyright©1995 Capintec, Inc.
CAPTUS®is a registered trademark of Capintec, Inc.
ALL RIGHTS RESERVED
Manual Stock No. 9250-0072
Rev. M – November 2007


CAPINTEC, INC. CAPTUS®600
TABLE OF CONTENTS
PREFACE
SAFETY................................................................................................................................1-1
SYMBOL DEFINITIONS....................................................................................................1-1
WARNING AND CAUTION LABELS.................................................................................1-2
CAUTIONS AND NOTES ..................................................................................................1-5
GENERAL SAFETY TIPS..................................................................................................1-6
FUNCTIONAL & TECHNICAL DESCRIPTION....................................................................2-1
GENERAL..........................................................................................................................2-1
TECHNICAL DESCRIPTION.............................................................................................2-1
Detector .........................................................................................................................2-2
Multichannel Analyzer....................................................................................................2-2
TECHNICAL SPECIFICATIONS .......................................................................................2-2
Thyroid Uptake Stand....................................................................................................2-2
System Power Requirements ........................................................................................2-3
Environment Requirements ...........................................................................................2-3
Operational................................................................................................................2-3
Storage......................................................................................................................2-3
Isolation Transformer.....................................................................................................2-4
Printer ............................................................................................................................2-4
Readout Unit..................................................................................................................2-4
Nuclear Data..................................................................................................................2-4
Calibration Source Data.................................................................................................2-5
Multichannel Pulse Height Analyzer ..............................................................................2-5
Diagnostics and Test .....................................................................................................2-5
Utilities ...........................................................................................................................2-5
Display...........................................................................................................................2-5
Two Separate Inputs......................................................................................................2-5
Probe Detector...............................................................................................................2-6
Well Detector – Optional................................................................................................2-6
Regulatory Listings ........................................................................................................2-6
EMC...........................................................................................................................2-6
Electrical....................................................................................................................2-6
ETL Listed .................................................................................................................2-6
FUNCTIONAL DESCRIPTION ..........................................................................................2-7
Thyroid Uptake...............................................................................................................2-7
Thyroid Uptake Nuclides................................................................................................2-7
Pre-Dose Measurement.................................................................................................2-7
Thyroid Uptake Protocols...............................................................................................2-7
One Capsule Method.................................................................................................2-8
Standard Capsule Method.........................................................................................2-8
Two Capsule Method.................................................................................................2-8
Multi-Capsules...........................................................................................................2-9
Wipe Tests.....................................................................................................................2-9
November 07 TABLE OF CONTENTS TOC-i

CAPINTEC, INC. CAPTUS®600
Lab Tests – Urine......................................................................................................2-9
Lab Tests – Blood...................................................................................................2-10
Staff Bioassay..............................................................................................................2-10
GENERAL OPERATING OVERVIEW..................................................................................3-1
TURN ON SYSTEM ..........................................................................................................3-1
LOG-ON PASSWORD ......................................................................................................3-1
SYSTEM FUNCTIONS......................................................................................................3-2
PROBE SCREEN..............................................................................................................3-3
WELL SCREEN.................................................................................................................3-3
HOME Key ........................................................................................................................3-4
Changing from PROBE to WELL Functions......................................................................3-4
Changing from WELL to PROBE Functions......................................................................3-4
CONTRAST.......................................................................................................................3-4
SCREEN DIMMER............................................................................................................3-4
FUNCTION Keys...............................................................................................................3-5
PRINT Key ........................................................................................................................3-5
UTIL Key ...........................................................................................................................3-5
KEYBOARD ......................................................................................................................3-6
BSP Key............................................................................................................................3-6
PROMPT...........................................................................................................................3-6
ARROW Keys....................................................................................................................3-6
ALPHANUMERIC ENTRIES .............................................................................................3-7
Entries with Number and Letters ...................................................................................3-7
SPACE Key ...................................................................................................................3-7
Decimal Point.................................................................................................................3-7
BLANK ENTRIES..............................................................................................................3-8
INVALID VALUES .............................................................................................................3-8
ENTERING DATE and TIME.............................................................................................3-8
SETUP PASSWORD ........................................................................................................3-8
YES or NO Answers..........................................................................................................3-9
COUNTING METHODS ....................................................................................................3-9
MEASUREMENT ERRORS............................................................................................3-10
HIGH VOLTAGE ADJUSTMENT ....................................................................................3-10
OFFSET VOLTAGE ADJUSTMENT...............................................................................3-10
DISPLAY CONTRAST ADJUSTMENT ...........................................................................3-11
DATA ENTRY INPUT LIMITS .........................................................................................3-11
Thyroid Uptake Protocols ............................................................................................3-11
Thyroid Uptake Measurements....................................................................................3-11
Cs137 Standard...........................................................................................................3-11
Bioassay Protocol........................................................................................................3-12
Additional Nuclide........................................................................................................3-12
Blood Volume ..............................................................................................................3-12
RBC Survival Measurement ........................................................................................3-12
Patient Demographics .................................................................................................3-12
Wipe ............................................................................................................................3-12
Description..............................................................................................................3-12
Background Counting..............................................................................................3-12
Generic Conversion Factors ...................................................................................3-12
Trigger Levels .........................................................................................................3-13
TOC-ii TABLE OF CONTENTS November 07

CAPINTEC, INC. CAPTUS®600
Staff Data.....................................................................................................................3-13
SYSTEM SETUP...................................................................................................................4-1
INSTALLATION.................................................................................................................4-1
ENVIRONMENT REQUIREMENTS ..................................................................................4-1
Operational ....................................................................................................................4-1
Storage ..........................................................................................................................4-1
POWER REQUIREMENTS ...............................................................................................4-2
TURN ON PROCEDURE ..................................................................................................4-2
GENERAL OPERATIONAL SETUP..................................................................................4-3
ACCEPTANCE TESTING..................................................................................................4-3
SETTING UP FOR THYROID UPTAKES OR WIPE TESTS ............................................4-4
SETTING UP FOR LAB TESTS (URINE & BLOOD).........................................................4-4
SETUP FOR STAFF BIOASSAY.......................................................................................4-4
SETUP FOR MULTICHANNEL ANALYZER .....................................................................4-4
SYSTEM INITIALIZATION....................................................................................................5-1
DATE & TIME SETUP.......................................................................................................5-1
Date Entry Format..........................................................................................................5-1
Changing the Time and Date Setting.............................................................................5-2
SETTING HOSPITAL OR FACILITY NAME......................................................................5-3
CURIES / BECQUEREL SETUP.......................................................................................5-4
Cs137 STANDARD SETUP...............................................................................................5-7
CHI-SQUARE TEST SETUP.............................................................................................5-9
CONVERSION FACTORS ..............................................................................................5-11
Entering Nuclide Name................................................................................................5-13
Selecting Measure or Input CF....................................................................................5-14
Input CF...................................................................................................................5-14
Measure CF.............................................................................................................5-15
ACCEPTANCE & QUALITY ASSURANCE TESTS.............................................................6-1
ACCEPTANCE TESTS......................................................................................................6-1
QUALITY ASSURANCE TESTS .......................................................................................6-1
Daily Tests.....................................................................................................................6-1
Monthly Tests.................................................................................................................6-1
Quarterly Tests ..............................................................................................................6-2
CALIBRATION TEST.........................................................................................................6-2
CHI-SQUARE TEST..........................................................................................................6-5
AUTO CALIBRATION TEST..............................................................................................6-5
MDA TEST.........................................................................................................................6-9
DIAGNOSTICS................................................................................................................6-12
Diagnostics Printout.....................................................................................................6-13
System Configuration..............................................................................................6-13
Nuclide Data............................................................................................................6-13
Additional Nuclides..................................................................................................6-14
User Keys................................................................................................................6-14
Standard Source......................................................................................................6-14
PROM Values..........................................................................................................6-14
Memory Tests..........................................................................................................6-14
November 07 TABLE OF CONTENTS TOC-iii

CAPINTEC, INC. CAPTUS®600
THYROID UPTAKE SETUP.................................................................................................7-1
COUNTING METHODS ....................................................................................................7-1
PROTOCOLS....................................................................................................................7-4
Adding a Protocol ..........................................................................................................7-6
Editing a Protocol.........................................................................................................7-11
Deleting a Protocol ......................................................................................................7-11
Printing Protocol Information .......................................................................................7-11
THYROID UPTAKE GENERAL INSTRUCTIONS................................................................8-1
ACCESSING THYROID UPTAKE STUDIES....................................................................8-1
ENTERING PATIENT DEMOGRAPHICS.........................................................................8-2
SELECTING PROTOCOL.................................................................................................8-3
SELECTING PATIENT......................................................................................................8-4
PRINTING THYROID UPTAKE PATIENT SUMMARY.....................................................8-5
VIEWING PATIENT INFORMATION.................................................................................8-5
SELECTING PATIENT TO MEASURE.............................................................................8-6
MEASURING THE NECK AND THIGH.............................................................................8-7
PRE-DOSE MEASUREMENT...........................................................................................8-9
PRINTING RESULTS......................................................................................................8-11
THYROID UPTAKE: ONE CAPSULE METHOD .................................................................9-1
GENERAL PROCEDURE .................................................................................................9-1
PRE-DOSE MEASUREMENT...........................................................................................9-1
ENTERING CAPSULE DATA............................................................................................9-2
MEASURING THE CAPSULE(S)......................................................................................9-3
ADMINISTERED DOSE....................................................................................................9-6
Administering the Dose Later ........................................................................................9-6
Entering Administered Dose Data .................................................................................9-7
UPTAKE MEASUREMENT...............................................................................................9-8
VIEWING RESULTS .........................................................................................................9-9
THYROID UPTAKE: TWO CAPSULE METHOD...............................................................10-1
GENERAL PROCEDURE ...............................................................................................10-1
PRE-DOSE MEASUREMENT.........................................................................................10-1
ENTERING ADMINISTERED DOSE...............................................................................10-1
MEASURING THE REFERENCE CAPSULE(S).............................................................10-3
UPTAKE MEASUREMENT.............................................................................................10-6
VIEWING RESULTS .......................................................................................................10-7
THYROID UPTAKE: STANDARD CAPSULE METHOD...................................................11-1
GENERAL PROCEDURE ...............................................................................................11-1
ACCESSING TU STANDARD.........................................................................................11-1
ADDING A TU STANDARD.............................................................................................11-2
EDITING A TU STANDARD............................................................................................11-3
DELETING A TU STANDARD.........................................................................................11-3
PRINTING TU STANDARD DATA..................................................................................11-4
MEASURING TU STANDARD........................................................................................11-4
SELECTING STANDARD FOR TU STUDY....................................................................11-6
PRE-DOSE MEASUREMENT.........................................................................................11-7
ENTERING ADMINISTERED DOSE...............................................................................11-7
TOC-iv TABLE OF CONTENTS November 07

CAPINTEC, INC. CAPTUS®600
UPTAKE MEASUREMENT .............................................................................................11-9
VIEWING RESULTS........................................................................................................11-9
WIPE TEST SETUP............................................................................................................12-1
INTRODUCTION.............................................................................................................12-1
WIPE DESCRIPTIONS....................................................................................................12-2
Entering Data for a Wipe Location...............................................................................12-2
Adding a Location........................................................................................................12-3
Editing or Deleting a Location......................................................................................12-4
TRIGGER LEVEL/COUNT TIME.....................................................................................12-4
GENERIC CONVERSION FACTORS.............................................................................12-6
ADDITIONAL NUCLIDES................................................................................................12-7
USER KEY ASSIGNMENT..............................................................................................12-9
WIPE TEST.........................................................................................................................13-1
INTRODUCTION.............................................................................................................13-1
MEASURING BACKGROUND........................................................................................13-1
MEASURING A WIPE .....................................................................................................13-3
WIPE RESULTS..............................................................................................................13-5
WIPE GRAPHICS SCREEN............................................................................................13-6
Scale/Cursor Mode......................................................................................................13-7
Activity of a Nuclide......................................................................................................13-7
Activity for Generic Conversion Factor ........................................................................13-7
WIPE REPORT................................................................................................................13-7
INTERPRETING WIPE RESULTS ..................................................................................13-9
LAB TESTS SETUP ...........................................................................................................14-1
INTRODUCTION.............................................................................................................14-1
GENERAL PROCEDURE................................................................................................14-1
STARTING LAB TESTS ..................................................................................................14-2
ENTERING PATIENT DEMOGRAPHICS .......................................................................14-3
MEASUREMENT PROCEDURE.....................................................................................14-4
LAB TESTS: URINE TEST.................................................................................................15-1
INTRODUCTION.............................................................................................................15-1
PERFORMING SCHILLING TEST MEASUREMENTS...................................................15-1
LAB TESTS: BLOOD TESTS.............................................................................................16-1
INTRODUCTION.............................................................................................................16-1
BLOOD VOLUME (Cr51).................................................................................................16-1
BLOOD VOLUME (I125)..................................................................................................16-6
RED BLOOD CELL SURVIVAL.....................................................................................16-10
STAFF BIOASSAY.............................................................................................................17-1
STAFF BIOASSAY DATA................................................................................................17-1
Adding Staff Member...................................................................................................17-2
Editing Staff Member Data...........................................................................................17-3
Deleting a Staff Member..............................................................................................17-4
ENTERING THE BIOASSAY PROTOCOL......................................................................17-4
X-RAY/GAMMA RATIO...................................................................................................17-5
November 07 TABLE OF CONTENTS TOC-v

CAPINTEC, INC. CAPTUS®600
MAKING BIOASSAY MEASUREMENTS........................................................................17-5
PRINTING STAFF BIOASSAY RESULTS......................................................................17-7
MULTICHANNEL ANALYZER SETUP..............................................................................18-1
INTRODUCTION.............................................................................................................18-1
MCA SETUPS.................................................................................................................18-1
Adding a Setup............................................................................................................18-3
Editing a Setup ............................................................................................................18-6
Deleting a Setup..........................................................................................................18-6
Printing All the Setups .................................................................................................18-6
MULTICHANNEL ANALYZER OPERATION ....................................................................19-1
MCA MEASUREMENTS.................................................................................................19-1
Calibration....................................................................................................................19-2
MEASURING BACKGROUND........................................................................................19-2
MEASURING A SOURCE...............................................................................................19-4
Activity of a Nuclide .....................................................................................................19-8
Exiting MCA Measurement..........................................................................................19-8
CLEANING AND MAINTENANCE.....................................................................................20-1
INTRODUCTION.............................................................................................................20-1
CLEANING INSTRUCTIONS..........................................................................................20-1
PREVENTATIVE MAINTENANCE..................................................................................20-2
Collimator Block Inspection .........................................................................................20-3
SERVICING.....................................................................................................................20-3
DISPOSAL ......................................................................................................................20-4
FUSE SERVICING..........................................................................................................20-5
Power Line Fuses........................................................................................................20-5
Readout Unit Fuses.....................................................................................................20-6
Printer Fuse.................................................................................................................20-6
ACCESSORIES AND REPLACEMENT PARTS.............................................................20-7
SHIPPING .......................................................................................................................20-7
INDEX
WARRANTY
DECLARATION OF CONFORMITY
TOC-vi TABLE OF CONTENTS November 07

CAPINTEC, INC. CAPTUS®600
PREFACE
Thank you for purchasing the CAPINTEC line of Thyroid Uptake Measurement Systems.
Every effort has been made to insure that the information in this document is complete,
accurate, and up-to-date. CAPINTEC assumes no responsibility for the results of errors
beyond its control. Mention of products manufactured by other companies does not
necessarily constitute endorsement by CAPINTEC, Inc.
CAPTUS®600 and CII are registered trademarks of CAPINTEC, Inc.
OKIDATA and MICROLINE are registered trademarks of Oki Electric Industry Company, Ltd.
DESCRIPTION
Your new CAPTUS®600 is available in several configurations. Please take a moment to
review your system to determine exactly the configuration you have received. The basic
Thyroid Uptake System consists of the following:
•Thyroid Uptake Stand and NaI (sodium iodide) Detector with Collimator, which
includes an Isolation Transformer attached to stand.
Note: Two versions of detectors are available: Flat Face Probe or Drilled Probe.
•Readout Unit.
•Printer (Various options are available)
•Cable Set for connecting detector, printer, etc.
•Well Counter with drilled Well detector (Optional purchase).
YEAR 2000 COMPLIANCE
The CAPTUS®600 measurement systems contain information technology that accurately
processes date and time data between the years 1999 and 2000. These products, when
used in combination with products purchased from other manufacturers, whose products
properly exchange date and time information, will accurately process the date and time. All
future products are committed to meeting the same Year 2000 Compliance.
MEDICAL EQUIPMENT SAFETY CLASSIFICATION
•CLASS I EQUIPMENT energized from an external power source.
•TYPE B EQUIPMENT with no applied parts to the patient.
•Ordinary EQUIPMENT without protection against the ingress of water.
November 07 PREFACE p-1

CAPINTEC, INC. CAPTUS®600
p-2 PREFACE November 07
•Suitable for CONTINUOUS OPERATION.
•NOT suitable for use in an OXYGEN or a FLAMMABLE ENVIRONMENT.
ELECTROMAGNETIC INTERFERENCE POTENTIAL
This equipment complies fully with interference immunity requirements of the standard EN
60601-1-2 “Medical electrical equipment: Part 1: General safety requirements; 2.
Supplementary standard: Electromagnetic compatibility”.
This equipment generates radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to nearby devices.
However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference, the user is encouraged to try to correct the
interference by one of the following measures:
•Increase the separation between the equipment and the affected device.
•Plug the unit into an outlet on a circuit different from that which the affected device is
connected.
If this fails to correct the problem, please contact Capintec’s Authorized Service Center.
INSTALLATION
Your system has been shipped from the factory pre-configured and fully tested. Installation
consists of placing the components on the stand and securing them in place. Detailed
installation and assembly instructions are provided in the CAPTUS®600 Unpacking and
Assembly Instructions.
IMPORTANT SAFETY INFORMATION
The CAPTUS®600 measurement system has been carefully designed to give you years of
safe and reliable performance. As with all electrical equipment, however, there are basic
precautions you must observe to avoid injuring yourself, the patient or damaging the
equipment.
•Follow the unpacking and assembly instructions document, and read this manual
carefully before using this equipment. Be sure to save all provided documents for
future reference.
•Understand all warning and caution labels as explained in CHAPTER 1: SAFETY
chapter before operating this equipment.

CAPINTEC, INC. CAPTUS®600
CHAPTER 1
SAFETY
These warnings and instructions for use form an integral part of the CAPTUS®600 and must
therefore be kept available for consultation at all times. Precise compliance with the warnings
and instructions is an essential condition for normal use, correct application and thus safety
of patients and user.
SYMBOL DEFINITIONS
Dangerous Voltage Present
Protective Earth Ground
Attention: consult owner’s manual for further information
Fuse
AC Voltage
“ON” (power)
“OFF” (power)
0413 CE Mark
Waste in Electrical and Electronic Equipment (WEEE) –
This symbol indicates that the waste of electrical and
electronic equipment must not be disposed as unsorted
municipal waste and must be collected separately.
Environmentally Friendly Use Period (EFUP) – 20 years
from the date of manufacture – Toxic or hazardous
substances or elements contained in the unit will not leak
or mutate under normal operating conditions resulting in
any environmental pollution, bodily injury or damage to
assets.
November 07 SAFETY 1-1

CAPINTEC, INC. CAPTUS®600
WARNING AND CAUTION LABELS
Located in the center section on both sides of the spring-arm of the floor stand is a warning
label, (Figure 1-1), concerning the hazard of removing the collimator when the spring-arm is
in the wrong position. When the weight of the collimator is removed from the spring-arm, the
arm can forcefully spring upward and cause serious bodily injury to the operator. The spring-
arm must be fully extended in the upright position to prevent this hazardous condition.
Figure 1-1
Located in the front section on both sides of the spring-arm of the floor stand is a warning
label, (Figure 1-2), concerning the hazard of improperly installing the collimator onto the
spring-arm. If the collimator support pin is not fully inserted and latched into the receptacle
block of spring-arm, the collimator may detach from the spring-arm and fall onto the patient.
In addition, the arm can, again, forcefully swing upward and cause serious bodily injury to the
operator. The collimator is safely installed on the spring-arm when the red warning ring is
NOT visible between the collimator and the receptacle block. Refer to the UNPACKING AND
ASSEMBLY INSTRUCTIONS for illustrations.
Figure 1-2
Located on the top of the Isolation Transformer is a label, (Figure 1-3), providing the system
power requirements, the voltage and serial number of the system and the replacement fuse
values for both power line voltages.
Please reference CHAPTER 20: CLEANING AND MAINTENANCE for instructions on how to
change the fuses on the CAPTUS®600 Isolation Transformer.
A fire hazard may exist if the wrong size of fuse is installed.
1-2 SAFETY November 07

CAPINTEC, INC. CAPTUS®600
Figure 1-3
Located on the left side of the Isolation Transformer is a label (Figure 1-4), with two caution
statements pertaining to the main power cord of the Isolation Transformer. The statement,
READ INSTRUCTIONS CAREFULLY BEFORE CONNECTING TO OUTLET, directs the
user to the owner’s manual before connecting the power. The second statement alerts the
user to select an electrical power receptacle that is designated as HOSPITAL ONLY or
HOSPITAL GRADE, to insure reliable grounding is provided to the CAPTUS®system for
electrical safety requirements.
Figure 1-4
Located on the back panel of the CAPTUS®600 Readout Unit is the following label (Figure 1-
5), stating the proper fuse ratings. A fire hazard may exist if the wrong size of fuse is
installed.
November 07 SAFETY 1-3

CAPINTEC, INC. CAPTUS®600
Figure 1-5
Located on the connector guard on the back panel of the CAPTUS®600 Readout Unit is a
label (Figure 1-6), with the caution statement pertaining to safety of the operator. This
statement, DO NOT REMOVE CONNECTOR GUARD, is necessary because of the
presence of high voltage on the connectors and on the inside of the Instrument.
Figure 1-6
This “Consult Accompanying Document” symbol (Figure 1-7) refers to the CHAPTER 3:
GENERAL OPERATION OVERVIEW, for instructions on how to use the Probe Offset, Well
Offset and Contrast adjustments on the CAPTUS®600 Readout Unit.
Figure 1-7
This “Consult Accompanying Document” symbol (Figure 1-8) refers to the Chapter 3:
GENERAL OPERATION OVERVIEW chapter in this manual, for instructions on how to use
Probe and Well High Voltage adjustments on the CAPTUS®600 Readout Unit.
Figure 1-8
1-4 SAFETY November 07

CAPINTEC, INC. CAPTUS®600
CAUTIONS AND NOTES CAUTION
Only qualified/trained personnel should operate or service this unit.
CAUTION
If the equipment is used in a manner not specified in this manual, the protection provided by
the equipment may be impaired.
CAUTION
High voltage is present inside the Detectors (up to 1000 Volts). Due to the presence of these
high voltages, opening the covers with the system plugged in may be hazardous. Refer all
servicing to qualified personnel.
CAUTION
No internal adjustments inside the Detector or Readout may be performed by the user within
the conditions of the warranty, except for changing the fuse. Due to the presence of high
voltages, opening the cover with the system plugged in may be hazardous. Refer all
servicing to qualified personnel.
CAUTION
Never use the Well Detector without the well liner in place. Liners are inexpensive and easy
to replace. A contaminated Detector is a very costly mistake. If unit becomes contaminated
remove liner and clean unit as stated in CHAPTER 20: CLEANING AND MAINTENANCE
before operating.
CAUTION
Care must be exercised when moving the system or when maintenance is performed. The
system must not be inclined more than a 15-degree angle or tip over may occur.
CAUTION
Verify that the Collimator is correctly installed as stated in the Unpacking and Assembly
Instructions. Failure to follow these instructions could result in injury or product malfunction.
CAUTION
This equipment generates radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful Electromagnetic Interference (EMI) to
nearby devices. However, there is no guarantee that interference will not occur in a particular
installation.
CAUTION
If any printer other than the model supplied by Capintec is used, the safety of the unit may be
compromised or Electromagnetic Interference (EMI) may be introduced into other devices
located in the same general area as the CAPTUS®600 may become susceptible to EMI.
CAUTION
The unit contains lead. The unit should be disposed of in accordance with local and national
regulations.
November 07 SAFETY 1-5

CAPINTEC, INC. CAPTUS®600
CAUTION
The unit contains a Lithium Battery. This should be disposed of in accordance with local and
national regulations.
CAUTION
The user should always verify the validity of any measurement or test result in order to
minimize measurement errors.
NOTE
It is recommended that periodic (every five years) re-calibration of the unit be performed by
Capintec’s Authorized Service Center (reference CHAPTER 20: CLEANING AND
MAINTENANCE) to guarantee that the instrument's high reliability is maintained.
GENERAL SAFETY TIPS
•Unplug the equipment before cleaning it. Use only a damp cloth; do not use solvents
or aerosol cleaners.
•To protect the equipment from overheating, make sure none of the openings (vents)
on the Readout Unit or the Printer are blocked. Do not use the equipment directly in
front of a radiator or heat register.
•The print head, in the printer, can get quite hot when it’s been printing for a length of
time. Do not touch the print head until it has had a chance to cool.
•Do not use the equipment near water, or spill liquids of any kind into the equipment.
•Be sure that your power source matches the rating listed on the Isolation
Transformer.
•The CAPTUS®600 power cord has a grounded, 3-prong plug as a safety feature, and
it will only fit into a grounded outlet. Do not use an adapter to defeat the grounding.
•To avoid damaging the power cord, do not place anything on it or place it where it will
be walked on. If the cord becomes damaged, replace it immediately.
•Aside from the routine maintenance described in this manual, do not try to service
this equipment yourself. Do not make any adjustments other than those outlined in
this manual, as you may in-validate the calibration or cause damage requiring
extensive repair work. Refer servicing to qualified service personnel.
•Do not subject the Detectors to rough handling, or rapid temperature changes, as the
NaI Scintillation Crystal can be damaged. Do not use the equipment directly in front of
a radiator, heat register, or air conditioning duct.
1-6 SAFETY November 07

CAPINTEC, INC. CAPTUS®600
CHAPTER 2
FUNCTIONAL & TECHNICAL
DESCRIPTION
GENERAL
The CAPTUS®600 Thyroid Uptake / Well System is intended to be used by trained Nuclear
Medicine Technologists for performing Thyroid Uptake, Thyroid Bioassay, Wipe Test, Blood
Volume Test, Shillings Test and RBC Survival Test.
The CAPTUS®600 also performs Pulse Height Analysis, isotope identification, radioisotope
decay corrections, efficiency calculations, and maintains appropriate storage and retrieval
files for reference data, and patient demographic data.
Your new CAPTUS®600 is available in several configurations. Please take a moment to
review your system to determine exactly the configuration you have received. The basic
thyroid uptake system consists of the following:
•Thyroid Uptake Stand and NaI (sodium iodide) Detector with Collimator, which
includes an isolation transformer, attached to stand.
Note: Two types of probe are available: Flat Face Probe or Drilled Probe.
•Readout Unit.
•Printer.
•Cable Set for connecting detector, printer, etc.
•Well Counter with drilled well detector (Optional purchase).
TECHNICAL DESCRIPTION
The CAPTUS®600 contains a 256 channel microcontroller based pulse height analyzer.
Lower level discriminators are selectable and the firmware permits predefined or selectable
ROI’s (maximum width of 256 channels). The LCD displays alphanumeric data, live spectra,
and graphics. Corrections are automatically calculated for dead time loss. Real time and live
time may be displayed. Coarse high voltage adjustments are made from analog controls at
the back of the console. Fine high voltage adjustments and all other parameters are
controlled through the keypad.
November 07 FUNCTIONAL & TECHNICAL DESCRIPTION 2-1

CAPINTEC, INC. CAPTUS®600
Detector
Your CAPTUS®600 system is equipped with a Probe (and possibly a Well), which consists of
a NaI (Tl) (thallium-activated sodium iodide) crystal and photomultiplier tube detector
assemblies. When a gamma or x-ray from a radioactive source reaches the NaI crystal, it
collides with the electrons in the crystal matrix and transfers energy. The excited electron
releases this energy as visible light in direct proportion to its absorbed energy. For this
reason, NaI detectors are classified as scintillation detectors. These sequential emissions of
light for each incident ray generated from the collisions appear to the measuring system as a
single light pulse. The intensity of the pulse is directly proportional to the energy of the
incident gamma or x-ray.
The light is eventually directed toward the photomultiplier tube. The photomultiplier tube,
which is optically connected to the sodium iodide crystal, converts the visible light pulses into
electrical energy and amplifies the signal through a series of dynodes. The resultant
electrical signal is directly proportional to the intensity of the incident light pulse. Therefore,
the pulse height generated by the system is directly proportional to the energy of the incident
gamma or x-ray detected by the system.
Multichannel Analyzer
A multichannel analyzer (MCA) records the number and height of the electrical pulses
generated by the system, divides the pulses into groups based upon pulse height, and
provides a visible picture or spectrum of the pulses as well as count information. The
CAPTUS®600 divides the pulses into 4 energy bands (200, 400, 800 and 1600 keV)
consisting of 256 channels each. Since pulse height is directly related to the energy of the
incident photon, the spectrum displays the pulse heights as a graph of pulses or counts on
the vertical scale and energy on the horizontal scale. By matching the displayed spectrum to
the decay scheme of an isotope, positive identification of the radioactive source can be
achieved.
TECHNICAL SPECIFICATIONS
Thyroid Uptake Stand
Rugged floor stand with heavy duty locking casters and an articulating arm for the positioning
of detector/collimator. Individual shelves provided for the CAPTUS®600 readout unit and the
printer. Mounting for the isolation transformer and the optional well detector is provided on
the base of the floor stand.
Dimensions: Height 175.3 cm (69")
Width 71.1 cm (28")
Depth 78.7 cm (31")
Weight 181.8 kg (400 lb.)
(with optional well) 213.6 kg (470 lb.)
2-2 FUNCTIONAL & TECHNICAL DESCRIPTION November 07

CAPINTEC, INC. CAPTUS®600
System Power Requirements
CAUTION
If the input voltage to the following items is not within the stated limits,
the unit may not function correctly or may be damaged.
Line Voltage: 115 or 230 VAC depending on model ordered
Line Frequency: 60 or 50 Hz
Current: 0.7 Ampere @ 115V
0.4 Ampere @ 230V
Fuse: For 115V use T 3.15AH, 250V, (Time Delay)
For 230V use T 2.5AH, 250V, (Time Delay)
Line Filter: Line filter is provided internally.
Use of a filtered line is recommended if excessive line noise is anticipated.
Environment Requirements
Operational
The instrument should be located where the level of the background radiation is as
low and as constant as possible.
The instrument should be located where the temperature is stable within a range of
+50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing to warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of
27–31 inches of mercury (91–105 kilopascals).
Storage
The instrument should be stored where the temperature is stable and the range is
from +39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90%
non-condensing to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of
15–33 inches of mercury (51–112 kilopascals).
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
November 07 FUNCTIONAL & TECHNICAL DESCRIPTION 2-3

CAPINTEC, INC. CAPTUS®600
Isolation Transformer
An isolation transformer is provided with the CAPTUS®600 system to ensure that the
leakage current meets EN60601-1-1 Medical Equipment Safety Standard. A detachable
4.6m (15 ft) power cord, with a hospital grade plug, is provided to connect the isolation
transformer to the power source. Dual power line fuses are provided at the power entry
module for quick replacement.
Dimensions: Height 13.3 cm (5.25")
Width 17.5 cm (6.88")
Depth 10.9 cm (4.31")
Printer
Okidata 184 Turbo
80 column dot matrix printing mode.
Graphics and near letter quality text printing
Power Requirements:115V 0.4A 50/60Hz
Dimensions: Height 8.1 cm (3.2")
Width 36.0 cm (14.2")
Depth 27.4 cm (10.8")
Weight 4.5 kg (10 lb.)
Readout Unit
256 channel MCA with LCD display
Power Requirements:30 Watts
115V 0.3A 50/60Hz
Power Conditioning: Power Line Filter
Transient Voltage Suppressers
Dual Power Line Fuses
Dimensions: Height 16.5 cm (6.5")
Width 26.7 cm (10.5")
Depth 35.9 cm (14.125")
Weight 4.5 kg (10 lb.)
Nuclear Data
9 Pre-set nuclide data keys
5 User assignable nuclide data keys
Over 60 nuclides in memory (major gamma-ray energies and intensities, half-life)
2-4 FUNCTIONAL & TECHNICAL DESCRIPTION November 07
Table of contents
Other Capintec Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Otto Bock
Otto Bock 13E135 Instructions for use

Bien-Air Dental
Bien-Air Dental MC2 Instructions for use

Mercoframes Optical Corp
Mercoframes Optical Corp Volk iNview quick start guide

BLACK & BLACK SURGICAL
BLACK & BLACK SURGICAL B89055 Service manual

AVIDICARE
AVIDICARE OPRAGON Technical manual

GlobalMIND
GlobalMIND VenenWalker basic instruction manual