Lode Excalibur Sport 2019 User manual

m965910/20 - Manual Excalibur Sport 2019
Sports Medicine
Lode BV
Zernikepark 16
NL 9747 AN
Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
www.lode.nl
M965910 / VERSION: 1.5 - 2021-12-22

44
44
88
99
9
1010
10
10
11
11
1212
12
1313
14
14
15
1515
15
16
16
1616
1717
1818
1919
1919
2020
Table of contents
1 - Intended Use1 - Intended Use
2 - Precautions2 - Precautions
3 - Contra Indication3 - Contra Indication
4 - Validity4 - Validity
4.1 - Description of parts
5 - Installation5 - Installation
5.1 - Install the device
5.2 - Free area
5.3 - Connecting cables
5.4 - Moving the ergometer
6 - Preparation6 - Preparation
6.1 - Saddle and handlebar adjustment in Control Unit
7 - Exchange parts of the Excalibur7 - Exchange parts of the Excalibur
7.1 - Exchange saddle
7.2 - Exchange handle bar
7.3 - Exchange pedals
8 - Terminal - external control8 - Terminal - external control
8.1 - External control with ECG or Ergospirometry system
8.2 - External control with Lode Ergometry Manager
8.3 - Display
9 - Excalibur PFM (965920)9 - Excalibur PFM (965920)
10 - List of symbols used10 - List of symbols used
11 - Maintenance11 - Maintenance
12 - Cleaning12 - Cleaning
13 - Calibration13 - Calibration
14 - Accessories14 - Accessories

1 - Intended Use
This device is intended to be used as a stress test device in a medical environment, primarily to create
reproducible stress tests and measuring the effect. With this device typically workload (watt) is
imposed.
The intended use is further enhanced to monitor:
- the diastolic and systolic blood pressure, and/or
- oxygen saturation of the blood, and/or
- pulse rate of the test subject,
when the measurement is executed with blood pressure and SpO2 measurement.
This device is designed for both manual control and external control by means of ergometry software,
external ECG and pulmonary equipment.
The device has to be operated under the supervision of well-trained medical specialists1 in the field of
use.
All users should read the IFU (available in corresponding language) and should have knowledge and/or
experience of exercise protocols, contra indications for exercise test, knowledge of stop criteria and
cleaning of the device.
The maintenance and service of the medical device has to be done by a well-trained service engineer
who is working for a Lode distributor and is specifically trained regarding the MD by the International
Service Coordinator of Lode.
1These specialists hold a bachelor and/or master degree in the medical field, they can be sport
physicians, exercise physiologists, or (specialized) physical therapists.
2 - Precautions
Read all warnings posted on the device.
Inspect the device for worn or loose components prior to
use. Tighten / replace any loose or worn components prior to
use.
4/23M965910 / VERSION: 1.5 - 2021-12-22

The operator should not touch accessible parts and the
patient simultaneously.
Read this manual before using the device and follow it
carefully.
The operator should instruct the test subject prior to
performing an exercise protocol. If, at any time during
exercise, the test subject feels faint, dizzy, or experiences
pain, stop the test and he or she should be consulted by the
physician.
Replacement of parts can only be done by the manufacturer
or designated service personnel.
Set up and operate the device on a solid level surface.
The equipment has a safety earth (ground) connection and
must be connected to a (grounded) wall socket with
protective earth to avoid the risk of electric shock. The
functional earth connection is for potential equalisation only.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
5/23M965910 / VERSION: 1.5 - 2021-12-22

Be sure that no cables are in the area of the crank shaft.
Check for completeness of all warning labels on the
equipment.
A regular check of the fixation and guiding of the power cord
is needed.
The operator should clean the device (handgrips, cushions)
after each exercise test.
Do not use when damaged.
Heart rate monitoring systems, BP measurements or SpO2
monitoring may be inaccurate. Over exercise may result in
serious injury or death. If the test person feels faint stop
exercising immediately.
The installation should only be done by authorized persons
who follow the installation instructions.
Do not spill any liquid over the device.
Only use the device at normal operating temperature and air
pressure (see specification).
6/23M965910 / VERSION: 1.5 - 2021-12-22

Portable RF communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part
of the ergometer, including cables specified by the
manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Any serious incident that has occurred in relation to this
device should be reported to the manufacturer and the
competent authority of your Member State.
The ergometer is to be operated by classified personnel only.
The medical device is heavy: take care for clamping of foot or
hands when moving or tilting the device.
When replacing cranks, pedals or handle bar be sure to
mount the parts in the correct position and tighten the
connection firmly.
7/23M965910 / VERSION: 1.5 - 2021-12-22

3 - Contra Indication
Although the ergometer does not have any contra indication other than the maximum weight and
length of the test subject, in respect of exercise testing the following contra indication should taken
into account:
-Acute myocardial infarction (within 2 days)
-Unstable angina not previously stabilized by medical therapy
-Uncontrolled cardiac arrhythmias causing symptoms or hemodynamic compromise
-Symptomatic severe aortic stenosis
-Uncontrolled symptomatic heart failure
-Acute pulmonary embolus or pulmonary infarction
-Acute myocarditis or pericarditis
-Acute aortic dissection
The device is to be operated by classified personnel only. As stated in the intended use, the device is
intended to be used in a medical environment. During the intended use the test subject will deliver
energy. Application of the wrong dosis of energy could lead to permanent damage of the test subject
health. Only use this device under supervision of a medical physician.
8/23M965910 / VERSION: 1.5 - 2021-12-22

4 - Validity
This manual covers all Excalibur versions starting with the following serial number:
965910 Excalibur Sport: S/N 20190001 -
965920 Excalibur Sport PFM: S/N 20190001 -
Date of first issue: 2019-07-01
4.1 - Description of parts
1. Display (APPLIED PART)
2. Handle bar (APPLIED PART)
3. Saddle (APPLIED PART)
4. Pedal (APPLIED PART)
5. Symbol plate
6. Network Module
7. Type plate
8. Communication Interface
Module
9. Potential Equalization
Conductor
10. Service port (not to use)
11. On/Off switch
9/23M965910 / VERSION: 1.5 - 2021-12-22

Be careful for toes and back when moving the ergometer, the device is heavy.
5 - Installation
5.1 - Install the device
- Unpack the device while leaving the transport straps in place.
- Place the device in the location where it will be used.
- Remove transport straps and remaining packaging materials.
- Connect the power cord to the line output at the required voltage (see type plate).
- Remove all objects from the device.
- Switch on the unit with the on/off button and wait until the self-test is carried out. The product is
ready for use after you have heard a series of “beeps”.
- Use the on/off button in order to switch the ergometer off.
5.2 - Free area
The free area around the
ergometer shall be not less than
0,6 m greater than the training
area in the directions from
which the equipment is
accessed. The free area must
also include the area for
emergency dismount. Where
equipment is positioned
adjacent to each other the
value of the free area may be
shared.
10/23M965910 / VERSION: 1.5 - 2021-12-22

5.3 - Connecting cables
The connection for the power
cord and optional connectors
for external devices are placed
at the front side of the
ergometer, so they are easy to
reach.
The operator is instructed not
to connect any device to the
optional connectors unless it
possesses an approval to
60601-1 or 60950 issued by an
independent third party.
5.4 - Moving the ergometer
The Excalibur has 2 wheels in the front. Moving the Excalibur can be done by lifting the back end using
the transport wheel and ride the Excalibur to its location. Mind your back when tilting the ergometer, it
is heavy.
11/23M965910 / VERSION: 1.5 - 2021-12-22

Due to safety reasons the saddle and handlebar adjustment cannotcannot be done during a
test or training. Be sure the settings are convenient before the test/training starts.
6 - Preparation
The easiest and most convenient way for a test subject to get on the Excalibur Sport is putting the
saddle and the handle bar in the estimated position first. The saddle and handlebar adjustments are
electrical and can be done in the control unit of the Excalibur or in the Lode Ergometry Manager
software (see manual LEM). When the test subject sits on the Excalibur and slowly pedals without a
load the saddle and handlebar settings can be fine tuned.
In the control unit changes can be made in the test view before the test start.
6.1 - Saddle and handlebar adjustment in Control Unit
Saddle and handlebar adjustment in Control Unit
Before a test is started, in
manual or terminal mode, in
the left side of the screen an
icon (red box) is visible for the
saddle and handlebar
adjustments. If you press on
the icon the screen "Ergometer
Adjustments" will open.
Changing the saddle and
handlebar settings is only
possible with a load of zero
watt, so due to safety reasons
the icon will disappear when
the test is started. No further changes can be made during the test.
12/23M965910 / VERSION: 1.5 - 2021-12-22

Vertical saddle adjustment
Horizontal saddle adjustment
Vertical handle bar adjustment
Horizontal handle bar adjustment
"Ergometer Adjustment" is used
to change saddle and
handlebar position. The button
in the middle is used to select in
which direction (horizontal or
vertical) the saddle or
handlebar needs to be
changed. Press on the button
until the correct direction is
selected. Use the arrows to
move the saddle or handlebar
to the desired position.
If all settings are done press the exit button.
For a more detailed explanation you can read the display manual.
7 - Exchange parts of the Excalibur
In order to test the athletes in the best way, some parts of the Excalibur can be exchanged to come to
the best configuration. The exchangeable parts are saddle, handle bar and pedals. For all three Allen
keys are used, #6 for saddle and pedal and #4 for the handle bar (delivered with the Excalibur).
13/23M965910 / VERSION: 1.5 - 2021-12-22

7.1 - Exchange saddle
Use Allen key #6 to loosen the
screw. When possible remove
the saddle and place the
desired saddle. Tighten the
screw again.
7.2 - Exchange handle bar
Use Allen key #4 to loosen the
four screws and remove the
small frontpanel. Remove the
handle bar and replace the
desired handle bar. Tighten all
four screws again, be aware
that the angle of the handle bar
is as wished.
Note that it's very important
that the handle bar is fixated
properly, always check if this is
the case after switching the
handle bar.
14/23M965910 / VERSION: 1.5 - 2021-12-22

Always use the correct shoes and pedals in respect to the foot size.
7.3 - Exchange pedals
Use Allen key #6 to loosen the
screw at the end of the crank (if
the crank is at twelve o'clock it
is easy to reach the screw from
the other side). Loosen the
pedal and exchange it for the
desired pedal.
8 - Terminal - external control
8.1 - External control with ECG or Ergospirometry system
When your device is fitted with a communication card, you can connect to many external devices:
1. Connect the device with the supplied USB cable to the external device (optionally RS232 cable can
used (930911)). You can control the device with e.g. an exercise ECG device or Metabolic Cart.
2. Download the interfacing instructions from the internet at www.lode.nl after you have registered
your device.
3. Select “Terminal mode” on the control unit to allow external commands (see for detailed instructions
the user manual of your external device).
15/23M965910 / VERSION: 1.5 - 2021-12-22

The RS232 connection and the USB connection on the Excalibur can only be used for
external CPET devices. Never use those when the Excalibur should be connected with
LEM.
8.2 - External control with Lode Ergometry Manager
Your device is enabled to communicate with the Lode Ergometry Manager (LEM). LEM gives the
opportunity to work with a database, create protocols and to perform a variety of tests:
1. Use the network module connection cable (930930, UTP cable) to connect your device to the pc, plug
the RJ45 connector in the ergometer and the USB in an USB port of your pc.
2. Start the LEM Software on your PC.
3. Select on the display of the ergometer “Terminal” mode to allow external commands. The terminal
mode is automatically recognized by LEM (see for detailed instructions the user manual of LEM).
Note: Lode Ergometry Manager (LEM) is a PC software program having it's own medical CE mark.
8.3 - Display
When your device is executed with a single display in display mode, it will only display data. It is not
able to steer the ergometer. It is intended to be used in terminal mode. You can change the settings
for communication and language of the device.
To be able to control the ergometer you will need the optional 945834 Control Unit with touch screen
7" for ergometer or 945835 Programmable Control Unit for ergometer.
For the functioning of the display and/or control units, we refer to the manuals M945810, M945834
and M945835 which can be found on the supplied CD with additional information. The most recent
version in many other languages can always be downloaded from the Lode website.
9 - Excalibur PFM (965920)
In case your Excalibur is an Excalibur with Pedal Force Measurement (Excalicur PFM), you have to use
the Lode Ergometry Manager (LEM) software to collect and analyse the PFM data. In the manual
Manual LEM 10 PFM all information about the use of the PFM is described.
16/23M965910 / VERSION: 1.5 - 2021-12-22

General warning
Read manufacturer’s guide, advices and instructions and manual
Alternating Current
IEC 60417-5032
The equipment has a safety earth (ground) connection and must be connected to an
earthed (grounded) wall socket.
Potential Equalization Conductor
Caution
"ON" position of the on/off switch: IEC 60417-5007 “ON” (power)
"OFF" position of on/off switch: IEC 60417-5008 “OFF” (power)
Type B applied part
USB
Remote Control
10 - List of symbols used
17/23M965910 / VERSION: 1.5 - 2021-12-22

RS232
Network (Proprietary Lode)
CE and notified body
FCC 18 RF ISM tested
Symbol for collection, treatment, recycling and disposal of waste electrical and
electronic equipment (WEEE) as set out in Directive 2002/96/EC of 27 January 2003 of
the European Parliament and of the Council on waste electrical an electronic
equipment are necessary to reduce the waste management problems linked to the
heavy metals concerned and the flame retardants concerned
11 - Maintenance
Maintenance should be carried out on a regular and planned basis. We recommend to check the unit
annually. This may be done by your local dealer. It is also recommended that a record of the service
history is kept for all activities relating to service and maintenance.
Maintenance and all repairs should only be carried out by an authorized agency. The manufacturer will
not be held responsible for the results of maintenance or repairs by unauthorized persons.
The check up and/or technical maintenance must be carried out conform the procedure described in
the service manual of the unit.
Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to
warranty.
Lode will make available on request circuit diagrams, component part lists, descriptions, calibration
instructions, or other information that will assist authorized service personnel to repair those parts of
the ergometer that are designated by Lode as repairable by authorized service personnel.
18/23M965910 / VERSION: 1.5 - 2021-12-22

12 - Cleaning
Before cleaning and other maintenance of any part of the device, first switch off the device and
disconnect it from the mains. The surfaces can be cleaned with a damp cloth and a liquid (non-
abrasive) cleaning product.
General Cleaning
Wipe the device surface down with a cloth moistened with soap water or a disinfectant. The cloth
should not be dripping wet; do not allow liquids to enter the device.
Cleaning the Saddle
Clean the saddle with a soft and dry or moist cloth. Disinfectants used should not contain any alcohol.
Cleaning the Upholstery (e.g. couch ergometer)
Wipe the upholstery down with a soft cloth moistened with soap water. The cloth should only be moist
and not dripping wet. If the cleaning agents and disinfectants used are caustic or contain alcohol, they
may damage and/or discolor the upholstery.
If you are not certain about the discolouring effect of a cleaning agent, you may try a little on a part of
the product that is not visible during normal use.
Lode cannot advise a specific cleaning agent, since local recipes may differ.
13 - Calibration
The ergometer should be calibrated once a year. In case any damage is observed of the ergometer
Lode B.V. or his representative should be informed in order to execute the necessary repair(s). Service
of the ergometer is restricted to factory-trained personnel only.
19/23M965910 / VERSION: 1.5 - 2021-12-22

14 -
Accessories
Add program function
to 7" touch screen for
ergometer
Partnumber: U945835Partnumber: U945835
Easily programmable
Add Program function to
Control Unit
Colour Display 3.5" -
2nd screen
Partnumber: 945819Partnumber: 945819
Multifunctionality
Colour Display 3.5"
External Blood
Pressure monitor for
bicycle ergometer
Partnumber: 945500Partnumber: 945500
Reliable BPM during
exercise stress testing
External BPM for bicycle
ergometer
SpO2 for control unit
with touch panel (extra
long ...
Partnumber: 945822Partnumber: 945822
Oxygen saturation
SpO2 for control unit with
touch panel
Bluetooth Smart heart
rate
Partnumber: 945833Partnumber: 945833
Heartrate available
within an extreme wide
range
Bluetooth Smart heart rate
Ambient sensor pack
Partnumber: 945856Partnumber: 945856
Check environmental
conditions during test
Ambient sensor pack for
Excalibur
0-Watt start-up system
Partnumber: 960805Partnumber: 960805
Lowest possible startup
power
Motor drive operated 0-Watt
start-up
Adjustable sports
cranks incl. pediatric
range
Partnumber: 925818Partnumber: 925818
Optimal force
application, allows for
different leg/arm
length
Adjustable sports cranks incl.
pediatric range
Saddle for children
Partnumber: 401066Partnumber: 401066
Versatile ergometry
Saddle for children
Remote Control for
ergometers
Partnumber: 945832Partnumber: 945832
Ease of use
Remote Control for
ergometers
20/23M965910 / VERSION: 1.5 - 2021-12-22
This manual suits for next models
2
Table of contents
Other Lode Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Soredex
Soredex Cranex Novus e Service manual

ultraMEDIC
ultraMEDIC ultraBASKET STRETCHER Operating manual and service instructions

Huntleigh
Huntleigh Hydroven 3 Instructions for use

Dr. Mach
Dr. Mach LED 6MC user manual

B.A. International
B.A. International Ultimate BA101 Operation manual

Thuasne
Thuasne LombaStab Dorso manual