Lumex DSC260 User manual

DSC250 SCALE
for LF1050 / LF1090 Patient Lift
DSC260 SCALE
for LF1030 / LF1031 / LF1031FP /
LF1040 / LF1041 Patient Lift
USER MANUAL
Read these instructions before using the Lumex Lift Scale.
Save these instructions for future reference.
DSC250-INS-LAB-RevC15

2
CONTENTS
INTRODUCTION...................................................................................................3
IMPORTANT SAFETY INFORMATION.........................................................3
INTENDED USE............................................................................................4
SCALE INSTALLATION ON PATIENT LIFT ..........................................................5
DSC250 SCALE COMPONENTS .........................................................................6
DSC250 SCALE INSTALLATION ON LF1050 / LF1090 PATIENT LIFT...............7
DSC260 SCALE COMPONENTS .........................................................................8
DSC260 SCALE INSTALLATION ON LF1030 / LF1031 / LF1031FP / LF1040 /
LF1041 PATIENT LIFT ..........................................................................................9
LIFT SCALE
OPERATION INSTRUCTIONS ............................................................10
KEYBOARD ILLUSTRATION......................................................................10
DISPLAY ILLUSTRATION...........................................................................10
INSTRUCTIONS FOR USE ........................................................................10
LIFT SCALE MAINTENANCE.............................................................................12
CLEANING..................................................................................................12
BATTERY REPLACEMENT........................................................................12
LIFT SCALE SPECIFICATIONS..........................................................................13
LIMITED WARRANTY.........................................................................................14

3
INTRODUCTION
Important Safety Information
WARNING: Important! Read and understand
these instructions before assembling or using
the Lumex DSC250 / DSC260 Lift Scale. If you
do not understand any part of these warnings or
instructions, contact a healthcare professional
for direction in the use of this product. If the Lift
Scale is not properly assembled, personal injury
and damage to the Lift Scale could result.
WARNING: DO NOT use this product
without proper instruction from a healthcare
professional.
WARNING: If components are damaged or
missing, contact your GF distributor immediately.
DO NOT use substitute parts. Use only Lumex
replacement parts. Non-Lumex replacement parts
could cause personal injury and damage to the
Lift Scale.

4
WARNING: This product has a maximum weight
capacity of 600 lb (300 kg). If patient weight
exceeds maximum weight capacity, the scale will
designate overload by displaying “-oL-”. DO NOT
exceed the maximum weight capacity of the lift
being used for measurement.
WARNING: California Proposition 65 WARNING:
This product contains a chemical known to
the State of California to cause cancer and
reproductive or developmental harm.
WARNING: GF Health Products, Inc. assumes no
responsibility for any damage or injury caused
by improper assembly, installation or use of this
product.
Intended Use
The intended use of the DSC250 / DSC260 Lift Scale is
the measurement of patient weight.

5
SCALE INSTALLATION ON PATIENT LIFT
1.
Use the following instructions (DSC250 or DSC260)
to attach the Lift Scale to the appropriate Patient Lift
with the included mounting hardware. Do not use
any other mounting hardware than that provided.
2. Keep the scale in a cool, dry place. If operation
temperature is over 104°F (40°C), the tolerance
will not be accurate within 0.2 lb (0.1 kg).
Recommended storage temperature is 32°F ~158°F
(0°C ~ 70°C).
3. When the low battery indicator ( ) appears in the
lower right corner of the display, replace the
battery (see MAINTENANCE section).
4. If the scale is not used for an extended period of
time, remove the battery.
5. The scale maximum weight capacity is 600 lb
(300 kg). If patient weight exceeds maximum weight
capacity, the scale will designate overload by
displaying “-oL-”.

6
DSC250 SCALE COMPONENTS
Ensure all components shown below are included in
the DSC250 Scale package.
Item Qty Picture Description
1 1 Scale
2 1 Bushing 1220
3 1 Bushing 1020
4 1 Universal bracket
5 2 Pin
6 2 Key ring
7 2 Washer

7
DSC250 SCALE INSTALLATION ON LF1050 / LF1090
PATIENT LIFT
Install DSC250 Scale on LF1050 / LF1090 Patient Lift as
shown above. Item numbers shown are those listed in
the page 6 table.
WARNING: Ensure scale is securely mounted as
shown above before proceeding.

8
DSC260 SCALE COMPONENTS
Ensure all components shown below are included in
the DSC260 Scale package.
Item Qty Picture Description
1 1 Scale
2 1 Bushing 1220
3 1 Bushing 1020
4 1 Universal bracket
5 1 Pin
6 1 Key ring
7 1 Screw
8 2 Washer
9 2 #10 Washer
10 1 Nut

9
DSC260 SCALE INSTALLATION ON LF1030 / LF1031 /
LF1031FP / LF1040 / LF1041 PATIENT LIFT
Install DSC260 Scale on LF1030 / LF1031 / LF1031FP /
LF1040 / LF1041 Patient Lift as shown above. Item
numbers shown are those listed in the page 8 table.
WARNING: Ensure scale is securely mounted as
shown above before proceeding.

10
LIFT SCALE
OPERATION INSTRUCTIONS
Keyboard Illustration
Display Illustration
Instructions for Use
1. Place sling to be used onto spreader bar without
patient/resident/user.
2. Press the scale faceplate On/Zero button to zero-
out the sling weight (this weight will remain zeroed
out until the Off button is pressed).
3. Press the Kg/lb button to select mode you prefer.

11
4. Carefully lift the patient/resident/user in the sling,
performing the procedure so that all of the weight
is supported by the lift. Stabilize the movement of
the sling after lifting.
5. Weight will display on the scale. Press the
Operate/Hold button to hold the weight displayed.
To reset the scale to zero, press the Operate/Hold
button again.
6. Lower patient/resident/user safely.
7. If you are weighing another individual using the
same sling, perform the lift and their weight will
display. Press Operate/Hold to maintain their
weight on the scale display.
8. Press the Off button when the weighing procedure
is complete.
9. The scale will automatically power off ve minutes
after measurement has been completed.

12
LIFT SCALE MAINTENANCE
Cleaning
As needed, wipe scale gently, using mild soap or a mild
disinfectant.
sNOTICE: DO NOT wash under water pressure
or steam clean. DO NOT use solvents, abrasive
cleaners, or scouring pads to clean scale.
Battery Replacement
When the low battery indicator ( ) appears in the
display’s right corner, replace the battery:
1. Open the scale back battery compartment.
2. Remove battery and replace with a 9V alkaline
battery. Ensure battery polarity is correct.
3. Close the scale back battery compartment.
If the scale is not used for an extended period of time,
remove the battery.

13
LIFT SCALE SPECIFICATIONS
Model Number DSC250 / DSC260
Capacity 600 lb (300 kg)
Resolution 1/3,000 ±1d
Division 0.2 lb (0.1 kg)
Power Source 9V alkaline battery x 1
Temperature Operating 32°F ~ 104°F (0°C ~ 40°C)
Storage 32°F ~ 158°F (0°C ~ 70°C)
Dimensions (W x H x D) 3.54 x 2.83 x 2.17 in. (90 x 72 x 55 mm)
Product Weight 24 oz ± .35 oz / 1.5 lb ± .02 lb (675 g ±10 g)
LCD Display 2.60 x 1.28 in. (66.0 x 32.4 mm)
Display Description Five digit indication
Kg/lb selectable unit mode
Low power alarm battery symbol
S1 “Zero” indication
S4 “Hold” indication
(S2 and S3 are not used)

14
LIMITED WARRANTY
SCOPE OF WARRANTY
GF Health Products, Inc. (“GF”) warrants to the original purchaser only that it will
replace or repair components, at GF’s sole discretion, that are defective in material
or workmanship under normal use and service. All warranties are conditioned upon
the proper use of the products strictly in accordance with good commercial practice
and applicable GF instructions and manuals, including proper use and maintenance.
To the extent that a component is warranted by a third party, GF conveys all of its
rights under that warranty to the original purchaser, to the extent permitted. This
limited warranty shall only apply to defects that are reported to GF’s customer
service team within the applicable warranty period and which, upon examination
by GF or its authorized representative, prove to be a warranty item. This limited
warranty is not transferable. Within the guidelines set forth in this document,
this product is warranted for two (2) years. The applicable warranty period shall
commence from date of shipment to the original customer, unless there is an
expiration date on the component in which case the warranty shall expire on the
earlier of warranty period or the expiration date.
OBTAINING WARRANTY SERVICE
This limited warranty shall only apply to defects that are reported to the Distributor
from whom the Customer purchased the product within the applicable warranty
period. If there is not a Distributor, you must contact GF directly by calling
770-368-4700, sending a fax request to 770-368-2386, or by e-mailing a request to
cs@grahameld.com. Specic directions will be provided by the Customer Service
Representative. Failure to abide by the specic directions will result in denial of the
warranty claim.
EXCLUSIONS
The warranty does not cover and GF shall not be liable for the following:
1) Defects, damage, or other conditions caused, in whole or in part, by misuse,
abuse, negligence, alteration, accident, freight damage, tampering or failure to
seek and obtain repair or replacement in a timely manner;
2) Products which are not installed, used, or properly cleaned and maintained as
required in the ofcial manual for the applicable product;
3) Products considered to be of a non-durable nature including, but not limited to:
casters, lters, fuses, gaskets, lubricants, and charts;
4) Accessories or parts not provided by GF;

15
5) Charges by anyone for adjustments, repairs, replacement parts, installation or
other work performed upon or in connection with such products which are not
expressly authorized in writing, in advance, by GF;
6)
Any labor or shipping charges incurred in the replacement part installation or repair;
7) Costs and expenses of regular maintenance and cleaning; and
8) Representations and warranties made by any person or entity other than GF.
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES
DISCLAIMER
THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL
OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF
THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY
CONFORM TO THE MODEL OR SAMPLE IN ALL RESPECTS. THIS WARRANTY IS LIMITED
TO THE REPAIR OR REPLACEMENT OF THE DEFECTIVE PARTS. GF SHALL NOT BE
LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL,
EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO:
DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER, OR
EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS.
The warranties contained herein contain all the representations and warranties with
respect to the subject matter of this document, and supersede all prior negotiations,
agreements and understandings with respect thereto. The recipient of this document
hereby acknowledges and represents that it has not relied on any representation,
assertion, guarantee, warranty, collateral contract or other assurance, except those
set out in this document.
For additional information on this product or this warranty, please contact a GF
Customer Service Representative.
NOTES:
1) Additional terms and conditions may apply.
2)
Freight claims must be notated on the appropriate shipping documents and must
be made with immediacy. International, federal and state regulations govern
specic requirements for freight claims. Failure to abide by those regulations may
result in a denial of the freight claim. GF will assist you in ling the freight claim.
3)
Claims for any short shipment must be made within three (3) days of the invoice date.

This manual suits for next models
1
Table of contents