LUXO LHH G2 User manual

Luxo LHH G2/LHH G2 CCT
User manual

English
Our products are subject to the Directive
2002/96/EC (Waste Electrical and
Electronic Equipment - WEEE) and should at
the end of their lifespan always be collected
separately and brought to the appropriate
collection point in your community or region.
Norsk
Våre produkter er underlagt direktiv
2002/96/EC (Waste Electrical and
Electronic Equipment - WEEE) og skal etter
endt levetid leveres til mottak for slikt avfall
eller til en forhandler av slikt avfall.
Eesti
Meie toodete kohta kohaldatakse
2002/96/EC WEEE (elektri-ja
elektroonikaseadmetest tekkinud jäätmete)
direktiivi ning nad peavad olema
eluea lõppedes eraldi kokku kogutud
ning toimetatud piirkonna jäätmete
vastuvõtupunkti.
Nederlands
Onze producten zijn onderworpen aan de
Richtlijn 2002/96/EC (Waste Electrical
and Electronic Equipment -WEEE) en moeten
aan het einde van hun levensduur apart
ingezameld en naar het juiste inzamelpunt
gebracht worden in uw gemeente of regio.
Dansk
Vores produkter er underlagt direktiv
2002/96/EC (Waste Electrical and
Electronic Equipment -WEEE) og skal efter
endt levetid leveres til modtagere for sådan
type affald eller til en forhandler af sådan
affald.
Svenska
Våra produkter omfattas av direktivet
2002/96/EC (Waste Electrical and
Electronic Equipment – WEEE) och skall
vid slutet av sin livsängd alltid samlas in
och levereras till återvinningsstation i din
kommun eller region.
Deutsch
Unsere Produkte unterliegen der Richtlinie
2002/96/EG (Waste Electrical and
Electronic Equipment - WEEE) und sollten
am Ende ihrer Lebensdauer immer getrennt
gesammelt und an einem entsprechenden
Sammelpunkt in Ihrer Gemeinde oder
Region entsorgt werden.
Suomi
Tuote kuuluu elektronisten laitteiden
jätteenkäsittelyä (WEEE) koskevan direktiivin
2002/96/EC piiriin ja elinkaaren lopussa
oleva tuote tulee toimittaa asiaankuuluvaan
keräyspisteeseen.
RISK GROUP 2
CAUTION: Possibly hazardous
optical radiation emitted from this
product. Do not stare at operating
lamp. May be harmful to the eye.
2

1050
335
144
Loosening
CCT 3500/4000/4500 K
ON/OFF/DIM
Tightening
LHH G2/LHH G2 CCT
This product contains a spring loaded arm. In order to avoid
damage or possible injury, first mount table bracket or floor stand.
Insert product and remove protective plastic sleeve or rubber band,
and carefully unfold the product.
Use only brackets supplied by Glamox or a Glamox representative.
Glamox cannot guarantee the lifetime and functionality if third part
accessories are used.
Night light 1%
ON/OFF/DIM
Instructions for use in your local language, please visit www.glamox.com
3

LHH G2/LHH G2 CCT
General information
This luminaire is intended for use as a general purpose examination luminaire. If the product is used during treatment there
should be NO contact with the patient.
This product are not designed and approved for home use.
Optimum working distance between the light source and illuminated object or surface, using a standard lamp, is 50 cm.
Installation
This product contains a spring loaded arm. In order to avoid damage or possible injury, first mount table bracket or floor
stand. Insert product and remove protective plastic sleeve or rubber band, and carefully unfold the product.
Place the supply cable so it not will be damaged.
Always use enclosed table bracket, optional floor stand or wall brackets.
For accessories available for this product, please visit our website www.glamox.com
Luxo cannot guarantee the lifetime and functionality if third party accessories are used.
Cleaning
Blow and dry off any dust and dirt.
Using of different cleaning agents for cleaning this product:
Avoid use of alcohol, acids or any kind of organic solvents (Isopropanol etc.)
Only use the weak mix of ordinary soap and water without any perfumes.
Maintenance
Cleaning of product is the only maintenance needed. No requirements on frequency of cleaning other than when the
products get degraded functions on movements or light.
There are no parts requiring special inspection or maintenance on the product.
Transportation
The product should be set in transportation mode when moved. Fold upper arm down in parallel with the lower arm, and
secure them together with a rubber band or strips.
4
Spare parts
S P A 0 26 011 Cable 3x0.75+/3,0m BS1363A White
SPA026706 AC/DC converter 24V class I
SPA026707 LED module LHH LED G2
SPA026708 Cable 3x0.75+/3,0m Schuko White
SPA026709 Cable 3x18 AWG/3,0m NA White
SPA026710 Fuse T1.25AH/250V
Discription to exchange spare parts will be supplied upon request to service personell

LHH G2/LHH G2 CCT
Troubleshooting for qualified electrical fitters and other similarly qualified personnel
• Replaces the fuses
- Remove the lamp from mains supply.
- Loosen the screw holding the bottom lamp cover
- Exchange fuses
- Place bottom cover, and tighten the screw
• Replaces the mains cable
- Instruction for changing cable will be shipped with the spare part
If there should arise other problems please contact your local Glamox representative
5
Classification
LHH G2/ LHH G2 CCT
Medical class: Class I
Permitted leakage current
Protection reliability Must not exceed 0.5 mA
Power supply Resistance in earth circuit – earth connection 0.1
Power consumption 100-240VAC 50/60 Hz
Mode of operation LHH G2 max 60VA
LHH G2 CCT max 50VA
Protection against risk of explosion Not protected
Protection against liquid penetration Not protected
Degree of mobility Mobile table
Type of approval MEDICAL Standard
Safety test in line with EN 60601-1 :2012
EMC test in line with EN 60601-1-2 :2007
Primary fuse IEC 127-5X20mm, 2xT1.25AH 250V
Operating life 50000 Hours (L70B50)
G2 CCT:
Central illuminance 4500K 28500lx at 1m
Central illuminance 4000K 27900lx at 1m
Central illuminance 3500K 25900lx at 1m
G2:
Central illuminance 3700K 35200lx at 1m

Symbol explanations:
Do not throw away. Electronic waste.
The product must be recycled
Self declaration symbol
Approved by Nemko
Attention, read instructions before use
Warning:
• Avoid spilling liquid on the lamp head
• Have minimum 400 Lux light, when reading the labels on the product
• The power cable must only be replaced by a qualified electrical fitter or other similarly qualified professional.
• The switch disengages the low voltage to the lamp only. To disconnect the mains supply, unplug the plug from the
socket outlet.
• To avoid risk of electric shock, this equipment must only be connected to a supply main with protective earth.
• This ME EQUIPMENT could cause risk of reciprocal interference during specific investigations or treatments.
• Position the product so that it’s easy to disconnect the device when the product is used as isolation means.
• To terminate operation, turn the product off with the switch. For storage, unplug the device and fold the arm together.
• Any repairs of the arm must be performed by Glamox or Glamox approved personnel. The arm contains high tensions
springs.
• Do not modify this equipment without authorization of the manufacturer.
• This product complies to IEC/EN 60601-1-2, electromagnetic compatibility for medical devices. Electromagnetic or
other interference between this equipment and other devices should not occur. In the event of interference,
please move the equipments apart from each other.
• The switch disengages the low voltage to the lamp only. To disconnect the mains supply, unplug the plug from the
socket outlet.
• When abnormal use Glamox declines all responsibilities regarding the product
• Users of this form of lighting should be aware that the manufacturer of the product can only accept liability for the
lighting when maintenance, repairs or changes are carried out by the manufacturer or by companies which the
manufacturer has approved to carry out such work, and when components that influence the safety of the lighting are
replaced with original spare parts. The lighting is not intended for use in areas where there is a risk of explosion.
Disposal
Disposal the product at a special collection centre for recycling of electronic and waste.
Environmentally conditions
Transport and storage
Ambient temperature .................................................................... -20° C to 70° C
Moisture content .......................................................... 10 % to 100 % (keep dry)
Atmospheric pressure ............................................................ 500hPa til 1060hPa
Operate
Ambient temperature ....................................................................... 5° C to 40° C
Moisture content ................................................................................ 30 % to 75 %
Atmospheric pressure ............................................................ 700hPA Max 3000m
6
LHH G2/LHH G2 CCT

Declaration of Conformity
(Directive 93/42/EEC as amended by directive 2007/47/EE
and RoHS directive 2011/65/EU, EcoDesign directive 2012/1194/EU)
Manufacturer: Glamox AS
Adress: Bruksvegen 7
2260 Kirkenær
Norway
Product: Ceiling, trolley, wall and Table standing LED
luminaire for medical
Product name: LHH G2 / LHH G2 CCT
We declare under sole responsibility that above listed products confirms with the standards
listed.
Reference Date of issue Name
EN-60601-1 2012 General requirements (lighting)
EN-60601-1-6 2010 Collateral standard: Usability
EN 60601-2-41 2009 Particular requirements for the safety of
surgical luminaires and luminaires for
diagnosis
EN 62366 2008 Application of usability engineering to
medical devices
EN 60601-1-2 2014 Electromagnetic compability –
Requirments and tests.
Place and date: Kirkenær, 20 April 2016
Name and signature of
authorized persons:
Lars Andersson Wehlin Geir Sylte
Factory Manager Laboratory Manager
Doc.No: PME-305-0044.Rev0
7

Please refer to our websit
e
for information about
our 5-year warranty.
Please visit our web site for contact information
www.glamox.com
© Glamox March 2016 Item no: 05-900-713-00 Doc. no: BRU-303-0112 Rev. no. 0
Manufactured by:
Glamox AS, BU Luxo production Kirkenær
Industrifeltet, 2260 Kirkenær
TEL. (47) 22 02 11 00, FAX (47) 22 02 11 02