Lynton promax lipo User manual


2

1.Description of ProMax Lipo …………………………………… 5
1.1 General Information …………………………………… 5
1.2 Description of Applicators …………………………………… 6
1.3 Installation …………………………………… 8
1.4 Start Up and Shut Down Procedure …………………………………… 9
1.5 Cleaning of Hand-pieces …………………………………… 10
1.6 Filter Cleaning …………………………………… 11
1.7 Statement of Intended Use …………………………………… 13
2. Safety …………………………………… 14
2.1 General Safety …………………………………… 14
3. Using the Touchscreen …………………………………… 15
3.1 Operating the ProMax Lipo …………………………………… 15
3.2 Operating the ProMax Lipo –Mode A …………………………………… 16
3.3 Cavitation (Body) …………………………………… 17
3.4 RF & Vacuum (Body) …………………………………… 18
3.5 RF & Vacuum (Face) …………………………………… 19
3.6 Operating the ProMax Lipo –Mode B …………………………………… 20
4. Treatment & Protocols …………………………………… 21
4.1 Consultation and Client Preparation …………………………………… 21
4.2 Facial Treatments –Summary Sheet …………………………………… 22
4.2.1 Facial Movements …………………………………… 23
4.3 Body Treatments –Summary Sheet …………………………………… 24
4.3.1 Abdomen Movements …………………………………… 25
4.3.2 Waist Side Line Movements …………………………………… 26
4.3.3 Back Movement …………………………………… 27
4.3.4 Thigh and Buttocks Movements …………………………………… 28
4.3.5 Calf and Arm Movements …………………………………… 29
5. Cautions …………………………………… 30
5.1 Contra-indications …………………………………… 30
5.2 Cautions …………………………………… 31
5.3 Maintenance …………………………………… 32
6. Warranty …………………………………… 32
6.1 Warranty Coverage …………………………………… 32
6.2 Exclusions …………………………………… 32
7. Specifications …………………………………… 33
Contents
Operating Manual
3

4
DISCLAIMER
Do not treat or use the system if any part is damaged or not functioning properly.
Lynton cannot be held responsible for any adverse reaction caused during the use
of this system.
The Company (Lynton Lasers Ltd) accepts no liability whatsoever arising from the
use to which this Product is put or from the non-use or attempted use thereof or
for the consequences arising therefrom.

1. Description of ProMax Lipo
1.1 General Description
Body
LCD
Hand-pieces
Start Key
Wheels
Hand-piece Holder
Hand-piece
Hose Holder
This manual contains instructions for the safe operation of the ProMax Lipo system
provided by Lynton Lasers Ltd.
The picture below indicates the main features of the ProMax Lipo system.
5

1.2 Description of Applicators –Mode A
• Treatment of cellulite and skin tightening
• Simultaneously delivers multi-polar RF and Vacuum
• Dynamic pulse function
• Red LED (630nm)
• Treatment of face, eye and neck for lifting
• Lymph drainage technology
• Simultaneously delivers bi-polar RF and
Vacuum
• 4 pulse modes
Body (RF & Vacuum)
RF : 1MHz
Face (RF & Vacuum)
RF : 2MHz
Body (Cavitation)
• Treatment of fat and cellulite
• 80mm diameter hand-piece
• 4 pulse modes
• Light signal function
1. Description of ProMax Lipo
6

•BodyApplicator
• Multi-polar RF Technology
• Used to treat larger and wider areas
• Red LED (630nm)
•Arm / LegApplicator
• Multi-polar RF Technology
• Used to treat smaller and targeted areas
• Red LED (630nm)
•Face Applicator
• Multi-polar RF Technology
• Used to treat face
• Blue LED (430nm)
1.2 Description of Applicators –Mode B
Arm / Leg
Face (RF 2 MHz)
Body (RF 1 MHz)
1. Description of ProMax Lipo
7

1.3 Installation
1. Screw the accessory holder into the machine.
2. Place the hand-pieces in the accessory holder.
3. Connect hand-pieces into the correct sockets on the back
of the machine.
NOTE: If the accessory holder doesn’t align so that the
holder faces towards the back of the machine, release the
screw on the top of the assembly and set the orientation
correctly.
NOTE : Take care to connect the correct hand-piece in the
correct socket. Failure to do so can result in damage to the
connector and hand-piece.
4. Hand-pieces now connected properly.
1. Description of ProMax Lipo
8

Press
1.4 Start Up and Shut Down Procedure
1. Press the power switch on the
rear of the system.
2.Turn the key switch.
3. The blue LED will light up on the side of ProMax Lipo.
1. Description of ProMax Lipo
To Shut Down
1. Leave the system for 2-3 minutes after treatment to allow internal fans
to cool down the system.
2. Whilst the system cools, clean the hand-pieces (see section 1.5).
3. Turn off at the key switch.
4. Turn off the power switch at the rear of the machine.
To Start Up
9

1.5 Cleaning of hand-pieces
1. It is essential to clean all the hand-pieces following
treatment. It is particularly important to ensure that the
vacuum Body & Face Accessories are cleaned with the
cleaning brush after EVERY treatment to avoid stray
lotion being sucked into the compressor and damaging
it.
2. Clean the vacuum inlet of the Body hand-piece by
inserting the cleaning brush.
3. Clean the vacuum inlet of Face hand-piece by inserting
the cleaning brush.
4. Clean other parts using sterilizing wipes.
5. Clean the Cavitation hand-piece with sterilizing wipes.
1. Description of ProMax Lipo
10

1.6 Filter Cleaning
1. Description of ProMax Lipo
11
1. You need to check the filters regularly to avoid system
blockage.
2. Open the panel on the rear of the machine by unscrewing
the four screws by hand.
3. Empty the filter container before the oil reaches the red line.
4. Remove the filter container using the tool provided. Twist the
container in the red arrow direction to remove.
5. After cleaning, replace the container by twisting in arrow
direction as shown.
6. Replace the panel on the back of the system.
For ProMax Lipo System Serial Numbers PROM-001 to PROM-026 do the following:

1. You need to check the filters regularly to avoid system
blockage.
2. Open the panel on the rear of the machine by unscrewing
the three screws by hand.
3. Empty the filter container before the oil reaches the red line.
4. To remove the filter, release the locking catch by pushing
downwards and rotating the filter clockwise from the locked to
unlocked position, denoted by the open and closed padlock
symbols.
5. After cleaning, replace the container by turning the filter
counter clockwise, ensuring the canister is located in the
locked position, (alignment arrow at the locked padlock
position).
6. Replace the panel on the back of the system.
1.6 Filter Cleaning (continued)
1. Description of ProMax Lipo
12
For ProMax Lipo System Serial Numbers PROM-027 and onwards do the following:

The ProMax Lipo body shaping system safely promotes fat metabolism,
enhances blood flow and promotes collagenesis with a combination of bipolar
and multipolar Radiofrequency, high frequency ultrasound cavitation and
vacuum massage.
The treatments are suitable for inch loss, skin tightening and cellulite reduction
on the body and skin tightening and rejuvenation on the face.
1.7 Statement of Intended Use
1. Description of ProMax Lipo
13

CAUTION: In the UK, the ProMax Lipo body shaping system should be used at all
times in accordance with the Health and Safety at WorkAct 1974.
CAUTION: Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous output.
(i) Handling:
The ProMax Lipo System, complete with hand-pieces, weighs in excess of 50kg. It
is therefore recommended that lifting or carrying the ProMax Lipo System from
place to place should be undertaken by two people, each person holding one of the
side handles.
(ii) Read the User Manual:
You are strongly advised to read this User Manual carefully before operating the
ProMax Lipo body shaping system for the first time. Systems may be electrically
damaged if not operated systematically and with a thorough knowledge of the
function of each component.
CAUTION : Non-Professional Users of the ProMax Lipo body shaping system
should undertake a course of Training before carrying out any actual treatments.
Lynton Lasers Ltd does not accept any responsibility for the actions of
inadequately trained users.
(iii) Keep Records:
You are further recommended to keep a regular record of all maintenance or repair
procedures carried out on the ProMax Lipo system. By doing this, the causes of
sudden performance loss can be more immediately diagnosed.
2. Safety
2.1 General Safety
14

3.1 Operating the ProMax Lipo
•Select the hand-pieces you want to use on the LCD screen.
3. Using the Touchscreen
Mode A: Cavitation / RF & Vacuum Mode B: Multipolar RF
15

3.2 Operating the ProMax Lipo - Mode A
RF & Vacuum (Face)
RF & Vacuum (Body)
Cavitation (Body)
Conversion button to Mode B screen
* Select and press the hand-piece from LCD screen
Choose Cavitation / RF & Vacuum for body treatment.
Choose RF & Vacuum for face treatment.
3. Using the Touchscreen
16

3.3 Ultrasound Cavitation Controls (Body)
⑧Start button
①Type of
Hand-piece
⑦Select Hand-piece type
②Cavitation
③Control
Time Button
⑤Preset Timer
①Type of Hand-piece: Displays the current operation Hand-piece.
②Cavitation: Cavitation operating button
③Control Time Button: Adjusts operation time button (1~30 min)
④Pulse Program Button: Selects the pulse program. (4 modes: C, P1 ,P2 ,P3)
C - Continuous P1 - 4:1(on time:off time) P2 - 3:2 P3 - 1:1
⑤Preset Timer: Shows total treatment time
⑥Display Power Level: Shows the current power level of cavitation. (Level : 1~10)
⑦Hand-piece change button
⑧Start and Stop Button
⑨Conversion button to Mode B screen
NB: When using Cavitation
Continuous mode should be used when the hand-piece is in motion.
The pulsed modes can be used whilst holding the hand-piece still on stubborn areas.
⑥Display Power Level
④C, P1, P2, P3
⑨Conversion button
to Mode B screen
3. Using the Touchscreen
17

①Type of Hand-piece: Displays the current Hand-piece type.
②Vacuum: Operating Vacuum Control button (Level: 1~ 5)
③Control Time Button: Adjusts operation time button (1~30 min)
④Vacuum Pulse: Choose the pulse (4 modes).
* Pulse type < C:Continous P1:Slow Pulse P2:Medium Pulse P3:Fast Pulse >
⑤Preset Timer: Shows total treatment time
⑥Multipolar RF: RF operating button (Level : 1~10)
⑦Select Hand-piece Type
⑧Display Power Level: Shows the current power level of Vacuum and RF
⑨Start and Stop button
⑩Hand-piece change button
⑪Conversion button to Mode B Screen
3.4 RF & Vacuum (Body)-1MHz
⑨Start Button
①Type of
Hand-piece
④Vacuum Pulse < C, P1 , P2 , P3 >
⑦Select Handpiece Type
②Set Vacuum level
⑥Multipolar RF
⑤Preset Timer
③Control Time Button
⑧Display Power Level
⑪Conversion Button
to Mode B screen
⑩Hand-piece change Button
3. Using the Touchscreen
18

3.5 RF & Vacuum (Face)-2MHz
⑨Start Button
①Type of
Hand-piece
Vacuum Pulse < C, P1, P2, P3 >
⑩Hand-piece change button
②Bipolar RF
level
⑥Set Vacuum
Level
⑤Preset Timer ③Control
Time Button
⑧Display Power Level
⑦Select Handpiece
Type
①Type of Hand-piece: Displays the current Hand-piece type.
②Bipolar RF: RF operating button (Level: 1~10)
③Control Time Button: Operation time control button (1~30 Min)
④Vacuum Pulse: Choose the pulse (4 modes).
* Pulse type < C:Continous P1:Slow Pulse P2:Medium Pulse P3:Fast Pulse >
⑤Preset Timer: Shows total treatment time
⑥Vacuum: Operating Vacuum Control button (Level: 1~ 5)
⑦Select Hand-piece Type
⑧Display Power Level: Shows the current power level of Vacuum and RF
⑨Start and Stop button
⑩Hand-piece change button
⑪Conversion button to Mode B Screen
⑪Conversion button to Mode B screen
3. Using the Touchscreen
19

1. Choose the hand-piece you wish to use
2. Adjust the power output and time as required
Set the level
* Preset output level : 5 (Adjustable from 1 to 10)
3.6 Operating the ProMax Lipo –Mode B
Arm / Leg Body (1 MHz) Face (2 MHz)
Conversion button
to A screen
3. Using the Touchscreen
Notes:
For Arm, leg and body treatments, use higher settings (from 5 upwards)
For face, use lower settings (2-3, increasing if necessary)
Important –keep good contact at all times to avoid RF arcing
20