EC/EU DECLARATION OF CONFORMITY
FOR MAGNAFLUX MEASURING INSTRUMENTS
We hereby declare that this equipment complies with the follow-
ing EU standards. The device has been tested and approved. Any
modification made to the device without our written consent will
invalidate this declaration.
Equipment name: FSM-2 Field Strength Meter
Applicable standards:
DIN EN ISO 9934-1:2017-03
DIN EN ISO 9934-3:2015-12
DIN EN 61010-1:2011-07
Authorised person for documentation:
Mr. Georg Koch, Quality Management Ocer
Address:
Magnaflux GmbH, Bahnhofstr. 94-98, D-73457 Essingen, Germany
Person responsible: Signature:
Sales/Design Manager
Mr. Silvio Georgi
Date: 01.01.2017
4. DISPOSING OF THE DEVICE
Before disposing of the FSM-2, check that the power supply is disconnected and
the battery is removed.
Refer to your local waste disposal regulations as chargers, batteries, and button
cells are hazardous waste and pose risks to health and the environment.
Procedure:
• Take the machine out of service.
• Separate the components of the FSM-2 into metals, plastics, composite materials,
electrical components, batteries.
• Dispose of all parts in accordance with local regulations.
If you have any questions, please contact Magnaflux GmbH.
3.5 CALIBRATION
Using long-life robust components reduces the need for daily calibrating. If
required, we can supply a reference standard of 200 A/cm (part number 133042)
as an accessory to help you check the proper functioning of the device.
NOTE: if you order a reference standard once you have started using the FSM-2,
you will need to send the entire unit to Magnaflux GmbH for calibration.
3.5.1 Tangential field probe FSM-2 with reference standard (optional)
The display accuracy of the FSM-2 should be periodically checked using a refer-
ence standard. To do this, lay the probe flat and flush with the groove of the refer-
ence standard, with the white point pointing upwards. There should be a reading
of 200 A/cm ± 4% in the display.
If the displayed value diers by more than 4% from the target value of 200 A/cm,
send the FSM-2, with the probe and reference standard, back to Magnaflux GmbH
for calibration. We recommend that you do this on an annual basis anyway.
3.5.2 Tangential field probe FSM-2 without reference standard
We recommend that the FSM-2 meter, along with its probe, is returned to Magna-
flux GmbH once a year for recalibration.
6