Equipment required
•A calibrated PCCU
•The 3005 station with the Soldering iron and Hot air pencil
•The Soldering iron and Hot air pencil calibration jigs
•The RJ-12 to RJ-12 connecting cable
Note: The calibration procedure must be carried out in a closed environment, as the
slightest breeze will affect the accuracy.
Procedure
•Connect the PCCU with the RJ-12 cable to the 3005 Station.
•Connect the thermocouple plugs to the correct inputs
•Connect the Soldering iron and Hot air pencil to the Station and insert into their
respective jig’s.
•Switch on the PCCU
•Switch on the 3005 Station
•The dots will disappear within two seconds but the C or F will remain.
•If this does not happen the connection or the cable is faulty.
•Press the Function button on the PCCU until the calibrate LED is lit.
•Press the Execute button on the PCCU.
•If all the LED’s on the PCCU light up, then there is a fault and the PCCU has to
be recalibrated.
•If everything is ok then the Busy LED will light and the calibration procedure will
start.
•Calibration consists of six phases: Warming up, and calibrating of the following
temperature points, 100 °C, 200 °C, 300 °C, 400 °C and 500 °C.
•The duration of the procedure is determined by the factory
(± 20 minutes)
•During calibration the display on the Station cycles through three values :
•The setpoint temperature (shown when the dots are flashing)
•The temperature of the Soldering iron measured by the PCCU (shown when the
Soldering iron LED is on)
•The temperature of the Hot air pencil measured by the PCCU (shown when the Hot
air pencil LED is on)
•The procedure will stop if the temperature measured by the PCCU is out of
specification.
•If the Calibrate, Configure, Busy and Complete LED light up, then the Soldering
iron is out of specification.
•If the Calibrate, Reset to factory setting, Busy and Complete LED light up then
the Hot air pencil is out of specification.
file:///C|/Documents%20and%20Settings/Lauren...her%20stuff/magnum/3005%20word%20booklet.htm (10 of 12) [03/03/2003 10:16:31 PM]