Mangar Airflo 12 User manual

Simple solutions for everyday independence
English
User Instrucons
Airo 12

Airo 12
2
Intended Use
The Airo 12 by Mangar Health is a low pressure
air compressor, which has been designed to
inate a range of Mangar Health pneumac
liing equipment.
Applied Parts (Accessories) for the Airo 12:
Bathing Cushion
Handy Pillow Li
Sit-u-Up Pillow Li
Leglier
Surfer Bather
WARNINGS
• Be sure that you have read and understood
these instrucons and those provided with
any other Mangar Health product, before
operang the equipment.
• To avoid the risk of electric shock, NEVER
recharge the baery in the bathroom,
outdoors or in a damp environment.
• Always disconnect from the baery charger
before cleaning.
• Use only the supplied baery charger.
• Do not operate the Airo 12 compressor
in the presence of ammable anaesthec
mixtures with air, oxygen or nitrous oxide,
as these mixtures can create a combusble
environment leading to the potenal for re
or even an explosion in the presence of a
spark.
• Ensure that when in use, the Airo 12 and
its connected hose and hand control lead
are laid out in such a way as to avoid any risk
of tripping or strangulaon.
• Ensure that children and pets do not
interfere with the safe operaon of the
Airo 12 and inatable product.
• Keep the Airo 12 and inatable product
away from sources of heat.
• Disconnect the Airo 12 compressor from
the inatable product before transporng so
as to avoid accidental inaon.
• The Airo 12 should only be used for its
stated purpose and should not be modied
in any way.
This product and associated packaging does
not contain latex.
Contents
Intended Use Pg 2
Warnings Pg 2
Standard Package Contents Pg 3
Baery Status Indicator Pg 3
Baery Charging Pg 4
Operaon Pg 5
Cleaning & Storage Pg 6
Maintenance Pg 6
Trouble shoong guide Pg 7
Specicaons Pg 7
Symbols used on Airo 12 and Charger Pg 8
Symbols used on packaging Pg 9
Warranty Pg 9
Declaraon of Conformity Pg 10

Simple solutions for everyday independence
Airo 12
3
Standard Package Contents - g. 1
1 x Airo 12 compressor (A) ed with a 12v -
2200mAh NiMH baery pack.
1 x Baery charger (with adaptors) (B).
1 x 2-Way hand control (C)
1 x User Instrucons.
1
Baery Status Indicator
Geng Started
The Airo 12 does not have an ‘on/o’ switch. To
operate the compressor, press the ‘Up’ buon
on the Airo or on the hand control.
Up Down
A
B
C
NOTE: The baery must be put on charge immediately as soon as the red indicator light appears.
Running the Airo 12 on RED unl the baery is exhausted will damage the baery.
When the compressor is running, the baery status indicator lights will only indicate the state of
baery charge as it was at the start of the li. When the compressor stops running, the baery
status indicator will show a boom green light (this will be red if the baery charge has become
low). To determine the true baery state following a li, press the ‘Baery Status Indicator’ buon
and the lights will indicate the baery charge state.
IMPORTANT: To maximize the life and liing performance of the baery, fully recharge the
baery for 16 hours. When the baery is new, it will not reach its peak performance and
capacity unl it has been discharged and recharged at least 3 to 5 mes.
Fully charge the baery for a minimum of 16 hours before using the Airo 12 for the rst me
and when a new baery has been ed.
To determine the baery charge state before use, press the ‘Baery Status Indicator’ buon to
acvate the baery status indicator lights:
RED
Recharge
Immediately
GREEN
OK to Use
GREEN
OK to Use
GREEN
OK to Use
GREEN
OK to Use
Baery Status
Indicator
Buon

Airo 12
4
2
First 2
Hours
16
Hours
NOTE: The charger is not capable of providing sucient power to charge the baeries while the
compressor is running.
To maintain baery condion, it is recommended that the Airo 12 is le connuously on charge
when not in use. The baeries cannot be overcharged using this charger.
WARNING: DO NOT recharge the baeries in a bathroom or in a damp environment.
To disconnect the charger from the mains power supply, unplug the charger from the mains power
socket.
Baery Charging - gs. 2
• Check the baery charger casing and lead
for signs of damage and if damaged DO NOT
USE.
• Fit the appropriate mains plug adaptor to
the baery charger (A).
• Plug the baery charger into a mains power
socket and connect the charger output lead
to the Airo 12 charging socket (B).
NOTE: The baeries must be recharged for a
minimum of 16 hours to achieve full capacity.
When the baery is being charged:
• For the rst 2 hours, the boom light will
show a steady red and the next light will be
ashing green. Aer 2 hours the lights will
change to two steady green and one ashing
green. Aer 16 hours all lights will become
steady green.
NOTE: If charging is momentarily interrupted
(a power cut, switching the power socket o
or disconnecng the charging lead), on re-
connecon, the charging process will start at
the beginning and follow the sequence above
for the next 16 hours.
B
A

Simple solutions for everyday independence
Airo 12
5
C
Operaon - g. 4
Disconnect the Airo 12 from the baery
charger and place it in a safe and convenient
posion.
• Connect the Hand Control plug (A) to the
Airo 12 hand control socket (B).
• Connect the hose from the Mangar
pneumac liing equipment to the Airo
12 air outlet socket (C). The connector will
‘click’ into place when rmly pushed onto
the socket.
• Press the ‘Up’ buon (D) to raise.
• Press the ‘Down’ buon (E) to lower.
When inang Mangar Health products, the
compressor will stop automacally once the
product is fully inated.
NOTE: When either the ‘Up’ or ‘Down’ buons
are pressed, the internal microprocessor carries
out a baery voltage check and if the voltage
becomes too low, the ‘Down’ buon will not
work. This feature will always ensure that the
seat of a Mangar Health bathing product cannot
be lowered into a bath if there is insucient
baery power to raise it again.
B
A
D
E
4
Mains Power Adaptor (oponal)
An oponal mains power adaptor is available to
power the Airo 12 directly from a mains power
socket. Connect the mains power adaptor
between a mains power socket and the adaptor
inlet socket on the end of the Airo 12 (Fig. 3).
With the mains power adaptor connected and
the power switched ‘on’, all lights on the baery
state indicator will show steady green.
NOTE: The Airo baeries are not being charged
when the mains power adaptor is connected.
3

Airo 12
6
Maintenance - g. 5
The Airo 12 has been designed to be maintenance
free and apart from replacing the baery, the Airo
12 contains no other user serviceable parts.
Baery replacement:
• Turn the Airo 12 onto its end containing the
air outlet socket, remove the screw (A) and
slide the end panel (B) away from the housing.
• Disconnect the wire from the baery pack
and remove the baery pack from the baery
compartment.
• Fit the replacement baery pack, ensuring that
the connector is secure, ret the end panel and
secure the screw (A).
A
B
5
Disposal
When the Airo 12 has reached the end of it’s working life, remove the baery for safe disposal
in accordance with local regulaons, and dispose of the Airo 12 at your local household waste
recycling centre (civic amenity site).
Cleaning
Unless contaminaon is suspected, periodic cleaning is all that is recommended. The Airo 12
should be cleaned/disinfected as follows:
• Unplug the Airo 12 from the baery charger. Clean the external surfaces using a non abrasive
damp cloth with warm water. Wipe dry thoroughly with a dry towel before reconnecng to the
baery charger. Do not allow moisture to enter any part of the compressor.
• Do not use solvent-based products.
Disinfecng
• Pre-clean using a damp cloth and then disinfect using a hard surface alcohol impregnated wipe.
Storage
• When not in use, store in a dry place together with these instrucons. To maintain the baeries,
when being stored for long periods, put the Airo 12 on charge for 16 hours every 3 months.
NOTE: Ensure that the baeries are fully charged before storing the Airo 12 for long periods.
When it becomes apparent that the Airo 12 is producing fewer than expected liing cycles, the
baery is probably reaching the end of its working life and should be replaced.
If a fault should develop or if you require assistance in seng up using or maintaining the equipment,
please contact our Customer Service Department on Freephone 0800 2800 485 (UK only. Outside
of the UK please contact your supplier).

Simple solutions for everyday independence
Airo 12
7
Trouble shoong guide
Compressor does not run when ‘up’ is
selected on the 2-way hand control.
Ensure that the air hose from the product
being inated is connected to the Airo.
Compressor keeps stopping when trying
to inate the liing aid.
Check that there are no kinks in the air
hose.
Airo 12
Weight: 3.7Kg (8 lbs)
Output pressure: 0.25 bar (3.75 psi)
Baeries: 1 x 12v - 2200mAh Ni-MH baery pack
Baery Life: Greater than 500 charging cycles (2 - 5 years depending on usage)
Environmental protecon:
(Airo 12 housing) IP51 (Protecon against large dust deposits & vercal falling drops
of water)
(Push-buon hand control) IP67 (Totally protected against dust & immersion in water)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislaon
Charger: Model number - HX1506SR
Input: AC 100v - 240v, 50-60Hz, 0.5A Max
Output: DC 15v, 0.35A
Class II Double insulated equipment
Check that there are no kinks in the air
hose.
This happens when the compressor’s
internal temperature cut-out has tripped.
Leave the Airo 12 for 60-90 minutes to
cool.
Boom Baery Status Indicator light
ashing ‘Red’.
Compressor causes electrical
interference with another piece of
electrical equipment.
Move either the compressor or other
item further away from each other unl
the interference stops.
Specicaons

Airo 12
8
Symbols used on Airo 12 and Charger labels
For indoor use only. Do Not
connect to mains power in a
bathroom or outdoors.
‘Up’ buon
‘Down’ buon
Alternang current (AC)
Fuse (with rang)
Direct current (DC)
Type BF equipment (protecon
against electric shock)
Max/Min operang environment
temperature
Aenon consult instrucons
Warning
Class II Double Insulated equipment
Dangerous voltage
Duty cycle
I = Operang me
O = Non operang me
(Max connuous operang me = 5
minutes)
Recycle
Do not dispose of in household waste
Date of manufacture
Baery Status Indicator Buon

Simple solutions for everyday independence
Airo 12
9
Symbols used on packaging
Keep packaging dry
Handle with care
Max/Min storage temperature
Net & gross weights
Stack no more than ve high
Recycle packaging
Do not dispose of in household waste
Airo 12 serial number: .............................
Date of purchase: .............................
Invoice/Delivery note number: .........................
Warranty
The warranty is given against defects in material and workmanship and comprises piece, parts and
labour. The warranty is not valid in the case of fair wear and tear, wilful damage, common neglect,
misuse or unauthorised alteraon or repair by unqualied persons.
The following statements do not aect the consumer’s statutory rights. This product is sold on the
understanding that in the event of any defect of manufacture or material appearing within the
warranty period, the defect will be reced free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the claim is being
made within the warranty period.
(b) The defect is not aributable to accidental damage (either in transit or otherwise), misuse or
unauthorised repair.
We strive to provide our customers with reliable and durable products of the highest quality.
In addion to our standard warranty, and for added peace of mind, our Customer Service Department
oers an Annual Service and Clean package.
To nd out more about this service please call our Customer Service Department on
Freephone 0800 2800 485 (UK only).
Max/Min humidity when transporng
or in storage
Max/Min pressure when transporng or
in storage
Max/Min temperature when
transporng or in storage
Mangar Internaonal Ltd, trading as Mangar Health, reserves the right to alter product specicaons and/or any of the informaon contained within
this document without noce.

Airo 12
10
DECLARATION OF CONFORMITY
This is to cerfy that :-
The
Airo 12 Compressor -
complies with:-
The requirements for Class 1 devices according to Annex VII of the Medical Devices
Direcve 93/42/EEC as amended by 2007/47/EC
IEC 60601-1
Medical Electrical Equipment - General requirements for basic safety and essenal
performance
IEC 60601-1-11
Medical Electrical Equipment - Requirements for use in the home care
environment
IEC 60601-1-2
Medical Electrical Equipment - Electromagnec compatability
Mangar Internaonal Ltd
Presteigne, Powys, Wales,UK. LD8 2UF
tel: +44 (0)1544 267674
fax: +44 (0)1544 260287
heado[email protected]
www.mangarhealth.com
Signed:
Date: April 2017
C. Butcher (Finance Director)

Simple solutions for everyday independence
Airo 12
11
Please note that these instrucons are available in other languages and can be downloaded from
our website by vising www.mangarhealth.com

Simple solutions for everyday independence
MI0353 issue 1
(04/17)
Mangar Internaonal Ltd
Presteigne, Powys, Wales, LD8 2UF
United Kingdom
Tel: +44 (0)1544 267674
Fax: +44 (0)1544 260287
E-mail: sales@mangarhealth.com
www.mangarhealth.com
Other manuals for Airflo 12