Matsushima PFM-KCU12 User manual

TPFM-005E
Rev7:Jul.14,2017
Contents
1.Outline・・・・・・・・・・・・・・・・・2
2.Specification・・・・・・・・・・・・・・3
3.Shape・・・・・・・・・・・・・・・・・・4
4.Mountingmethod・・・・・・・・・・・・・5
5.Howtoconnect・・・・・・・・・・・・・・6,7
6.Partnamesandfunctions・・・・・・・・・8
7.Startup・・・・・・・・・・・・・・9〜11
8.SetupexampleofParameter・・・・・・・・12
9.Maintenance・・・・・・・・・・・・・・・12
10.Troubleshooting・・・・・・・・・・13,14
INSTRUCTION MANUAL
FOR
Dust Monitor with
Density compensation
TYPE
PFM-KCU12
(
Transducer
)
PFM-M11P
(
Sensor
)
PFM-M11PT
(
High-temperature type
)
1-8-18Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu807-0837Japan
PhoneNo.(8193)691-3731FaxNo.(8193)691-3735
http://www.matsushima-m-tech.com
E-mailsales@matsushima-m-tech.com
※TheoperatorshouldreadthisInstructionManualcarefullyand
handlethedevicecorrectly.

-
1
-
Safety Precautions
Be sure to thoroughly read the instruction manual before using the products.
Keep the instruction manual in a safe, convenient location for future reference.
All or part of the contents described in this manual may be changed without any notice.
Due to our constant striving for further improvement of products, parts or products that differ from those
described in this manual may be substituted.
WARNING
(Failure to observe this WARNING may cause a fatal or serious injury.)
Be sure to confirm that any peripheral equipment does not move before installation work.
In addition, observe safety requirements for installation work where high-place work is expected.
Be sure to turn off the power source before wiring, mounting and transportation work. (Failure to observe
this WARNING may result in an electric shock/ injury or equipment damage due to short-circuit.)
Carry out wiring work correctly with reference to a proper drawing.
Never disassemble the equipment. (Failure to observe this WARNING may result in an electric shock.)
Do not open the cover under an explosive environmental condition when power is entered. (Failure to
observe this WARNING may result in an injury or equipment damage.)
Do not place or store the equipment in any hostile environmental place where it will be subjected to direct
sunlight, rain, water droplet, hazardous gas/water, etc..
CAUTION
(Failure to observe this CAUTION may cause a moderate injury or equipment damage.)
Do not use the equipment for any purpose other than the original purpose of use.
Be sure to confirm the specification of equipment and use the equipment within the range of specification.
(Mounting conditions such as temperature, power source, frequency, etc.)
Make sure a correct wiring before applying power source.
Do not have a shock or strong impact to the equipment.
(Failure to observe this CAUTION may result in equipment damage.)
Be sure to connect necessary terminals (grounding, etc.).
Remove all wiring to the equipment before doing electrical welding work near the equipment.
Do not forcedly bend or pull the lead wire also do not use unnecessarily long wire.
Tighten the cover, lead outlet, etc. properly so that dust, rainwater, etc. do not enter inside the equipment.
Do not use the equipment under a corrosive condition (NH
3
, SO
2
, CI
2
, etc.).
Be sure to tighten the cable grand so that outer air does not enter inside the equipment.
When applying piping connection such as conduit, etc. instead of cable grand, apply putty or equivalents
on the cable entry so that outer air does not enter inside the equipment.
●
Do not directly touch the probe with hands regardless of power ON or OFF.
If you unavoidably have to touch the probe with hands when installation or maintenance process, be sure
to turn OFF the power and use non-conductive gloves such as rubber gloves during the process.
*Touching the probe with hand directly may cause a failure due to static electricity charge in a human body.
●
MAINTENANCE: When you clean the probe, be sure to turn OFF the power, put on non-conductive
gloves such as rubber gloves to clean the probe with moist cloth.*Touching the probe with hand directly
may cause a failure due to static electricity charge in a human body.Cleaning the probe with dry cloth may
also cause a failure due to friction charge generated between the cloth and probe.
IMPORTANT
(indicates notes or information to help customers.)
Limitations of Warranty:
Warranty period shall be one year from the date of delivery (ex-factory).
Any damage of any other products that have occurred for use of the equipment is not covered by
this warranty. Also any loss induced by failure or malfunction of the equipment is not covered by
this warranty.
Failure or malfunction caused by following are not covered by this warranty:
a. Modification or repair by a party other than MATSUSHIMA’s authorized personnel,
or replacement of parts not recommended by MATSUSHIMA.
b. Inadequate storage, installation, use, inspection or maintenance that does not comply
with specifications.
c. Cause for any peripheral equipment or device.
d. Accident beyond control and force majeure (fire, earthquake, flood, riots, etc.).
Lack of instructions to MATSUSHIMA for information or safety requirements that can be predicted only
by customers’ side.
This warranty conditions do not limit customers’ legal right.
Price for the equipment does not include any charge for services such as commissioning,
supervising, etc..
!
!
!

-
2
-
1.Outline
Electricchargewillgeneratewhenanumberofcorpusclecontainedingaspassthroughand/or
getintouchwithprobeofdustmonitorequippedwithduct,plumbingpipe,etc.
Thedustmonitoramplifiesthis"Displacementofthecharge"withthesensor,andsendsthe
amplifiedslightsignaltotheoutputunit.
Thiselectricsignal"Displacementofthecharge"isbothfilteredandamplifiedandthenoutput
instrumentationsignalof4〜20mAinproportiontodustamountandcontactsignalaswell.
Important:ViewfordensityofdustmonitorperFrictionElectrificationMechanism
OutputvalueofFrictionElectricChargeMechanicdustmonitorwillbevariabledueto
variousenvironmentalfactorssuchas「kindofparticle」「particlediameter」「specificgravity
「currentspeed」throughmeasureprinciple.
Underaregularenvironment,however,linearcharacteristicwillnotofcoursevaryunderthat
environment.
AstoconsideringdensityregardingtoDustMonitorofFrictionElectricChargeMechanism,
itrequiresyoutoconsiderinfluenceduetovariationofparticlecharacteristicsasafore
mentionedanditdoesnotalwaysmeetdensityspecified.
Densityindicatedonspecificationisfordesigndevelopmentofmachineryandispublicized
foryoucustomerstouseasaguidelineindecidingtherange.
Whenyoucustomersdecidetherange,youpleasefindtherangeandsetitinwhichalarmlevel
iseasytosetstrictlyfrom0−100%signalandbecomespulsesignal.
Bydoingso,youcanrealizeyourmanagementofexhaustdrift.
Ifthereisaneedherebytoexecutedensitymanagement,youarerecommendedtodecidedensity
equivalentto0-100%.
Ondecidingabove,youpleaseinvestigatedensityofpulseandyoucanreadsignalasdensity
from current signal of machinery by jointly utilizing compensation coefficient, integral
function,etctogettherangeacquiringsignalnearesttoactualdensityof 0−100%.
(Thevalueyoucangetiscompensatedandisarelativeconversionvaluebutnotabsolute.)
!

-
3
-
2.Specification
2-1.Powersupply
2-1-1.Powerpressure AC110Vor220V-15%/+10%(specifyondemand)50/60Hz
2-1-2.Powerconsumed 7VA
2-2.Materialtomeasure Solidparticleingas
2-3.Qualityofparticle Allkindofsolidparticle
2-4.Particlesize ≧0.3μm
2-5.Densityofparticle 0.1mg/m3〜1000mg/m3
2-6.Measureprinciple Frictionelectricchargedetectionmethod
2-7.Measurecondition(processcondition)
2-7-1.Temperature ≦250℃(PFM-M11PT:≦400℃)
2-7-2.Pressure ≦200kPa
2-7-3.Speed ≧4m/s
2-7-4.Humidity ≦40vol%
2-8.Environmentalcondition
2-8-1.Temperature-20〜65℃(Transducer:50℃)※1
2-8-2.Humidity ≦95%RH(Condition:Stateofcasesealingup)
2-9.Keepingcondition
2-9-1.Temperature -20〜65℃
2-9-2.Humidity ≦95%RH
2-10.Outputsignal
2-10-1.Currentoutput DC4〜20mA(isolatedoutput,loadresistance:500Ω)
(max.23.3mA)
2-10-2.Contactoutput Alarmcontactoutput×1(250VAC,2A/30VDC,2A)
Upperlimitcontact(limit)output×2(250VAC,2A/30VDC,2A)
2-11.Fittingmethod R1fitscrew(1"socket),(option:flangefitting)
2-12.Sensitivityswitch ・9stepsswitchcando
・Foreachrange(sensitive),0.1〜2.0doubleoutput
・compensationcando
2-13.Protectionstructure IP65(cover,inaconditionofcableglandtightening)
IP20(Transducer)
※1:Themaximumtemperatureisdifferentaccordingtothemeasuringrange.
(Pleaserefertothechart.)
Range
PFM-M11P PFM-M11PT
Fig.1.Graphoftemperatureandhumidity

-
4
-
3.Shape(roughdrawing)
Fig.2.Shape(roughdrawing)
L2(※2)
80
66
60
68
(85)(30)
φ18
φ10
(90)(57)L1(※2,3)
※1
146
35
DINRail
Railsocket
4-φ5
Mtg.Holes
Signalcable
LeadoutletG1/2(※6)
Earthterminal:M4screw(※4)
Housing:ADC
M5Bolt,Nut(※5)
Mtg.BossR1
1-socketL=43mm
Insulator:PTFE
Probe:SUS316L(※3)
160
Transducer:PFM-KCU12 Sensor:PFM-M11P
80
66
60
68
146
35
DINRail
Railsocket
4-φ5
Mtg.Holes
Signalcable
160
φ18
φ10
(90)(122)L2(※2)L1(※2,3)
※1
Earthterminal
:
M4screw(※4)
Housing:ADC
M5Bolt,Nut(※5)
Mtg.BossR1
1-socket
L=43mm
Insulator:
Ceramic
Probe:SUS316L(※3)
Radiationfin:
Alminium
φ98
(30) (85)
LeadoutletG1/2(※6)
Transducer:PFM-KCU12 Sensor:PFM-M11PT
※1)PleasedetachAbetweensocketandinsulator10mmormore.
※2)RefertothespecificationsforlengthL1,L2.
※3)Theprobeshouldbetoucheddirectlybyhandwithoutgrounding
whilebeingenergized.
※4)Pleasebesuretoconnectearthterminal.
※5)PleasetightentheBoltandNutwhenyouinstallit.
※6)Pleasemakesurethattheleadoutletisfirmlyclosedsothat
theoutsideairdoesnotinvadeinsidetheunit.
Whereitisconnectedbypipingconstructionsuchasconduitsystem,
pleaseapplyputtyandsoontotheleadcableentrysothatthe
outsideairdoesnotinvadeinsidetheunit.

-
5
-
D
ustmonitor
4.Mountingmethod
TomountDustMonitor,1-socketisprovidedwith.
Wledthesocketontothepipe.Installthedustmonitorsoitishorizontalorvertical.
Weldingshouldexecuteallaroundthepipetoshutoffgas.
4-1.Precautionformounting
DustMonitormaymal-functioninanenvironmentaslistedhereunder.
MountingMonitorinsuchenvironmentshouldberefrainedfrom,please.
(1) Plumbingpipetobeextremelysqueezedorclosetoaplacebending.
(2) Mountingpositionwithindiameterofductor1.5ofverticaldimension.
(3) Nearafan
(4) Fittingplumbingpipe・shortpipe・ductarenonconductive.
(5) Measureenvironmentandconditionsareoutofspecification.
(6) Incaseraisingbulkupwithshortpipeinaplacewheredustsaccumulatebetweenshort
pipeandprobe.
(7) Placewheretipofprobeiswithin20mmfromplumbingpipe.
Important:Wiringofpowerandsignalshouldbeindependent,please.
Capisfortransportprotectionandisnotaraincoat.
Attention:・Exclusivecableistransmittingveryweaksignal.
Cableshouldbeprotectedwith,byallmeans,steelmadeelectricwirepipe
independentlysothatinductivenoiseformambientareadoesnotbothersignal.
・Securelyclosetheleadoutlettopreventoutsideairfromenteringinthemain
unit.
DonotremovetheaccessorySilicagel.(fordehumidification)
・Whereusingconduitpipingandsooninsteadofcablegrand,besuretoapply
puttyorlikeatcableentrysothatambientairdoesnotcomeintotheunit.
・Isolatetheevaluationunitfromtheouternoisesourcesandtakeanymeasure
againstnoiseiftheremaybeanynoisefrompowersupply/signallines.
!
!
※ClearanceA≧10mm
※
Clearance
B
≧
2
0mm
Important:
ForinstallationofSensor,
PlaceR1Attachmentscrew,
theninsertProbeandtighten
M5boltandnutonthesideof
R1Attachmentscrew.
Formaintenance,by loosening
boltandnut,youcanpullout
Sensor.
!
Fig.4.Example)Installationnotgood
Fig.3.Installation
Insulation
Probe
Directionofflow
Probe
Direction
of
flow
Directionofflow
Exhaustgas
R1Attachmentscrew
Probe
Fig.5.Installationreferencefigure

-
6
-
5.Howtoconnect
①
②
② AttachingScrewCover
AttachingScrewCover
Upper:Mainblockpart
Lower:Terminalblockpart
ThereareAttachingScrewCoversatupperandlower
portionsofasurfaceofaTransducer.
Pleaseremovethecoverswithanegativedriverorthe
like.
Pleasekeepthecoverswhichyouremoved.
Pleaseattachthecoverwhichyoukeptafterwiring.
Fig.6.Removingcover
Aterminalblockpartisjoinedtoamainblockpart
withfourpointsofscrewsandconnectors.
Youcanseeascrewjoiningtotheterminalblockpart
underholesforTransducerattachingwhenyouremove
theattachingscrewcovers.
Pleaseloosenthesejoiningscrewsofthe
terminalblockpartwithaPhillipsdriver.
YoucanseetheterminalblockoftheTransducerwhen
youremovethemainblockpartfromtheterminalblock
partafterhavingloosenedtheseJoiningScrews.
Joining
Screws
Mtg.Hole
Mtg.Hole
Mtg.Hole
Mtg.Hole
Fig.7.Joiningscrewsofterminalandmainblockpart
Fig.8.Terminalpart
Board(Terminal)
Powerterminal
DC4-20mterminal
Sensorterminal
Connector
(mainblockpart)
Relayterminal
Whenyouwiretheterminalblockoftheconverter,you
mustnotmakeanyfalsewiring.
Afterwiring,pleaseattachthemainblockpartwith
theprocedurethereverseofhowtoremovethemain
blockpart.

-
7
-
Fig.9.Connection
Important:EMC-cableglandisusedforSensor.
Again,whenwiring. Figure.10.becomespointsofwiring.
Sensor
Substratefixing
screw
COMNONC
(1)Powersupply
(2)Alarmcontact(SPDT)
(3)Upperlimit1contact(SPDT)
(4)Upperlimit2contact(SPDT)
(5)4-20mA(Analogoutputsignal)
※PleaseearththeterminalFG.
(
3
)(
2
)
(
1
)(
4
)(
5
)
Tocablegrand
White/Brown/Black/Green/Yellow
Exclusiveline
Shildedcable
Transducer
Line-connectingterminal:
Max.1.5mm2
!
(Componentpart)
Packing Body
Shield-ring
Rubberbush
Clampingmember
Cap
(
2
)
Completiondrawing
(1)
85
(
3
)
5
(
4
)
Packing
(1)
PleasepassCap,Clampingmember,andRubber
bushthroughthecable.
Pleasepeeloffthecablecoveringfrom
thevergebythelengthof85mm.
(2)
Pleasemaketheshieldlineasunder.
Pleasewrapanasundershieldline
asshowninleftfigure.
(3)
PleasepassShield-ringthroughthecable.
Pleasepeeloffthecoverofthecorewireby5mm.
Please use solderless terminal with insulating
coatingofStick-typebythree(brown,yellow,and
white)corewires.
Agreencorewiremustusethepressureterminalof
Rtype(R1.25-4).
(4)
PleasepassBodyandPackingthroughthecable,and
installitsothattheshieldlinemaysurelycome
incontactwiththeShield-ring.
Fig.10.becomespointsofwiring.

-
8
-
6.Partnamesandfunctions
(1) RANGE: measurerange(1、2、3、4、5、6、7、8、9)set
(2) L1ON: setONpointoflimit1(setrange:0〜100%)
(3) L1OFF: setOFFpointoflimit1(setrange:0〜100%)
(4) L2ON: setONpointoflimit2(setrange:0〜100%)
(5) L2OFF: setOFFpointoflimit2(setrange:0〜100%)
MinimumrangeofHysteresisisstill1%.
(6) INTEGRATIONTIME: Integraltimeset(setrange:0〜30seconds)
(7) CORRECTION : Compensationvalueset(setrange:0.1〜2.0)
【forindication】
POWER LED:Lightonaspoweron
FAULT LED:CPUonwhenabnormal
L1ONLED:Lightonoverlimit1set
L2ONLED:Lightonoverlimit1set
4beamsindicateLED:detectiondensity
willindicateper0-100%complyingwith
setup.
【forbutton】
→: use when changing beam of
number/value.
↑: usewhenincreasingnumber/value.
MOD
: usewhenswitchingparameterset
modeandmeasuremode.
Pushingaftersetchangewill
memorizeparameterandreturntomeasure
mode.
SELECT
:
Usewhenswitchingsetitems.
F−RESET
:
Usewhensettingmeasureagain.
%
Fig.11.FrontpanelofOutputunit
MatsushimaMeasureTechCo.,Ltd.
MADEINJAPAN

-
9
-
7.Startup
7-1.Warmupoperation
DustMonitorrequireswarmupoperationforabout1hourafterswitchon.
7-2.Measurerangeset
Afterfinishingwarmupoperation,youpleasesetmostappropriatemeasurerange.
Youpleasesettherangeat1-4iftheconditioninwhichnormalflowspeedisregular,
filteriscorrectandmeasureenvironmentisappropriate.
Incaserequiringobservationofsmallerchange,youpleasesettherangeat5〜9so
thatyoucanobservesmallerchange.
(Incasedensityishighhowever,youmaygetoverrange.)
4mA(0%) 20mA(100%) 4mA(0%) 20mA(100%)
Measurerange1 : 0.1 〜 1000mg/m3Measurerange6 : 0.1 〜 20mg/m3
Measurerange2 : 0.1 〜 500mg/m3Measurerange7 : 0.1 〜 10mg/m3
Measurerange3 : 0.1 〜 200mg/m3Measurerange8 : 0.1 〜 5mg/m3
Measurerange4 : 0.1 〜 100mg/m3Measurerange9 : 0.1 〜 2mg/m3
Measurerange5 : 0.1 〜 50mg/m3
Thisrange(density),utilizing「Talc」asstandardpowder,isdesignedtoget20mAoutput
underanenvironmentofflowspeedof10m/sandthemaximumrangeof1000mg/m3.
Astootherranges,themaximumvalueof20mAofeachrangeisdesignedtooutputdue
toitsoutputlinear.
【Reference:Talc】
(Talc)isnonorganicpowdermadefromorecalledTalcbeingcrushedinto
graywhiteparticleandsmoothpowder.
Talc,withitschemicalname:hydrousmagnesiumsilicate(Mg3Si4O10(OH))containing
mainingredientofSiO2approx60%,MgOapprox30%andcombinedwaterof4.88%.
Asitsphysicalnature,Talcoftruespecificgravity2.7〜2.8isasubstancehaving
lowesthardness,superiorinheatresistanceinadditiontochemically stablenature
amongnonorganicmineral.
7-3.Integralset
Integrationwillevenoffmeasuresignalpertimefixednumberuptomaximum30seconds.
(optionalset)
※ Troubleoffiltersystemwillindicatethroughsuddenriseofoutputsignal.
Whendoinganalysisoffiltersystem,setintegralat0second,please.
Bysettingintegralat0second,youcanmonitorsuddenchangeofdustdensity.
Asmeasureexample,bagfilterobservationexampleisindicatedforyouinFig.12
inwhichDustMonitorisusedfor13lines.
Dustexhaustisindicatingitsrespectiveamounttotransactas1〜6lines,and12,
13linesarelowbut7〜11linesprettyhighpeak.
Youcandecidewhenyouchangefilterbasedonyourobservation.
Filter No.
Density
of
dust
Time
Warning
point
Fig.12.Filterobservationpicture(example)

-
10
-
7-4.Limitset
Limitsetcanbedone2differentupperlimits(alarm)intherangeof4〜20mA(signal
level0〜100%)
Incasesignalchangegetsseverehowever,chatteringmightriseoften.
Insuchcaseyoumayberecommendedtouseintegralfunctionordelayfunctionat/from
higherposition(Customersexpertise)respectively.
【Limit1,2andactionchartoftroublealarm】
Limitoutput(upperlimitalarm)
PowerOFFtoopenCOM-NO
Troublealarm
PowerOFFtocloseCOM-NO
20mA
4mA
Time
Rangetoset
c
u
r
r
e
n
t
o
u
t
p
u
t
電
流
Limitsetvalue(ON)
Limitsetvalue(OFF)
Hysteresis
Trouble Devicenormal
Devicenormal
COM-NO close
LEDOn
COM-NO open
LEDOff
COM-NO open
LEDOff
COM-NO close
LEDon
COM-NO open
LEDoff
COM-NO open
LEDoff
Fig.13.Operationchart(upperlimitalarm)
Fig.14.Operationchart(Troublealarm)

-
11
-
7-5.Compensationvalueset
Compensationvaluecanoutputinarangeof0.1〜2.0(double)ascompensation
coefficient.
Insuchcaseasoutputistoohighortoolowforactualdensitywhenstandardcoefficient
is1.0,thisfunctioncanoutputnearestdensitytoactualdensitybymultiplyingoutput.
Utilizingthisfunction,outputcanbecompensatedasshownhereunder.
Thisfunctionis0.1inincrementsofinputintherangeof0.1〜2.0(defaultis1.0).
Ifmanagementisonlyfortrendofdustexhaust,youcandowithoutusingthisfunction
(=setvalue1.0).Incasemanagementrequiresthenearestvaluetoactualdensity,youseize
actualdensitysuchascounterwashpulsetime,etcbymanualanalysisfirstandthenset
counterwashpulseshownbydeviceverymuchclosetovalueofmanualanalysis.
Compensationvalue1.0 trend
Foractualdensity,
detectdensityoutputshigh.
Compensationvalue0.5 trend
Foractualdensity,
Detectdensityoutputsalmost
allalikeandunderstood
compensationiseffective.
Detectiondensity
Realdensity
Detectiondensity
Realdensity

-
12
-
8.Setexampleofparameter
8-1.Flowofparameterset
Parametershouldsetaccordingtotheflowasshownhereunder.
※
Incasesettingupvalueoverinputrangeexceedingspecification,setupvalue
ofparameterwillindicate「Err」onindicationLED,letyouknowsetupvalue
iswronganddonotacceptnextoperation.
※
Inthiscase,pleasechangetocorrectsetvalueagain.
8-2.Exampleofparameterset
Followingisasetexampletosetintegraltimeat15second.
(Providingthatsetvaluewas0secondbeforechange.)
(1) PushMODbutton1time・・・・・・・・・・movetoparameterwindowandLED(orange).
writtenRangeontheleftwilllighton.
(2) PushSELbutton5times・・・・・・・・LED(orange)litgoesdowninturn
andINTEGRATIONTIMELEDwilllight.
(3) Byusingboth→buttonand↑button,
setIndicationLEDat「15」.
(4) PushMODbutton1time・・・・・・・・・returntomeasurewindowandstartmeasure
againatintegral15second.
Youoperateotherparameterswithsameprocedure,please.
Incasesettingotherparameterscontinuously,at(4)aboveinsteadofMODyoupush
SELandcanperformbothsettingparameterandmovingtopluralitemsatthesametime.
Reference:ParameterandlightupofcorrespondingLED
RANGE:onlyRANGELEDlightup
L1ON : onlyL1ONLEDlightup
L1OFF : onlyL1OFFLEDlightup
L2ON : onlyL2ONLEDlightup
L2OFF : onlyL2OFFLEDlightup
INTEGRATIONTIME: onlyINTEGRATIONTIMELEDlightup
CORRECTION : bothRANGELEDandCORRECTIONLEDlightupsametime
9.Maintenance
Measureprobeshouldregularlybewipedandcleanedwithdryrag.Youpleasekeepitcleaned
onceinevery3monthsbutmoreoftenbecauseofchangingmediaormeasureenvironment,you
see.Whencleaning,youpleaseswitchoffpower.Temperatureofpipemaybehighanddustmonitor
mightbeheatedaccordingly.Becareful!Oryoumightgetyourhandsburned.
Caution:Theearthlineshouldnotberemovedevenatthemaintenanceincase
itmaycauseafault.
Measure
window Parametermode ValuesetParameterselect
MODbutton
Pressdown
(to parameter
mode)
SELbutton
Pushdown
Respective
parameterselect
→Move beam by
button
↑ Each button
pushdownchange
value
PushdownMODbutton
(afterparameterset,tomeasure
window)
SELbuttonpushdown
(aftersetparametermove
tootherparameteritem)
!

-
13
-
10.Troubleshooting
Whenyourdevicehasanyproblem,pleaserefertothefollowingparticularsandsolveitby
yourselvesfirst. Ifthedevicedoesnotrecovertonormalafterthoseconfirmationslisted
ontherightbelowhavebeendone,pleasekindlycontactMatsushima.
Contents Causethinkable Itemstoconfirm
PowerLEDand
OtherLEDs
areoff.
・Powerisnotsupplied.
・Powerspecificationischangedfromthe
orderedspecification.
・CPUabnormal.
・Confirmpowerlineiswired
toTerminalAC.
Ifnotsupplied,supplypower.
・
Confirmthevoltageonnameplate
ofTransduceristhesameasthe
suppliedpower.
Ifitisdifferent,returnus
Transducer,thenthepowerspec.
willbechangedatoutsite.
・PowerOFF,thenON.
Output
Directionis
unstable.
・Warmupoperationaftersetupisnotcomplete.
・Groundingisnotprovided.
・ThewiringbetweenSensorandTransduceris
eitherofforwrong.
・MaterialstickingonProbeelectrode.
・Probeelectrodehasdewcondensation.
・Atinyamountofmaterialisflowing.
・
Makewarmupoperationfor1hour
andcheckoutputagain.
・ConfirmDclassgrounding
(100ohmorless)isdone.
・Confirmwiringofexclusive
cable.
(SeePage6,5.Howtoconnect.)
If wiring is wrong, rewire
correctly.
・RemoveSensorandcheckifany
stickingonProbeelectrode.
Ifyes,poweroffandwipethem
offwithmoistcloth.(SeeNOTE)
・RemoveSensorandcheckifProbe
electrodehasdewcondensation.
Ifyes,poweroffandwipeit
offwithmoistcloth.(SeeNOTE)
・Checkifanapplicableamount
isflowing.Ifnecessary,
changethemeasurementrangeand
integralsetting.
Outputindication
goesoffscaleand
doesnotrecover.
・Ductorpipeisshort-circuitedwithProbe
electrodeduetothestickingmaterial.
・Probeelectrodehasdewcondensation.
・Probeelectrodeisincontactwithpipeinside
(GND).
・Deviceisintrouble.
・Alargeamountofmaterialisflowing.
・CPUabnormal.
・RemoveSensorandcheckifany
stickingonProbeelectrode.
Ifyes,poweroffandwipethem
offwithmoistcloth.(SeeNOTE)
・ConfirmifInsulationisinserted
morethan10mmintoDuctorPipe.
Insulationshouldbeextended
ifnecessary.(SeeNOTE)
・
RemoveSensorandcheckifProbe
electrodehasdewcondensation.
Ifyes,poweroffandwipeit
offwithmoistcloth.(SeeNOTE)
・Measureinsidediameterand
Probeelectrodelengthandcheck
iftheyarenotincontact.
Ifyes,changeelectrodelength.
(SeeNOTE)
・ConsultwithSalesDept.to
checkthesoundnessofDevice.
・Checkifapplicableamountof
materialisflowing.
ShortenProbeelectrodeif
necessary.(SeeNOTE)
・PowerOFF,thenON.
Table1.Troubleshooting

-
14
-
Outputindication
doesnotchange.
・Powerisnotsupplied.
・ThewiringbetweenSensorandTransduceris
eitherofforwrong.
・MaterialstickingonProbeelectrode.
・Probeelectrodeisincontactwithpipeinside
(GND).
・Deviceisintrouble.
・CPUabnormal.
・Confirmpowerlineiswired
toTerminalAC.
Ifnotsupplied,supplypower.
・
Checkwiringofexclusivecable.
(SeePage6,5.Howtoconnect.)
If wiring is wrong, rewire
correctly.
・RemoveSensorandcheckifany
stickingonProbeelectrode.
Ifyes,poweroffandwipethem
offwithmoistcloth.(SeeNOTE)
・Measureinsidediameterand
Probeelectrodelengthandcheck
iftheyarenotincontact.
Ifyes,changeelectrodelength.
(SeeNOTE)
・ConsultwithSalesDept.to
checkthesoundnessofDevice.
・PowerOFF,thenON.
FaultLEDis
lighted.
・Wiringiswrong.
・Thesuppliedpowersupplyisoutofallowable
range.
・CPUabnormal.
・Checkwiringofpower/output
signallines,exclusivecable.
(SeePage6,5.Howtoconnect.)
Ifwiringiswrong,rewire
correctly.
・
CheckpowersupplywithaTester
etc.
Ifitis out oftheallowable
range,applytheonewithinthe
range.
・PowerOFF,thenON.
[NOTE]
(1)WhenremovingSensor,loosenM5boltsandnuts.
Securelytightenthematinstallationaftercleaningit.
(2)Ifinsulationextensionisrequired,consultwithSalesDept.ofMatsushima.(non-free)
(3)Donotcutprobelengthtooshortincaseanyelectricchargemaynotbedetected.
This manual suits for next models