manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. medaxis
  6. •
  7. Medical Equipment
  8. •
  9. medaxis debritom+ Parts list manual

medaxis debritom+ Parts list manual

debritom+QuickCard - USA
1. Turn on debritom+
debritom+ON Activate foot pedal 1x to
connect with debritom+
2. debritom+set-up
2a
„click“
2b
2c 2d 2e
Insert pump as shown until it stops/turn clockwise until click
Attach the luer connection - make sure the
connection is tighten strongly
Attach rinsing fluid, connect the connecting tube to pump
Connecting tube: close clamp/spike into rinsing fluid/open venting cover
Open clamp
3. Test of water jet
Press foot on/off switch
•For a visual test of water jet (flat or
point, water jet thickness)
•To test the water jet intensity on your
hand – start with a distance of 10“
(approx. twice the hanpiece-length).
Guide the jet starting at the wound
edge, and move the jet slightly back
and forth.
•If something doesn‘t work, replace the
handpiece
4. Optimal working parameters
Intensity levels at START for each handpiece : Level 4
(Level 1-5/weak-strong)
debritom+handpiece standard jet REF 2000.0003
Working distance recommended 6-8“/15-20cm
debritom+handpiece flat jet narrow REF 2000.0004
Working distance recommended 1.5-4“/4-10cm
debritom+handpiece flat jet wide REF 2000.0005
Working distance recommended 1.2-3“/3-8cm
Working angle: 3 – 45 degrees to wound surface
TIP flat jet: The flat jet should be applied like a spatula to the wound.
5. Intensity-
Level
Intensity
Intensity
6. Patient change / end of the treatment
6a Patient change, pump one day
6b End of the treatment
Close clamp Remove handpiece and dispose of it Attach LuerCap
For next patient
remove LuerCap
=> 2d
Remove connecting tube/Remove pump and handpiece as one unit and dispose of it
debritom+OFF
Contraindication
––Malignant tumors
––Open injuries to blood vessels, and unprotected blood vessels
––Eyes, ears, nose
––Delicate vessels and structures, such as neurovascular structures
––Complex and/or highly contaminated wounds
––Patients with HIV, Hepatitis C, and other contagious diseases
A warning with regard to the use of the device for the following conditions, since
they are associated with higher risk. If treatment is indicated for the following
conditions, they should only
be performed under medical guidance and supervision:
––Patients with increased tendency to hemorrhage and arterial ulcers
––Infected wounds
This QuickCard does not replace the Instruction for use REF 9000.5504
swiss medical technology
debritom+QuickCard - USA
Medaxis LLC - 4512 Legacy Drive Suite 100 Plano, TX 75024
Phone 833 MEDAXIS / (833) 633-2947, [email protected], www.medaxis.us
medaxis AG / REF 9000.5505_B/ 20 .07.2020
swiss medical technology
7. Error messages
NOTE: The on/standby button lights up red as an error message occurs.This button can be used to acknowledge the fault after rectification.
Battery foot on/off switch almost empty
Troubleshooting: Replace batteries (Battery type 2 x LR03/AAA, chapter 11)
Pump in use for too long time period
Troubleshooting: Replace pump
Handpiece in use for too long or trying to use with different single use pump
Troubleshooting: Replace handpiece
Handpiece missing or handpiece not recognized
Troubleshooting: Replace handpiece. If the error can not be resolved, contact the Medaxis service department immediately.
Nozzle blocked or motor overheated
Troubleshooting: Change handpiece and pump if necessary. If the error can not be resolved, contact the Medaxis
service department immediately.
Pump not inserted correctly
Troubleshooting: Replace pump and reinsert it correctly, or replace it.
General system error
Troubleshooting: Please contact the Medaxis service department immediately.
8. Symbols on debritom+display
press
confirm
4
Search for Bluetooth device for pairing
Pressing the on/standby and ‚+‘ buttons for > 1 s at debritom+ will show the ‚Search for Bluetooth device‘
symbol for pairing a new or replacement of the foot on/off switch.
Bluetooth signal of foot on/off switch detected - confirm
As soon as debritom+has detected the Bluetooth signal of the foot on/off switch, this symbol appears on the display.
debritom+connected with foot on/off switch
This symbol appears on the display as the foot on/off switch is connected to debritom+.
Intensity levels
The intensity levels can be adjusted between 1 - 5. (1 weakest/5 strongest)
No foot on/off switch detected
Troubleshooting: Press the foot on/off switch. If foot on/off switch cannot be detected, replace the batteries (see IFU
chapter 11) and operate the foot pedal. If the foot on/off switch is still not detected, repeat the pairing procedure (see
IFU chapter 11)
Foot on/off switch time-out
The symbol appears if the foot on/off switch is pressed too slowly or the foot on/off switch is not pressed all the way
down. Troubleshooting: Release the foot on/off switch completely to acknowledge the error. Press the foot pedal all the
way down to start the water jet. If the fault cannot be rectified, contact the Medaxis service department immediately.

Other medaxis Medical Equipment manuals

medaxis debritom+ Parts list manual

medaxis

medaxis debritom+ Parts list manual

medaxis debritom+ Parts list manual

medaxis

medaxis debritom+ Parts list manual

medaxis debritom+ User manual

medaxis

medaxis debritom+ User manual

Popular Medical Equipment manuals by other brands

inhealth Blom-Singer Instructions for use

inhealth

inhealth Blom-Singer Instructions for use

Mindray Trio operating instructions

Mindray

Mindray Trio operating instructions

EDI VERIS EMS IV Installation & service manual

EDI

EDI VERIS EMS IV Installation & service manual

iFIT IFAHR120 quick start guide

iFIT

iFIT IFAHR120 quick start guide

VOCARE vitals360 user manual

VOCARE

VOCARE vitals360 user manual

Orliman OCR200 INSTRUCTIONS FOR USE AND PRESERVATION

Orliman

Orliman OCR200 INSTRUCTIONS FOR USE AND PRESERVATION

Osprey Medical R-RRL Instructions for use

Osprey Medical

Osprey Medical R-RRL Instructions for use

Atmos Chair M 2 operating instructions

Atmos

Atmos Chair M 2 operating instructions

Stryker Flosteady Arthroscopy Pump 200 manual

Stryker

Stryker Flosteady Arthroscopy Pump 200 manual

McKesson LED Exam Light Installation & operation manual

McKesson

McKesson LED Exam Light Installation & operation manual

Konica Minolta REGIUS CS-2 brochure

Konica Minolta

Konica Minolta REGIUS CS-2 brochure

Leica FL400 user manual

Leica

Leica FL400 user manual

Maquet HANAULUX blue 80 Hospital operating instructions

Maquet

Maquet HANAULUX blue 80 Hospital operating instructions

Sound Oasis SP-101 instruction manual

Sound Oasis

Sound Oasis SP-101 instruction manual

DeVilbiss Healthcare Vacu-Aide 7314 Series Service manual

DeVilbiss Healthcare

DeVilbiss Healthcare Vacu-Aide 7314 Series Service manual

Stryker Performance-PRO XT Operation & maintenance manual

Stryker

Stryker Performance-PRO XT Operation & maintenance manual

Invacare Medley Ergo user manual

Invacare

Invacare Medley Ergo user manual

SECHRIST 21465 Service manual

SECHRIST

SECHRIST 21465 Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.