Mediair SUF01 Series User manual

759001 (Rev.4) 10/2008
Portable AC/DC Suction Pump
User Manual


759001 (Rev.4) 10/2008
INDEX
Important Safeguards........................................................................................1-2
International Travel............................................................................................... 2
Introduction........................................................................................................... 2
Accessory/Replacement items............................................................................. 3
Set-Up .................................................................................................................. 4
How to Operate your Portable AC/DC Unit........................................................5-8
Battery Charging................................................................................................... 8
Check the battery quality...................................................................................... 9
Replace the battery .........................................................................................9-10
Cleaning Instructions.......................................................................................... 10
Maintenance....................................................................................................... 10
Troubleshooting...................................................................................................11
Specifications/Classifications ............................................................................. 12
Warranty............................................................................................................. 13
Declaration of Conformity................................................................................... 14




759001 (Rev.4) 10/2008 4
SET UP
SETUP
Connect either end of the 12” tubing to the tubing connector then connect the1. other end to the bacteria filter. Ensure that the clear side of the bacteria filter is
toward elbow and bottle when installing/re-installing. Do not reverse direction
of the filter.
The bacteria filter should then be connected to the 902. oelbow connection, and
the 90oconnection should be connected to the top of the canister lid where it
says <Vacuum>.
The 72” patient tubing should be connected to the canister lid at the outlet3. labeled <Patient>.
Please assure that all connections are secure and without leaks before using.4. Verify that unit is at desired suction level before beginning patient suction.5.
14” Connection Tubing with
Bacteria Filter
P/N AI5785-14C
72” Patient Tubing
P/N AI5785-07C

5759001 (Rev.4) 10/2008
OPERATION
HOW TO OPERATE’ YOUR PORTABLE SUCTION PUMP
Before connecting the unit to the AC adapter or 12V DC cord, make sure that1. the power switch located on the side of the unit is in the ’Off’ position. Select
power source desired.
LED EXPLANATIONS:
POWER - Green/Yellow: When external power is supplied to the unit fromAC power
source or DC cord, the power light will be illuminated as green. When internal
power is supplied to the unit from an internal rechargeable battery, the power
light will be illuminated as yellow.
CHARGING - Yellow: When the battery is being charged, the yellow light on the front
cover will be illuminated. At the same time, the red light on the AC adapter will
also be illuminated. When the battery is fully charged, the red light on the AC
adapter will turn to green.
LOW BATTERY- Red Low Battery: Seek another power source and charge the
battery as soon as possible when the red light remains flashing and the audio
alarm sounds.

759001 (Rev.4) 10/2008 6
OPERATION
Select Power Source Desired2.
AC OPERATION: Plug the small connector of the AC adapter into the DC power
input located on the side of the unit. Plug the AC end into a grounded wall-outlet
power source.
NOTE: The power supply may become warm to the touch during charging or
running of the unit. This is normal.
12V DC OPERATION: (such as acar lighter receptacle.) Plugthe small connector
of the DC power cord into the DC power input on the side of the unit. Plug the
large connector into the 12V DC power receptacle of the automobile.
BATTERYOPERATION: Verifythatyourunithasaninternalrechargeablebattery;
To ensure proper operation from internal battery, fully charge the battery for 1-1.5
hours (depending on the depth of discharge) as explained in the Battery Charging
section. To operate the unit from an internal rechargeable battery, ensure that no
external power sources are plugged into the DC power input on the side of the
unit.
Once the power source is selected, simply turn the unit “On” using the power3. switch located on the side of the unit. The power indicator light (displayed on
the top of the unit) will indicate which power source is being used by staying
continuously lit when external power is connected.

7759001 (Rev.4) 10/2008
OPERATION
WARNING
If the unit does not receive an external power source or the battery does not
get recharged immediately, the low battery indicator light will remain on and the
performance of the unit will drop off rapidly. Switch to another power source
immediately to avoid an interrupted suction procedure.
Adjust the vacuum level from 2 to 24 in Hg by turning the vacuum regulator4. knoblocatedon the sideof the unit(clockwiseto increase vacuumand counter-
clockwise to decrease vacuum). The desired level of vacuum can be set by
using the gauge located on the top of the unit near the handle. To accurately
read the gauge, block the patient end of the hose or cap off the canister and
allow the gauge to reach a stable vacuum reading.
WARNING
Tightening the knob above 24 in Hg (600 mmHg) will damage the valve.





759001 (Rev.4) 10/2008 12
SPECIFICATIONS / CLASSIFICATIONS
SPECIFICATIONS/CLASSIFICATIONS
Dimensions: (L) 11.5 x (W) 9.8 x (H) 9.8 in
(293 x 249 x 250 mm)
Weight: 7.5 lb (3.3 kg)
Maximum Vacuum: 24 in Hg (600 mmHg)
Flow Range (Free air Flow): >30 LPM
Sound Level (at 1m,, 24 in Hg): <60 dBa
Electrical Requirements: AC 100-240V, 50-60 Hz, 1.5A max
DC 14V, 3.2A max (45W max)
Battery Type: Lithium Ion
Average Run Time: 45 minutes
Bacteria Filter port size: 0.45 microns
Canister Capacity: 800cc (ml)
Operation Temperature: 32o- 104oF (0o- 40oC
Operating Humidity: 0-95% RH
Operating Atmospheric Pressure: 10.2 - 15.4 psi (70 - 106 kPa)
Storage Temperature: -40o- 149oF (-40o- 65oC)
Storage Humidity: 0 - 95%
Storage Atmospheric Pressure: 7.3 - 15.4 psi (50 - 106 kPa)
Equipment Classifications: With respect to protection from electric
shock Class I
Mode of Operation: Intermittent operation 30 minutes ON,
30 minutes OFF
NOTE: Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or
nitrous oxide.

13 759001 (Rev.4) 10/2008
WARRANTY
WARRANTY
TWO-YEAR LIMITED WARRANTY
The compressor portion of the Suction Unit is warranted to be free from defective
workmanship and materials for a period of two years from date of purchase.
Any defective part(s) will be repaired or replaced at our option if the unit has not
been tampered with or used improperly during that period. Make certain that any
malfunction is not due to inadequate cleaning or failure to follow the instructions.
If repair is necessary, contact your medical provider or our service department for
instructions.
NOTE
This warranty does not cover providing a loaner unit, compensating for costs
incurred in rental while said unit is under repair, or costs for labor incurred in
repairing or replacing defective part(s).
THERE IS NO OTHER EXPRESS WARRANTY, IMPLIED WARRANTIES,
INCLUDINGTHOSEOFMERCHANTABILITYANDFITNESSFORAPARTICULAR
PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED
WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL
IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY
AND LIABILITY FOR CONSEQUENTIALAND INCIDENTAL DAMAGES UNDER
ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION
IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS
ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR
EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE
ABOVE LIMITATION OR EXCLUSION MAY NOTAPPLY TO YOU.
This warranty gives you specific legal rights, and you may also have other rights
which vary from state to state.




Table of contents
Other Mediair Water Pump manuals
Popular Water Pump manuals by other brands

MSW
MSW MSW-M-BP-2000 user manual

Graco
Graco 224826 instructions

K2 Pumps
K2 Pumps WPS05001K owner's manual

Profi-pumpe
Profi-pumpe OASIS 500-3 LI+LED operating instructions

KNF
KNF N 816 TRANSLATION OF ORIGINAL OPERATING AND INSTALLATION INSTRUCTIONS

Wayne
Wayne Submersible Sump Pump CDUCAP Series Operating and parts manual

BUSCH
BUSCH COBRA ATEX NC 0100-0300 B instruction manual

PYD Electrobombas
PYD Electrobombas MOTORES 6 Manual Installation

Raider
Raider RD-WP40 user manual

Oerlikon Barmag
Oerlikon Barmag GM manual

Pentair
Pentair Myers MSK60 Installation and operator's manual

Well Pumps
Well Pumps 3"WPS-CP Installation and operation instructions