Medist MEVACS M90 User manual

D4O-0085-2-0
MEDIST s.r.o., Petrusovskeho 4, 066 01 Humenne, Slovakia
T-QN-04/05-01-1
2070
Medical electric suction unit
Ordering number:
MEVACS M90 - 72001.02
User's Manual

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
2 / 15
Dear Customers,
Dear Customers,Dear Customers,
Dear Customers,
Thank you for the decision to buy the MEDIST medical electric suction
Thank you for the decision to buy the MEDIST medical electric suction Thank you for the decision to buy the MEDIST medical electric suction
Thank you for the decision to buy the MEDIST medical electric suction
unit.
unit.unit.
unit.
Read the manual carefully, please
Read the manual carefully, pleaseRead the manual carefully, please
Read the manual carefully, please, in order to be able to utilise all its
, in order to be able to utilise all its , in order to be able to utilise all its
, in order to be able to utilise all its
features.
features.features.
features.
General instructions
•This user’s manual is inseparable component of the device. It should be stored so that it can be at
disposal at any moment.
•Especially important instructions are bounded in this manner
•Device MEVACS M90 is labelled with a mark according to MDD 93/42/EEC outline and meets
requests of the attachment 1 of the outline.
•Device meet requests of security specifications according to EN 60601-1, EN ISO 10079-1 and
interference requests EN 60601-1-2.
•Failure-free operation is guaranteed when
- parts and accessories used are listed into this operation manual and they are approved by
firm MEDIST
- device installation, its set-up, variations, service extension are made by MEDIST company or
authorized service organizations
- the device is used in accordance with this instruction manual
•Use only microbiological filters supplied by manufacturer, because these filters are
hydrophobic (prevent penetration of fluid into vacuum pump and protected it from damage)
with very high bacterial efficiencies up to 99,99999% particles bigger than 0,027 micron (which
is smaller than Hepatitis A, B and C).
•Company MEDIST does not warrant for secure operation and failure-free operation if uncertified
accessories were used.
•Duration and guarantee conditions follow standard clauses listed into warranty. Duration of warranty
is extended in the repair time in the case of warranted repair.
•Quality management system used by the company MEDIST meets international standards
EN ISO 9001 and EN ISO 13485.
Security warning
SECURITY
•The device can only by operated by an operator advised with the instruction manual and device
application methodic. Needed level of qualification – staff instructed how to operate with electric
equipment.
•Disposal of cover material perform in accordance with valid standards.
•Suction unit is not intended for explosive conditions. These conditions can arise by using flammable
anaesthetic gases and disinfecting skin agents. The instrument and foot-operated manipulator must
be placed outside this area.
•Suction hose must be always connected with filtering catheter, filtering nozzle or filtering set and
there must be no direct contact with the filtering place.
•Do not open the unit cover in order to avoid the electric shock.
CE
1293

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
3 / 15
•Device is protected by a safety valve into the lid of jar and safety valve into the lid of safety jar, and
hydrophobic filter against secret sucking into the vacuum pump. It in spite of this secret is sucked,
further using is possible as late as a control was made. The control can only be provided by a
authorized service organization.
•Before connecting the mains cord ensure that the local electric mains voltage is 230 V/ 50 Hz.
•Protect the mains cord against mechanical damage, especially against cutting and short-circuiting.
•Do not bend the mains cord too much.
•Pull the plug of the mains cord when unplugging.
•Never leave the mains cord plugged in electric mains, when unplugged from the unit socket.
•Attendant must make certain before every use of suction unit about its good condition and device
ready.
•
Don’t expose the collection vessels for direct UV ray of the germicidal lamps
Installation
•The device should have sufficient air circulation in order to avoid the overheating.
•Do not put the device on subjects, which could block the vents on the unit bottom part.
•Do not use or locate the device in the vicinity of heat sources.
Note
•Do not forget regular replacement of filters.
•When you are interconnecting the vessels, keep the rule that the vessel input leads through a
direct connector and the vessel output leads through the connector under which the safety
valve is located.
•When you are autoclaving, the vessels must stand freely upside down, they must not touch each
other and must not be loaded in order to avoid distortions. Max. temperature is 121°C.

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
4 / 15
Contents
1. INTRODUCTION...................................................................................................................................................5
1.1.
A
REAS OF USE
.....................................................................................................................................................5
2. EQUIPMENT..........................................................................................................................................................5
2.1.
B
ASIC EQUIPMENT
...............................................................................................................................................5
2.2.
S
UPPLEMENTING EQUIPMENT
..............................................................................................................................5
2.3.
S
AMPLE OF SOME VERSION SUCTION UNIT
:..........................................................................................................6
3. INSTALLATION....................................................................................................................................................7
3.1.
U
NIT LOCATION
...................................................................................................................................................7
4. PRODUCT DESCRIPTION ..................................................................................................................................8
4.1.
S
YMBOLS DESCRIPTION
.......................................................................................................................................8
5.
U
NIT INTERCONNECTION
........................................................................................................................................9
5.1.
I
NTERCONNECTION OF THE UNIT WITH BASIC EQUIPMENT
...................................................................................9
6. SWITCHING ON AND OPERATION .................................................................................................................9
6.1.
S
UCTION UNIT OPERATIONAL SET
-
OUT
................................................................................................................9
6.2.
V
ESSEL PREPARATION FOR SECRET
.....................................................................................................................9
6.3.
S
WITCHING ON
..................................................................................................................................................10
6.4.
F
UNCTIONAL EXAMINATION OF SUCTION UNIT
..................................................................................................10
6.4.1. Inner suction circuit ..................................................................................................................................10
6.4.2. Outer suction circuit..................................................................................................................................10
6.5.
C
HANGING OF THE SECRET VESSEL
....................................................................................................................11
6.6.
C
HANGING OF BACTERIOLOGIC FILTER
..............................................................................................................11
7. TECHNICAL DATA ............................................................................................................................................12
8. BASIC MAINTENANCE AND CLEANING.....................................................................................................12
8.1.
U
NIT ATTENDING AND CLEANING
......................................................................................................................12
8.2.
S
MALL FAULTS CORRECTION
,
UNIT TROUBLE SHOOTING
...................................................................................13
8.3.
R
EGULAR CHECK
...............................................................................................................................................14
9. MAIN ASSEMBLIES, SUBASSEMBLIES, COMPONENTS AND SPARE PARTS....................................14

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
5 / 15
1. Introduction
High-power fly suction unit MEVACS M90 is available for using into all kinds of surgery,
gynaecology and liposuction. Oilless vacuum-pump guarantees long time for no service-need
operation. Microbiological filter, silicone hoses and unbreakable vessels guard high hygiene at
work. Collecting secret vessels are autoclave
Main advantages of the suction unit are the following:
1. Oil-free vacuum pump,maintenance-free.
2. High level of vacuum - 92,4 kPa / 693 mmHg ( 92,4 % vacuum).
3. High suction output vacuum pump - 108 lit / min
4. Long-term operation - several days.
5. Simple control of vacuum level by regulator valve
6. Very low noise level.
7. Two suction circuits
8. Possibility switching of suction circuit
9. Suction volume of up to 8 litres without bottle change
10.Quadruple overflow protection
♦safety jar
♦overflow valve in lid of safety jar
♦overflow valve in lid of suction jar
♦hydrophobic bacteriological filter
11.Protection of the unit, personnel and patient against infection by means of fitted
bacteriologic filter.
12.The possibility of all functions operation by means of foot control.
13.Wide range of additional equipment:
•unbreakable autoclavable vessels for secretion with volume 2 L and 4 L
•suction adapters, single-use suction bags, cylinders ...
•various diameters of silicone suction hoses
♦vacuum-extractors, curettage adapters
1.1. Areas of use
a) Surgery - Plastic surgery - Accident surgery - Orthopaedics
b) Gynaecology (suction curettage, vacuum extraction)
2. Equipment
2.1. Basic equipment
You can find the following in the transport packing of the unit of basic performance delivery:
suction unit MEVACS, on which the following is fitted:
⇒plastic polycarbonate vessel for secretion 2.0 L - 2 pcs
⇒safety vessel - 2 pcs
⇒suction catheter container 50/400 mm
⇒suction microbiologic filter - 2 pcs
⇒suction silicone hose ∅8 x 3mm - long 2 m
electric cord
2.2. Supplementing equipment
No Name/type Ordering
number
1.
Suction hydrophobic microbiologic filter 2200/902 42101.04
2.
Silicone hose 10 x 3 40100.04
3.
Silicone hose 8 x 3 40100.00

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
6 / 15
4.
Suction adapter with suction interrupting for adults 42205.00
5.
Suction adapter with suction interrupting for children 42206.00
6.
STOP valve for adults - disposable 42201.05
7.
STOP valve for children - disposable 42201.04
8.
2.0 L jar TM, polycarbonate, with pressure-fit lid and metal tubing nipple 42011.11
9.
2.0 L jar TM, polysulfone, with pressure-fit lid and metal tubing nipple 42011.09
10.
2.0 L jar TM, polycarbonate, with silicone ring lid and metal tubing nipple 42011.15
11.
2.0 L jar TM, polysulfone, with silicone ring lid and metal tubing nipple 42011.13
12.
2.0 L jar TM, polycarbonate, with screw-fit metal lid and metal tubing nipple 42011.14
13.
2.0 L jar TM, polysulfone, with screw-fit metal lid and metal tubing nipple 42011.21
14.
4.0 L jar TM, polycarbonate, with pressure-fit lid and metal tubing nipple 42011.12
15.
4.0 L jar TM, polysulfone, with pressure-fit lid and metal tubing nipple 42011.10
16.
MONOVAC suction liner – capacity 1.4 L 42002.01
17.
EURO universal rail adapter TM - plastic 70105.16
18.
Direct hose connector 7-7 42803.07
19.
Direct hose connector 10-10 42803.06
20.
Direct hose connector 4-7 42804.04
21.
Direct hose connector reduced 4-10 42804.05
22.
Direct hose connector reduced 7-10 42804.06
23.
Direct hose connector reduced 6.5-15-6.5 42804.07
24.
T hose connector 5-5-5 42801.02
25.
T hose connector 7-7-7 42801.03
26.
Y hose connector 5-5-5 42805.04
27.
Y hose connector 7-7-7 42805.02
28.
Y hose connector 9-9-9 42805.03
29.
Suction catheter container 50/400 mm - TM 42003.01
30.
Footswitch for MEVACS M 73000.04
31.
Silicone vacuum extraction cup 50mm 70101.18
32.
Silicone vacuum extraction cup 60mm 70101.08
33.
Silicone vacuum extraction cup 70mm 70101.09
2.3. Sample of some version suction unit:
MEVACS M90 - BASIC - 2 x
2,0Lit jars
MEVACS M90 - 4 x 2,0Lit jars
MEVACS M90 - 3 x 2,0Lit jars

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
7 / 15
MEVACS M90 - 1 x 4,0Lit jar + 1
x 2,0Lit jar
MEVACS M90 - 2 x 4,0Lit jars
MEVACS M90 - 2 x 4,0Lit jars +
1 x 2,0Lit jar
MEVACS M90 - 2 x 4,0Lit jars +
2 x 2,0Lit jars
MEVACS M90 - 2 x 4,0Lit jars +
1 x 2,0Lit jar + foot switch
3. Installation
The MEVACS suction unit is a compact unit. There is no needed assembly.
3.1. Unit location
WARNING!
Suction unit is not intended for explosive
conditions. The instrument and foot-operated manipulator
must be placed outside this area „M“.

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
8 / 15
4. Product description
The suction unit consists of the following:
1. vacuum meter
2. regulator valve
3. mains switch
4. 2.0 L jar for secretion
5. suction catheter container
6. 4.0 L jar for secretion
7. wheels with brake
8. foot control
9. change valve
10. microbiologic hydrophobic filter
11. EURO rail adapter
12. safety jar
13. safety overflow valve
14. hose connector for connection of vacuum
branch
15. safety overflow valve
16. electric socket
17. connector for foot switch
18. electric cord holder
4.1. Symbols description
Symbol Description Location
Type B Equipment Serialized Label
Year of Manufacturing Serialized Label
1
9
2
3
4
5
6
7
8
10
12
13
15
14
11
16
17
18

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
9 / 15
0 / I
Switch On-Switch Off Control Panel
Turn the right - the vacuum is rising Control Panel
Attention, see the User’s Manual Control Panel
Foot Control Electric Label
Fuse Electric Label
5. Unit interconnection
5.1. Interconnection of the unit with basic equipment
Procedure:
1. Connect the silicone hose (2) to the hose
connector VACUUM (1).
2. Connect the bacteriologic filter (3) to the hose
(2). The side of filter marked „IN“ was directed
to the safety jar (6).
3. Shift the hose (4) onto the other bacteriologic
filter connector.
4. Shift the hose (4) to the output connector (5)
(safety valve under the connector is safety
valve (7)) of safety jar.
5. Shift the hose (10) on the input connector (9)
(stream-line tip (8) is under connector) of safety
jar (6).
6. Shift the loose end of the hose (10) on the
output connector (12) of collecting jar - a safety
valve is under connector - it is labelled as
“VACUUM”.
7. Shift the suction hose (11) on the input
connector (13) (without safety valve) of the
collecting jar - it is labelled as “PATIENT”.
6. Switching on and operation
Attention:
Ensure so as before each suction by new patient was realized sterilization of suction
hose with suction adapter, vessel for secretion, safety vessel and interconnection hose to
microbiological filter.
6.1. Suction unit operational set-out
Connect a suction unit with a vessel exit connector to secret – it is labelled as “VACUUM”. Hoses must not be
damaged, cracked or messed up.
Important!!!
Check the bacteriological filter from side “IN” and if the filter is glazed replace its.
6.2. Vessel preparation for secret
Vessels must be controlled for possible breaks, granularity or damage before use. Any defect can produce leakage or
split of the vessel (implosion of the vessel on secret is possible at vacuum).
Before sucking we recommend to infuse into the secret vessel a little water or disinfecting agent for easier cleaning.
1
2
3
4
5
10
8
7
6
11
12
13
9

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
10 / 15
6.3. Switching on
1. The unit is switched on by means of ON-OFF switch.
•position „0“ - the unit is off
•position „I“ - the unit is on (the switch illuminates)
2. Suction circuit - option and switching on
You can choose left or right suction circuit. It is possible to use both suction circuits at the same time.
3. Select the respective required value by regulator valve.
Vacuum is increasing by turning round of a regulating vent with the sun. Value of vacuum is displayed only if filtering
perimeter is closed, e.g. by cranking the end of the filtering hose.
Vacuum-meter scale is into bars.
-1bar = -100 kPa = -100% vacuum
6.4. Functional examination of suction unit
6.4.1. Inner suction circuit
•turn the suction unit on
•cover the connector (1) on the suction unit by your finger
•turn the vacuum regulating vent with the sun till it stop
•maximum vacuum value must be displayed on the vacuum-meter (see point “Technical data”)
Status of components and leakage inside of suction unit is controlled by the test.
6.4.2. Outer suction circuit
•turn the suction unit on
•crank or cover free end of the suction hose (11) by your finger
•turn the vacuum regulating vent with the sun till it stop
•vacuum-meter must display maximum value of vacuum (see point “Technical data”) after
some time (it depends of the secret vessel volume)
Status of components and leakage inside of suction unit outside its own body is controlled by the test.
Suction circuit off Suction circuit on
VACUOMETER SCALE

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
11 / 15
Note:
If the suction unit meets the first test but not the second, leakage must be into outer suction circuit
(incorrectly mounted a vessel lid, defective filter, hoses put on deficiently, released connectors,…)
6.5. Changing of the secret vessel
Secret vessels must be replaced or cleared when they are filled up to 4/5 of their volume. We recommend
having always a spare vessel in order to exchange the vessel quickly. Plastic vessel covers are equipped
with a security float vent protecting it against over-flowing. The vent must be removed during clearing.
6.6. Changing of bacteriologic filter
Suction microbiologic filter eliminated risk of contamination with
bacteria, viruses and infected particle from patient to suction pump
or central vacuum distribution. Suction microbiological filter is
hydrophobic with very high bacterial efficiencies up to 99,99999%
particles bigger than 0,027 micron (which is smaller than Hepatitis
A, B and C).
Applications:
Suction filter located behind output of the jar in front of the vacuum source. The filter put on so as the
side of filter marked „IN“ was directed to the jar and the side without marked to the vacuum source.
IN
IN
ON-OFF Valve
on suction unit Air flow form
safety jar
IN
IN
Important!!!
Check the
bacteriological filter from side “IN”
and if the filter is glazed replace its

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
12 / 15
Exchange:
The filter is necessary replaced, as it is glazed with dirtiness or by liquid sucked. Maximum using time is
30 days. After ending of suction of infection patient we recommended immediately exchanged the filter.
It prevents spread of infection.
7. Technical data
Typology MDD/93/43/EEC Medical Device CLASS IIa
Model
MEVACS M90
Specification by EN ISO 10079-1 High Vacuum - High Flow
Electrical safety Class I
Direct/indirect contact safety Type B
Substances penetration protection IP 20
Safety level in presence of anaesthetic/explosive mixture Not protected device
Power consumption 220-230V ± 10% / 50 Hz
Input power 420W + 15%
Electric protection 2 fuses Ø 5 x 20 T - 3,15 A
Electric motor Induction mono-phase with thermic protection
Vacuum pump Maintenance-free piston vacuum pump
Air flow rate of pump 108 L/min
Air flow rate on socket of suction
unit 89 L/min
Air flow rate on the end of suction
tube in the base version 76 L/min
Maximum vacuum up to - 92,4 kPa (693mmHg)
Jars 1or 2 x 2,0 / 4,0 LT polycarbonate jars with
overflow valve
Suction
Suction tube Silicone - Ø 8 x 3 mm - 2,0 m
Operation Continuous
Filtration Hydrophobic suction microbiological filter with
very high bacterial efficiencies up to
99,99999% particles bigger than 0,027 micron
Noise level 45 +15% dB (A) @1m
(by ISO 7779)
Weight 33 kg
Dimension (L x W x H) apr. - 500x500x1080 mm
Supply cord 4 m
Vacuum control valve
Control vacuum gauge (precision ± 2.5%)
Polycarbonate jar ml. graduated - autoclavable
Special features (optional):
-operation with foot control (IPX8)
Surrounding environment
temperature 5 ÷ 35
o
C
Working condition Relative humidity 30 ÷ 75%
Surrounding environment
temperature - 40 ÷ 70
o
C
Storage condition Relative humidity 10 ÷ 100%
Packing - wrappage 5 laminated cardboard carton
International Standards:
MDD 93/42/EEC; EN 60601-1; EN 60061-1-2; EN ISO 10079-1; EN ISO 14971
Technical life: 10 years
8. Basic maintenance and cleaning
8.1. Unit attending and cleaning
After each use of the MEVACS unit the cleaning and disinfection of secretion vessel and suction hose is
necessary to be performed and you have to perform sterilization. The filter replacement when it has
blocked with liquid or dirty over time. After ending of suction of infection patient we recommended

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
13 / 15
immediately exchanged the filter and used up filter to disposal. It prevents spread of infection. The vessel
content is decontaminated in a way utilised on site and the vessels including hoses are submerged into
the disinfection solution. Keep the following rules during maintenance of the polycarbonate secretion
reservoir and silicone hoses:
•Do not use abrasives
•Check the compatibility of polycarbonate and detergents used in the hospital. Polycarbonate
enters into reaction with aldehyde residues, which are always present after use of disinfection
solutions containing formaldehyde, glyoxale, glutaraldehyde
•Since we cannot know the whole range of cleaning and sterilization agents used in hospitals, we
recommend to contact your suppliers and consult the compatibility of their agents and
polycarbonate. Disinfection agents proved to be successful in practice are sekusept and
chloramine (Formalin and organic solvents are prohibited)
•Before putting into autoclave wash the bottle thoroughly and dry in order to remove all residues of
disinfection agents and water
•Put the bottle into autoclave always upside down without contact with other components
•Autoclave the bottle and the cover always separately
•Do not put heavy subjects on the bottle in the autoclave to prevent its deformation
•Autoclave the polycarbonate and silicone hoses at temperatures not exceeding 121°C max. for
the period of 20 min.), autoclave mode „RUBBER“.
The microbiologic filter is disposable. The filter replacement when it has blocked with liquid or dirty over
time. The microbiologic filter protected in the main patient or operating personnel into the face of infection
and protect together suction unit for overflow secretion.
The plastic case, holders, truck and other equipment can be cleaned with soft, wet cloth moistened by
cleaning agent, which does not react with the plastic and its surface finish,(SEKUSEPT). Agents with
organic solvents are prohibited.
Also cleaning agents based on synthetic spirit (OKENA) are suitable, but also disinfection agents on the
basis of sprays can be used.
Important! It is necessary to disconnect the mains cord before cleaning the unit. During cleaning the
electric connectors, switch and the mains socket must be protected against water or cleaning agents, do
not submerge the unit into water.
Note: No lubrication is necessary.
8.2. Small faults correction, unit trouble shooting
As each unit of our company, also MEVACS suction unit was tested and found trouble-free. Despite
that fact some faults can occur, which follow either from blocking protection systems or from incorrect
use.
For better orientation the faults, you can correct yourselves, are shown in the following table.
Damage symptom Possible reason Repairing action
Vacuum pump engine
does not operate
•Switch is not on
⇒Power supply
•Instrument is not connected
⇒Fuses
•Air pump engine is damaged
•Turn the switch into the position “I“
⇒Repair electric socket into the room
•Connect a power cord into the instrument
⇒Replace fuses into the instrument socket
•Call service technician
Inadequate sucking
⇒Leakage into the vacuum pump or inner
suction perimeter
•Hoses are damaged or set up incorrectly
⇒Glazed filter
•Damaged filter body
⇒Damaged connectors or “O” rings on the
vessel lid
⇒Damaged gasket into the vessel lid
•Damaged vessel, ripped or torn boundaries
⇒Determine failure source according to point
6.4.1. Call service technician.
•Control and push hoses
⇒Replace the filter
•Replace the filter
⇒Replace a vessel lid, call service technician
⇒Replace a vessel lid, call service technician
•Replace the vessel

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
14 / 15
Vacuum pump engine
operates, however a
suction unit does not
suction and vacuum-
meter displays vacuum
•Leech security float into the vessel lid
⇒Hoses blocking into the inner suction system
•Absorbed liquid into the hydrophobic filter
•Turn the suction unit off and disconnect a
hose carefully from the suction unit socket
⇒Determine failure source according to point
6.4.1. Call service technician.
•Replace the filter
•Since it is not possible to specify all possibilities of faults, which can occur during the unit
operation, in case of not specified faults contact either qualified persons of service company,
or directly the manufacturer:
MEDIST s.r.o., Petrusovskeho 4, 066 01 Humenne, Slovakia
8.3. Regular check
Suction flask MEVACS is no service-needed device. We recommend perform regular check minimally
once a year for mistake-free operation. Only service technicians can perform check and repair. They have
authority from the company MEDIST s.r.o., Humenne.
9. Main assemblies, subassemblies, components and spare parts
No Ord. number
Name No Ord. number Name
1 40100.00 Silicone hose 8x3 - 2 m 5 70105.35-X Safety jar right or left
2 42101.04 Suction filter 6 40100.00 Connecting silicone hose 8x3
mm
3 70105.16
EURO rail adapter
7 13801.00 Electric cord
4 42003.01 Suction catheter container
2
1
3
4
5
6

UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
15 / 15
Jars:
2.0 L jar TM with silicone ring lid
☛polycarbonate, autoclavable 121
O
C (No: 42011.15)
☛polysulfone, autoclavable 134
O
C (No: 42011.13)
☛safety valve in lid
☛metal connectors
2.0 L jar TM with screw metal lid
☛polycarbonate, autoclavable 121
O
C (No: 42011.14)
☛polysulfone, autoclavable 134
O
C (No: 42011.21)
☛safety valve in lid
☛metal connectors
4.0 L jar TM with pressure-fit lid
☛polycarbonate, autoclavable 121
O
C (No: 42011.12)
☛polysulfone, autoclavable 134
O
C (No: 42011.10)
☛safety valve in lid
☛metal connectors
MONOV C suction liner
☛MONOVAC 1.4 L (No: 42001.31)
☛safety valve in lid
B-15/16
B-17/18
B-22/23
B-27/28
Table of contents
Popular Medical Equipment manuals by other brands

AccuRelief
AccuRelief ACRL-9100 user manual

Magic Care
Magic Care MIZAR Instructions for use manual

Stryker
Stryker 6253 Operation & maintenance manual

Aspire
Aspire Community & Homecare Bed Range user manual

Sissel
Sissel Portable Massage Table Robust Instructions for use

bort medical
bort medical ActiveColor quick guide