Mesalabs IBP HDM97Pocket Series User manual

www.ibpmedical.com
HDM97Pocket Reference Meter Series
Safety Instructions and Quick Start Guide
HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ
This document describes the major functions of the HDM97Pocket instrument series. The complete manual can
be found at http://www.ibpmt.com/medical_e/meter_documents.shtm and on the installation CD. For reading
the document Adobe®PDF it’s necessary to have Adobe®PDF Reader installed. Adobe®Reader for Microsoft®
Windows®is included in several language versions on the CD.
1
HDM97Pocket Reference Meter Series
Safety Instructions and Quick Start Guide
HDM97
BL
HDM97
BM&BH
HDM97
BN
HDM97
BO
HDM97
BP
HDM97
BQ
This document describes the major functions of the HDM97Pocket instrument series. The complete manual
can be found at http://www.ibpmt.com/medical_e/meter_documents.shtm and on the installation CD. For
reading the document Adobe
®
PDF it's necessary to have Adobe
®
PDF Reader installed. Adobe
®
Reader for
Microsoft
®
Windows
®
is included in several language versions on the CD.
Display
Reset / T
are
pressure / Adjust (pH)
and navigation key "OK"
Special functions
and navigation key
"next"
Select fu
nctions
and navigation key
"previous"
ON / OFF
For OFF press 3 sec.
Channel select
and navigation key "ESCAPE"
Display
Reset / tare pressure / adjust (pH)
and navigation key “OK”
Special functions and navigation key
“next”
Select functions and navigation key
“previous”
ON / OFF
For OFF press 3 sec.
Channel select
and navigation key “ESCAPE”

HDM97Pocket Quick Reference \\ 2www.ibpmedical.com
Safety Instructions
For Your and Your Patients Safety Read and Follow Carefully the Safety Instructions Below
• Read the entire manual (not only this guide) carefully before using the HDM97Pocket.
• Keep the instrument away from unauthorized persons.
• Only use the HDM97Pocket as reference instrument to measure Conductivity, Temperature, Pressure, Flow and pH depending
to it’s equipped features.
• Never use the HDM97Pocket as a replacement of sensors of a medical device.
• Do not use the HDM97Pocket in conjunction with a medical device during a treatment.
• Only use the instrument in a dry environment and do not touch it with wet hands.
• Make sure that no liquids intrude inside the instrument or it’s inlets / outlets.
• Always use a clean protection filter for the pressure transducer inlet.
• Prevent every mechanical overstraining of electrical wires. Never buckle or pull the cable of the conductivity probe.
• If the acquired values seem to be not believable, make sure that the HDM97Pocket is not defective.
• Prevent electrostatic discharge on the connectors. This can lead to substantial damage of the instrument. Make sure to be
completely discharged before touching the connectors or cables connected to the instrument.
• Adjust the meter only, if you are familiar with the consequences of the adjustment. Consider the hints for reference solutions
in the user manual.
• Potentials above 42V against earth ground are dangerous. This potentials can lead to electrical shock and therefore to
health hazards. Make sure that none of the connectors has higher voltage than mentioned before.
• The direct connection of earth potential to any connector of the instrument is not applicable for safety purposes and may
substantially damage the instrument. Make sure that the medical device itself is properly connected to earth ground
according to its instruction manual.
• Make sure that the instrument never heats over 60°C / 140°F. Prevent direct sunlight.
• Never sterilize the instrument using an autoclave. Danger of explosion of the battery!
• Never throw the meter into a fire. Danger of explosion of the battery!
• Never open the meter. There are no serviceable parts inside.
• Never try to replace or repair the internal battery. Handled in the wrong way these part is dangerous. Extreme risk of fire
and explosion!

HDM97Pocket Quick Reference \\ 3www.ibpmedical.com
Access Connectors
Depending on the configuration the appliance part consists of a BNC inlet for pH electrode connection, a male connector
for pressure transducer inlet, a connector for the flow probe and a directly attached cable which leads to the conductivity /
temperature probe or connector for the probe. The appliance part is isolated against the USB-Connector withstanding 4000V
AC or 6000V DC for one minute minimum according to ISO EN 60601-1.
USB CONNECTOR / CHARGING THE BATTERY
The USB connector is mounted on the right side of the instrument. It is implemented as a
Mini-USB-A receptacle. The current consumption of the HDM97Pocket (all versions) can be up
to 500 mA while charging the battery. The USB connector can be used to charge the battery and
communicate with the instrument at the same time. The period for charging an empty battery to
100% is approximately 3 hours.
ATTENTION: The USB connector is not part of the appliance part.
HDM97BL HDM97BH
HDM97BN HDM97BO
Flow Pressure
Conductivity
Temperature
Conductivity
Temperature
Pressure
Conductivity
Temperature
Conductivity
Temperature
Pressure
Pressure
pH Flow
Conductivity
Temperature
Pressure
Flow
HDM97BP HDM97BQ
HDM97BQ with Option C (conductivity probe connector)

HDM97Pocket Quick Reference \\ 4www.ibpmedical.com
Basic Settings
BASIC SETTINGS
After switching on the instrument immediately press “F” key to invoke the basic settings menu. The parameter to change can
be selected using the navigation keys down arrow and up arrow. By pressing the navigation key (the “ok” button) you can
adjust the value of each parameter.
These parameters are as follows:
Parameter Meter Versions Description
AUTO-OFF All Set the delay time fort he auto power off function or disables it
BEEP All Enables / disables the keypad acknowledge beep
CAL-RES All Overwrites the user calibration dataset with the device initial calibration
dataset
LANGUAGE All Selectionof display language.
PHT BM,BH, BO,BP,BQ Target time adjustment for pressure holt test (PHT)
TIMER All Enables / disables the stopwatch function
Flow BL, BQ Enables / disables the Flow channel
pH BP enables / disables the pH channel
TEMP BN, BO, BP, BQ Enables / disables the sole readout of the temperature channel
PR HR-Res BM, BO, BP, BQ Enables / disables the high resolution for pressure channel
AUTO-OFF
OK
BEEP
OK
OK
ESC
OK
ESC
OFF
30 MIN
60 MIN
90 MIN
disables function
30 minute delay
60 minute delay
90 minutes delay
ON
OFF
keypad beep enabled
keypad been disabled
Then to apply and save setting,
or to discard this setting.
CAL-RES
OK Press 5 seconds until the counter in the display reaches zero. Then a calibration dataset reset will
be preformed. This takes a couple of seonds. The instruments performs a restart afterwards.
LANGUAGE
OK DEUTSCH
ENGLISH
German language
English language
OK
ESC
Then to apply and save setting,
or to discard this setting.

HDM97Pocket Quick Reference \\ 5www.ibpmedical.com
Then to apply and save setting,
or to discard this setting.
TIMER
OK
OK
ESC
ON
OFF
stopwatch function enabled
stopwatch function disabled
Then to apply and save setting,
or to discard this setting.
PHT
OK
Pressure Hold Test
increases the value
lowers the value
OK
ESC
Then to apply and save setting,
or to discard this setting.
pH
OK
Only available with version BP
ON
OFF
pH enabled
pH enabled
OK
ESC
Then to apply and save setting,
or to discard this setting.
FLOW
OK
Only availble with version BQ
ON
OFF
FLOW enabled
FLOW disabled
OK
ESC
Then to apply and save setting,
or to discard this setting.
TEMP
OK
Only availble with version BN, BO, BP and BQ
ON
OFF
sole readout of temperature enabled
sole readout of temperature disabled
OK
ESC
PR H-RES
OK
Only availble with version BM, BH, BO, BP and BQ
ON
OFF
sole readout of temperature
enabled
OK
ESC
Attention: The higher resolution does not give a higher measuring accuracy.
EXIT
OK A save dialog will appear, if changes have been made.
ON
OFF
OK
ESC
SAVE
to apply and save all settings by exiting
to discard all settings by exiting

HDM97Pocket Quick Reference \\ 6www.ibpmedical.com
Press 3 seconds
Versions
The HDM97Pocket meters are equipped with up to 5 channels, depending on the version.
Channel HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ
Conductivity
Temperature
Timer
Pressure
pH
Flow
Operating
The functions of the keys are structured hierarchically from left to right.
X X X X
X X X X
X X X X
X X X X
X
X X
CD/TEMP (Available in versions BN, BO, BP and BQ)
Returns automatically back to measuring mode after 5 sec.
Selection of conductivity temperature coefficient
Reading min. Value Reading max. Value
Reset of both the min. And max. Value
Return to measuring mode
ESC
FF
Channel Select
No function
TEMP
F
(Available in versions bn, bo, bp and bq, if enabled in the basic settings)
Selection between units °c or °f
Returns automatically back to measuring mode after 5 sec.
Reading min. Value Reading max. Value
F
Reset of both the min. And max. Value
Return to measuring mode
ESC
No function
PRES
F
(Only available in versions bm, bh, bo, bp and bq)
Selection between units mmhg, mbar, kpa and psi
Pressure drop measurement
OK
ESC
Start countdown timer
Return to measuring mode
FReading min. Value FReading max. Value
Returns automatically back to measuring mode after 5 sec.
Reset of both the min. And max. Value
ESC Return to measuring display
Tare (shown with “zero” in display afterwards)

HDM97Pocket Quick Reference \\ 7www.ibpmedical.com
pH (Available only in versino BP, if enabled in the Basic Settings)
Returns automatically back to measuring mode after 5 sec.
Selection between compensation 20°C, 25°C, 37°C and measured temperature
Reading min. value
F F Reading max. value
Flow
F
No function
Reset of both the min. And max. Value
ESC Return to measuring display
(Available only in versino BL, BQ, if enabled in the Basic Settings)
Attenuation
Reading min. value FReading max. Value
Returns automatically back to measuring mode after 5 sec.
Reset of both the min. And max. Value
ESC Return to measuring display
Reset accumulated volume to 0
Timer (If enabled in the Basic Settings)
OK
OK
OK
Afterwards
Afterwards
Timer
Timer
Timer
The cycle “start, stop and reset” can start again
No function
No function
Multifunction key: The first push starts the timer, the next push stops the stopwatch and pushing it agian
F
Timer Function
The reading of the timer is as follows.
TIMER The display changes to MM/SS/DS automatically after 2 seconds
MM/SS/DS
00.00.0
minutes.seconds.tenthseconds
After 59 minutes and 59.9 seconds the display form changes so that hours, minutes and seconds are being displayed. The
tenth second display is not available then. After 9 hours, 59 minutes and 59 seconds the stopwatch stops itself.
HH/MM/SS
0.00.00
hours.minutes.seconds

HDM97Pocket Quick Reference \\ 8www.ibpmedical.com
The adapter is suitable for measuring pressure, temperature and conductivity in flow-
though mode. With immersion usage of the probe it is suitable for measuring temperature
and conductivity concurrently. The preferred orientation of the adapter is vertical. The axial
connector must point down (see picture). In flow-through mode the axial connector is to be
used as inlet meanwhile the side connector is to be used as outlet. The connection to the
pressure transducer inside the HDM97Pocket is accomplished by the lateral connection
above the inlet. Before measurement it is necessary to shake the adapter slightly, so that
air bubbles trapped inside can pass off through the outlet.
For immersion usage the probe must be unscrewed from the flow-through adapter. To
accomplish that do not turn the upper shank, the probe may be damaged. Only turn the
nut as shown in the picture left.
Never touch the carbon ring electrode surfaces with your fingers or other objects. Do
not buckle the probe cable. The minimum allowed bending radius is 3 cm or 1¼ inches.
IBPVIEW DATA ACQUISITION
For details read the user manual HDM97Pocket. Templates for the HDM97Pocket meters are
available on the CD “HDM97Pocket”.
All brand names mentioned in this document are property of their respective owners.
Flow-through Adapter for Conductivity
and Temperature Measurement
For complete details on IBP products visit:
www.ibpmedical.com
IBP Medical GmbH
Ikarusalle 15
D 30179 Hannover
Germany
Phone: +49 511 651647
Fax: +49 511 652284
This manual suits for next models
6
Table of contents
Other Mesalabs Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Bacharach
Bacharach CO2 Analyzer 3150 Operation & maintenance instructions

Leica
Leica PINMASTER II PRO instructions

Coatmaster
Coatmaster 3D user manual

Swan Analytical Instruments
Swan Analytical Instruments AMI Sodium P Operator's manual

ABB
ABB AquaMaster 3 manual

Büttner Elektronik
Büttner Elektronik Dometic MT iQ Basic Pro Installation and operating manual